
SERVICE MANUAL
TP600 & 700 Series T/PUMP
19
Operating
Temperatures, Flow
and Over-Temperature
Safety Circuit Check
This section provides a complete check of all T/Pump functions.
Refer to Functional Check and Safety Inspection Form to record
test data. Follow the steps in this section carefully, paying particular
attention to each step, and its expected result. If at any time the
expected result can not be veried, press the On/Standby button to
stop the test, then restart the procedure. If after a second attempt
the expected result can not be veried, press the On/Standby button
to stop the test, then unplug the T/Pump and call your dealer or
Gaymar's Technical Service Department for assistance.
1. Connect the T/Pump to the TPT9 and Mul•T•Pad as shown in
Figure 9a, 9b. Unkink the pad and hose. Open the hose clamps.
2. Open the ll cap on top of the pump.
3. Fill with room temperature water to the Heating water line.
4. Plug the T/Pump into a properly grounded Hospital Grade
receptacle.
The T/Pump performs its self test of the lights and audible
alarm, with light pattern and short audible beep.
The T/Pump goes to Standby mode with only the Standby
light on.
5. Press the On/Standby button.
6. Set the temperature to 107ºF (42ºC). For TP700 Series pumps,
select Continuous Therapy Mode. Allow temperature to stabilize.
Stabilization is indicated by a steady (not ashing) light at the
temperature setpoint.
7. After stabilization, allow 5 minutes before proceeding. The T/
Pump controls to 107°F (42°C) at the inlet to the pad for the
duration of the test.
Note: If at any time the Function
test needs to be stopped, press
the On/Standby button.
Flow & Operating
Temperature Test
Note: Stay present for the full duration of the test to observe all
readings and to prevent damage to the T/Pump. Perform this test on
a fully assembled unit.
FUNCTIONAL CHECK
Physical Condition Check 1. Examine the plug and line cord along its entire length for physical
damage, such as cuts or cracked insulation. A damaged line cord
should be replaced rather than repaired. Check the quality of the
strain relief of the line cord.
2. Examine membrane panel for cuts or cracks. If damaged, re-
place.
3. Visually inspect pump. Check for cracked or damaged plastic
parts.
Be sure unit is unplugged. Remove four (4) screws secur-
ing upper housings to tray. (reference #13 Parts Illustration - (1)
screw by power cord, (1) screw by hose strap and (2) screws in
front corners). Perform visual inspection of all internal parts. Re-
move any accumulated dirt with a vacuum cleaner or compressed
air hose.
4. Check connectors for cracks, missing O rings, or other damage.
Replace connectors if necessary.