
7
Section 4 - Operating Instructions (cont’d)
Section 5 - Routine Maintenance
NOTE:
•The air temperature surrounding the patient
will vary at each setting, depending upon
room air temperature and the quality of the
insulating blanket or sheet that is used over
the Quilt.
•Operate the Power Unit only in the specied
ambienttemperatureandvoltageranges(see
Specications, page 9).
7. Placeyourhandunderthequilttoverifythat
the Power Unit is warming when set to the
32°C, 38°C, 43°C temperature settings.
8. Place a sheet or blanket over the inated
Quilt to maximize the eciency and
minimize the heat loss from the system.
Monitorthepatient'stemperature,vital
signs, and skin condition regularly
(every 20 minutes or as prescribed by
a physician). If patient's temperature
reaches the normal range, reduce air
temperature or discontinue therapy. If
vital signs become unstable, notify the
attendingphysicianimmediately.Failure
to follow these precautions could
result in death or serious injury.
9. If the REMOVE FROM USE indicator lights,
discontinueuse. Referservicingtoqualied
personnel.
4.1 Shutdown Procedure
When therapy has been completed, turn the
unit oand disconnect the hosefrom the Quilt.
Discard the Quilt.
5.0 Routine Maintenance
Thefollowingroutinemaintenanceprocedures
will help insure dependable operation:
5.1 Cleaning the Power Unit
Clean the control panel, enclosure exterior,
and hose with a soft cloth lightly dampened
with a nonstaining hospital disinfectant. Clean
accumulated dirt from the air vents with a
vacuum cleaner.
5.2 Storing the Hose
Slide the hose under the bed rail hooks on the
back of the Power Unit when not in use.
5.3 Storing the Power Cord
Coil the power cord. Hold it in place with the
strap on theside of thePowerUnit. Or, hangthe
powercordfromthe bracketon theback ofthe
accessory stand.
5.4 Quilt Care
RefertotheinstructionsincludedwitheachQuilt
fordetails on correct application and use. Small
rips or tears in Quilt material may be repaired
with adhesive tape.
Quilts are not sterile. Quilts are intended
for single patient use only. Quilts should be
discarded after use.
5.5 Servicing the Equipment
Repairs to the Power Unit should be
performedbyqualiedpersonnelsuch
as certied biomedical electronics
technicians or clinical engineers
familiar with repair practices for
servicing medical devices, and in
accordance with the TC3249 Service
Manual. Damage to the Power Unit or
malfunction may otherwise result.
WARNING
WARNING