
2.4 START-UP PROCEDURE
Once the humidification system is operational, the incubator may be placed into operation. Preset
the controls as follows:
Overtemp Safety Thermostat ...............10 (Fully Clockwise)
Undertemp Safety Thermostat .............10 (Fully Clockwise)
Main Power Switch..........................................................ON
Humidity Controller....................................Desired Setpoint
Temperature Controller...............................Desired Setpoint
CO2Flowmeter .............................................................. OFF
CO2Purge Control ......................................................... OFF
For best overall performance, the refrigeration switch should be placed in the "ON" position for
most applications. When running Low/No humidity, at high temperatures, the refrigeration switch
may be placed in the "OFF" position.
The defrost switch should be placed in the "ON" position when the temperature setpoint is 10°C or
below.
CAUTION! Failure to turn the defrost switch "ON" when required will cause icing of the
refrigeration coil, which in turn will impair performance.
2.5 SETTING THE OVERTEMP SAFETY THERMOSTAT
After the chamber temperature and humidity have stabilized, the overtemp safety thermostat may
be set as follows:
1) Turn the overtemp control knob slowly counterclockwise until the audible alarm sounds and
the overtemp indicator light comes on.
2) Turn the overtemp control knob clockwise approximately two units on the arbitrary scale.
The alarm should be silenced and the overtemp indicator light should go out.
The overtemp safety thermostat is now set a few degrees above the control temperature setpoint. If
the chamber temperature should rise to the overtemp control point, the alarm system will be
activated, power to the heaters will be interrupted, and the chamber temperature will be maintained
at the overtemp control point.
If an overtemp condition should occur, the cause must be determined and corrected before normal
operation via the main temperature controller can be resumed.
NOTE: Whenever the chamber temperature control setpoint is changed, the overtemp safety
thermostat must be reset to accommodate the change.