Unpacking and Setting Up
04-6923 page 1 of 2 Issue1-4 February 2014
Unpacking
The evaporator and accessories and the vacuum pump
are packed in two separate boxes. Upon receipt of the
equipment, please check the packaging for damage
before signing for it. If there is any damage, contact
the Genevac Service department or your local
distributor immediately. Refer to the Genevac website:
www.genevac.com for up to date contact details of
Genevac representatives.
Handling information
Approx weight of packed crate *
129 kg
Approx weight of EZ-2 Elite *
80 kg
Weight of vacuum pump
26.2
*Actual weights vary according to build options.
Packaging
To unpack the wooden crate, undo the lid retaining
screws and remove the lid. Undo the screws from the
base of the timber sleeve and carefully lift the sleeve
clear of the evaporator.
Lifting
Make sure enough people are present to share the
lifting safely. Carefully lift the EZ-2 Elite slightly at first
to feel the weight and to find the centre of gravity.
When ready, manually lift the EZ-2 Elite into position
using the handholds in the base. Alternatively, use
mechanical lifting equipment such as a lifting table or
forklift.
Left side hand holds
Right side hand holds
Caution: The hand holds are for
manual lifting only. Do not insert lifting
forks.
Installation site
The evaporator must be placed on a level and sturdy
work surface; there must be a 50 mm (2”) air gap
between the evaporator and the edge of the bench or
any other object.
Dimensions (w x d x h) 570 x 645 x 700 mm
The vacuum pump is connected to the evaporator by
a vacuum hose and electrical control cable which are
approximately three metres in length.
The waste solvent container must be positioned to
allow unimpeded drainage of solvent from the
evaporator and must be easily accessible for
emptying. Once a suitable location has been
decided, the drain hose can be cut to length.
There must also be provision to connect the exhaust
(PTFE: 8 mm OD, 6 mm ID) to a suitable laboratory
fume extraction system and to connect two power
cables to mains power outlets.
Note: The EZ-2 Elite may be placed in a fume
cupboard. To maintain a vapour free environment
around the EZ-2 Elite, do not store vessels
containing solvents or acids within the fume
cupboard.
Genevac recommend that evaporators equipped with
inert gas purge, are installed by an authorised
Genevac representative.
50 mm
2”
50 mm
2”
50 mm
(2”)
50 mm
(2”)
Genevac EZ-2 Elite
Unpacking and Setting Up
04-6923 page 2 of 2 Issue1-4 February 2014
Setting Up
Place the drip-plate in the recess to the right of the lid.
Vent filter
Insert the vent filter into
the hole at the bottom of
the side service panel
and screw it onto the
threaded connector (take
care to avoid dropping
the vent filter inside the
casing).
Vacuum pump
Set the pump gas balast
control to position “1” (as
shown).
Check an O-ring is fitted
to the exhaust adapter
and the adapter firmly
into the pump exhaust
outlet. The adapter may
be twisted so the outlet
faces the reqired
direction.
Gas balast
Exhaust adapter
Using the pump control cable, electrically connect the
vacuum pump to the evaporator.
Using Klein Flange connectors, connect the three
metre vacuum hose to the vacuum pump inlet via the
combined centring ring/strainer, and to the evaporator
outlet.
Waste solvent drain hose
PTFE hose (6 mm ID, 8 mm OD) is
supplied to form the waste solvent drain
and exhaust hoses. Connections are
made using a finger nut and shouldered
comprssion ring.
Note: The drain hose must allow unimpeded
drainage of waste solvents into the waste container,
by gravity. The waste container must be below the
level of the waste outlet, there must be no sharp
bends, loops or upward inclination of the drain hoses
and the waste container must have sufficient
capacity to handle solvent throughput.
Exhaust hose
Using PTFE hose, connect the pump exhaust to the
waste solvent container. Connect the third port of
the waste solvent container to the laboratory fume
extraction system.
To prevent drain failure or the
possibility of waste solvent
being drawn back up the drain
hose, make sure the ends of
the hoses are above the
maximum waste solvent level.
Using the mains cable supplied, connect two power
inlets, one on the back and one on the side of the
evaporator, to two separate mains power outlets.
Further instructions are provided in the User Manual.
Genevac Ltd, The Sovereign Centre, Farthing Road, Ipswich, IP1 5AP, UK
Service: +44 (0) 1473 243000
Email:service@genevac.co.uk
Pump control cable

Other manuals for EZ-2 Elite

1

Other Genevac Laboratory Equipment manuals

Genevac EZ-2 Elite User manual

Genevac

Genevac EZ-2 Elite User manual

Genevac HT-24 User manual

Genevac

Genevac HT-24 User manual

Genevac HT-8SII User manual

Genevac

Genevac HT-8SII User manual

Genevac EZ-2 Series User instructions

Genevac

Genevac EZ-2 Series User instructions

Genevac 04-6682 User manual

Genevac

Genevac 04-6682 User manual

Genevac HT Series User manual

Genevac

Genevac HT Series User manual

Genevac Rocket 4D User manual

Genevac

Genevac Rocket 4D User manual

Genevac HT-4X User manual

Genevac

Genevac HT-4X User manual

Genevac EZ-2 Series User manual

Genevac

Genevac EZ-2 Series User manual

Genevac II Series User manual

Genevac

Genevac II Series User manual

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.