manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Ferno
  6. •
  7. Medical Equipment
  8. •
  9. Ferno PEZO2 User manual

Ferno PEZO2 User manual

1. Device to be used on or by the order of a physician.
2. Before use, carefully read the manual.
3. Follow all local protocols and care provider standards.
4. Do not use the pulse oximeter in an MRI or CT
environment.
5. Do not use the pulse oximeter in situations where
alarms are required. The device has no alarms.
6. Explosion hazard: Do not use the pulse oximeter in
an explosive atmosphere.
7. The pulse oximeter is intended only as an adjunct in
patient assessment. It must be used in conjunction
with other methods of assessing clinical signs and
symptoms. Unit is designed for temporary checking;
not permanent attachment.
8. Check the pulse oximeter sensor application site
frequently to determine the positioning of the sensor
and circulation and skin sensitivity of the patient.
9. Do not use adhesive tape to apply or secure the unit.
10. The pulse oximeter has no SpO2 alarms; it is not for
continuous monitoring, as indicated by the symbol.
11. Prolonged use or the patient’s condition may require
changing the sensor site periodically.
Change sensor site and check skin integrity,
circulatory status, and correct alignment at least
every 4 hours.
12. Inaccurate measurements may be caused by
autoclaving, ethylene oxide sterilizing, or immersing
the sensors in liquid may cause inaccurate readings.
13. Significant levels of dysfunctional hemoglobin’s
(such as carbonxy-hemoglobin or methemoglobin)
14. Intravascular dyes such as indocyanine green or
methylene blue.
15. SpO2 measurements may be adversely affected in
the presence of high ambient light or direct sunlight.
Shield the sensor area from the light source
(example: shield with a towel).
16. Excessive patient movement.
17. High-frequency electrosurgical and electromagnetic
interference and defibrillators.
18. Venous pulsations
19. Placement of a sensor on an extremity with a blood
pressure cuff, arterial catheter, or intravascular line.
20. The patient has hypotension, severe
vasoconstriction, severe anemia, or hypothermia.
21. The patient is in cardiac arrest or is in shock
22. Fingernail polish or false fingernails may cause
inaccurate SpO2 readings.
Follow local ordinances and recycling instructions
regarding disposal or recycling of the device and
device components, including batteries.
1 Operation of the product is simple and convenient
2 The product is small in volume, light in weight (total
weight is about 50g including batteries) and can be
conveniently carried.
3 Power consumption of the product is low and the two
originally-equipped AAA batteries can be operated
continuously for approximately 30 hours.
4 Low voltage warning will be indicated in the visual
window when battery voltage is so low that normal
operation of the oximeter might be influenced.
5 The product will automatically be powered off when
no signal is detected for longer than 8 seconds.
The finger type Oximeter can be used to measure
human Hemoglobin Saturation and heart rate through a
finger. The product is suitable for use in family, hospital
(including clinical use in internist/ surgery, anesthesia,
pediatrics, intensive care, etc.) With medical direction,
the unit can be used in sports medicine. The unit can
be used before or after sports. Operation in sport
procedure is not recommended, etc. The product is not
suitable to monitor patient continuously during play or
activity.
1 Remove batter cover. Install two AAA batteries into
the battery cassette. Replace cover.
2 Patients finger to be monitored should be clean. Use
medical alcohol to clean the test finger. Medical
alcohol can be used to wipe down the rubber pads
inside the clamp zone. Clean the unit before and
after each use.
3. Nip (squeeze) the clamp as shown in the diagram.
4. Plug one patient finger into the rubber hole of the
Oximeter. It is best to fully engage the finger into the
unit before releasing the clamp.
5 Press the switch button once on front panel.
6 Hold the patient’s finger steady while the unit is taking
its reading. Body movement is not recommended
during operation.
7 Read correspondent data from display screen.
8 There are six (6) display modes.
1. General Description
2. Measurement principle
Diagram of Operation Principle
3. Precautions for use
1. Red and Infrared-ray Emission
2. Red and Infrared-ray Receipt
4. Product Properties
5. Product Operation Scope
6. Operation Instructions
1.
2.
3.
4.
5.
6.
When the patients finger is inserted into the
Oximeter, the fingernail surface must be
facing upward.
Hemoglobin Saturation is a percentage of the amount
of oxygen combined with hemoglobin in proportion to
the amount of oxygen the hemoglobin is capable of
carrying.
It is expressed as a percentage of a ratio:
content/capacity.
In other words, it is the existence of oxygen in the
blood, which is known as Oxyhemoglobin. It is a very
important physiological parameter for Respiratory
circulation system. Many respiratory diseases can
result in hemoglobin saturation being lowered in human
blood. Moreover, the following factors can also lead to
problems in oxygen supply, potentially causing human
hemoglobin saturation to be reduced:
Automatic Organic Regulation Malfunction caused
by Anesthesia, Intensive Postoperative Trauma,
injuries resulting from medical examinations, etc. or in
the situation of illnesses, such as light headedness,
asthenia, vomiting, etc., might happen to the patients
and possibly endanger the patient’s life.
Therefore, it is very important to know the patient’s
hemoglobin saturation. The Finger type Pulse Oxime-
ter features, a large display, low power consumption,
convenient operation all in a compact portable unit. The
patient simply puts one finger into the photoelectric
sensor for diagnosis, and the display screen will
directly show the measured value of hemoglobin
Saturation.
Principle of the Oximeter is as follows: An experience
formula of data process is established taking use of
Lambert Beer Law according to Spectrum Absorption
Characteristics of Reductive hemoglobin (R Hb) and
Oxyhemoglobin (O2 Hb) in glow and near-infrared
zones. The operation principle of the instrument is
Photoelectric Oxyhemoglobin inspection technology is
adopted in accordance with Capacity Pulse Scanning
and Recording Technology, so that two beams of
different wavelength of lights (660nm glow and 940nm
near infrared light) can be focused onto a human nail
tip through perspective clamp finger-type sensor.
The measured signal is obtained by a photosensitive
element. The information acquired will be shown on
two groups of LEDs through a process in the
electronic circuits and microprocessor.
After turn on the oximeter, each time you
press the power switch, the oximeter will
switch to another display mode, there are 6
display modes shown as follows:
When you press the power switch for a long
time (more than on second), the brightness
of the oximeter will be changed by degrees,
there are 10 levels on brightness; the default
level is level four.
Declaration: Please use the medical alcohol to
clean the rubber touching the finger inside of
Oximeter, and clean the test finger using alcohol
before an after each test. (The rubber inside os
the Oximeter belongs medical rubber, which has
no toxin, and no harmful to the skin of human
being).
7. Brief Description of Front Panel
The heart rate bar graph display corresponds with
pulse rate.
Instruction Manual
Finger type
Pulse Oximeter
No.
T10565
Low power indicator
SpO
2%
PRbpm
98 77
SPO2 waveform
SpO
2%
PR
bpm
98
77
PR Bar graph
SPO2
PR
Power
Switch
FERNO2
FERNO2
FERNO2
FERNO2
FERNO2
FERNO2
FERNO2
FERNO2
8. Product Accessories
1 One hang lace
2 Two AAA batteries
3 One Instruction manual
9. Battery Installation
Remove batter cover. Install two (2) AAA batteries
into the battery cassette with the proper polarity
alignment.
Replace cover horizontally along the arrow shown
as below:
Notes: Battery polarities must be correctly
installed.
Otherwise, damage to the device may occur.
Install or remove batteries with care, and in the
right position, so as to not damage the device.
Batteries should be removed from the Oximeter
if the unit will not be used for an extended period
of time (battery corrosion may occur).
10. Hang Lace Installations
1. Thread thinner end of the hang lace
through the eyelet on end of the unit.
2. Thread thicker end of the lace through the
loop of the thread that was just inserted in eyelet.
Pull thicker thread firm. Do not over tighten.
11. Maintenance and Storage
12. Calibrating the pulse oximeter
14. Detailed descriptions of product functions
13. Declaration
15. Possible Problems and Resolutions
Problems Possible
reason Solution
The Oximeter
can not be
powered on
Display screen
is suddenly off 1. The product
is automatically
powered off
when no signal
is detected
longer than 8
seconds
2. Power level
of batteries is
inadequate.
1. Normal
2. Replace the
batteries
“Error3” or
“Error4”
displayed on
screen
1 .Low power
2. Receiving
tube being
shielded or
damaged
together with
broken
connector.
3. Mechanical
misplace for
receive-emissi
on tube
4. Amp circuit
malfunction.
1. Change
battery
2. Contact your
local service
center
3. Contact your
local service
center
4. Contact your
local service
center
17. Symbol Definitions
16. Guidance and manufacture’s declaration
- electronmagnetic emissions -
for all Equipmnet and Systems
1. The functional tester cannot be used to
assess the accuracy of the oximeter.
2. Index 2 that is made by Bioteck company is a
function tester. Set Tech to 1. R curve to 2, then
user can use this particular calibration curve to
measure the oximeter.
3. The test methods used to establish the SpO2
accracy is clinical testing. The oximeter used to
measure the arterial hemoglobin oxygen
saturatin levels and these levels are to be
compared to the levels determined from arterial
blood sampling with a CO-oximeter.
EMC of this product comply with IEC60601-1-1-2
standards.
The materials which the user comes in contact with
are not toxic and no adverse effects on tissues. The
unit complies with ISO10993-1 -5, -10.
1. Display Type: OLED display
2. SPO2 :
Measurement range: 0-100%
Accuracy: ±2 on the range of 70% - 100%
< 70% no definition
3. Pulse Rate:
Measure range: 30 to 235 BPM (Beats Per
Minute)
Accuracy: ±2 BPM
4. Power Requirements: Two AAA 1.5V Alkaline
Batteries
Power consumption: less than 40mA
Low power indication:
Battery Life: Two AAA 1.5V, 600mAh alkaline
batteries can be continuously operated for
approximately 30 hours.
5. Dimension:
Length: 58mm (2.28 inches)
Width: 32mm (1.25 inches)
Height: 34mm (1.33 inches)
Weight: 50g / 1.7 oz. (including two AAA batteries)
6. Environment Requirements:
Operation Temperature: 5 to 40°C (41 to 104° F)
Storage Temperature: -10 to 40°C (14 to 104° F)
Ambient Humidity: 15% to 80% in operation
10% to 80% in storage
7. Declaration: EMC of this product complies with
IEC60601-1-1-2 standard.
8. Measurement Performance in Weak
Filling Condition: required the test equipment
(BIO-TEK INDEX Pulse Oximeter tester) the
pulse wave is available without failure when
the simulation pulse wave amplitude is at 6%.
9. Interference Resistance Capacity aginst
Ambient Light: Device work normally when mixed
noise produced by BIO-TEK INDEX Pulse
Oximeter tester.
SpO2 or PR
can not be
displayed
normally
SpO2 or PR is
showed
unstably
“Error7”
displayed on
screen
Guidance and manufacture’s declaration -
electromagnetic emission
The Pulse Oximeter is intended for use in the
electromagnetic environment specified below. The
customer of the user of the Pulse Oximeter should
assure that it is used in such an environment.
Emission test Compliance Electromagnetic
environment-guldance
RF emissions
CISPR 11
Group 1 The Pulse Oximeter
uses RF energy only
for its internal function.
Therefore, it’s RF
emissions are very low
and are not likely to
cause any interference
in nearby electronic
equipment.
RF emission
CISPR 11
Class B The Pulse Oximeter is
suitable for use in all
establishments,
including domestic
establishments and
those directly
connected to the public
low-voltage power
supply network that
supplies buildings use
for domestic purposes.
Mfg. For:
Spiracle Technology
16520 Harbor Blvd. #D
Fountain Valley, CA 92708 U.S.A.
Ph:(714) 418-1091
Fax:(714) 418-1095
www.spiracle.com
email:[email protected]
EUROPEAN CONTACT:
Ferno Washington Italia s.r.l.
via Benedetto Zallone n.26
40066, Pieve di Cento (Bo) - Italy
P +39 051 6860028 / F +39 051 6861508
Revision: July 2013
Symbol Definition
The equipment type
is BF
Refer to user manual
before application
Hemoglobin saturation
Heart rate (BPM)
Low power indication
SN Serial No. (Attached
inside of the unit )
Notes: Please remove the battery when product
will not be in used for period of time.
No SpO2alarm
SpO2%
PRbpm
1. Replace the batteries when the low voltage lamp
is displayed on the screen.
2. Clean surface of the finger type Oximeter before it
is used in diagnosis for patients.
3. Remove the batteries inside the battery cassette
if the Oximeter will not be operated for an
extended period of time.
4. It is best to store the product in a place where
ambient temperatures are –10 to 40°C (14 to
104°F) and humidity is 10% to 80%.
5. It is recommended that the product be kept in a
dry environment for storage. A wet and excessive
humid environment might affect the units
performance, damage the product or effect its
useful life.
6. Please follow the local laws and protocols for
proper disposable of used batteries.
1.Retry by
inserting the
finger until fully
engaged
2. Repeat test
method. If
problem
continues, follow
local protocols for
patient
assessment
and treatment.
1. Finger is not
plugged in
correctly
2. Patient’s
Oxyhemoglobin
value is too
low to be
measured
1. Finger might
not be plugged
deep enough
2. Finger is
trembling or
patient’s body
is in movement
status
1.Retry by
inserting the
finger until fully
engaged
2. Try not to
move
1. Change
batteries
2. Contact your
local service
center
3 Contact your
local service
center
1. Low power
2. Emission
tube damaged.
3. Current
control circuit
malfunction.
NOTIFIED BODY: DET NORSKE VERITAS 0434
1. Replace
Batteries.
2. Remove and
reinstall the
batteries
3. Contact your
local service
center
1. Low or dead
batteries.
2. Batteries
might be
installed
incorrectly
3. The Oximeter
might be
damaged

This manual suits for next models

1

Other Ferno Medical Equipment manuals

Ferno 28-Z PROFlexx User manual

Ferno

Ferno 28-Z PROFlexx User manual

Ferno EZ Glide 59-T User manual

Ferno

Ferno EZ Glide 59-T User manual

Ferno One Man 27-1 User manual

Ferno

Ferno One Man 27-1 User manual

Ferno 11-T Stretcher User manual

Ferno

Ferno 11-T Stretcher User manual

Ferno Stretcher Series User manual

Ferno

Ferno Stretcher Series User manual

Ferno Compact 2 Track Chair User manual

Ferno

Ferno Compact 2 Track Chair User manual

Ferno 175 Series Reference manual

Ferno

Ferno 175 Series Reference manual

Ferno 24 MAXX User manual

Ferno

Ferno 24 MAXX User manual

Ferno Mobile Transporter 35A Series User manual

Ferno

Ferno Mobile Transporter 35A Series User manual

Ferno Harrier LT-LBS User manual

Ferno

Ferno Harrier LT-LBS User manual

Ferno 101-H User manual

Ferno

Ferno 101-H User manual

Ferno 35A-NM User manual

Ferno

Ferno 35A-NM User manual

Ferno 50-E User manual

Ferno

Ferno 50-E User manual

Ferno PRO F1 User manual

Ferno

Ferno PRO F1 User manual

Ferno F2 User manual

Ferno

Ferno F2 User manual

Ferno RS5 User manual

Ferno

Ferno RS5 User manual

Ferno 23 User manual

Ferno

Ferno 23 User manual

Ferno Pedi Mate 678 User manual

Ferno

Ferno Pedi Mate 678 User manual

Ferno Pedi-Mate 678 User manual

Ferno

Ferno Pedi-Mate 678 User manual

Ferno Integrated Patient Transport & Loading System... User manual

Ferno

Ferno Integrated Patient Transport & Loading System... User manual

Ferno Tirol Kit Operating instructions

Ferno

Ferno Tirol Kit Operating instructions

Ferno Pegasus User manual

Ferno

Ferno Pegasus User manual

Ferno ScoopEXL Operating instructions

Ferno

Ferno ScoopEXL Operating instructions

Ferno Mobile Transporter 35A Series User manual

Ferno

Ferno Mobile Transporter 35A Series User manual

Popular Medical Equipment manuals by other brands

Nasco LF03500U instruction manual

Nasco

Nasco LF03500U instruction manual

Barco AMM240ED Operation manual

Barco

Barco AMM240ED Operation manual

BIONIME Rightest GD500 Getting started guide

BIONIME

BIONIME Rightest GD500 Getting started guide

Hollister 9650 Guide for Use

Hollister

Hollister 9650 Guide for Use

3M Littmann Classic II S.E. manual

3M

3M Littmann Classic II S.E. manual

EDAN iM8 user manual

EDAN

EDAN iM8 user manual

Moretti EASYRED EM210 instruction manual

Moretti

Moretti EASYRED EM210 instruction manual

smirthwaite Strato user manual

smirthwaite

smirthwaite Strato user manual

Elektron Henson 9000 user manual

Elektron

Elektron Henson 9000 user manual

Medivance ARCTIC SUN 5000 Service manual

Medivance

Medivance ARCTIC SUN 5000 Service manual

StimTec NEO manual

StimTec

StimTec NEO manual

Linktop HC-21 user manual

Linktop

Linktop HC-21 user manual

Datex-Ohmeda 3770 Service manual

Datex-Ohmeda

Datex-Ohmeda 3770 Service manual

ResMed Mirage SoftGel manual

ResMed

ResMed Mirage SoftGel manual

Sedecal X Service manual

Sedecal

Sedecal X Service manual

Medin Miniflow user manual

Medin

Medin Miniflow user manual

HumaPen MS9673 Instructions for use

HumaPen

HumaPen MS9673 Instructions for use

Top shelf Orthopedics Knight OA Instructions for use

Top shelf Orthopedics

Top shelf Orthopedics Knight OA Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.