1. General Safety
2. Product Description
3. Installation
4. Cleaning
5. Annual Maintenance
6. Troubleshooting
7. Technical Description
8. Guarantees and Warranties

3
1. General Safety
Before you use your Eclipse Seating system, please familiarise yourself
with the information in this user manual. Keep this manual in a handy place
for future reference.
)RU\RXUVDIHW\DQGWKHHI¿FLHQWSHUIRUPDQFHRIWKHV\VWHPWKHIROORZLQJ
precautions should be observed:
- MAXIMUM PATIENT WEIGHT 25 stones / 159 kgs
- Do not expose the pump to liquids of any kind
$OZD\VSURWHFWWKHV\VWHPVIURPRSHQÀDPHV
- Avoid the use of Phenol based solutions when cleaning
- Store the system in a clean and dry environment
- Do not store in direct sunlight
Protected against solid foreign objects of 12.5mm and greater
Class II
“BF” symbol, indicate this product is according to the degree of
protecting against electric shock for type BF equipment
Disposal of Electrical & Electronic Equipment (WEEE):
This product should be handed over to an applicable collection
point for the recycling of electrical and electronic equipment. For
more detailed information about the recycling of this product, please
FRQWDFW\RXUORFDOFLW\RI¿FHKRXVHKROGZDVWHGLVSRVDOVHUYLFHRUWKH
retail store where you purchased this product.
Refer to instruction manual/booket
NOTE on ME EQUIPMENT ‘Follow instructions for use’
Attention, should read the instructions!
Manufacturer
Authorised representative in the European community
Notes

Notes
12
4
5
Fire Warning
7KLVDFWLYHVHDWLQJV\VWHPFRPSOLHVZLWKDOO8.¿UHVDIHW\VWDQGDUGV
,QRUGHUWRUHGXFHWKHULVNRI¿UH
1. DO NOT SMOKE - Smoking will contaminate the product
and is NOT permitted around or on the seating system. This
LVDFRPPRQFDXVHRIIDWDO¿UHV7KLVV\VWHPXVHVURRPDLU
for circulation through the cushion. A cigarette could burn a
hole in the seating surface and cause damage to the seating
system. Patient clothing, bed sheets and other items, may
EHFRPEXVWLEOHDQGFRXOGFDWFK¿UH)DLOXUHWRREVHUYH
WKLVZDUQLQJFRXOGUHVXOWLQDVHYHUH¿UHSURSHUW\GDPDJH
physical injury or death.
2. DO NOT - use candles on or around the seating system
3. DO NOT - use electric blankets with this seating system
4. DO - keep hot equipment off the seating surface. E.g. Hair
dryer, curling tong, etc.
5. DO - keep matches and lighters away from children
6. DO - keep heaters away from the cushion system
Follow all manufacturer’s instructions and warning. In case of
¿UHH[LWDQGFDOO

4
5
WARNING
ELECTRICAL EQUIPMENT CAN BE HAZARDOUS IF MISUSED.
ONLY AUTHORISED TECHNICAL PERSONNEL SHOULD
REMOVE THE PUMP FOR MAINTENANCE.
WARNING
DO NOT USE THIS PUMP IN THE PRESENCE OF FLAMMABLE
GASES.
WARNING
BEFORE CLEANING THE UNIT ENSURE THAT THE
ELECTRICAL SUPPLY TO THE PUMP HAS BEEN
DISCONNECTED. PLEASE REMOVE THE PLUG FROM
THE POWER SUPPLY.
WARNING
IF THE PUMP IS SWITCHED OFF
THE SEAT WILL DEFLATE WHILST CONNECTED TO THE PUMP.
CAUTION
DO NOT USE THIS SYSTEM AS A MEANS FOR LIFTING THE
PATIENT.
TRANSPORT & STORAGE CONDITIONS
DO NOT GO OUTSIDE THE FOLLOWING RANGES IN A
PERIOD NOT EXCEEDING 15 WEEKS.
AN AMBIENT TEMPERATURE RANGE OF - 40 ° C TO + 70° C.
A RELATIVE HUMIDITY RANGE OF 10% TO 100% INCLUDING
CONDENSATION.
12
8. Guarantees And Warranties
Pump Unit - All new pumps have a guarantee for a period of 2 years following
the date of dispatch.
Cushion - All new cushions have a guarantee for a period of 2 years following
the date of dispatch.
Guarantee - Drive DeVilbiss Healthcare guarantees to repair or place
any equipment issued to its customers, which is found to be faulty
during the relevant guarantee or warranty period.
Drive DeVilbiss Healthcare guarantees are subject to the following
conditions:
a) That the usage has been on a ‘fair wear and tear’ basis.
b) That the cleaning / disinfection guidelines have been followed.
c) That Drive DeVilbiss Healthcare maintenance guidelines have
been followed.
d) That maintenance has been carried out by a Drive DeVilbiss
Healthcare approved engineer.
Warranty - Extended warranties can be purchased from Drive
DeVilbiss Healthcare, for more information contact Customer Services on
0845 0600 333.
Claims relating to Guarantee and Warranty
In the event of a fault being discovered within the warranty period,
the customer must notify Drive DeVilbiss Healthcare at the earliest
opportunity.
If upon inspection, Drive DeVilbiss Healthcare accepts liability then the
equipment shall be repaired or replaced immediately.
If Drive DeVilbiss Healthcare does not accept liability it shall inform the
customer of its reasons for declination and provide the customer with an
estimate on either the repair or replacement cost.
Drive DeVilbiss Healthcare reserves the right to alter or amend this
document without prior notice.

10
11
7. Technical Description
Product Code 1006100001
Pump 6SHFL¿FDWLRQ
Air outlet 2 air outlet
Power Supply AC230V / 50/60Hz, 0.17A
Fuse Rating T1AL, 250V
Cycle Time 8 min / 60Hz
9.6 min / 50Hz
Dimensions (L x W x H) 30 x 9.5 x 19cm or 11.8” x 3.7” x 7.5”
Weight 2.8 kg or 6.2 lb
Environment
Temperature
Operation:
Storage:
Shipping:
10°C to 40°C (50°F to 104°F)
-15°C to 50°C (5°F to 122°F)
-15°C to 70°C (5°F to 158°F)
Humidity Operation:
Storage:
10% to 90% non-condensing
10% to 90% non-condensing
&ODVVL¿FDWLRQ • Class II, Type BF, IP2X
6
7
2. Product Description
The Eclipse Seating System is an alternating cushion that can be used in the
prevention and treatment of pressure ulcers. By using the established principles of
alternating therapy the Eclipse Seating System offers patients a comfortable and
relaxing support which can both prevent breakdown and enhance healing.
The Eclipse power unit is both lightweight and compact. It features: a low pressure
alarm, a visual low pressure indicator and a manual pressure control function.
Cells are individually replaceable in the event of damage. The seat has a vapour
permeable P.U. coated stretch nylon cover.
3. Installation
1. Unpack the system and place the pump at the foot of the chair.
2. Place the cushion on the top of the chair cushion with the cell and /
or vapour permeable cover uppermost.
3. Connect the tubes leading from the cushion to the outlet ports on the
pump.
127(7KH(FOLSVH3XPSPXVWRQO\EHXVHGZLWKHLWKHUWKH(FOLSVH0DWWUHVV
Eclipse Seat or Eclipse Seat & Lumbar
4. Refer to the comfort control setting and adjust the dial according to
the patient’s weight. (See support setting procedure)
5. Connect the pump to the mains power supply and switch on the
pump. At this stage both the mains power and low pressure
indicators will illuminate.
6. When the low pressure indicator has extinguished, patient may be
seated on the cushion.
7. When normal pressure has been reached press the alarm reset button
located on the side of the pump to activate it. The alarm will sound should
the pump go into low pressure. The alarm can be silenced by pressing the
alarm reset button again.
NOTE: the alarm will not sound if the alarm reset button is in the off position

6
7
Operating Functions
a) Transport Mode
If required, disconnect the hoses from the pump and join together.
The system will equalise and retain seat pressure.
b) Infection Control
Should the seat become infected then please refer to our
disinfection protocol for appropriate cleaning processes.
Support Setting Procedure
It is important to follow the correct support setting procedure to ensure the patient
receives adequate support (lift) while achieving maximum pressure relief and
comfort. Failure to do this could result in the patient being put at risk.
1. Reset pressure dial on pump according to patient weight guide. Ensure the
correct pressure setting has been selected for sitting position.
2. Sit the patient on the cushion.
MAXIMUM PATIENT WEIGHT 159 kgs / 25 stones
Weight in KG Weight in lb Weight in Stones
& lb
Pressure
Range
31 - 85kg
85 - 113kg
113 - 141kg
141 - 159kg
70 - 187lb
187 - 249lb
249 - 310lb
310 - 350lb
5 stone - 13 stone
33% turn away
from soft
66% turn away
from soft
Firm
Soft
13 stone - 17 stone
11lb
17 stone 11lb - 22
stone 2lb
22 stone 2lb - 25
stone
WEIGHT CONVERSION TABLE
10
11
UNTWIST ANY
KINKS FOUND
ENSURE BOTH FEED PIPES
ARE CONNECTED TO THE
PUMP
FAULT
DESCRIPTION THINGS TO CHECK SOLUTION TO
PROBLEM
IF STILL NOT
WORKING
THE PUMP IS
SHOWING NO
INDICATORS TO SAY
THAT IT IS ON
MAINS IS PLUGGED IN PLUG IN
RETURN SYSTEM
DRIVE DEVILBISS
HEALTHCARE
THE LOW PRESSURE
LIGHT TURNS ON AND
THE ALARM IS
SOUNDING
THE PUMP IS
ON BUT NOT
INFLATING THE
CUSHION
RETURN
SYSTEM TO
DRIVE DEVILBISS
HEALTHCARE
RETURN
SYSTEM TO
DRIVE DEVILBISS
HEALTHCARE
MAINS SOCKET IS
SWITCHED ON TURN ON
FRONT SWITCH IS ON TURN ON
FUSES ON THE SIDE
PUMP ARE NOT BLOWN
REPLACE WITH A 1 AMP
FUSE
FUSE IN THE PLUG
BLOWN
REPLACE WITH A 3 AMP
FUSE
THE HOSES ARE ALL
CONNECTED RELOCATE HOSES
THERE ARE NO LEAKS IN
THE SYSTEM
RETURN CUSHION TO
DRIVE DEVILBISS
HEALTHCARE
DISCONNECT HOSES
FROM PUMP TO SEE IF
AIR IS COMING OUT
RETURN PUMP TO
DRIVE DEVILBISS
HEALTHCARE
THE HOSES ARE ALL
CONNECTED
RECONNECT HOSES
THERE ARE NO OBVIOUS
AIR LEAKS IN THE SYSTEM
RETURN YOUR SYSTEM
TO DRIVE DEVILBISS
HEALTHCARE
CHECK THE PRESSURE
IS IN THE SUGGESTED
RANGE FOR THE WEIGHT
OF THE PATIENT
TURN THE PRESSURE UP
TO WITHIN THE CORRECT
WEIGHT RANGE
THERE ARE NO KINKS
IN THE TUBING RUNNING
DOWN THE SIDE OF
THE CUSHION
UNTWIST ANY KINKS
FOUND
THE SYSTEM
DOES NOT
APPEAR TO BE
ALTERNATING
THERE ARE NO
KINKS IN THE
TUBING RUNNING
DOWN THE THE SIDE OF
THE CUSHION
RETURN SYSTEM TO
DRIVE DEVILBISS
HEALTHCARE
THE PUMP IS
OPERATING
NOISILY
THE PUMP IS NOT
RESTING AGAINST
A SOLID SURFACE
RETURN PUMP TO
DRIVE DEVILBISS
HEALTHCARE
6. Troubleshooting
If your problem cannot be resolved using the above information. Please
contact Drive DeVilbiss Healthcare.

8
9
Frequent or prolonged exposure to higher concentration disinfectant
solutions may prematurely age the fabric cover. Surfaces must be protected
during use and rinsed and thoroughly dried after application of a disinfectant.
Laundering
Before laundering covers should be completely removed. Where required
covers can be laundered as follows:
Pre wash 80ºC + 15 minutes
Main wash 80ºC + 15 minutes
This should be followed by a cold rinse and extraction.
Drying
Covers should be hung from a line or bar and drip dried in a clean indoor
HQYLURQPHQW&RYHUVPXVWEHFRPSOHWHO\GULHGEHIRUHUH¿WWLQJWRWKHFXVKLRQ
Covers can be tumble dried on a low heat setting for 90 minutes. Drying
temperature must not exceed 40°C. Exceeding this temperature can cause
VLJQL¿FDQWGDPDJHWRWKHFRYHU
During cleaning procedures suitable protective clothing should be worn.
Ensure that mains power supply to the pump has been disconnected prior
to cleaning.
5. Annual Maintenance
,QVSHFWDOO¿OWHUV,I¿OWHULVGLUW\LWFDQEHFOHDQHGZLWKGHWHUJHQWDQGZDWHU
0XVWEHFRPSOHWHO\GU\EHIRUHEHLQJSODFHGEDFNLQWKHFRPSUHVVRU,I¿OWHU
is still dirty then replace with a new one.
NOTE: FURTHER INFORMATION IS AVAILABLE FROM Drive DeVilbiss
Healthcare
8
9
4. Cleaning
The following guidelines are suggested by Drive DeVilbiss Healthcare as
being suitable infection control procedures. Further information is available.
Every 12 Months Every 6 Months Every 3 Months
Low particle density
Medium particle density
High particle density
PUMP UNIT
General cleaning may be affected by using a cloth dampened with a mild
detergent and water solution. This approach may be followed either by
wiping with a sodium hypochlorite solution to a dilute of 1000 ppm or by
using an alcoholic wipe.
Do not use hyper carbonate or phenol based cleaning solutions.
Do not use any abrasive compounds or cleaning pads.
CUSHION
General Cleaning
Using a single use wipe, clean the cover with a solution of neutral detergent
and hand hot water. Rinse thoroughly with clean water and a damp single
use wipe. NOTE: The top, bottom and all four sides of the cushion, including
XQGHUWKH]LSÀDSV0867EHFOHDQHG
Disinfecting the cover
,IWKHFRYHULVKHDYLO\VRLOHGRUKDVEHHQH[SRVHGWRERGLO\ÀXLGVVXFKDV
blood, it will require a more thorough cleaning procedure:
Wipe the cover using a single use wipe and a 0.1% Chlorine Solution
(1,000ppm) and cold water. If required a 1% Chlorine Solution (10,000ppm)
and cold water can be used. Rinse thoroughly with clean water and a damp
VLQJOHXVHZLSH0DNHVXUHWKHFRYHULVFRPSOHWHO\GULHGEHIRUHUH¿WWLQJWR
the cushion.
Table of contents
Other Drive DeVilbiss Medical Equipment manuals

Drive DeVilbiss
Drive DeVilbiss Med Aire FlexWave User manual

Drive DeVilbiss
Drive DeVilbiss Casalift 18001 User manual

Drive DeVilbiss
Drive DeVilbiss 1025 Series Manual

Drive DeVilbiss
Drive DeVilbiss iGo 2 User manual

Drive DeVilbiss
Drive DeVilbiss sidhil Innov8 User manual

Drive DeVilbiss
Drive DeVilbiss 123S User manual

Drive DeVilbiss
Drive DeVilbiss BONSAI VELOCITY User manual

Drive DeVilbiss
Drive DeVilbiss Selectis User manual

Drive DeVilbiss
Drive DeVilbiss HbO-3000 User manual