Operating Manual for JJ-B Series of Electronic High Precision Balances
(for following models: JJ100B, JJ200B, JJ1500B, JJ2000B)
1. Overview
Manufactured and well-designed by G&G Measurement Plant, adopting
industrial-class high precision sensors, measurement circuit, a “G&G”
dedicated single-chip computer system, and controlling system, JJ-A series
electronic balances are high-quality electronic weighing machines. Advanced
technology, well-selected, exquisite processing and strict testing make JJ-A
series of electronic balances be possession of following strong points:
- Simple operations.
- Integrated water level on the scale
- High precision, hypersensitivity and fast reaction speed
- High quality, strong anti-interference capability, long life, long-term
using stability, adapting to the adverse environment and long hours
continuous work
- With auto calibration
- Unit conversion feature: in seven units between arbitrary conversions
- Balance is targeting output data interface and can be connected directly
to the printer and computer to collect and analyze the data. The balance can
even remote controlled by computer.
In general, with high precision, high stability, high reliability and high
efficiency, the “G&G” electronic balances has widely replace the traditional
mechanical, tray balances. The “G&G” electronic balances have been widely
used in all callings, in different kinds of measurement and test. Those have
played an important role in improving products quality, cost-effectiveness
and efficiency.
2. Specifications and Performance indexes
Depending on the difference of range and resolution, G&G Brand JJ-B
series electronic scales are classified to 4 types as shown
Model JJ100B JJ200B JJ1500B JJ2000B
Capacity 110g 200g 1500g 2000g
Resolution Ratio 0.001g 0.001g 0.01g 0.01g
Calibration Graduation 10d 10d 10d 10d
Tare Scope 2000g 3000g
Calibrating Weight 2000g 2000g
Dimension of Platform Ф135mm
Volume 325(W)X 211(D)X 111(H) mm
Power Supply AC Adapter (AC 230V 15V 300 mA 50Hz-60Hz)
Use Temperature 0-40°C
Use Humidity ≤80%R.H
P.S. Only use the provided adapter or the same specifications of the
adapter, otherwise there will be dangerous.
3. Operational Method
3.1 Turn on the power. Open the switch. The displaying windows
displays from “F----0” to “F----9” and after a certain time for stabilizing,
“0.0” appears; Then preheats for 15 minutes. Instability of the displaying
at every beginning is normal; after a certain time, the displaying will stabilize.
3.2 With empty scale platform, the displaying deviates from zero,
pressing “TARE” key, then the displaying will return to zero.
3.3 If the scale has not been used for a long time or just bought, the
scale should be calibrated. Firstly, the scale is fully preheated (over 15 min.)
with empty scale platform. Then press “CAL”, the displaying window shows
“CXXX” and the balance will automatic calibrate (XXX should be the
weight of standard weights). At the same time, only place calibrating weights
on the scale platform. After stabilizing, the scale displays the weight of the
calibrating weights; till now the calibration is completely finished and the
scale can be normally used. If press “CAL”, but display “C----F”, which
indicates the Zero point is not stable. Then re-press “TARE” to make “zero”
point stable; and re-press the “CAL” to carry out calibration.

3 4
“G&G”
3.4 The units can be changed between “g”, “ct”, “ozt”, “oz”, “GN”,
“dwt”, “lb”. Press “Units” to change the units during weighting
3.5 If needs to tare, first place the container on the platform, after
stabilizing, press “TARE”; the scale displays “zero”. Then place the article
in container, at this time, it shows the net weight of the article. Remove the
article and the container. The scale shows a negative value of container
weight. Still press “TARE” to return to “zero”
3.6 .Count function
3.6.1 Change the balance modus into count modus according to the
Attachment.
3.6.2 Choose samples. To precisely count the articles, firstly the quantity
of samples for counting should be chosen in according with the weights of
the articles. The quantity of samples can be chosen from “1-10-20-50-100”.
For a small number of articles, the larger quantity of samples, the more
precise.
3.6.3 Place the samples on the pan, the balance will show the total
weights of samples. Then press “Count”, balance will show “1”, and unit
will show “pcs”, which indicate that balance is now counting. And all the
samples are one unit. Then press “Count”, the displaying will show between
1-10-20-50-100. The quantity of articles must be the same with the quantity
of samples, and place the similar articles. The total amount of objects is
showed. Press the “Count” again back to the normal situation.
4. Attachment: settings and meanings
Turn on the Scale by pressing “CAL” key, in order to change the con-
figuration and function. the display shows C1 and the relevant number,
which can be changed. Pressing “CAL” can change the parameters; pressing
“TARE” can change the values of parameters.
4.1 C1------ sensitivity setting 0 1 2 3 4, the larger number is, the less
sensible but the better stable. Default setting is 2 or 1
4.2 C2------filtering strength setting 0 1 2 3, the larger number is, the
lower reaction speed, but the better stable. Default setting is 2 or 1
4.3 C3------Baud rate setting 2(600) 3(1200) 4(2400) 5(4800) 6(9600).
Default setting is 2
4.4 C4------communication setting. “TARE” changes the lower unit.
“COUNT” changes the higher units. This is the first data during commun-
ication. Default setting is 27.
4.5 C5------units change. 0 is for units changing and 1 for printing.
Default setting is 0. Counting is 2.
4.6 Re-pressing “CAL” after setting. Then the balance is auto calibr-
ated to 0.
4.7 Press “TARE” and do not loose till “F----3” (Balance should be
switched on).
All the parameters can be reset to default settings. Re-calibration before
using.
5. Data Transmission
Balance is targeting data output interface RS232C and can be directly
connected to the printer. The following soft wares are needed, when connects
to computer:
3.8.1 Transmission format: serial asynchronous
3.8.2 Serial communication protocol: Baud rat: 600, data: 8 units,
starting: 1 unit, ending: 1unit
3.8.3 Data format:
Starting spaces data symbol ending enter
2unit 7units 3unit 1 unit 1 unit
3.8.4 Check bits: no check bits
3.8.5 Output pins content (9-core socket): 2 pins: RXD 3 pins: TXD
5pins: GND
3.8.6 Computer can control balance though RS232. The orders to
balance are:
1BH+70H: sending data

65
1BH+71H: calibration (like “CAL”)
1BH+73H: Units change
1BH+74H: Tare
Sampling demo, connection method of balance and computer, please
download from our website: http://www.gandg.de
6. Structure
1. Glass Windshield 2. Scale pan
3. Part from Windshield 4. ON/OFF key
5. TARE key 6. Multi-function key
7. CAL key 8. Water level
9. RS 232 interface 10. Power supply
7. Cautions for operation
7.1 Using environment: indoor, max. Altitude: 2000m, temperature:
0-40°C, humidity ≤80%, Power fluctuations less than ±10%
7.2 The electronic scale is a precision instrument which needs to be
gently handled during the weighing of articles. Avoid impact and heavy load
on top (never exceed its max. weighing limit). Any forms of impact or
overloading many cause permanent damage to the electronic balance, even
in the use of electronic balance without power supply.
7.3 Place the scale on a firm table for operation, and make sure that the
operating environment of the balance is free of great vibration and interfe-
rence of power supply, and of corrosive gases and liquids.
7.4 Ensure the warm-up time after power-on..
7.5 Warning: If the equipment is used in a manner not specified by
this instruction the perfection provided by the equipment may be impaired
7.6 Gentle handling in daily using, with dry and clean cotton cloth
cleaning.
8. Trouble Shooting
Measures to take
1.the ON/OFF switch must be changed
2. please contact our service
department
please contact our service department
please contact our service department
please contact our service department
Possible causes
1. The ON/OFF switch is
defect
2.Internal Circuit is defect
1.A/D-switch problem
2.Sensor is defect
Keyboard defect
1.Sensor is defect
2.A/D switch problem
Problems
Display shows
nothing
The message “F-2”
The message “F-3”
The message “F-5”

7 8
If electrical appliances are disposed of in landfills or dumps, hazardous
substances can leak into the groundwater and get into the food chain, dam-
aging your health and well-being.
When replacing old appliances with newones, the retailer is legally
obligated to take back your old appliance for disposals at least free of charge.
11. The Warranty
11.1 All product, which produced by the G&& Testing Instrument
Factory in Changchu city, have direct warranty provided by the producer,
this including refund, exchange and repair service.
11.2 Within one year after purchase, the producer offers a free repair
service to all G & G Product, which become defect under proper usage
without human damage. The customer are kindly requested, to mail the
product as well as its original package back to the producer, the producer
will send the repaired or exchanged product to the customer back within
one week.
11.3 There will be a slight charge for the repair service of those prod-
ucts, by which the warranty period expires.
11.4 Please attach the address, postal code and telephone number of
the buyer, so that we can mail the repaired product back in time.
12. Contact
Address: G&G GmbH
Koenigstr. 34
41460 Neuss
Germany
Tel: +49-2131-609-7271
Fax: +49-2131-609-7273
Email: [email protected]
Website: www.gandg.de
Remarks
Option by JJ100B; JJ200B
Quantity
1
1
1
1
1
Article Description
Operation manual
Scale pan
Power cable
Calibration weight
Wind shield
restore the fabric configuration
1.remove the object
2. restore the fabric configuration
restore the fabric configuration
Press the “TARE” key with empty pan,
the balance shows 0, then calibrieren.
1. remove the object, Press the “TARE”
key with empty pan, the balance shows
0, then calibrieren.
2. restore the fabric configuration
1.check the C3 and C4 configuration
2.check the cabel
3. please contact our service
department
1.Error by sensor
2.Error by calibration
1.Object weight beyond
capacity
3.Error by sensor
Error by sensor
Unstable zero or zero not
correctly set
1.Object weight beyond
capacity
2.Error by sensor
1.Parameter configuration
Error
2.Cable connection problem
or cable is defect
3.internal R232 Chip is
defect
The message “F-L”
The message “F-H”
The message “C-L”
by Calibration
The message “C-F”
by Calibration
The message “C-H”
by Calibration
Störung der R-232
Übertragung
10. Trash
Meaning of crossed–out wheeled dustbin:
Do not dispose of electrical appliances as unsorted municipal waste,
use separate collection facilities.
Contact you local government for in formation regarding the collection
systems available.
9. Packing List
This manual suits for next models
3
Table of contents
Other G&G Scale manuals
Popular Scale manuals by other brands

KERN&SOHN
KERN&SOHN TMPN 200K-1HM-A operating instructions

Yamato
Yamato DKS-3002 Operation manual

Medisana
Medisana BS 430 Connect quick start guide

CAS
CAS WEIGHING SOLUTION SWII PLUS Series owner's manual

Seca
Seca 656 Instruction manual and guarantee

Little Balance
Little Balance GREEN POWER 8192 user manual