Genevac EZ-2 Series User instructions

Genevac EZ-2 Series
Unpacking and Setting Up
04-5365 page 1 of 2 Issue1-9 February 2014
Unpacking
Upon receipt of the equipment, please check the
packaging for damage before signing for it. If there is
any damage, contact the Genevac Service department
or your local distributor immediately. Refer to the
Genevac website: www.genevac.com for up to date
contact details of Genevac representatives.
Handling information
Approx weight of packed crate *
125 kg
Approx weight of EZ-2 *
88 kg
*depends on build options.
Packaging
To unpack the wooden crate, undo the lid retaining
screws and remove the lid. Undo the screws from the
base of the timber sleeve and carefully lift the sleeve
clear of the evaporator.
Lifting
Make sure there are enough people to share the lifting
safely. Carefully lift the EZ-2 slightly at first to feel the
weight and to find the centre of gravity. When ready,
manually lift the EZ-2 into position using the handholds
in the base.
Alternatively, use mechanical lifting equipment such as
a lifting table or fork lift.
Left side hand holds
Right side hand holds
Caution: The hand holds are for
manual lifting only. Do not insert lifting
forks.
Installation site
Place the EZ-2 on a level and sturdy work surface.
Make sure there is enough space for:
A 50 mm air gap between the EZ-2 and the edge
of the bench, a wall or any other equipment
Access to the exhaust catch-pot (to be fitted to the
right hand side of the EZ-2)
An exhaust line to be connected from the exhaust
catch-pot to a suitable laboratory fume extraction
system
The mains cable.
Dimensions (w x d x h) 570 x 845 x 700 mm
Note: The EZ-2 may be placed in a fume cupboard.
To maintain a vapour free environment around the
EZ-2, do not store vessels containing solvents or
acids within the fume cupboard.
50 mm
2”
50 mm
2”
Genevac EZ-2 Series
Unpacking and Setting Up
04-5365 page 2 of 2 Issue1-9 February 2014
Setting Up
Place the drip -plate in the moulding recess to the right
of the lid.
Fit the catch pot by screwing it into the retainer on the
side of the EZ-2.
Fit the condenser jar.
Make sure the jar is correctly aligned
Taking care to avoid dropping the vent filter inside the
EZ-2 cladding, carefully insert the vent filter into the
hole in the service panel and screw onto the threaded
connector.
Push the exhaust hose (supplied) to the spigot on top
of the exhaust catch-pot. Connect the other end of
the exhaust hose to an appropriate laboratory fume
extraction system.
Arrange the exhaust hose so it takes solvent vapours
away from the evaporator safely. Avoid forming U-
bends in the pipe as this could cause condensed
solvent to gather.
Connect the mains supply lead.
Before powering up, check all parts of the system are
installed correctly.
Genevac Ltd, The Sovereign Centre, Farthing Road, Ipswich, IP1 5AP, UK
Service: +44 (0) 1473 243000
Email:service@genevac.co.uk

Other manuals for EZ-2 Series

1

Other Genevac Laboratory Equipment manuals

Genevac EZ-2 Elite User instructions

Genevac

Genevac EZ-2 Elite User instructions

Genevac Rocket 4D User manual

Genevac

Genevac Rocket 4D User manual

Genevac EZ-2 Elite User manual

Genevac

Genevac EZ-2 Elite User manual

Genevac HT-4X User manual

Genevac

Genevac HT-4X User manual

Genevac HT-8SII User manual

Genevac

Genevac HT-8SII User manual

Genevac II Series User manual

Genevac

Genevac II Series User manual

Genevac HT Series User manual

Genevac

Genevac HT Series User manual

Genevac 04-6682 User manual

Genevac

Genevac 04-6682 User manual

Genevac HT-24 User manual

Genevac

Genevac HT-24 User manual

Genevac EZ-2 Series User manual

Genevac

Genevac EZ-2 Series User manual

Popular Laboratory Equipment manuals by other brands

PASCO SE-9638 Instruction manual and experiment guide

PASCO

PASCO SE-9638 Instruction manual and experiment guide

Qiagen QIAamplifier 96 user manual

Qiagen

Qiagen QIAamplifier 96 user manual

AMA Instruments The GC 5000 BTX operating manual

AMA Instruments

AMA Instruments The GC 5000 BTX operating manual

Extron electronics PI 130 installation guide

Extron electronics

Extron electronics PI 130 installation guide

Endress+Hauser Analytik Jena InnuPure C16 touch operating manual

Endress+Hauser

Endress+Hauser Analytik Jena InnuPure C16 touch operating manual

Hanil HyperVAC HVC-2124 user manual

Hanil

Hanil HyperVAC HVC-2124 user manual

Selecta FLOCUMATIC 4 PLAZAS/SAMPLES LED manual

Selecta

Selecta FLOCUMATIC 4 PLAZAS/SAMPLES LED manual

Scientifica Multiphoton Scanhead Setup and operation manual

Scientifica

Scientifica Multiphoton Scanhead Setup and operation manual

Heidolph MR Hei-End operating manual

Heidolph

Heidolph MR Hei-End operating manual

Optika Italy Modular Series instruction manual

Optika Italy

Optika Italy Modular Series instruction manual

HBK FUSION HF-16 quick guide

HBK

HBK FUSION HF-16 quick guide

VWR avantor 730-1461 Short user manual

VWR

VWR avantor 730-1461 Short user manual

Biosan Bio RS-24 operating manual

Biosan

Biosan Bio RS-24 operating manual

Sartorius Secura Quintix 26 Installation and operating instructions

Sartorius

Sartorius Secura Quintix 26 Installation and operating instructions

Vernier STIR quick start guide

Vernier

Vernier STIR quick start guide

BIO RAD MyCycler instruction manual

BIO RAD

BIO RAD MyCycler instruction manual

Waters Extraction+ user manual

Waters

Waters Extraction+ user manual

THORLABS CM401 user guide

THORLABS

THORLABS CM401 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.