11
12
CLEANING, SANITIZATION AND DISINFECTION
Turn o the unit before any cleaning and disconnect the power cord from the mains.
Werecommendapersonaluseofboththenebulizerandaccessoriestoavoidanyriskof
infection or contagion.
CLEANING THE DEVICE AND THE TUBE’S OUTER SURFACE
Onlyuseaclothdampenedwithantibacterialdetergent(non-abrasiveandfreeofsolvents
ofanykind).
SANITIZATION AND DISINFECTION OF THE ACCESSORIES
Openthenebulizerbyrotatingtheupperpart(C5.1)counter-clockwise,detachthenozzle
bypressingasindicatedbythe2arrowsinthe“Assemblydiagram”sectionC5.
SANITIZATION
Beforeandaftereachuse,sanitizethenebulizerandaccessoriesaccordingtooneofthe
methods descrive below.
(Method A): The accessories that can be sanitized with the method A are (C1-C2-
C3-C4-C5.1-C5.2-C5.3-C5.4-C6). Sanitizethe accessoriesin warm (approximately40°C)
drinkingwaterwithmilddishwashingliquid(nonabrasive)orinthedishwasherwithhot
cycle.
(Method B):The accessories that can be sanitized with the method B are (C1-C2-C3-C4-
C5.1-C5.2-C5.3-C5.4-C6).Sanitizetheaccessoriesbyimmersioninasolutionconsistingof
50%waterand50%whitevinegar.Finallyrinsethoroughlywithwarm(approximately40°
C)drinkingwater.
Aftersanitizationshaketheaccessoriesvigorouslyandplacethemonapapertowel,or
alternatively,drythemwithajetofhotair(e.g.usingahairdryer).
DISINFECTION
The accessories that can be disinfected are (C1-C2-C3-C4-C5.1-C5.2-C5.3-C5.4-C6)
Thedisinfectionproceduredescribedinthisparagraphistobeperformedbeforeusing
the accessories and is eective only if all the prescriptions herein are met and all the
components involved are previously sanitized.The disinfectant to be used must be of
electrolyticoxidizingchlorinetype,(activeingredient:sodiumhypochlorite),specicfor
disinfection and is available in all pharmacies.
Procedure:
-Fillacontainer,suitabletocontainalltheindividualcomponentstobedisinfected,with
asolutionofdrinkingwateranddisinfectant,whilerespectingtheproportionsindicated
onthepackagingofthedisinfectantitself.
-Completelyimmerse eachindividual componentin thesolution, takingcare toavoid
the formation of air bubbles in contact with the components. Leave the components
immersed for the period of time indicated on the packaging of the disinfectant, and
associated with the concentration chosen to prepare the solution.
-Removethecomponentsnowdisinfectedandrinsethoroughlywithlukewarmdrinking
water.
-Afterdisinfectionshaketheaccessoriesvigorouslyandplacethemonapapertowel,or
alternatively,drythemwithajetofhotair(e.g.usingahairdryer).
-Disposeofthesolutionasrecommendedbythedisinfectantmanufacturer.
SoftTouch Mask SoftTouchmaskshaveanouteredgemadeofsoft
biocompatible material that ensures excellent
adherence to the face, and is also equipped with
an innovative Dispersion Limiting Device.These
distinctive elements that distinguish it allow great-
er sedimentation of medication in the patient and
also limit dispersion.
Duringtheinspiratoryphase,the
tab,actingas aDispersion Lim-
iting Device, bends inwards to-
wardsthemask.
Duringtheexpiratoryphase,the
tab,actingasaDispersion Lim-
iting Device, bends outwards
fromthemask.
Soft
biocompatible
material
Dispersion
Limiting Device
Tospeeduptheinhalationtherapy,movethespeedselectorbutton(C5.4)
bypressingonMAXwithyournger.
To increasetheeectiveness oftheinhalation therapy, presswith your
ngeronthewordingMINofthespeedselector(C5.4).Inthisposition,the
speed selector acts as a valve and allows nebulizing the optimal amount
ofdrugtothelowerrespiratorytract,reducingitsdispersionintheenvi-
ronment.
USE OF NEBULIZATION MANUAL CONTROL
Toachievecontinuousnebulisationaction youshouldnotusethe manualnebulisation
control(C6),especiallyinthecaseofchildrenorpersonswithreducedphysical,sensory,or
mentalcapabilities.Themanualnebulisationcontrolisusefulforlimitingdispersionofthe
medication in the surrounding environment.
To start nebulizing close with
a nger the hole of the nebu-
lizer manual control (C6) and
breathe in gently; we recom-
mend to hold your breath for
a moment so that the inhaled
aerosol droplets can be depos-
ited,
meanwhile, to disable
nebulizing, remove your
nger from the hole of
the nebulizer manual
control to avoid the
waste of drug, optimiz-
ing its acquisition. Then
exhaleslowly.
tubefromthenebulizeranddryitusingthedevice’scompressorfan;thisoperationwill
prevent mold from forming inside the tube.
How to use the“RF7 Dual Speed” nebulizer with the speed selector
Professional and fast, this device is suitable for administering all types of medications,
includingmoreexpensiveones,eveninpatientswithchronicdiseases.Thegeometryof
theinternallinesoftheRF7DualSpeednebulizerensurestheidealgranulometryforactive
treatmentallthewaydowntothelowerrespiratorytract.