The centrifuge rotor can separate samples whose density is lower than 2.0g/ml. If the separated sample’s density is over 2.0g/ml, please
calculate the allowable speed using the following formula:
Allowable Speed (rpm) = Maximum Speed × (2�0(g/ml) / Sample Density (g/ml))1/2
GCC-MP / Multi-Purpose
Clinical Centrifuge
Quick Start Guide
Rotor Tube Adapter Maximum Maximum Centrifugal Maximum
Type Type Required Speed (rpm) Radius (cm) RCF (×g)
2.0/1.5mL x 30 — 6000 10 4020
0.2mLx30 GCM-24-AD2 6000 8.5 3415
0.5mLx30 GCM-24-AD5 6000 9 3618
2.0/1.5mL x 60 — 6000 10 4020
0.2mLx60 GCM-24-AD2 6000 8.5 3415
0.5mLx60 GCM-24-AD5 6000 9 3618
8-PCR x 12 — 6000 10 4020
0.2mL x 96 — 6000 10 4020
50mL x 6 — 6000 10.7 4306
15mL x 30 — 4500 R1 = 14.2 3210
R2 = 12.2 2760
15mL x 16 — 5000 12.4 3460
50mL x 8 — 5000 12.4 3460
100mLx4 GCC-MP-SWING-100ML4 4000 15.9 2840
50mLx4 GCC-MP-SWING-50ML8 4000 15.9 2840
10mL x 16
12mLx16 GCC-MP-SWING-15ML16 4000 15.1 2700
15mL x 16
3mL x 36
7mL x 36 GCC-MP-SWING-7ML36 4000 14.7 2630
MTP — 4000 12.1 2160
(128 x 85.6 x 15)
Cell Culture — 4000 12.1 2160
(128 x 85.6 x 21)
DWP — 4000 12.1 2160
(128 x 85.6 x 45)
Kits — 4000 12.1 2160
(128 x 85.6 x 60)
GCC-MP-2ML60
GCC-MP-2ML30
GCC-MP-PCR
GCC-MP-50ML6
GCC-MP-15ML
GCC-MP-50ML8
GCC-MP-SWING
GCC-MP-SWING-
PLATE