
D-5759-2018
Disassembly/Assembly Flow Sensor
Open the ap by lifting the lower
edge upwards.
Close the ap.
Push the ow sensor as far as it
possible go to the left.
Push the ow sensor to
the right into the socket.
Remove the expiratory valve
from the tting.
Turn the locking ring as far as
possible to the left.
Remove and empty the water
trap container.
Connect the water trap container.
Remove the diaphragm.
Fit the ow sensor
into the socket.
Remove the ow sensor
from the socket.
Remove the sensor sleeve
from the expiratory valve.
91 03 106 | 19.11-1 | AW | LE | Subject to modications | © 2019 Drägerwerk AG & Co. KGaA
Attach the diaphragm onto the
edge of the expiratory valve housing.
Make sure that it tts properly.
Attach the ow sensor sleeve
on the expiratory valve.
Turn the locking ring as far
as possible to the right
until it clicks into place.
Turn the locking ring as far as
possible to the left, push the
expiratory valve into the tting.
Dräger Evita®V-Series
Flow Sensor and Expiratory Valve
Disassembly/Assembly
Disassembly/Assembly Expiratory Valve*
* The Neonatal Expiratory Valve is
also working with Evita® V-Series.
For manual reprocessing of the
Neonatal Expiratory Valve please
refer to Babylog guideline.
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.