
 
1592023040 XM668D GB r1.1 2011.07.05.doc XM668D 5/14 
 
 The Adr parameter cannot be duplicated because in this case the defrost 
cannot be correctly managed. 
 
BEGIN 
Press for 3 sec, the rtC or other will be showed. 
The measurement unit blinks. 
Find Adr 
Press more than once the DOWN arrow to find the 
Adr parameter, the press SET. 
Modify 
Adr or 
Set the value of Adr parameter, then press SET to 
confirm the parameter. 
EXIT 
Press the two keys together to exit from menu or 
wait for about 10 sec. 
 
The LSn and LAn parameter are only to show the actual settings (read only). Se the 
following example of configuration: 
 
 
 
DAILY DEFROST FROM RTC: [CbP = Y] and [EdF = rtC] 
 
idF Parameter: for safety reason force the value of idF at +1 respect to the interval 
between two Ld parameters. The idF timer is reinitialized after defrost and at every 
power-on. 
DEFROST START: at the time selected by the parameters Ld1 to Ld6 or Sd1 to 
Sd6. 
DEFROST END: if the probes reach the dtE temperature or for maximum MdF time. 
SAFETY and rtC or rtF ALARM: with clock alarm the device will use the parameter 
idF, dtE and MdF. 
 
WARNING: don’t set [EdF=rtC] and [CPb=n]. 
 
MULTIMASTER DEFROST: all the probes with clock 
 
Table for example 
n N + 1 N + 2 
rtC (clock) rtC (clock) rtC (clock) 
Y Y Y 
9 hours safety 9 hours safety 9 hours safety 
45 min safety 45 min safety 45 min safety 
12°C safety 12°C safety 12°C safety 
06:00 1° 06:00 1° 06:00 1° 
14:00 2° 14:00 2° 14:00 2° 
22:00 3° 22:00 3° 22:00 3° 
10. DISPLAY MESSAGES 
 
 
1 nod 
No display: the keyboard is trying 
to work with another board that is 
not working or not present 
Press for 3 sec UP arrow, 
the SEC menu and select LOC 
entry. 
2 Pon Keyboard is unlocked 
3 PoF Keyboard is locked 
4 rSt Alarm reset Alarm output deactivated 
5 noP, nP
nA 
Not present (configuration) 
Not available (evaluation) 
 
6 
P2 
P3 
P4 
P5 
P6 
PPF 
CPF 
Sensor brake down, value out of 
range or sensor incorrectly 
configured P1C, P2C to P6C. 
 
PPF can be showed by slaves of 
pressure that don’t receive the 
value of pressure. 
 
CPF is showed when the remote 
probe 4 is not working. 
P1: the cooling output works 
with Con and COF, 
With defrost probe on error the 
defrost is performed only at 
interval. 
 
For P5, P6 and PPF: the 
percentage of the valve 
opening is fixed at PEO value. 
 
7 HA 
Temperature alarm from parameter 
ALU on probe rAL. 
8 LA 
Temperature alarm from parameter 
ALL on probe rAL. 
9 HAd 
probe defrost probe [dPa / dPb]. 
10
LAd 
probe defrost probe [dPa / dPb]. 
11
HAF 
probe defrost probe [FPa / FPb]. 
12
LAF 
probe defrost probe [FPa / FPb]. 
 
13
dA 
Door open alarm from input 
i2F or i3F = after delay d1d, d2d 
or d3d. 
Cooling relay and fan follow the 
odc parameter. Cooling restart 
as specified on rrd parameter. 
14
EA Generic alarm from digital input 
i1F,i2F,i3F=EAL. 
15
CA Severe alarm of regulation lock 
from digital input i1F,i2F,i3F=bAL. Regulation output OFF. 
16
PAL 
= PAL. All the outputs are OFF. 
 
17
LOP Minimum operating pressure 
threshold from LOP parameter. 
The valve output increases its 
opening of dML quantity every 
second. 
18
MOP Maximum operating pressure 
threshold from MOP parameter. 
The valve output decreases its 
opening of dML quantity every 
second. 
19
LSH Low superheating from LSH 
parameter and SHd delay. 
The valve will be closed; the 
alarm will be showed after SHd 
delay. 
20
HSH 
parameter and SHd delay. Only display. 
 
21
rtC Clock settings lost. 
Defrost will be performed with 
IdF till restoring the settings of 
RTC. 
22
rtF Clock damaged. Defrost will be performed with 
IdF. 
 
EEPROM serious problem. Output OFF. 
24
Err Error with upload/download 
parameters. Repeat the operation. 
25
End Parameters have been correctly 
transferred. 
10.1 ALLARM RECOVERY 
Probe alarms P1, P2, P3 and P4 start some seconds after the fault in the related 
probe; they automatically stop some seconds after the probe restarts normal 
operation. Check connections before replacing the probe. 
Temperature alarms HA, LA, HA2 and LA2 automatically stop as soon as the 
temperature returns to normal values. 
Alarms EA and CA (with i1F = bAL)recover as soon as the digital input is disabled. 
Alarm CA (with i1F = PAL) recovers only by switching off and on the instrument. 
11. USE OF THE PROGRAMMING “HOT KEY“ 
The XM units can UPLOAD or DOWNLOAD the parameter list from its own E2 
internal memory to the HOT-KEY and vice-versa through a TTL connector. Using 
HOT-KEY the Adr will not changed. 
11.1 DOWNLOAD (FROM THE HOT-KEY TO THE INSTRUMENT) 
1. Turn OFF the instrument by means of the ON/OFF key, insert the HOT-KEY 
and then turn the unit ON. 
2. Automatically the parameter list of the HOT-KEY is downloaded into the 
controller memory: the doL message is blinking. After 10 seconds the 
instrument will restart working with the new parameters. At the end of the data 
transfer phase the instrument displays the following messages: End for right 
programming. The instrument starts regularly with the new programming. Err 
for failed programming. In this case turn the unit off and then on if you want to 
restart the download again or remove the HOT-KEY to abort the operation. 
11.2 UPLOAD (FROM THE INSTRUMENT TO THE “HOT KEY”) 
1. When the XM unit is ON, insert the HOT-KEY and push è key; the uPL 
message appears. 
2. The UPLOAD begins; the uPL message is blinking. 
3. Remove the HOT-KEY.At the end of the data transfer phase the instrument 
displays the following messages: 
 End = right programming; 
 Err = failed programming. In this case push SET key if you want to restart the 
programming again or remove the not programmed HOT-KEY. 
12. CONTROLLING LOADS 
12.1 THE COOLING OUTPUT 
The regulation is performed according to the temperature measured by the 
thermostat probe that can be physical probe or virtual probe obtained by a weighted 
average between two probes following the formula: 
 
value_for_room_regulation = (rPA*rPE + rPb*(100-rPE))/100 
 
If the temperature increases and reaches set point plus differential the solenoid valve 
is opened and then it is closed when the temperature reaches the set point value 
again.In case of fault in the thermostat probe the opening and closing time of 
solenoid valve is configured by Con and CoF parameters. 
12.2 STANDARD REGULATION AND CONTINUOUS REGULATION 
The regulation can be performed in three ways: the goal of the first way (standard 
regulation) is reaching the best superheat via a classic temperature regulation 
obtained using hysteresis. The second way permits to use the valve to realize an high 
performance temperature regulation with a good factor of superheat precision. This