FLIR IGM CM275 User manual

QUICKSTART
FLIRMODELCM275
ImagingClampMeterwithIGMTMandBluetooth®

FLIRCM275–QUICKSTARTDocumentIdentifier:CM275‐QS‐en‐US_AB
CM275QUICKSTART–EN
ThelatestversionsofthedetailedUserManualandQuickStart,withtranslations,arelocatedonthe
support.flir.comwebsite(downloadtab).PleaseRegisterforExtendedWarrantyandProductUpdatesat
www.flir.com/testwarranty
QUICKVIEW(FRONT)
1. Jawopeningtrigger
2. NavigationPad/OKbutton
3. MODEbutton
4. RANGEbutton
5. DataHold(shortpress)/WorkLight(longpress)
6. Rotaryfunctionswitch
7. COM(negative‐)ProbeInputjack
8. Positive(+)ProbeInputjack
9. Image/DisplaySavebutton(shortpress)
10. Return/Exitbutton
11. IGMTMbutton(shortpress)
12. ColorTFTDisplay
13. Clampjaw
QUICKVIEW(BACK)
14. Worklights
15. Laserpointerlens
16. ThermalImaginglens
17. Tripodmount
18. Batterycompartment
19. Batterycompartmentlock/unlock
Note:AmicroUSBportislocatedinthebatterycompartmentfortransferringimagesanddatalogstoPC
THERMALIMAGINGQUICKSTEPS
a. Movetherotaryswitch(6)toanyposition.
b. Shortpress(11)toopenthethermalimager.
c. Pointthelens(16)towardtheareaundertest.
d. Usethecrosshairsandlaserpointertotargetasurfacespot(seeThermalSettingsMenusection).
e. Readthespottemperatureontheupperleftcornerofthedisplay.
f. PresstheImageSavebutton(9)tocaptureathermalimagewithfullradiometricdata.
g. View,delete,andtransfer(Bluetooth®)savedimages(seeGalleryMenusection).
METERPOWER
TheCM275usesthree‘AA’batteries(rearcompartment)ortheoptionalTA04RechargeableLi‐Polymer
Batterysystemforpower.Themeterenterssleepmodeafter10minutesofinactivity.SetAPOtimertoOFF,
1,2,5or10minutesintheGeneralSettingsmenu(SeeGeneralSettingsMenusection).

FLIRCM275–QUICKSTARTDocumentIdentifier:CM275‐QS‐en‐US_AB
THERMALIMAGEDESCRIPTION
1. IRTemperaturemeasurementoftargetedspot
2. StatusIconBar
3. ClampMeterMeasurements
4. Crosshairsfortargeting
5. Thermalimage
6. MainMenu
7. Lowestreadingmeasuredinthecurrentframe
8. Thermalscalerangeofcolorsforthethermalimage
9. Highestreadingmeasuredinthecurrentframe.
NotestheDistance‐to‐Spotratiofortheimageris30:1meaning
thatthespotis30timessmallerthanthemeter‐to‐spotdistance.
Thethermalimager’sFieldofView(FOV)is50oby38.6o.
MENUSYSTEM
PressOKtoshowtheMainmenu:
UseNavigationPadarrowstohighlighticonsandtheOKbuttontoopenthem.Lefttorighttheiconsare
ImageMode,ThermalSettings,Image/DataLogGallery,AdvancedFunctionMenu,andGeneralSettings.
PressOKtosetanoptionONorOFF.WhenanoptionisON,abluedotwillappearnextitsicon.
UsetheReturn buttontoexitmenulevelsandtoreturntothenormaldisplaymode.
Theselectedfunctionormodedetermineswhichiconsareavailableforuse.
IMAGEMODEMENU
TheImagemodehastwooptions: Image+Clamp:DisplayshowsClampmeterdataonthethermal
images. Image‐only:Displaydoesnotshowclampmeterdata.PressOKontheImagemodeiconand
usethearrowbuttonstoselectthedesiredoption.Thismenuisonlyavailableinthethermalimagemode.
THERMALSETTINGSMENU
PressOKtoopenthemainmenu.PressOKagainontheThermalSettingsmenuicon.Thismenuisonly
availableinthethermalimagemode.Refertothetablebelow.
IconDescriptionOperation
PalettePressOKtoselectacolorpalette(Iron,Rainbow,orGray)
Emissivity
PressOKandusearrowstoscrolltoapreset(0.95,0.85,0.75,or0.65)ortothe
finetuningԐicon.Tofinetune,pressOKatthefine‐tuneiconandusethearrows
toset;pressOKtoconfirm.(Range:0.10to0.99in0.01steps)
UnitsPressOKtotogglethetemperatureunits(oC/oF)
LaserPressOKtotogglethelaserpointerONorOFF
CrosshairsPressOKtoswitchthecrosshairsONorOFF
GALLERYMENU
1. PressOKattheGalleryicontoviewthesavedimagesanddatalogsets.
2. Usethearrowstoscrollthroughdatalogs(upperarea)andimages(lowerarea).
3. PressOKtoopenalogoranimage.

FLIRCM275–QUICKSTARTDocumentIdentifier:CM275‐QS‐en‐US_AB
4. UsethearrowsandOKbuttontonavigateandactivatethe‘fillscreen’,transfer,anddeletefunctions
(iconsshownbelowopenedimageordatalog);seeUserManualforspecificinstructions.
ADVANCEDFUNCTIONMENU
PressOKattheAdvancedmenuicon.HighlightafunctionusingthearrowsandpressOKtoselectorde‐
selectit.RefertotheClampMeterBasicssectionlaterinthisguideandtheUserManualfordetails.
(lowpassfilter):FilterhighfrequencynoisefromACmeasurements
MAX‐MINReadings:Viewhighestandlowestreadings
InrushCurrent:Capturestart‐upACcurrent(in100mswindowaftertriggerpoint)
DCAZero:NulldisplaybeforetakingDCCurrentmeasurements
Datalogger:Automaticallyrecordupto40,000readingsin10sets
GENERALSETTINGSMENU
PressOKtoopenthemainmenuandpressOKattheSettingsicon .Seethemenuoptionsbelow:
IconDescriptionOperation
DiodePressOKtotoggleSMART/CLASSICdiodemodes
APOAutoPower
OFF
PressOKtoopen.ScrolltoatimeroptionOFF,1,2,5,or10minutes
andpressOK.ThenpressRETURN
LoggersampleratePressOKtoaccesstheselector.Usethearrowbuttonstosetrate
from1~99seconds;pressOKtoconfirm
Real‐timeclockPressOKtoopen.Usethearrowbuttonstoscrollandtoselectthe
currentdateandtime;pressOKtoconfirm
BatterytypePressOKtoselectbatterytypeinuse:AlkalineorLithium
Bluetooth®PressOKtotoggleBluetooth®ON/OFF.RefertotheUserManualfor
completedetails.
ButtonpresstonePressOKtotogglethebuttonpresstoneON/OFF
Language DisplaylanguagePressOKtoopenmenu.ScrolltothedesiredlanguageandpressOK.
PressRETURNtoexit
ClearlogreadingsPressOKtodeletealldataloggerrecords
ClearImagesPressOKtodeleteallsavedimages
HelpScreenPressOKtoviewFLIRsupportcontactinformation
InformationPressOKtoseefirmware,Lepton,Bluetooth®,andlaser
icons.PressOKatthedesiredicontoviewitsinformation.
STATUSBARDISPLAYICONS
Lefttoright,row1:Datalogger,DataHold,AutoRange,Laser,WorkLight,Bluetooth®,APO,Battery
Lefttoright,row2:Inrushcurrent,DCAZero,FlexClamp(iconandrange),VFD,LowImpedancemode
NotethatDiodeandContinuitysymbolsalsoappearintheStatusbarareaofthedisplay.

FLIRCM275–QUICKSTARTDocumentIdentifier:CM275‐QS‐en‐US_AB
CLAMPMETERBASICS
AUTO‐MANUALRANGE
InAutomode(default),themeterautomaticallyselectstheoptimalscale.InManual,therangeisuser‐
selected.InAutomode,the indicatorappears.ToswitchtoManualmodefromAuto,shortpress;
thenshortpress repeatedlytochangethescale.ToreturntoAutomode,longpress.
OUT‐OF‐RANGE(OL)MEASUREMENTS
InManualrange,iftheinputisaboveorbelowtheselectedrange‘OL’isshown;inthiscaseswitchto
anotherrange.InAutomode,theinputisbeyondthemeter’smeasurementcapability.
DATAHOLD/WORKLIGHTBUTTON
InDataHoldmode,thedisplayedreadingorthermalimageisfrozen.Toenter/exitDataHoldmode,short
press.InHoldmode,the indicatorappears.LongpresstoswitchworklightsON/OFF.
VFDMODE
VFDeliminateshigh‐frequencynoisefromACcurrent/voltagemeasurementsusingalow‐passfilter.
1. PressOKtoaccesstheMainmenu
2. PressOKattheAdvancedMenuicon
3. PressOKattheVFDicon;thebluedotnexttotheiconandtheVFDdisplayiconwillappear
4. De‐selecttheVFDmodebypressingOKagain.
MAX‐MINMEMORY
Displaymaximumandminimumreadings,updatingonlywhenahigher/lowervalueisregistered.
1. PressOKtoaccesstheMainmenu
2. PressOKattheAdvancedMenuicon
3. PressOKattheMAX‐MINicontoshowMaximum andMinimum readings.
INRUSHCURRENT
InInrushCurrentmode(),themeterdisplaystheACRMScurrentreadingforthefirst100msperiodafter
thetriggerpoint(currentdetectionthreshold)isreached.PleaserefertotheUserManualfortheCurrent
detectionthresholdspecifications.
DCCURRENTZERO
TheDCzerofeatureremovesoffsetandimprovesDCCurrentaccuracy.Beforeexecutingthestepsbelow,set
theclampmetertotheDCCurrentmeasurementmode.
1. PressOKtoaccesstheMainmenu
2. PressOKattheAdvancedMenu icon
3. PressOKattheDCAZeroicon,thedisplaywillzero
DATALOGGER
Logupto40,000totalreadingsovertenmemory‘sets’.Eachtimethedataloggerisstartedanew‘set’is
created.PressOKtoaccesstheMainmenu,pressOKattheAdvancedMenu icon,andpressOKatthe
dataloggericontobeginlogging.RefertotheUserManualandtotheGallerySettingsMenu,General
SettingsMenuandAdvancedMenusectionsearlierinthisQuickStartforadditionalinformationon
dataloggerfunctionality.

FLIRCM275–QUICKSTARTDocumentIdentifier:CM275‐QS‐en‐US_AB
TESTLEADMEASUREMENTSCURRENTCLAMPMEASUREMENTS
FLEXCLAMPADAPTORMEASUREMENTSSMARTDIODEBASICS
InSMARTDiodemode,themetersendsa
polarity‐alternatingtestsignalthroughthe
diode(noneedtoreversepolaritymanually).
1. Select‘SMART’diodeasshowninthe
GeneralSettingsMenusection.
2. Setthefunctionswitchtothediode
positionandusetheMODEbuttonto
selectdiode(ifnecessary).
3. Connectredandblacktestleadsto
meterandmeasureacrossan
unpowereddiode.
Themeterwillshow±0.400~0.800Vfora
gooddiode,‘BAD’forashorteddiode,and
‘O.L’foranopeneddiode.Fordetailed
information,pleaserefertotheUser
Manual.

FLIRCM275–QUICKSTART DocumentIdentifier:CM275‐QS‐en‐US_AB
CorporateHeadquarters
FLIRSystems,Inc.
2770SWParkwayAvenue
Wilsonville,OR97070USA
Telephone:+1503‐498‐3547
CustomerSupport
TechnicalSupportWebsite http://support.flir.com
TechnicalSupportEmail [email protected]
CustomerSupportTelephone +1855‐499‐3662option3(tollfree)
CertificateofConformityandISO9001QualityManagementSystem
FLIRCommercialSystems,Inc.,anISO9001certifiedcompany,herebycertifythatourinstrumentsmeetorexceedthe
specificationsstatedintheusermanuals.
ThisinstrumentwasproducedunderthestringentguidelinesofFLIRCommercialSystem’scertifiedISO9001Quality
ManagementSystemfromdesigntodelivery.Allinstrumentsarecalibratedandinspectedtomeetthestated
specifications.
FLIRCommercialSystems,Inc.inspectsitsincomingshipmentsusinganapprovedsamplingplanwithanAcceptable
QualityLevel(AQL).AllincominginspectionsareperformedusingtestequipmentthatistraceabletoNational
Standards.OurISO9001QualityManagementSystemextendstoaftermarketsupportwheretestedqualityprocedures
areappliedtocustomersupport,service,calibrationandrepairprocesses.
ThisinstrumentwillprovideyearsofreliableservicewhenusedinaccordancewiththeinstructionsprovidedintheUser
Manual.Tohavethisdevicecertifiedpleasecontactyourlocalservicecenter.
QualityAssuranceManagement
PublicationIdentificationNo.:CM275‐QS
Releaseversion: AB
ReleaseDate:November2017
Language:en‐US
Copyright©2017FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform.
www.flir.com
Table of contents
Other FLIR Measuring Instrument manuals

FLIR
FLIR CM55 User manual

FLIR
FLIR TA72 User manual

FLIR
FLIR MR04 User manual

FLIR
FLIR EXTECH SDL310 User manual

FLIR
FLIR EXTECH HDV-TX1 Dimensions

FLIR
FLIR CM42 User manual

FLIR
FLIR Extech Instruments RH520A User manual

FLIR
FLIR MSX MR277 User manual

FLIR
FLIR Extech Instruments EN300 User manual

FLIR
FLIR MR176 User manual