Short circuit current greater than 1 mA and less than 2 mA
Bar graph range 0 to 1 TΩ
Insulation test voltage -0 % to +10 % at 1 mA load current
accuracy
Induced ac mains 2 mA maximum
current rejection
Charging rate for 5 seconds per µF
capacitive load
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com/
©2003 Fluke Corporation. All rights reserved.
Trademarks are the property of their respective owners.
Printed in U.S.A. 10/2003 1629685 D-ENG-N Rev D
Fluke. Keeping your world
up and running.
Specifications
Electrical specifications
The meter’s accuracy is specified for one year
after calibration at operating temperatures of 0
°C to 35 °C. For operating temperatures outside
the range (-20 °C to 0 °C and 35 °C to 50 °C),
add +/- .25 % per °C, except on the 20 % bands
add +/- 1 % per °C.
Insulation resistance measurement
Test Voltage Range Accuracy
(dc) (± reading)
250 V 200 kΩto 5 GΩ5 %
5 GΩto 50 GΩ20 %
500 V 200 kΩto 10 GΩ5 %
10 GΩto 100 GΩ20 %
1000 V 200 kΩto 20 GΩ5 %
20 GΩto 200 GΩ20 %
2500 V 200 kΩto 50 GΩ5 %
50 GΩto 500 GΩ20 %
5000 V 200 kΩto 100 GΩ5 %
100 GΩto 1 TΩ20 %
Timer 0 to 99 minutes; settable in 1 minute increments;
indicated to within 1 second
Ramp 0 % to 100 % of selected test voltage, or until
breakdown
Mechanical specifications
Dimensions (L x W x H) 170 mm x 242 mm x 330 mm
(6.7 in x 9.5 in x 13.0 in)
Weight (with battery) 4 kg (8.8 lbs)
Power
Battery type 12 volt, lead-acid, rechargeable
Charger input (AC) 85 to 250 V ac 50/60 Hz 50 VA
Battery life (typical) Test voltages Number of tests
Note: At higher temperatures, 250 V 4138
the battery requires more 500 V 3913
frequent charging. 1 kV 3462
2.5 kV 2043
5 kV 1000
Environmental specifications
Operating temperature -20 °C to 50 °C (-4 °F to 122 °F)
Storage temperature -20 °C to 65 °C (-4 °F to 149 °F)
Relative humidity 80 % at 31°C decreasing linearly
to 50 % at 50 °C
Operating altitude 0 to 2000 meters
Enclosure sealing IP-40
Shock, vibration MIL-PRF-28800F, CLASS II
EMI immunity EN 61326
Safety EN 61010-1, CAT III, 600 V,
EN61557, PARTS 1 & 2
Input overload protection 600 V ac continuous
Software specifications
Quicklink 1550B Software requires a PC running
Windows®95, 98, Me, 2000, NT4.0
Two-year warranty
Included accessories
Test leads and low leakage probes
Alligator clips
Interface adapter and cable
Quicklink 1550B PC software
Line cord
Soft carrying case with
waterproof bottom
User’s manual CD with 12 languages
Getting Started Guide
Optional accessories
25 foot extension leads
Range Accuracy
Leakage current 1 nA to 2 mA ± (5 % + 2 nA)
Capacitance 0.01 uF to 15.00 µF ± (15 % rdg + 0.03 µF)
measurement
Live circuit 30 V to 600 V ± (5 % + 2 V)
indicator ac/dc, 50/60 Hz
Authorized Distributor
Optimum Energy Products Ltd.
333, 11979 - 40th Street S.E.
Calgary, Alberta T2Z 4M3
Toll Free 877-766-5412
Main 403-256-3636
Fax 403-256-3431