manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. GE
  6. •
  7. Measuring Instrument
  8. •
  9. GE Sievers 900 User manual

GE Sievers 900 User manual

Sievers 900 and 5310 C
On-Line TOC Analyzers
Quick Start Guide
IMPORTANT
For more detailed instructions, you can download manuals for the Sievers 900 On-Line TOC Analyzer, the
Sievers 5310 C On-Line TOC Analyzer,or the Sievers Inorganic Carbon Remover (ICR) from our Web site. Go to
www.geinstruments.com, click Library, then Manuals, and make your selection.
Sievers 900 and 5310 C On-Line TOC Analyzer Installation Instructions (continued)
System Specications*
Linear range (900)
Linear range (5310 C)
Precision
Accuracy
Analysis time
0.03 ppb - 50 ppm TOC
4 ppb - 50 ppm TOC
<1% of RSD
±2% or ±0.5 ppb, whichever is greater
4 minutes
Sample ow rate
(nominal)
Power requirements
Fuses
Analysis mode: 0.5 mL/min
Fast Flush (between samples): 1.0 mL/min
100-240 ± 10% VAC, 50 watts, 50/60 Hz
No user-replaceable fuses; built-in circuit
breaker
Sample temperature
Ambient temperature
Sample pressure
(in iOS System)
Maximum relative
humidity
1 ºC to 95 ºC**
10 ºC to 40 ºC
Up to 250 psig**
Up to 95%, noncondensing
Maximum altitude 3,000 m (9,843 ft)
Installation/Overvoltage
category
Serial (RS-232), USB, parallel printer port,
4-20 mA output, two alarms, binary output
II
Calibration stability Typically stable for 12 months
Chemical reagents
(prepackaged)
300 mL acid; 300 mL or 150 mL oxidizer
Safety certicates CE, ETL listed. Conforms to UL Std. 61010-1
Certied to CSA C22.2 No. 61010-1
Pollution degree 2
Display Color, touch-sensitive LCD
Weight 16.9 kg (37.2 Ibs)
* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
** If the sample temperature and pressure are above 60 ºC and 100 psi, the 900 and 5310 C On-Line Analyzer models with the stainless steel iOS is required.
Outputs
Step 6 C Install the Analog outputs and alarms
(continued).
Step 9
Flush the Analyzer
Before placing the Analyzer into normal operation, you should
perform a reagent ush to remove any gas bubbles that may
have formed in the reagent lines; you should then leave the
Analyzer powered on for 12 hours to allow the DI water reservoir
to properly complete rinse down action.
Size 62.4 cm height x 45.2 cm width x 26.4 cm
depth (24.6 in x 17.8 in x 10.4 in)
Required sample line
ow rate
30 mL - 300 mL/min (for On-Line Mode)
Normal operating
environment
Intended for indoor use only; avoid
sunlight IP Rating IP 45
Step 7
Connect the inlet system and waste outlet.
1. Connect the 1/4” Teon tubing with the capsule or inline
lter to the sample inlet on the iOS System. Tighten 1/4 turn
past nger-tight with a 9/16” open-end wrench. Do not over-
tighten the nut.*
* For raw water and heavy particulates applications, additional
pre-ltration is required. Please contact GE Analytical Instruments
if you need recommendations.
2. Connect the 3/4” OD waste line tubing to the waste outlet on
the iOS System by sliding the tubing over the barb tting.
Table 3: Serial and 4-20 mA
Inputs and Outputs (TB3)
12 4-20 mA ( - output)
11 4-20 mA ( + output)
10 Not used
9 Not used
8 Reserved
7 Reserved
6 Reserved
5 Not used
4 Serial (In)
3 Serial (Out)
Pin Number
(from left)
Output
2 Serial (Ground)
1 Not used
Step 8
Congure basic Analyzer settings.
Before using the Analyzer, you will need to congure various basic
settings. Some of these settings will not need to be changed again,
unless you move the Analyzer or recongure the operational environ-
ment. Refer to the 900 or 5310 C On-Line Operation and Maintenance
Manual for complete details.
Step 8 B Enable password protection.
Step 8 C Set the clock and time zone.
Step 8 D Name the Analyzer location (optional).
Step 8 E Set reagent expiration dates.
Step 8 F Set up the data history.
Step 8 G Set up the printer (pptional).
Step 8 H Export or print constants.
Step 8 A Power the Analyzer on.
Step 8 I Set up data I/O.
©2016. General Electric Company. All Rights Reserved. DQS 96003-01 Rev A
Americas: T 800 255 6964, 303 444 2009 | [email protected]
Europe/Middle East/Africa: T 44 (0) 161 864 6800 | [email protected]
Asia Pacic: [email protected]
www.geinstruments.com
For more detailed instructions, you can download operation and maintenance manuals for the Sievers 900 On-Line Analyzer, the Sievers
5310 C On-Line TOC Analyzer, and the 900 Inorganic Carbon Remover (ICR) from our Web site. Go to www.geinstruments.com, click Library,
then Manuals, and make your selection.
Where to Find Operation and Maintenance Manuals
Sievers 900 and 5310 C Series On-Line TOC Analyzer Installation Instructions
To avoid exposure to the chemical reagents, wear
acid-resistant gloves, protective clothing, and safety
goggles or a face shield when changing the reagent
supplies.
Reagent containers are for single-use only –
DO NOT REFILL. Relling or reusing reagent containers
will void all Analyzer and parts warranties and nullify
any performance claims.
Operation of the Analyzer without the in-line lter
on the sample inlet line will damage the Analyzer
and void the warranty. To avoid damaging the
Analyzer, install the lter and replace the lter
element as needed.
See the separate “Warnings” document for
additional important information. Step 1
Unpack and inspect the Analyzer.
Step 2
Complete the identication records.
Step 3
Select a location for the Analyzer. Refer to the System
Specications chart.
WARNINGS
To avoid false TOC readings and possible damage to
the Analyzer, always make sure the sample is owing
through the iOS System and the DI water reservoir is
lled before starting analysis.
Figure 1 - Reagent cartridges
NOTE: Hazardous reagents (ammonium persulfate
and phosphoric acid) are used in the Analyzer.
Before installing the reagents, please read the
Material Safety Data Sheets (MSDS) contained in
a pouch on the top of the reagent shipping box
for proper handling precautions and spill or leak
procedures.
Step 5
Fill the DI water reservoir.
1. Remove the rubber inlet cover on the DI water reservoir.
2. Slide the bottle nozzle into the inlet hole, and gently squeeze the
bottle. Be sure to ll the reservoir until the water reaches the ll
line, just below the inlet.
Step 4
Install the reagent cartridges. (See Figure 1.)
The oxidizer and acid reagents are shipped from GE Analytical
Instruments in specic packaging for safety in transit. Carefully
read the attached MSDS sheets prior to opening the packaging.
Within the packaging are two reagent cartridges. Each reagent
cartridge is identied with labels that indicate which cartridge
contains ammonium persulfate and which cartridge contains
phosphoric acid.
1. Locate the acid supply line, indicated by the Acid label. Attach
the PEEK nut to the acid reagent container and tighten nger-
tight.
2. Slide the valve on the acid container to the open position by
pushing the green button all the way in. Refer to the labeling on
the container for proper positioning.
Step 6
Install accessories and cables.
If you will be using a 900 ICR unit with your conguration, make
sure it is installed. Follow the complete set of operation instructions
in the ICR Operation and Maintenance Manual. (See downloading
instructions on the back page of this Quick Start Guide.)
Step 6 A Install the 900 Inorganic Carbon Remover (ICR).
Installation of the Sievers 900/5310 C On-Line TOC Analyzer requires an
external source of AC power connected to the enclosure using a water-
tight conduit connector. The electrical connection should be performed
by a qualied electrician. An external switch or circuit breaker is recom-
mended to facilitate maintenance and servicing of the Analyzer. It should
be installed near the Analyzer and be clearly marked as the disconnecting
device for the Analyzer. (See Figure 2.)
NOTE: Before installing any wiring inside the Analyzer, put on a
grounding strap for ESD protection.
Step 6 B Connect to a power supply.
The Analyzer has three terminal blocks, arranged horizontally
in the electrical enclosure inside the Analyzer; remove the
enclosure cover by loosening the front screws. Consult Tables
1–3 for a list of functions on each terminal block. The output
and alarm connections should be installed by a qualied
electrician. Refer to the 900 or 5310 C On-Line Operations and
Maintenance Manual for complete details.
Step 6 C Install the analog outputs and alarms.
3. Slide the acid container into the reagent enclosure, with the
valve pointing down and toward the front of the reagent
enclosure.
4. Locate the oxidizer supply line, indicated by the Oxidizer
label. Attach the PEEK nut to the oxidizer container and
tighten nger-tight.
5. Slide the valve on the oxidizer container to the open
position by pushing the green button all the way in. Refer
to the labeling on the container for proper positioning.
6. Slide the oxidizer container into the reagent enclosure, with
the valve pointing down and toward the front of the reagent
enclosure.
Table 1: Inputs and Outputs (TB2)
10 Ground, for binary input
9 24 V (+ output, for binary input)
8 Start/Stop - (for binary input)
7 Start/Stop + (for binary input)
6 Alarm 2 (NO*) (output)
5 Alarm 2 (NC*) (output)
4 Alarm 2 (Common) (output)
3 Alarm 1 (NO*) (output)
2 Alarm 1 (NC*) (output)
1 Alarm 1 (Common) (output)
* NC = normally closed *NO = normally open
Pin Number
(from left)
Input/Output
Table 2: Inputs and Outputs (TB1)
8 900 Lab: Reserved
5310 C Lab: Stream 1 - (input) 0
7 900 Lab: Reserved
5310 C Lab: Stream 1 + (input)
6 End of Batch (NO*) (output)
5 End of Batch (NC*) (output)
4 End of Batch (Common) (output)
3 Reserved
2Reserved
1 Reserved
* NC = normally closed *NO = normally open
Pin Number
(from left)
Input/Output
Figure 2 - Wiring AC Power Conduit (Arrows indicate connection points.)

This manual suits for next models

1

Other GE Measuring Instrument manuals

GE XL Vu+ VideoProbe User manual

GE

GE XL Vu+ VideoProbe User manual

GE Phasor XS User manual

GE

GE Phasor XS User manual

GE Druck ADTS 405 User manual

GE

GE Druck ADTS 405 User manual

GE Sievers Innovox User manual

GE

GE Sievers Innovox User manual

GE Druck DPI 705 User manual

GE

GE Druck DPI 705 User manual

GE Aquant User manual

GE

GE Aquant User manual

GE PACE 1000 Instruction Manual

GE

GE PACE 1000 Instruction Manual

GE N60 User manual

GE

GE N60 User manual

GE Protimeter MMS2 User manual

GE

GE Protimeter MMS2 User manual

GE TransPort PT878GC User manual

GE

GE TransPort PT878GC User manual

GE Precision 500D User manual

GE

GE Precision 500D User manual

GE AquaTrans AT600 User manual

GE

GE AquaTrans AT600 User manual

GE RHM 12 User manual

GE

GE RHM 12 User manual

GE Phasor CV User manual

GE

GE Phasor CV User manual

GE Multilin EPM 2000 Series User manual

GE

GE Multilin EPM 2000 Series User manual

GE PM880 User manual

GE

GE PM880 User manual

GE EPM6000 User manual

GE

GE EPM6000 User manual

GE D60 User manual

GE

GE D60 User manual

GE SEER 1000 User manual

GE

GE SEER 1000 User manual

GE PanaFlow HT User manual

GE

GE PanaFlow HT User manual

GE TransPort PT900 User manual

GE

GE TransPort PT900 User manual

GE Mentor Visual iQ CCF28 User manual

GE

GE Mentor Visual iQ CCF28 User manual

GE DB-16 User manual

GE

GE DB-16 User manual

GE Ultrasonic USM Go User manual

GE

GE Ultrasonic USM Go User manual

Popular Measuring Instrument manuals by other brands

RF Technology Code Alert CA630 user guide

RF Technology

RF Technology Code Alert CA630 user guide

Delta OHM HD2010UC manual

Delta OHM

Delta OHM HD2010UC manual

Deltatherm DTC120-G Quick install guide

Deltatherm

Deltatherm DTC120-G Quick install guide

PerkinElmer Lambda 365 installation instructions

PerkinElmer

PerkinElmer Lambda 365 installation instructions

Delta LTL2000S user manual

Delta

Delta LTL2000S user manual

Delta-T Devices WET150 quick start guide

Delta-T Devices

Delta-T Devices WET150 quick start guide

LineEye LE-8500X quick start guide

LineEye

LineEye LE-8500X quick start guide

Unit UT204R user manual

Unit

Unit UT204R user manual

UseTECH FlowTimer+ Set XL operating manual

UseTECH

UseTECH FlowTimer+ Set XL operating manual

Tek-Trol Tek-Clamp 1200A instruction manual

Tek-Trol

Tek-Trol Tek-Clamp 1200A instruction manual

Unit UT201+ user manual

Unit

Unit UT201+ user manual

B meters HYDROCLIMA 2 installation manual

B meters

B meters HYDROCLIMA 2 installation manual

Almaco FLEXICOMPT AUTONOME+ operating guide

Almaco

Almaco FLEXICOMPT AUTONOME+ operating guide

Rohde & Schwarz FSW43 user manual

Rohde & Schwarz

Rohde & Schwarz FSW43 user manual

brennenstuhl PM 240 E Operation and safety notes

brennenstuhl

brennenstuhl PM 240 E Operation and safety notes

POSITECTOR 6000 manual

POSITECTOR

POSITECTOR 6000 manual

YOKOGAWA PZ4000 user manual

YOKOGAWA

YOKOGAWA PZ4000 user manual

Rae MultiRAE user guide

Rae

Rae MultiRAE user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.