To ensure the correct application of the scales of this series,
please read these instructions carefully and properly
preserved for future reference.

G&G GmbH | Technical data
2
Version 2.0
04/2015
EN
Contents
1Technical data ....................................................................................................................................4
2Overview............................................................................................................................................7
2.1 Proper Use ..................................................................................................................................7
2.2 Improper Use ..............................................................................................................................7
2.3 Control of inspection, measuring and test equipment...................................................................7
3Warranty............................................................................................................................................8
4Basic Safety Precautions .....................................................................................................................8
5Transportation and Storage ................................................................................................................9
5.1 Acceptance check ........................................................................................................................9
5.2 Packing........................................................................................................................................9
6Unpacking, Setup and Commissioning.................................................................................................9
6.1 Installation Site, Location of Use ..................................................................................................9
6.2 Unpacking / Installation ...............................................................................................................9
6.3 General power supply................................................................................................................10
6.3.1 Battery power supply..........................................................................................................10
6.3.2 Mains connection during rechargeable battery operation....................................................10
6.3.3 Battery exchange................................................................................................................10
6.4 Initial Commissioning.................................................................................................................10
6.5 Adjustment................................................................................................................................10
6.6 External calibration....................................................................................................................11
7Wind Protection Overview................................................................................................................12
8Operating elements..........................................................................................................................13
9Basic Operation ................................................................................................................................14
9.1 ON/OFF - [Power On/OFF] .........................................................................................................14
9.2 TARE..........................................................................................................................................14
9.3 LIGHT - [Display backlight]..........................................................................................................15
9.4 COUNT - [Pieces counting function]............................................................................................15
9.5 UNIT- [Weighing unit selection]..................................................................................................15
9.6 CAL - [Calibration with external weight] .....................................................................................15
10 Displayinformation / Error messages .............................................................................................16
11 Configuration & User Menu...........................................................................................................16

G&G GmbH | Technical data
3
12 RS 232C Interface..........................................................................................................................17
12.1 Technical Data ...........................................................................................................................17
12.2 RS232 Pinout (9-pin D-Sub) ........................................................................................................18
12.3 Description of the data transfer .................................................................................................18
12.4 Remote control instructions.......................................................................................................18
13 Maintenance and Servicing............................................................................................................19
13.1 Cleaning ....................................................................................................................................19
13.2 Maintenance and repair.............................................................................................................19
13.3 Disposal.....................................................................................................................................19
14 Instant help...................................................................................................................................20

G&G GmbH | Technical data
4
1Technical data
Model
E150Y
E300Y
E600Y
E1200Y
E3000Y
E6000Y
Weighing range (max)
150 g
300 g
600 g
1200 g
3000 g
6000 g
Readability (d)
5 mg
10 mg
10 mg
10 mg
0.1 g
0.1 g
Tare range
150 g
300 g
600 g
1200 g
3000 g
6000 g
Linearity
± 1 d
± 1 d
± 1 d
± 1 d
± 1 d
± 1 d
Reproducibility
2 d
2 d
2 d
2 d
2 d
2 d
Calibration weight
100 g
200 g
300 g
1000 g
2000 g
5000 g
Minimum weight*
10 d
10 d
10 d
10 d
10 d
10 d
Stabilization time
< 4 s
< 4 s
< 4 s
< 4 s
< 4 s
< 4 s
Reference weight min.
10 d
10 d
10 d
10 d
10 d
10 d
Minimum piece weight**
1 d
1 d
1 d
1 d
1 d
1 d
Warming up time
30 min.
30 min.
30 min.
30 min.
30 min.
30 min.
Reference piece numbers
for piece count
1, 10, 20, 50, 100
Weighing units
g (Gramm), ct (Karat), ozt (Feinunze), oz (Unze),
GN (Gran), dwt (Pennyweight), lb (Pfund)
LCD-Display Size
105 x 27 mm
Dimensions
(W x D x H)
290 x 200 x 82 mm
Weighing plate
(Ø / W x D)
Ø 135 mm
160 x 190 mm
Wind protection
Yes
No
Operating temperature
20°C ± 15°C
Humidity
≤ 80% R.H.
Interface
RS-232
Power supply***
Mains power adapter Input: AC 230V±10% 50Hz/60Hz
Output: DC 6V/300mA
4 x 1.5V AA batteries
Auto Power Off
Yes (see Chap. “Configuration & User Menu“)
* Hits: minimum weight for an accurate weighting result.
** Hits: minimum piece weight at counting.
*** Hits: only use the original supplied mains adapter. Otherwise there is a risk that the
Scale be impaired.

G&G GmbH | Technical data
5
Model
E150Y-C
E300Y-C
E600Y-C
E1200Y-C
E100YB-C
E200YB-C
Weighing range (max)
150 g
300 g
600 g
1200 g
100 g
200 g
Readability (d)
5 mg
10 mg
10 mg
10 mg
1 mg
1 mg
Tare range
150 g
300 g
600 g
1200 g
100 g
200 g
Linearity
± 1 d
± 1 d
± 1 d
± 1 d
± 1 d
± 1 d
Reproducibility
2 d
2 d
2 d
2 d
2 d
2 d
Calibration weight
100 g
200 g
300 g
1000 g
100 g
200 g
Minimum weight*
10 d
10 d
10 d
10 d
10 d
10 d
Stabilization time
< 4 s
< 4 s
< 4 s
< 4 s
< 4 s
< 4 s
Reference weight min.
10 d
10 d
10 d
10 d
10 d
10 d
Minimum piece
weight**
1 d
1 d
1 d
1 d
1 d
1 d
Warming up time
30 min.
30 min.
30 min.
30 min.
30 min.
30 min.
Reference piece
numbers for piece
count
1, 10, 20, 50, 100
Weighing units
g (Gramm), ct (Karat), ozt (Feinunze), oz (Unze),
GN (Gran), dwt (Pennyweight), lb (Pfund)
LCD-Display Size
font (105 x 27 mm), back (91 x 24 mm)
Dimensions
(W x D x H)
290 x 200 x 256 mm
Weighing plate
(Ø / W x D)
Ø 120 mm
Ø 85 mm
Wind protection
Yes
Operating temperature
20°C ± 15°C
Humidity
≤ 80% R.H.
Interface
RS-232
Power supply***
Mains power adapter Input: AC 230V±10% 50Hz/60Hz
Output: DC 6V/300mA
4 x 1.5V AA batteries
Auto Power Off
Yes (see Chap. “Configuration & User Menu“)
* Hits: minimum weight for an accurate weighting result.
** Hits: minimum piece weight at counting.
*** Hits: only use the original supplied mains adapter. Otherwise there is a risk that the
Scale be impaired.

G&G GmbH | Technical data
6
Model
E300YA
E600YA
E1200YA
E3000YA
E6000YA
E2000YB
E3000YB
Weighing range (max)
300 g
600 g
1200 g
3000 g
6000 g
2000 g
3000 g
Readability (d)
0,1 g
0,1 g
0,1 g
0,5 g
1 g
10 mg
10 mg
Tare range
300 g
600 g
1200 g
3000 g
6000 g
2000 g
3000 g
Linearity
± 1 d
± 1 d
± 1 d
± 1 d
± 1 d
± 1 d
± 1 d
Reproducibility
2 d
2 d
2 d
2 d
2 d
2 d
2 d
Calibration weight
200 g
500 g
1000 g
2000 g
5000 g
2000 g
2000 g
Minimum weight*
10 d
10 d
10 d
10 d
10 d
10 d
10 d
Stabilization time
< 4 s
< 4 s
< 4 s
< 4 s
< 4 s
< 4 s
< 4 s
Reference weight min.
10 d
10 d
10 d
10 d
10 d
10 d
10 d
Minimum piece
weight**
1 d
1 d
1 d
1 d
1 d
1 d
1 d
Warming up time
30 min.
30 min.
30 min.
30 min.
30 min.
30 min.
30 min.
Reference piece
numbers for piece
count
1, 10, 20, 50, 100
Weighing units
g (Gramm), ct (Karat), ozt (Feinunze), oz (Unze),
GN (Gran), dwt (Pennyweight), lb (Pfund)
LCD-Display Size
105 x 27 mm
Dimensions
(W x D x H)
290 x 200 x 82 mm
Weighing plate
(Ø / W x D)
Ø 135 mm
160 x 190 mm
Wind protection
Yes
No
Operating temperature
20°C ± 15°C
Humidity
≤ 80% R.H.
Interface
RS-232
Power supply***
Mains power adapter Input: AC 230V±10% 50Hz/60Hz
Output: DC 6V/300mA
4 x 1.5V AA batteries
Auto Power Off
Yes (see Chap. “Configuration & User Menu“)
* Hits: minimum weight for an accurate weighting result.
** Hits: minimum piece weight at counting.
*** Hits: only use the original supplied mains adapter. Otherwise there is a risk that the
Scale be impaired.

G&G GmbH | Overview
7
2Overview
The electronic balance PLC is a light, rugged precision balance suitable for different purposes. The
removable stainless steel pan allows the laboratory balance to be cleaned easily. The calibration of this
precision balance is done by means of external calibration weights (automatic calibration procedure).The
precision balance is equipped with four adjustable feet. If the balance is used in a laboratory, it can be
powered by the external mains adapter (included). In case the balance is used as a mobile weighing
instrument, it can be powered by AA batteries (not included).
Description:
designed for professional indoor or outdoor weighing.
High precision, sensitivity and rapid response time (<5 sec.).
DMS technology and therefore is also suitable for weighing of iron, nickel or other magnetic materials.
High quality materials and an optimized power supply.
High reliability and long service life.
An easy-cleaning, durable and stable shell [with wind protection]. (see chap. “Scope of delivery”)
Removable stainless steel weighing pan on a plastic support.
Four adjustable feet, a Level Indicator.
External calibration, zeroing searches, counting function, tare function.
Different measuring units can be selected via the keypad (g, ct, ozt, oz, GN, dwt, lb).
Large LCD display with backlight, digit height 15 mm.
Powered by the external mains adapter or battery (accumulator).
2.1 Proper Use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is
intended to be used as a “non-automatic” balance, i.e. the material to be weighed is manually and
carefully placed in the center of the weighing plate. As soon as a stable weighing value is reached the
weighing value can be read.
2.2 Improper Use
Do not use the balance for dynamic weighing. In the event that small quantities are removed or added
to the material to be weighed, incorrect weighing results can be displayed due to the “stability
compensation” in the balance. (Example: Slowly draining fluids from a container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the measuring system.
Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly
existing tare load, must be strictly avoided. Balance may be damage by this.
Never operate the balance in explosive environment. The balance is not explosion protected.
The structure of the balance can not be modified. This may lead to incorrect weighing results, safety-
related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of use must be
released by G&G in document.
2.3 Control of inspection, measuring and test equipment
As part of quality assurance the measuring-related properties of the balance and the available adjusting
weight must be checked regularly. The responsible user must define a suitable interval as well as the

G&G GmbH | Warranty
8
nature and scope of this test. Information regarding the test equipment of scales and of the required test
weights is available on the G & G Company - Homepage. (www.gandg.de)
3Warranty
The G & G Company is liable for the defects within the framework of existing legislation.
We warrant the product manufactured by us to be free from defective material or factory
workmanship and agree to repair or replace this product which, under normal use and service, disclose
the defect to Bethe fault of our manufacturing, with no charge for parts and service. If we are unable to
repair or replace this product, we will make a full refund of the purchase price.
Consult the user's manual for proper instruction regarding use of this instrument. Our obligation under
this warranty is limited to repairing, replacing or making refund of this test equipment which proves to
be defective within 24 months from the date of original purchase
Warranty claims shall be voided in case:
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
4Basic Safety Precautions
COUTION!:Carefully read this operation manual before setup and commissioning.
Environmental conditions: only use indoor, max. Altitude: 2000m, Operating temperature: 20°C ± 20°C,
humidity of air: <= 60% R.H., Power fluctuations less than ±10%.
Do not use the scale in any high temperature fluctuations and avoid strong airflows, vibrations and dust.
The balance could never get wet and make sure that the balance is not used in an environment with
extreme temperature and humidity.
Do not use the balance in explosive conditions.
Only use the extender cable, which with ground wire.
If the power line cable is damaged, the appliance must be unplugged immediately and the power line
cable be replaced.
Place the balance on a stable surface and set it horizontally with the help of the adjustable feet and the
spirit level.
To avoid to damage the device, do not use at low-battery condition.
The electronic balance is a precision measuring instrument and must be handled with great care.
Violation of limit of weighing range or impact can cause permanent damage to the scale.
The balance needs a warm up time for about 30 min. before using.
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with
mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth.
Do not overload the balance more than 20% of maximum load (Max). Do not press the pan with a hand!
By maintenance please notice the information in chapter “Maintenance and Servicing”.

G&G GmbH | Transportation and Storage
9
If the balance seems not to operate properly, unplug it from the mains and do not use until checked by
authorized service.
All repairs and necessary regulations can be made by authorized personnel only.
5Transportation and Storage
5.1 Acceptance check
When receiving the appliance, please check packaging immediately, and the appliance itself when
unpacking for possible visible damage.
5.2 Packing
Keep all parts of the original packaging for a possibly required return
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove loose/mobile parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the weighing platform, power unit etc.
against shifting and damage.
6Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of
use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Major display deviations (incorrect weighing results) may be experienced should electro-
magnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations
or instable power supply occur. Change location or remove source of interference.
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing next to a radiator or in the
direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapors and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold
appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected
appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
6.2 Unpacking / Installation
Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended
workstation.

G&G GmbH | Unpacking, Setup and Commissioning
10
1. Take the balance and supplied accessories (a feeder, pan elements) out of the box.
2. Place the balance on a stable ground not affected by mechanical vibrations and airflows.
3. Level the balance with foot screws so that the air bubble in water-level is in the middle.
4. Plug the adapter to the power socket at the back of the balance.
5. Connect the power adapter to an AC220V ± 10%, 50Hz/60Hz outlet.
Scope of delivery:
Balance
PVC Wind protection (for E150Y, E300Y, E600Y, E1200Y, E300YA, E600YA and E1200YA available)
Glass Wind protection (for E150Y-C, E300Y-C, E600Y-C, E1200Y-C, E100YB-C und E200YB-C available)
Weighing pan
AC adapter
Operating Instructions
6.3 General power supply
If the balance is used in a laboratory, it can be powered by the external mains adapter (included). In case
the balance is used as a mobile weighing instrument, it can be powered by AA batteries (not included) or
internal accumulator (included). The accumulator is charged via the delivered adapter.
To extant work time of accumulators (battery), it is possible to turn off display’s backlight and the balance
automatically. The settings of “Power Save” operation see chap. 9.3 and chap. 11.
Please only use the original supplied mains adapter. Otherwise there is a risk that the Balance be
impaired.
If the balance is not used for a long time, remove the battery/accumulator and store them separately.
Leaking battery liquid could damage the balance.
6.3.1 Battery power supply
The balances can be supply from mains power witch is delivered with the balance. It is also possible to use
normal AA batteries.
6.3.2 Mains connection during rechargeable battery operation
When the balance is connected with the mains adapter, the battery power will be disconnected
automatically. Therefore, there is no battery charging function under the battery power supply.
6.3.3 Battery exchange
Remove the battery compartment cover and insert six batteries (AA size) into the battery compartment,
taking extreme care that the polarities plus (+) and minus (-) is observed.
6.4 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have reached the
operating temperature (warming up time 30 min.). During this warming up time the balance must be
connected to the power supply (mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity.
Please read the information carefully in chap. “Adjustment”.
6.5 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be
coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration

G&G GmbH | Unpacking, Setup and Commissioning
11
due to gravity at its place of location (only if the balance has not already been adjusted to the location in
the factory). This adjustment process must be carried out for the first commissioning, after each change of
location as well as in case of fluctuating environment temperature. To receive accurate measuring values it
is also recommended to adjust the balance periodically in weighing operation.
6.6 External calibration
Adjustment with recommended adjustment weight (default setting) Weight value of the required
adjustment weight sees chap. 1 “Technical data”. Weights of different nominal values may be used for
adjustment but are not optimal for technical measuring. Info about adjustment weights can be found on
the Internet at: www.gandg.de
Operations sequence:
Observe stable environmental conditions. A warming up time (30 min.) is required for stabilization.
Press the [ON/OFF]-Key, turn on the balance.
Then press the [CAL]-Key, keep it pressed until the acoustic signal gets mute and the weighed value “C
XXX” for the required adjustment weight appears flashing. (e.g.: C 100 means 100 g)
During the flashing display put the required adjustment weight carefully in the center of the weighing
plate.
The flashing display disappears.
After successful adjustment the balance automatically returns to weighing mode.
Take away the adjustment weight, the display shows ZERO.
If an error message “C----F” is displayed in the event of an adjustment error or incorrect adjustment
weight. Wait until the balance is again in weighing mode and repeat the adjustment procedure.

G&G GmbH | Wind Protection Overview
12
7Wind Protection Overview
Note: The PVC wind protection
is supplied for model E150Y, E300Y, E600Y,
E1200Y, E300YA, E600YA und E1200YA.
(see chap. 1 "Technical data").
Note: The Glass wind protection is supplied for model E150Y-C, E300Y-C, E600Y-C, E1200Y-C, E100YB-C
und E200YB-C. (see chap. 1 "Technical data").

G&G GmbH | Operating elements
13
8Operating elements
Pic.1 E150Y, E300Y, E600Y, E1200Y, E300YA, E600YA and E1200YA
Pic.2 E3000Y, E6000Y, E3000YA, E6000YA, E2000YB and E3000YB

G&G GmbH | Basic Operation
14
Pic.3 E150Y-C, E300Y-C, E600Y-C, E1200Y-C, E100YB-C and E200YB-C
1. Weighing pan* 2. Level Indicator 3. ON/OFF-key 4. Calibration-key
5. Unit-key 6. Count-key 7. Light-key 8. Tare-key
9. Adjustable feet 10. LCD-Display 11. RS-232 Interface 12. Power socket
13. Pic.1 PVC wind protection 14. Stainless steel column 15. Wind protection ring
Pic.3 Glass wind protection
Note: The weighing pan is supplied depending on the model in a round or square version. (see chap. 1
"Technical data").
9Basic Operation
After new purchase of the balance or after a long storage, you have to calibrate the balance before the first
commissioning. Level the balance with the adjustable feet horizontal until the air bubble in the level is in
the center of the Level Indicator. Give the balance after changing the location a warm-up time min. 30 min.
before using. This time is necessary for the balance to adapt to the eventual temperature.
9.1 ON/OFF - [Power On/OFF]
Press the [ON/OFF] button to turn on the balance. The display shows the message "ON". Then the max.
weighing range is displayed. The balance will carry out a self-test (F----1 to F----9). When the "0.00g" in
display appears, the balance is ready for operation.
Press the [ON/OFF] button and keep it pressed, until the display shows the message “OFF”. The balance
will switch off then.
9.2 TARE
When the pan is empty and indication is different than zero press [TARE] key.

G&G GmbH | Basic Operation
15
Weighting with tare: place the weighing box and press the [TARE] button. After standstill control the zero
display appears. The weight of the container is now internally saved. Weigh the material, the net weight
will be indicated. The weight of the weighing container will be displayed as a minus number after removing
the weighing container. The tare weight is saved until it is deleted. Remove the load from the balance and
press the [TARE] button. The tare procedure can be repeated as many times as necessary, for example
with initial weighing of several components for a mix (add-on weighing). The limit is reached when the
total weighing range capacity is full.
9.3 LIGHT - [Display backlight]
Press the [LIGHT] key to change the display’s backlight mode.
[b ON] - Background illumination on. Display rich of contrast which can also be read in the darkness.
[b OFF] - Backlight switched off to save battery.
[b AUTO] - This display will appear backlit for weight value is not stable. When the weight value is stable,
the display will be extinguished 3 seconds later.
9.4 COUNT - [Pieces counting function]
Before the balance can count parts, it must know the average part weight (i.e. reference). Proceed by
putting on a certain number of the parts to be counted. The balance determines the total weight and
divides it by the number of parts, the so-called reference quantity (1, 10, 20, 50, or 100). Counting is then
carried out on the basis of the calculated average piece weight.
As a rule: The higher the reference quantity the higher the counting exactness.
Follow these steps:
a) Press [COUNT]-key to change the balance into counting mode.
b) Place as many pieces to add-up as required by the reference quantity. Press the [UNIT] button to enter
the setting menu. The display now shows the number "1" and the unit "Pcs".
c) Using the [UNIT] button select the reference quantity (1, 10, 20, 50, or 100).
d) After successful reference determination the current quantity is displayed and the unit “Pcs”.
e) Remove reference weight. The balance is now in parts counting mode counting all units on the
weighing plate.
9.5 UNIT- [Weighing unit selection]
Use the [UNIT]-key to choose weighing units: g (gram), ct (carat), ozt (troy ounce), oz (ounce), GN (gran),
dwt (pennyweight) and lb (pound).
1 g = 5 ct
= 0.032150747 ozt
= 0.035273962 oz
= 15.43235835 GN
= 0.643014931 dwt
= 0.0022046226 lb
9.6 CAL - [Calibration with external weight]
The balance has an external calibration function.
The adjustment procedures see chap. "External calibration”.

G&G GmbH | Displayinformation / Error messages
16
10 Displayinformation / Error messages
F----1Date over display scope
F----2A/D-switch disorder
F----3 Function key disorder
F----5 Sensor signal disorder / Load Cell defect
F----LLower zero position of sensor signal
F----HWeighing capacity is exceeded
C----F Calibration error, instability or not at zero point during calibrating
C----HCalibration error, items on the pan or at higher zero point during calibrating
E----- Storage calibration data error, re-calibration is necessary
------ Zero point of the scale disorder / mechanical defect of the scale
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
11 Configuration & User Menu
The settings of the balance can be changed in the setting menu. This way, the balance can be adjusted to
individual weighing requirements. By the factory the user menu has been set in a way that normally no
more changes must be made, only at special conditions of use.
Overview:
In the off state press the [CAL] button and keep it pressed, press the [ON/OFF] button turn on the balance.
Now the balance is accessed to the setting menu. Using the [CAL] button the individual menu items (C1-
C3) can be selected one after the other. Press the [TARE] button to change the setting value. After all
points have been configured, press [CAL] button returns the balance automatically into weighing mode.
a) C1 = sensitivity setting - values: 0 1 2 3 4
The higher the value, the lower the sensitivity and higher stability.
Filter 0: Setting for dispensing.
Filter 1/2: Sensitive and fast, very quiet using location.
Filter 3/4: Robust but slow, busy using location.
Default setting is 2 or 1.
b) C2 = filtering strength setting - values: 0 1 2 3
The higher the value, the lower the sensitivity and higher stability.
Filter 0: Setting for dispensing.
Filter 1/2: Sensitive and fast, very quiet using location.
Filter 3: Robust but slow, busy using location.
Default setting is 2 or 1.
c) C3 = serial port setting - transmission speed (RS-232)
2(600), 3(1200), 4(2400), 5(4800), 6(9600) - Default setting is 6.

G&G GmbH | RS 232C Interface
17
d) C4 = device number / communication setting (RS-232)
Press [TARE] button changes the value smaller. Press [COUNT] button changes the value bigger. The
number corresponds to the first data signal to the command control of the computer.
Default setting is 27 (1Bhex).
e) C5 = Autom. Abschaltung - values: 0 1 2 3 4 (*only valid for Battery- / Accumulator power supply)
0: disabled
1: after 15 min.
2: after 30 min.
3: after 45 min.
4: after 60 min.
Default setting is 2 or 1.
f) C6 = setting of Print-Key for balance
0 (no print-key), 1 (use [UNIT] as print-key), 2 (use [COUNT] as print-key) - Default setting is 0.
g) C7 = Backside Display Off (*only valid for PLC-300-C, PLC-600-C, PLC-1200-C, PLC-100B-C and PLC-
200B-C)
0 = Backside Display is disabled.
1 = Front- and backside Displays are enable.
Default setting is 1.
h) Save settings
After all points have been configured, press [CAL] button returns the balance automatically into
weighing mode.
i) Restore the factory settings
In the off state press the [TARE]button and press the [ON/OFF]button turn on the balance, keep the
[TARE] button pressed until the balance returns automatically into weighing mode. All the parameters
will be reset to default settings. Please re-calibration the balance before using.
12 RS 232C Interface
An integrated RS-232-interface enables you to automatically send the weighing data to PC. Relevant codes
and control commands are specified as following.
12.1 Technical Data
8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bits, no parity bit
Bit rates supported include: 600, 1200, 2400, 4800 and 9600 bit/s. (default setting is 600)
Miniature plug is necessary (9 pin D-Sub)
For operation with interface faultless operation is only ensured with the correct G&G –interface
cable (max. 2m)

G&G GmbH | RS 232C Interface
18
12.2 RS232 Pinout (9-pin D-Sub)
DE-9 plug pin-out (9-pin., male), as it is usually
available at DTE
DE-9 socket pin-out (9-pin., female), as it is
usually available at DCE
DTE - data terminal equipment (Terminals resp. computer)
DCE - data circuit-terminating equipment (Balance)
Signal
Pin
Origin
Abbreviation
Name
Typical purpose
DE-9
DTE (e.g. PC)
DCE (e.g. Balance)
RxD
Receive Data
Carries data from DCE to DTE.
2
Input
Output
TxD
Transmit Data
Carries data from DTE to DCE.
3
Output
Input
GND
Ground
5
-
-
12.3 Description of the data transfer
Symbols and composition of the data
-/+
2 bit
Negative-sign, if the value is positive a space is transmitted.
Data
7 bit
Value for weight / decimal point
Unit
3 bit
Weight unit
CR
1 bit
Carriage Return
LF
1 bit
Line Feed
Example:
The output-value: -12,345 kg
Output data, displayed as ASCII code
-/+
Data
UNIT
CR
LF
-
1
2
.
3
4
5
g
\r
\n
12.4 Remote control instructions
Via the RS232 data interface, the computer can control the balance with subsequent commands.
If the transmit / device setting is 27 (27=1Bhex, default setting), the remote control commands are sent to
the balance by HEX-code:
1B 70 (ACSII Code p): Send data (Print)
1B 71 (ACSII Code q): calibration function is active (CAL - Key)
1B 72 (ACSII Code r): Counting function is active
1B 73 (ACSII Code s): Weighing unit selection
1B 74 (ACSII Code t): Taring (TARE - Key)
Drivers and software can be free downloaded on our homepage.
www.gandg.de

G&G GmbH | Maintenance and Servicing
19
13 Maintenance and Servicing
13.1 Cleaning
Before cleaning, please disconnect the appliance from the operating voltage.
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild
soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth.
Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Remove spillages immediately.
13.2 Maintenance and repair
The appliance may only be opened by trained service technicians who are authorized by G&G.
Before opening, disconnect from power supply.
13.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to valid national or regional
law of the location where the appliance is used.
According to legal regulations it if forbidden to
dispose electronic equipment in waste containers.
WEEE-Reg.-Nr. DE
Please return wasted balance to the point of purchase or other company specialized in
recycling of wasted electronic components.

G&G GmbH | Instant help
20
14 Instant help
Failure
Possible Cause
The display does not
light up.
The balance is not switched on.
The mains supply connection has been interrupted (mains cable not plugged
in/faulty).
Power supply is interrupted.
(Rechargeable) batteries are inserted incorrectly or empty.
No (rechargeable) batteries are inserted.
The displayed weight is
constantly changing
Draught /air movement.
Table / floor vibrations.
Weighing plate has contact with other objects.
Electromagnetic fields / static charging (choose different location/switch off
interfering device if possible).
The weighing result is
obviously wrong
The display of the balance is not at zero.
Adjustment is no longer correct.
Great temperature fluctuations.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose different location/switch off
interfering device if possible).
F----L
Lower zero position of sensor signal (re- Adjustment).
F----H
Weighing capacity is exceeded.
C----L
Lower zero position of sensor signal (re- Adjustment).
C----F
Calibration error, instability or not at zero point during calibrating.
C----H
Calibration error, items on the pan or at higher zero point during calibrating.
More Information see chap. “Displayinformation / Error messages” )
Should other error messages occur, switch balance off and then on again. If the error message remains
inform manufacturer.
Address: G&G GmbH
Hammfelddamm 7
41460 Neuss
Germany
Telephone: +49 [0] 21 31 - 1 33 56 05
Fax: +49 [0] 21 31 - 1 33 56 07
E-mail: info@gandg.de