HD2106 - 10 - V2.4
χ.Ω.TDS (conductivity - resistivity - total dissolved solids - salinity) / ESC
key
Changes the main variable measurement between conductivity, resistivity, total dis-
solved solids (TDS) and salinity. The selected parameter is used for display, printing
and logging (model HD2106.2).
The instrument has an Auto-Hold function, which can be set in the MENU, that "freez-
es" the measurement automatically when it has been stable (within 1 digit) for over
10 seconds: the message HOLD is displayed.
To perform a new measurement, it is necessary to press the FUNC/ENTER key.
The HOLD message starts blinking, while the display follows the actual measurement
trend, until it stabilizes again and the HOLD message remains still.
In the menu, the key clears or cancels the active function (ESC).
REL/
key
In measurement mode, it displays the difference between the current value and that
measured on pressing the key. The REL message appears on the display; press the
key again to return to the current measurement.
When used in the menu, it decreases the current variable value.
MENU key
The first menu item is accessed by initially pressing on the MENU key; press ENTER to
go to the following items. To modify the item displayed, use the arrow keys (and
). The current value is confirmed by pressing the ENTER key and the display moves
on to the next parameter. If pressing ESC the setting is cancelled.
To exit the menu, press the MENU key at any time.
The menu items are listed in this order:
1) Management of memorized data (only HD2106.2): the message
">>>_LOG_DUMP_or_ERAS" (Transfer data or erase) is scrolled in the
comment line. The center figure reports the number of free memory pages
(FREE). Pressing SERIAL/EraseLOG permanently erases all memory data. By
pressing LOG/DumpLOG, the data transfer of the logged data on the serial port
is started: the "BAUD-RATE" must have previously been set to the maximum
value (please see the menu items described below and the paragraph "STOR-
ING AND TRANSFERRING DATA TO A PERSONAL COMPUTER" on page 35).
2) Print and log interval: sets the interval in seconds between two loggings or
data transfers to the serial port. The interval can be set at 0, 1s, 5s, 10s, 15s,
30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min),
1200s (20min), 1800s (30min) and 3600s (1hour). If the value 0 is set, SE-
RIAL works on command: the sending of data to the serial port is per-