GE AMX-II User manual

GE
Medical
Systems
Technical
Publications
Service
Manual
SM
A0654E
Revision
3-1684
AMX-IT
X-Ray
Unit
Product
Operating
Documentation

DAMAGE
IN
TRANSPORTATION
All
packages
should
be
closely
examined
at
time
of
delivery.
If
damage
is
apparent,
have
notation
of
“bad
order”
placed
by the
delivering
driver
on
ail
copies
of
the
freight
or
express
bill.
If
damage
is
of
a
concealed
nature,
notify
transportation
agent
as
soon
as
possible
to
make
an
“inspection
report
of
damage”
but
in
any
event
not
later
than
15
days
after
delivery.
A
trans-
portation
company
usually
will
not
pay
a
claim
for
concealed
damage
if
an
inspection
is
not
requested
within
this
15
day
period.
Complete
instructions
regarding
claim
procedure
are
found
in
section
‘’$’’
of
the
Policy
&
Proce-
dure
Bulletins.
If
shipment
was
handled
by
moving
van
service
-
uncrated
-
call
Traffic
-
Milwaukee
immediately
when
any
damage
is
found.
Do
not
attempt
to
call
any
local
agent.
At
this
time
be
ready
to
describe
type
of
damage,
type
of
equipment,
serial
numbers
and
if
possible
the
order
number.
The
above
paragraph
is
in
regard
to
equipment
requiring
installation
only,
and
does
not
apply
to
supply
items.
The
F.0.B.
point
for
these
items
is
as
shown
in
the
Price
Book.
GE
Medical
Systems
GE
Medical
Systems:
|
Telex
3797371
20
Box
414,
Milwaukee,
Wisconsin
53201
USA
(Asia,
Pacific,
latin
America,
North
America)
GE
Medical
Systems—Europe:
Telex
261794
Shortlands,
Hammersmith,
London
WE
SEX
UK

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-H
X-RAY
UNIT
3-1684
SM
A0654E
TABLE
OF
CONTENTS
CHAPTER
1
es
OPERATING
INSTRUCTIONS
OM
A0654E
CHAPTER2
..........................
DOES
NOT
APPLY
CHAPTER
3
..........................
INSTALLATION
CHAPTER
4
.........................
ADJUSTMENT
AND
CALIBRATION
CHAPTER5
..........................
FUNCTIONAL
CHECK
CHAPTER
6
..........................
THEORY
AND
SERVICE
СНАРТЕВ
7
.............
——
_
SCHEMATICS
CHAPTER8
.........................
PREVENTIVE
MAINTENANCE
CHAPTER
9
.........................
RENEWAL
PARTS


GE
Medical
Systems
Technical
Publications
OPERATING
MANUAL
OM
A0654E
3-1684
АМХ-П
X-Ray
Unit
Product
Operating
Documentation


OPERATING
MANUAL
OM
A0654E
3-1684
AMX-II
X-Ray
Unit
IMPORTANT!...
X-RAY
PROTECTION
X-ray
equipment
if
not
properly
used
may
cause
injury.
Accord-
ingly,
the
instructions
herein
con-
tained
should
be
thoroughly
read
and
understood
before
you
attempt
to
place
this
equipment
in
operation.
The
General
Electric
Company,
Medical
Systems
Group,
will
be
glad
to
assist
and
cooperate
in
placing
this
equip-
ment
in
use.
Although
this
apparatus
incor-
porates
a
high
degree
of
pro-
tection
against
x-radiation
other
than
the
useful
beam,
no
prac-
tical
design
of
equipment
can
provide
complete
protection.
Nor
can
any
practical
design
compel
the
operator
to
take
adequate
precautions
to
prevent
the
possi-
bility
of
any
persons
carelessly,
unwisely,
or
unknowingly
expos-
ing
themselves
or
others
to
radiation.
it
is
important
that
everyone
hav-
ing
anything
to
do
with
x-radia-
tion
be
properly
trained
and
fuily
acquainted
with
the
recommen-
dations
of
the
National
Council
on
Radiation
Protection
and
Mea-
surements
as
published
in
NCRP
Reports
available
from
NCRP
Publications,
7910
Woodmont
Avenue,
Room
1016,
Bethesda,
Maryland
20814,
and
of
the
Inter-
national
Commission
on
Radia-
tion
Protection,
and
take ade-
quate
steps
to
insure
protection
against
injury.
All
persons
authorized
to
use
the
equipment
must
be
cognizant
of
the
danger
of
excessive
exposure
to
x-radiation
and
the
equipment
is
sold
with
the
understanding
that
the
General
Electric
Com-
pany,
Medical Systems
Group,
its
agents,
and
representatives
have
no
responsibility
for
injury
or
damage
which
may
result
from
exposure
to
x-radiation.
Various
protective
material
and
devices
are
available.
It
is
urged
that
such
materials
or
devices
be
used.


GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-II
X-RAY
UNIT
3-1684
OM
A0654E
VERTICAL
COLUMN
X-RAY
TUBE
UNIT
HV
CABLES
SENTRY
COLLIMATOR
TELESCOPING
ARM
X-RAY
CONTROL
ACCESS
COVER
CASSETTE
TRAY
SAFETY
BUMPER
23561-P
10
AMX-II
X-RAY
WITH
TUBE
UNIT
AND
SENTRY
COLLIMATOR
ILLUSTRATION
1-1


GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-il
X-RAY
UNIT
1-981
OM
A0654E
AMX-11
X-RAY
UNIT
FOREWORD
The
AMX-II
X-Ray
Unit
consists
of
a
full
wave
500
Hz
single
phase
x-ray
generator
control
and
transformer,
rechargeable
battery,
tube
support
column
with
x-ray
tube
unit
and
collimator,
all
mounted
on
a
self-
propelled
mobile
base.
Operation
of
the
AMX-Il
X-Ray
Unit
consists
of
driving
the
mobile
unit,
positioning
and
making
an
x-ray
exposure,
and
recharging
the
battery.
An
understanding
of
the
specific
operative
steps
and
various
switch,
signaling,
and
lockout
devices
is
essential.
Therefore,
these
directions
should
be
read
carefully
and
thorough-
ly
understood
before
attempting
to
operate
the
AMX-il
X-Ray
Unit.
COMPONENT
MODEL
NUMBERS
NAMEPLATE
LOCATIONS
Basic
Unit,
Control
Mobile
Base,
Support
Column
115V
50/60
Hz
46-165600G
10,
G12,
G14,
Beneath
exposure
control
G15
Handswitch
bracket
on
front
of
unit.
230V
50/60
Hz
46-165600G11
X-Ray
Tube
Unit
46-155750G2,
G3,
G4
Cathode
trim
cap
housing
Rear
Trim
Cover
Collimator
46-177399G1-G2
or
Bottom
-
Rear
of
46-194759G1
Trim
Cover
HV
Transformer
46-150625G1
On
side
of
transformer
beneath
terminal
strip
Battery
(cell
group)
46-152812P
1
End
of
cell
group
case
Extension
Cylinder
(optional)
46-138573G1
On
body
of
cylinder
WARNING
Explosion
Hazard
-
Do
not
use
in
the
presence
of
flammable
anesthetics.
This
X-ray
equipment
is
dangerous
to
both
patient
and
operator
unless
safe
exposure
factors
are
observed.
Sold
only
to
be
used
by
or
on
the
prescription
of
a
licensed
physician.

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-il
X-RAY
UNIT
1-981
OM
A0654E
SAFETY
PRECAUTIONS
RADIATION:
Always
consider
providing
radiation
protection
for
your
patient
and
yourself
when
operating
the
AMX-II.
GOOD
OPERATING
PRACTICES
©
Wear
a
lead
apron
while
operating
the
x-ray
controls
of
the
AMX-II.
e
Step
back
at
least
6
ft.
(1.8m)
from
the
AMX-II
or
to
the
full
extension
of
the
handswitch
cord
before
making
an
exposure.
e
Always
use the
proper
field
sizes
and
technic
factors
for
each
procedure
to
minimize
x-ray
exposure
and
produce
the
best
diagnostic
results.
©
When
x-raying
bed
patients,
move
them
as
far
as
possible
from
nearby
beds.
e
Ask
visitors
to
step
outside
the
room
during
an
exposure.
e
Use
gonadal
shields
for
patients
whenever
possible.
EXPOSURE
PROFILE
Illustration
1-1A
shows
the
iso-exposure
profile
of
the
AMX-II.
This
profile
shows
the
typical
scattered
radiation
levels
around
an
AMX-H
using
chest
radiography
technics.
NOTE:
This
profile
is
only
an
indication
of
typical
scattered
radiation
levels.
Variations
in
equipment
performance,
patient
size,
technic,
geometry,
and
radiation
measuring
devices
may
cause
the
profile
to
vary
slightly
from
the
one
shown
here.
0.1
mR/exposure
1.0
mR/exposure
Contours
are
at
eye
level
for
a
5
ft.
8
inch
(1.73m)
person
Bone
and
tissue
equivalent
rubber
phantom
80
kVp,
5
mAs
0-5
mR
/exposure
40
în.
SID,
14
x
17
in.
field
>
A
κ
SCALE
X
x
33121-010
0.5
1.0
Ft
AMX-II
ISO-EXPOSURE
PROFILE
ILLUSTRATION
1-1A

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-II
X-RAY
UNIT
1-981
OM
A0654E
Under
most
conditions,
cumulative
radiation
dose
to
the
operator
will
not
exceed
recommended
maximum
permissible
levels.
However,
as
with
all
radiation-producing
devices,
a
gualified
radiation
expert
should
evaluate
situations
involving
frequent
exposures
using
high
kVp
and
mAs
technics
to
determine
if
extra
protective
devices
are
necessary.
/
NOTE:
Recommended
maximum
permissible
dose
to
the
eyes
is
100
mrem/week
(NCRP
no.
39,
p.
89),
where
a
dose
of
1
mrem
is
produced
by
an
exposure
to
1
mR
of
x-radiation.
ELECTRICAL
SHOCK
WARNING:
BATTERY
WARNING:
An
audible
tone
and/or
red
X-RAY
light
indicates
presence
of
radiation.
If
either
indicator
is
on
at
any
time
other
than
during
the
expected
X-RAY
exposure
interval,
immediately
release
the
handswitch
ROTOR
and/or
X-RAY
button
and
turn
off
main
circuit
breaker.
Do
not
energize
the
unit
again
until
it
has
been
checked
by
a
qualified
x-ray
serviceman.
This
unit
contains
a
120
volt
rechargeable
battery
inside
the
cabinet.
It
can
cause
electrical
shock
even
when
the
unit
is
not
connected
to
a
power
outlet.
"The
x-ray
control
access
door,
shown
in
Illustration
1-1,
can
be
opened
only
by
removing
the
cover.
Because
of
the
possible
electrical
shock
and
other
hazards
present
when
this
cover
is
removed,
it
should
be
removed
only
by
a
trained
x-ray
serviceman
when
the
unit
requires
servicing.
Unauthorized
persons
should
not
remove
the
cassette
tray
shown
in
Illustration
1-1.
The
cassette
tray
should
not
be
removed
except
by
or
in
the
presence
of
a
person
thoroughly
familiar
with
safe
nickel-cadmium
battery
service
pro-
cedures
given
in
Chapter
8
of
the
AMX-II
X-Ray
Unit Service
Manual.
Should
a
metallic
object
come
in
contact
with
the
battery
terminals,
a
small
one
would
vaporize
and
a
large
one
would
become
very
hot.
If
the
object
is,
for
example,
a
ring
or
a
wrist
watch,
it
may
weld
to
a
terminal
and
the
wearer
could
be
severely
burned
or
injured
by
moiten
metal.
The
Electrolyte
used
in
nickel-cadmium
batteries
is
a
caustic
solution
of
potassium
hydroxide.
Serious
burns
will
result
if
it
comes
in
contact
with
any
part
of
the
body.
If
Electrolyte
gets
on
the
skin,
wash
the
affected
areas
with
large
quantities
of
water;
then
neutralize
skin
with
vinegar,
lemon
juice,
or
3%
acetic
acid.
If
Electrolyte
gets
into
the
eyes,
flush
with
water
and
GET
IMMEDIATE
MEDICAL
ATTENTION.
When
recharging
the
battery,
the
x-ray
unit
must
be
located
in
a
large
ventila-
ted
room.
During
normal
charging
conditions
and
when
all
batteries
are
func-
tioning
properly,
there
is
little
or
no
gas
emitted.
Should
abnormal
conditions
develop,
there
could
be
an
accumulation
of
hydrogen
gas
which
if
confined
in
a
small
space
would
be
explosive
in
the
presence
of
a
flame
or
electrical
spark.
Whenever
the
batteries
are
being
charged,
a
fan
ventilates
the
battery
compart-
ment.
A
hazardous
situation
will
not
develop
where
the
overflow
from
this
compartment
is
into
a
large
room
or
a
well
ventilated
small
room.
3

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-Il
X-RAY
UNIT
1-981
OM
A0654E
UNIT
ON
CHARGE
ON
CHARGE
LAMP
LAMP
COMPLETE
LAMP
BATTERY
CONDITION
INDICATOR
KEYED
X-RAY
METER
GENERATOR
STANDBY
BATTERY
SWITCH
VOLTAGE
COMPENSATOR
KVP
SELECTOR
SWITCH
MAS
SELECTOR
SWITCH
“>
X-RAY
ON
LAMP
READY
LAMP
=
~ —
MAIN
POWER
hs
circuit
DIRECTION
ee
na
BREAKER
SELECTOR
ーー
パテ
ーー
с
e
UT
BRAKE
SWITCH
\
一
HANDLE
га
DRIVE
ASSIST
о
y
—
SPEED
BUTTON
ILLUSTRATION
1-2
23563210
CONTROL
FUNCTIONAL
DESCRIPTION
The
MAIN
POWER
CIRCUIT
BREAKER
disconnects
the
Battery.
It
must
be
ON
in
order
to
charge
the
unit,
drive
it
from
one
location
to
another,
and
make
exposures.
The
keyed
X-RAY
GENERATOR
STANDBY
SWITCH
is
turned
to
OFF
when
the
AMX-II
is
being
driven
or
is
not
in
use,
to
ON
for
exposures,
and
to
CHARGE
when
charging
the
batteries.
A
key
is
provided
for
this
switch
to
assure
that
only
authorized
personnel
operate
the
unit.
The
key
should
be
taken
out
of
the
socket
and
retained
by
the
designated
operator
immediately
after
a
scheduled
period
of
x-ray
operation.
The
BATTERY
CHARGER
recharges
the
internal
battery
from
any
115v
or
230v,
50-60
Hz
power
outlet
receptacle.
The
unit
may
not
be
driven
or
an
x-ray
exposure
cannot
be
made
when
the
battery
charger
plug
is
in
a
power
outlet
receptacle.
The
x-ray
unit
battery
should
be
charged
(as
described
under
OPERATION)
whenever
the
BATTERY
CONDITION
needle
enters
the
red
band.
It
should
be
allowed
to
charge
until
the
amber
CHARGE
COM-
PLETE
lamp
giows,
and
the
BATTERY
CONDITION
indicator
needle
is
in
the
green
band.
Typically,
the
battery
will
charge
from
too
low
(red
band)
to
full
charge
in
less
than
10
hours.
Since
the
charger
is
self-regulating,
it is
permissible
to
have
it
connected
to
power
indefinitely.
CAUTION:
ALWAYS
LOCATE
THE
UNIT
IN
A
WELL
VENTILATED
ROOM
WHEN
CHARGING
IT.
The
SPEED
CONTROL
will
vary
the
speed
of
the
unit
in
forward
or
reverse.
When
released,
it
will
return
to
the
zero
speed
position.
The
SAFETY
BUMPER,
located
on
the
front
of
the
unit,
will
disconnect
the
drive
motor
and
stop
the
unit
upon
impact
with
any
object.
The
DRIVE
ASSIST
BUTTON,
when
pressed,
will
increase
the
low
speed
torque
of
the
drive
motor.
It
is
a
spring
loaded
button
which
should
be
used
when
climbing
an
incline
or
when
moving
over
the
edge
of
carpeting
or
other
small
floor
discontinuities.

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
АМХ-И
X-RAY
UNIT
1-981
|
OM
A0654E
COLUMN
EXPOSURE
CONTROL
HAN
DSWITCH
BATTERY
CHARGER
CORD
&
PLUG
ILLUSTRATION
1-3.
23564P10
The
BRAKE
HANDLE
is
used
to
steer
the
unit,
and
automatically
engages
the
brakes
and
disconnects
the
drive
motor
when
released.
The
BATTERY
VOLTAGE
COMPENSATOR
SWITCH
adjusts
for
variations
in
battery
charge.
When
the
battery
meter
is
in
the
HIGH
region,
the
switch
should
be
in
the
HIGH
position.
As
the
battery
charge
is
reduced
to
the
LOW
region,
the
switch
should
be
in
the
LOW
position.
After
battery
charging
is
completed,
allow
the
following
condition
for
at
least
15
minutes
before
making
exposures:
Keep
the
MAIN
POWER
BREAKER
at
ON.
If
the
line
cord
is
left
connected,
set
the
KEY
SWITCH
at
CHARGE.
However,
if
the
line
cord
is
disconnected,
set
the
KEY
SWITCH
at
ON.
The
Remote
Exposure
Contro!
Handswitch
(Illustration
1-3)
has
ROTOR
and
X-RAY
buttons.
Provided
that
the
red
UNIT
ON
fight
is
glowing,
depressing
the
rotor
button
will
cause
x-ray
tube
anode
rotation
and
filament
preheating.
There
will
be
a
2.5
second
delay
for
these
functions
to
reach
the
operating
point.
After
that
delay,
the
white
READY
lamp
will
glow.
Then
the
x-ray
exposure
will
occur
immediately
after
the
handswitch
X-RAY
button
is
depressed.
If
the
ROTOR
and
X-RAY
buttons
are
depressed
simultaneously,
the
x-ray
exposure
will
occur
after
the
2.5
second
delay
and
as
the
READY
lamp
flashes
on.
In
order
to
obtain
an
exposure,
the
MAIN
POWER
CIRCUIT
BREAKER
must
be
ON,
the
X-RAY
GENERATOR
STANDBY
SWITCH
must
be
ON,
and
the
charger
cord
disconnected
from
the
power
outlet.
Glow
of
the
red
X-RAY
lamp
indicates
that
x-rays
are
being
emitted.
When
the
READY
lamp
is
glowing,
the
X-RAY
lamp
will
glow
immediately
after
the
handswitch
X-RAY
button
is
depressed
and
it
will
remain
5

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
|
|
|
AMX-il
X-RAY
UNIT
1-981
OM
A0654E
ARM
INTERLOCK
TELESCOPING
YOKE
ROTATION
TUBE
UNIT
COLUMN
BUTTON
ARM
LOCK
ROTATION
LOCK
X-RAY
.
TUBE
UNIT
LATCH
BAR
A
o
VERTICAL
LATCH
ILLUSTRATION
1-4
ARM
TELESCOPING
YOKE
COLLIMATOR
LOCK
ARM
LOCK
23565-P
10
on
for
the
time
interval
equal
to
the
MAS
selected
divided
by
100.
All
exposures
with
the
AMX
I!
will
be
at
100
milliamperes.
Therefore,
exposure
time
may
be
determined
by
dividing
the
MAS
selection
by
100.
For
example,
if
300
MAS
is
selected,
the
lamp
will
glow
for
three
seconds
or
if
2.0
MAS
is
selected
it
will
glow
for
0.02
seconds.
During
the
time
that
the
X-Ray
lamp
glows,
there
will
also
be an
audible
tone
signal
to
indicate
that
x-rays
are
being
produced.
At
the
low
MAS
selections,
this
signal
will
be
a
short
beep.
For
longer
exposures
there
will
be
a
sustained
tone.
WARNING:
|
If
either
the
X-RAY
lamp
glow
or
the
tone
signal
continues
longer
than
3
seconds under
any
circumstances,
immediately
release
handswitch
ROTOR
and
X-RAY
buttons
and
turn
the
main
power
circuit
breaker
OFF.
In
such
a
case,
another
exposure
should
not
be
attempted
until
the
unit
has
been
examined
by
a
qualified
x-ray
serviceman.
The
Vertical
Column
supports
the
telescoping
arm,
x-ray
tube
unit
and
collimator.
It
rotates
in
a
bearing
structure
in
the
base
of
the
unit.
To
rotate
the
vertical
column,
move
the
telescoping
arm
to
the
desired
position.
360°
of
movement
is
possible.
A
friction
brake
holds
the
vertical
column
in
place.
If
the
column
drifts
after
positioning,
call
for
service.
The
VERTICAL
ARM
LOCK,
Illustration
1-4,
will
secure
the
telescoping
arm
in
any
position.
If
the
arm
drifts
when
the
lock
is
released,
call
for
service.
If
the
counterbalance
system
should
fail,
a
safety
device.
locks
the
arm
in
position,
and
the
unit
must
be
serviced
before
it
can
be
used
again.
6

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-Il
X-RAY
UNIT
1-981
OM
A0654E
ADDITIONAL
E
1
and
2
MM
A
κ
ENS
mm
”
n
ROTATIONAL
DETENT
ALUMINUM
1
Ču
m
PLUNGER
KNOB
SELECTIVE
(ON
RIGHT
HAND
SIDE)
FILTERS
``
CALIBRATED
SCALE
DRUM
[~~
FIELD
LIGHT
BUTTON
|
i
i
[OX
BLADE
CONTROL
KNOBS
TRACKS
FOR
y=
一
一
一
EXTENSION
CYLINDER
————
MEASURING
TAPE
16942-P10
ILLUSTRATION
1-5
The
Telescoping
Arm
consists
of
two
movable
sections
and
one
stationary
section.
The
sections
extend
and
retract
equally
as
the
arm
is
repositioned.
The
TELESCOPING
ARM
LOCK,
located
beneath
the
middle
section,
will
secure
the
telescoping
arm
in
any
position.
The
YOKE
ROTATION
LOCK
allows
the
tube
to
be
rotated
+/-180°
about
the
horizontal
axis
and
locked
in
any
position.
The
ARM
INTERLOCK
BUTTON
will
maintain
the
telescoping
arm
in
the
maximum
retracted
position.
When
driving
the
AMX-!|
from
room
to
room,
the
telescoping
arm
is
secured
by
means
of
the
LATCH
and
LATCH
BAR.
The
telescoping arm
must
be
in
the
position
shown
in
Illustration
1-1.
The
LATCH
at
the
bottom
of
the
telescoping
arm
automatically
latches
to
the
LATCH
BAR
when
they
are
joined.
To
release,
rotate
the
latch
clockwise
and
raise
the
telescoping
arm.

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-ll
X-RAY
UNIT
1-981
OM
A0654E
The
yoke
supporting
the
Tube
Unit
can
be
rotated
180
degrees
about
the
horizontal
axis
in
either
direction
of
the
central
position
(x-ray
port
pointing
straight
downward).
It
can
be
positively
retained
in
any
inter-
mediate
position
by
the
YOKE
ROTATION
LOCK.
The
TUBE
UNIT
ROTATION
LOCK,
located
on
the
yoke
clamp,
locks
or
releases
the
Tube
Unit
for
rotation
in
the
yoke clamps
toward
or
away
from
the
column.
Two
index
lines,
one
on
the
Tube
Unit
and
one
on
the
yoke
clamp,
may
be
used
to
place
the
central
x-ray
beam
axis
in
the
vertical
plane.
The
X-Ray
Tube
Unit
is
an
HRT
-
O9N
with
a
1.0
mm
focal
spot.
The
Sentry
Manuai
Collimator
is
used
with
the
AMX-!|
X-Ray
Unit.
The
collimator
limits
patient
rad-
iation
exposure
to
a
desired
area
at
a
given
distance
from
the
x-ray
tube
focal
spot.
Minimum
Source-to-Skin
distance
is
limited
to
more
than
30
centimeters
by
the
tracks
on
the
bottom
of
the
collimator
for
mounting
the
accessory
Extension
Cylinder.
Area
of
coverage
is
indicated
by
light
from
a
GE
Quartz!ine®
high-intensity
lamp.
Distinct
shadow
cross
hairs,
which
indicate
the
field
center,
aid
in
accurate
radiographic
positioning.
The
collimator
has
a
fixed
minimum
filtration
equivalent
to
1.5mm
of
aluminum
measured
at
150
kvp,
so
that
a
total
filtration
of
2.5mm
of
aluminum
equivalent
results
from
the
tube-coilimator
combination.
By
inserting
the
1mm
and/or
2mm
filter
in
the
slot
on
the
left
side
of
the
collimator,
additional
selective
filtration
of
1,
2,
or
3mm
of
aluminum
equivalent
may
be
obtained.
Dial
numbers
on
the
Calibrated
Scale
Drum,
Illustration
1-5,
are
for
40,
48,
60,
and
72
inch
source-to-
image
distances
and
standard
film
sizes
of
5
through
17
inches
(13
cm
through
43
cm).
To
select
the
scale
for
one
of
the
four
source-to-image
distances,
rotate
the
drum
by
turning
the
SCALE
SELECTION
KNOB
until
the
desired
detent
position
is
reached.
Two
moving
pointers
sweep
across
the
scale
upon
actuation
of
the
BLADE
CONTROL
KNOBS
permitting
easy
selection
of
the
field
size.
The
upper
pointer
indicates
the
transverse
dimension
while
the
lower pointer
indicates
the
longitudinal
dimension.
Independent
adjustment
is
provided
by
the
BLADE
CONTROL
KNOBS
on
the
face
of
the
collimator.
They
allow
collimating
to
any
desired
rectangular
field
size
up
to
17”
x
17”
(43
cm
x
43
cm)
at
a
40”
source-to-image
distance.
The
right-hand
knob
controls
the
transverse
dimension,
while
the
left-hand
knob:
controls
the
longitudinal
dimension,
while
the
left-hand
knob
controls
the
longitudinal
dimension.
Depressing
the
FIELD
LITE
BUTTON
energizes
the
high-intensity
lamp.
A
timer
automatically
switches
the
light
off
in
approximately
15
to
20
seconds
to
keep
lamp
operation
at
a
minimum.
This
time
interval
can
be
adjusted
by
the
serviceman
if
the
user
so
desires.
NOTE:
An
interlock
allows
exposures
and turns
off
the
field
light
when
x-ray
rotor
handswitch
is
en-
ergized.
The
collimator
can
be
rotated and
locked
positively
at
0°,
+/-
159,
+/-45°,
and
+/-
90°
from
the
front
position
by
detent.
Depress
the
ROTATIONAL
DETENT
PLUNGER
KNOB
on
the
right
side
of
the
collimator
to
move
it
from
a
detent
position.
Keep
the
plunger
knob
depressed
until
the
approximate
position
desired
is
reached.
Then
release
the
button
and
rotate
the
collimator
slightly
about
this
position
until
the
lock
snaps
in
place.

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-il
X-RAY
UNIT
1-981
.
OM
A0654E
A
Measuring
Tape
is
provided
on
the
center
line
of
the
bottom
of
the
collimator
housing.
It
is
calibrated
in
inches
and
centimeters
for
measuring
source-to-image
distances.
The
correct
source-to-image
distance
is
indicated
by
having
the
bottom
of
the
collimator
housing
even with
the
marking
on the
tape
which
corresponds
to
the
desired
source-to-image
distance.
For
accurate
measurements,
have
the
tape
tab
on
the
end
of
the
tape
straight
down,
not
perpendicular
to
the
tape.
Parallel
tracks
on
the
bottom
of
the
collimator
housing
accept
an
optional
extension
cylinder.
Slide
the
extension
cylinder
into
the
tracks
from
the
front
of
the
collimator
until
the
cylinder
latch
snaps
into
the
slot
on
the
right
track.
Press
the
latch
to
release
the
cylinder
for
removal.
The
added
weight
of
the
extension
cylinder
will
affect
the
counterbalance
of
the
vertical
column.
Lock
the
vertical
arm
when
installing
or
removing
the
extension
cylinder
and
apply
extra
caution
when
positioning
the
unit
with
the
extension
cylinder
installed.

GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-li
X-RAY
UNIT
1-981
OM
A0654E
OPERATION
Before
operating
the
unit
read
the
Warning
on
page
1
and
the
Safety
Precautions
on
page
2.
Driving
the
Mobile
Unit
1.
Loosen
the
Telescoping
Arm
Lock
and
withdraw
the
arm
until
the
interlock
button
is
heard
to
click.
Tighten
the
Telescoping
Arm
Lock.
Remove
the
Extension
Cylinder,
if
used.
Rotate
the
column
until
the
arm
is
facing
the
operator.
Loosen
the
Vertical
Arm
Lock
and
slowly
lower
the
arm
until
the
latch
and
latch
bar
are
seen
to
engage.
Secure
the
Carriage
Lock.
5.
Check
that
the
Yoke
Rotation
Lock
and
Tube
Rotation
Lock
are
secure.
6.
Be
sure
the
charger
cable
plug
is
removed
from
a
power
outlet
and
retracted.
CAUTION:
WHEN
DRIVING
THE
UNIT
THROUGH
DOORS,
BE
SURE
TO
PARK
THE
TUBE
UNIT
AS
SHOWN
ON
ILLUSTRATION
1-1
TO
PREVENT
STRIKING
DOOR
CASINGS
AND
DAMAGING
THE
TELESCOPING
ARM.
Turn
the
Main
Circuit
Breaker
to
ON.
Move
the
Direction
Switch
to
FORWARD
or
REVERSE
as
desired.
Depress
the
Brake
Handle.
10.
Rotate
the
Speed
Control
toward
the
x-ray
control
to
increase
speed.
11.
Steer
the
unit
with
the
Brake
Handle.
12.
To
aid
in
climbing
ramps
or
moving
the
unit
onto
thick
carpeting,
press
the
Drive
Assist
Button
with
the
left
thumb
while
keeping
the
Brake
Handle
depressed.
CAUTION:
DRIVE
ASSIST
OPERATION
IS
NOT
RECOMMENDED
WHILE
POSITIONING
THE
UNIT
FOR
RADIOGRAPHS
OR
WHILE
MANEUVERING
THE
UNIT
THROUGH
DOORWAYS
OR
CONGESTED
CORRIDORS.
13.
If
the
Bumper
contacts
any
object
when
moving
forward,
the
unit
will
stop.
To
regain
drive
control,
the
unit
must
be
manually
pushed
back.
CAUTION:
DO
NOT
DEPRESS
THE
BRAKE
HANDLE
WHEN
THE
SPEED
CONTROL
IS
SET
TO
FULL
SPEED.
ALWAYS
START
AT
ZERO
SPEED.
Positioning
and
Making
an
Exposure
This
operative
procedure
presupposes
that
preliminary
steps
have
been
completed.
These
inciude
position-
ing
the
patient,
placing
the
cassette
in
position,
selecting
KVP,
MAS,
etc.
1.
Check
that
charger
cable
plug
is
removed
from
the
power
outlet.
2.
Release
the
LATCH
and
raise
the
telescoping
arm.
3.
Rotate
the
column
so
the
arm
points
in
the
desired
position.
4
Loosen
the
Telescoping
Arm
Lock,
push
the
interlock
button,
extend
the
arm
and
lock
it
into
position.
5.
Adjust
the
Telescoping
Arm
to
the
desired
source-to-image
distance.
Use
the
measuring
tape
in
the
collimator.
10
Table of contents
Other GE Medical Equipment manuals

GE
GE Venue Fit User manual

GE
GE Marquette 418793-001 User manual

GE
GE BK MEDICAL 8838 Series User manual

GE
GE Fluorostar User manual

GE
GE Voluson Swift User manual

GE
GE Responder AED Operation manual

GE
GE OPTIMA MR360 Manual

GE
GE SENOGRAPHE DMR+ User manual

GE
GE Voluson E8 User manual

GE
GE MAC 5000 User manual
Popular Medical Equipment manuals by other brands

TESTO
TESTO 520 operating instructions

Boston Scientific
Boston Scientific SC-5213 Directions for use

Otto Bock
Otto Bock AxonArm Hybrid 12K500 Instructions for use

Syneron
Syneron Candela VelaShape III System user manual

Otto Bock
Otto Bock Manu Immobil Long Instructions for use

Sonoscape
Sonoscape E3 Quick manual