5. Re-secure the clamp lock after clamping around a single
conductor.
6. Do not attempt to measure current higher > 3000 A.
7. Switch the meter ON and then switch power ON to the device
under test.
8. Read the current value on the display. OL will display if the
measured signal exceeds the range. The meter defaults to the
Auto range mode; the Auto range icon is visible on the dis-
play in Auto range mode.
9. To manually select the range, use the RANGE button (30.00 A /
300.0 A / 3000 A).
BLUETOOTH COMMUNICATION
To stream to an iOS / Android device, download the mobile applica-
tion METERLiNK®and follow the application on-screen prompts.
See the User Manual for complete details.
BASIC SPECIFICATIONS
Measurement ranges 30.0, 300.0, 3000 A AC
Accuracy (of reading) ± (3.0% + 5 digits)
Clamp coil diameter 7.5 mm (0.3 in.); Coil connecting ring is 13 mm
(0.5 in.)
Coil bend radius 80 mm (3.1 in.)
AC bandwidth 45 to 500 Hz
AC response True RMS
Battery power 2 x AAA batteries
Safety IP54 rated
For indoor use and in accordance with the re-
quirements for double insulation to EN61010-1,
EN61010-2-032, EN61326-1; CAT IV 600V, CAT
III 1000V, Pollution Degree 2, CE
Operating conditions 0 to 50℃(32 to 122℉); < 80% RH
Storage conditions -20 to 60℃(-4° to 140℉) without batteries; <
80% RH