
PG7607™ OPERATION AND MAINTENANCE MANUAL
© 2010 Fluke Calibration Page 6
Remove the piston retaining assembly (5). The PG7607 platform is delivered with the
piston retaining assembly (5) installed in the mounting post (7). Holding the piston
retaining assembly by the knurled surface (14), unscrew it (rotate CCW) fully from the
mounting post and remove it. Avoid touching the WHITE cylinder centering ring (15) as it
will eventually contact the critical inside surface of the cylinder.
Check that the lower piston sealing O-ring (10a) is installed. The PG7607 platform is
delivered with the lower and upper piston sealing O-rings (10a, 10b) and O-ring carriers
installed loosely in the recess on the mounting post. Remove the upper O-ring and O-
ring carrier (10b) and retain. Check that the lower O-ring and O-ring carrier (10a) are
present and properly seated. (If needed two spare o-rings are included in the PG7607
platform accessory kit).
Remove the piston (6) from its packing. Using lint free gloves or wipes to avoid direct
contact with the piston critical surface (polished appearance) remove the piston (6) from
its packing.
Caution
Never handle the piston or cylinder with bare hands. Oil and acids from
bare skin can damage the critical surfaces (polished appearance). The
cylinder is part of the mass load and contamination can change its mass.
Install the piston (6) on the mounting post (7). With the end of the piston etched with “X”
and/or the serial number UP (11) so that it will enter the cylinder first, slip the piston over the
mounting post and slide it down until it seats on O-ring (10a.) Check that the end of the piston
facing UP has “X” and/or the serial number marked on it.
Install the piston retaining assembly (5). Place the upper O-ring and carrier (10a) centered
on the top surface of the piston. Hold the piston retaining assembly (5) by the knurled ring (14);
avoid touching the white cylinder centering ring (15) as it will eventually contact the critical
inside surface of the cylinder. Thread the piston retaining assembly into the top of the
mounting post (7) and screw it down carefully assuring that upper O-ring carrier (10a) is not
pinched. The assembly’s shoulder (13) should meet the top of the piston (11) with no gap
between the two surfaces. There will be some resistance from the mounting post O-ring (12),
only hand tightening is needed. Do not use tools to install the piston retaining assembly.
Tighten lightly by hand until the assembly reaches the end of the engagement threads and
surfaces (13) and (11) meet.
Clean the piston (6) and the cylinder centering ring (15), consider cleaning the
cylinder (4). Using the glass cleaner method described in Section 4.2, clean the
piston (6) and the cylinder centering ring (15). If this is the first piston-cylinder installation,
also clean the cylinder (4). On subsequent installations consider cleaning the cylinder
before installing it.
Verify that the cylinder cap O-ring (3) is properly installed. Remove the cylinder (4) from
its packing. It is delivered with the cylinder cap O-ring (3) installed. Check that the O-ring is
present and properly seated. If not, install one (an extra one is included in the piston-cylinder
accessories). Use tweezers or another long narrow tool. DO NOT touch the inside of the
cylinder.
Verify that the cylinder adjustment mass (17) is properly installed. Check that the
cylinder adjusting mass (17) is present and completely threaded onto the cylinder
retaining screw (2). This mass is used to adjust the total cylinder mass to the value
reported in the calibration report.
Install the cylinder (4) on the piston (6). Gently slip the cylinder (4) over the piston (6).
The cylinder centering ring (15) aligns the cylinder on the piston facilitating the
insertion process. Allow the cylinder to slip all the way down the piston until it reaches
end of stroke.