manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Dräger
  6. •
  7. Control Unit
  8. •
  9. Dräger Evita V Series User manual

Dräger Evita V Series User manual

Other manuals for Evita V Series

6

Other Dräger Control Unit manuals

Dräger REGARD 2400 User manual

Dräger

Dräger REGARD 2400 User manual

Dräger Scio Four User manual

Dräger

Dräger Scio Four User manual

Dräger Plus Series User manual

Dräger

Dräger Plus Series User manual

Dräger REGARD 7000 User manual

Dräger

Dräger REGARD 7000 User manual

Dräger Regis 300 User manual

Dräger

Dräger Regis 300 User manual

Dräger PAS ASV User manual

Dräger

Dräger PAS ASV User manual

Dräger PAS ASV User manual

Dräger

Dräger PAS ASV User manual

Dräger RailGard-S1 User manual

Dräger

Dräger RailGard-S1 User manual

Dräger REGARD 4208797 Installation and operating instructions

Dräger

Dräger REGARD 4208797 Installation and operating instructions

Dräger RAB 7000 User manual

Dräger

Dräger RAB 7000 User manual

Dräger BS8468-1 User manual

Dräger

Dräger BS8468-1 User manual

Dräger REGARD 3900 Series User manual

Dräger

Dräger REGARD 3900 Series User manual

Dräger VarioGard User manual

Dräger

Dräger VarioGard User manual

Dräger PEX 1000 User manual

Dräger

Dräger PEX 1000 User manual

Dräger Area User manual

Dräger

Dräger Area User manual

Popular Control Unit manuals by other brands

Festo Compact Performance CP-FB6-E Brief description

Festo

Festo Compact Performance CP-FB6-E Brief description

Elo TouchSystems DMS-SA19P-EXTME Quick installation guide

Elo TouchSystems

Elo TouchSystems DMS-SA19P-EXTME Quick installation guide

JS Automation MPC3034A user manual

JS Automation

JS Automation MPC3034A user manual

JAUDT SW GII 6406 Series Translation of the original operating instructions

JAUDT

JAUDT SW GII 6406 Series Translation of the original operating instructions

Spektrum Air Module System manual

Spektrum

Spektrum Air Module System manual

BOC Edwards Q Series instruction manual

BOC Edwards

BOC Edwards Q Series instruction manual

KHADAS BT Magic quick start

KHADAS

KHADAS BT Magic quick start

Etherma eNEXHO-IL Assembly and operating instructions

Etherma

Etherma eNEXHO-IL Assembly and operating instructions

PMFoundations Attenuverter Assembly guide

PMFoundations

PMFoundations Attenuverter Assembly guide

GEA VARIVENT Operating	 instruction

GEA

GEA VARIVENT Operating instruction

Walther Systemtechnik VMS-05 Assembly instructions

Walther Systemtechnik

Walther Systemtechnik VMS-05 Assembly instructions

Altronix LINQ8PD Installation and programming manual

Altronix

Altronix LINQ8PD Installation and programming manual

Eaton Powerware Series user guide

Eaton

Eaton Powerware Series user guide

WAGO 750-407/040-000 manual

WAGO

WAGO 750-407/040-000 manual

Sony CCU-590 Operation manual

Sony

Sony CCU-590 Operation manual

Graco 25D614 instructions

Graco

Graco 25D614 instructions

Task Force Tips YE-RF-900 Instructions for installation, safe operation and maintenance

Task Force Tips

Task Force Tips YE-RF-900 Instructions for installation, safe operation and maintenance

Shini SPV-U Series manual

Shini

Shini SPV-U Series manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

D-5759-2018
Disassembly/Assembly Flow Sensor
Open the ap by lifting the lower
edge upwards.
Close the ap.
Push the ow sensor as far as it
possible go to the left.
Push the ow sensor to
the right into the socket.
Remove the expiratory valve
from the tting.
Turn the locking ring as far as
possible to the left.
Remove and empty the water
trap container.
Connect the water trap container.
Remove the diaphragm.
Fit the ow sensor
into the socket.
Remove the ow sensor
from the socket.
Remove the sensor sleeve
from the expiratory valve.
91 03 106 | 19.11-1 | AW | LE | Subject to modications | © 2019 Drägerwerk AG & Co. KGaA
Attach the diaphragm onto the
edge of the expiratory valve housing.
Make sure that it tts properly.
Attach the ow sensor sleeve
on the expiratory valve.
Turn the locking ring as far
as possible to the right
until it clicks into place.
Turn the locking ring as far as
possible to the left, push the
expiratory valve into the tting.
Dräger Evita®V-Series
Flow Sensor and Expiratory Valve
Disassembly/Assembly
Disassembly/Assembly Expiratory Valve*
* The Neonatal Expiratory Valve is
also working with Evita® V-Series.
For manual reprocessing of the
Neonatal Expiratory Valve please
refer to Babylog guideline.
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.
D-5759-2018
Prepare the cleaning and the disinfectant solution in
accordance with manufacturer’s instructions.
Dräger Evita®V-Series
Flow Sensor
Reprocessing instructions
Manual cleaning
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.
Swirl the ow sensor back and forth at least 3 times.
Make sure the cleaning agent reaches all surfaces
and interior spaces.
Place the ow sensor in the cleaning agent. Ensure that the cleaning solution ushes all surfaces
and interior spaces. At the end of the specied
contact time, swirl the ow sensor back and forth
again at least 3 times.
After the contact time, rinse the ow sensor in the
water bath (at least drinking water quality). Shake out residual water carefully.
Check the ow sensor for visible soiling and repeat
steps if necessary.
Check the ow sensor for visible damage, paying particular attention to its measuring wires and their pins,
and replace if necessary.
Allow the ow sensor to dry completely.
Do this until no more cleaning agent residue is visible.
D-5759-2018
Dräger Evita®V-Series
Disinfection
Machine cleaning
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.
Manufacturer:
Drägerwerk AG & Co. KGaA
Moislinger Allee 53–55
23558 Lübeck, Germany
Disinfection by immersion
Place the ow sensor in the disinfectant. Observe the specied contact time.
At the beginning of the contact time, swirl the ow sensor back and forth at least
3 times. Make sure the disinfectant reaches all surfaces and interiors.
At the end of the contact time, swirl the ow sensor back and forth again at least
3 times.
After the contact time, rinse the ow sensor in the water bath (at least drinking
water quality) until no more disinfectant residue can be detected.
Shake out residual water carefully. Allow the ow sensor to dry completely.
Check the ow sensor for visible damage, paying particular attention to its
measuring wires and their pins, and replace if necessary.
If you need visual descriptions, please look at the reprocessing instructions for manual cleaning.
Position the expiratory valve as shown. Position the components (ow sensor sleeve, diaphragm, water trap container)
to be stable.
Please ensure that all surfaces and interior spaces can be completely rinsed and the water can drain o freely.
Machine cleaning
6