GPI GM Series User manual


2
INSTALLATION
Installation should be performed only by qualified
personnel, and in accordance with local governing
regulations.
GENERAL INFORMATION
This manual will assist you in operating and maintaining
the GPI electronics supplied with your GPI meter or as an
accessory unit on both local and remote models. The GPI
Electronics can be used in indoor or outdoor applications
where occasional exposure to moisture is common.
• ThePulseOutwithDisplayisavailableintwoversions.
One indicates flowrate in units/minute and one in units/
hour. Both indicate flow totals in gallons and litres.
• The PulseOut with Display can be used on all GPI
models, including the Precision G series, the Industrial
Grade G2 series, the Commercial Grade A1 series, and
the positive displacement GM Series oval gear meters.
• ThePulseOutwithDisplaycanbebatterypoweredor
externally powered.
Product differences in this manual are identified by either,
Local or Remote as necessary.
The Display may come calibrated from the factory for gal-
lons “GAL” and litre “LTR” and be locked out “PRESET” to
prevent unwanted calibrations. Field calibrations are available
on “CAL B” and “CAL C.”
If the unit was purchased as an accessory or remote, the
calibration has not been entered and the end user will be re-
quired to calibrate the display. The gallon and litre calibration
may then be locked out using the configuration procedure.
To determine if the Pulse Out with Display has been configured
and calibrated at the factory to your specific meter, view the
display. If “GAL” is followed by the word “PRESET,” it has
been configured and calibrated. If “GAL” is not followed by
the word “PRESET,” the display must be calibrated.
• Whenmeasuringammableliquids,observeprecau-
tions against fire or explosion.
• When working in hazardous environments, always
exercise appropriate safety precautions.
• BesureO-ringsandsealsarekeptingoodrepair.
• Whenapplyingpower,useDCpower only!
• Disconnectexternalpowerbeforeattachingordetach-
ing input or output wires.
SAFETY INSTRUCTIONS
The following installation guidelines are separated by meter
series and mounting type.
Precision G Series:
• Local – The GG510 Pulse Out with Display mounts
directly to the 1 inch MNPT conduit connector.
• Remote – The GG500 Pulse Out with Display connects
via an output cable as shown in the Wiring Diagram.
Industrial Grade G2 Series:
• Local – The GG510 Pulse Out with Display requires the
GPI Conduit Connector Kit (Part #113437-01) for local
mounting to the G2 series meter.
• Remote – The GG500 Pulse Out with Display connects
via an output cable as shown in the Wiring Diagram.
Commercial Grade A1 Series:
• Local – The GG510 Pulse Out with Display requires the
GPI Conduit Connector Kit (Part #113437-01) for local
mounting to the A1 series meter.
• Remote – The GG500 Pulse Out with Display connects
via an output cable as shown in the Wiring Diagram.
GM Oval Gear Series:
• Local – The Pulse Out with Display is mounted directly
to the oval gear meter housing.
• Remote – The GG500 Pulse Out with Display connects
via an output cable as shown in the Wiring Diagram.
Connecting the Battery
G series and GM models are shipped with batteries. To save
power, the display is shipped with the battery disconnected.
To connect the battery:
1. Remove the two large and two small screws from the
battery coverplate located on the lower portion of the
meter face.
2. Insert the battery. When properly connected, numbers
will appear in the readout.
NOTE: To protect the terminals from corrosion, coat with
petroleum jelly.
3. Replace the gasket and battery cover, then tighten all
four screws.
Remote Installation
Choose a mounting location suitable for the Display. The ideal
mounting location is where the:
• owmeterisascloseaspossible.
• mountingsurfacehasminimalvibration.
• ambienttemperatureis+14°Fto140°F(–10°Cto60°C).
• cablelengthsareminimal.
CAUTION

Avoid mounting locations where the Display is:
• subjecttoconstantexposuretowaterorotherliquids
(occasional low-pressure splashing will not harm unit).
• subjectto>5gshockloading.
• facingthesundirectlyforlongperiodsoftime.
Mount the GPI Display Remote using bolts, screws or standard
U-bolts for pipes. Mounting options include:
• Wall
• Pipe
• Meter(1inchFNPTconduitconnectionrequired)
Some products come with 20 feet of shielded cable. GPI offers
an optional cable kit with 100 feet of cable (see spare parts
list). When wiring longer lengths of cable, be sure to connect
the shield to LOCAL-COM only! (Multiple shield connections
may cause ground-loop problems. Some trial and error may
be needed because of the wide variety of user conditions.
Try to keep cable lengths short. Individual installation sites
vary widely, contact GPI Customer Service with questions
regarding your specific needs.
This manual refers to various models of GPI flowmeters.
Determine what type of input the electronics will receive and
what type of output, if any, you require. If the unit is not already
wired, use the diagrams found in the wiring diagrams section
to correctly wire the system.
Connecting the Equipment
1. Remove the faceplate by removing the four corner
screws.
2. Attach wiring from your equipment according to the
following terminal connections and wiring instructions,
depending on your circumstances. (See Terminal Con-
nections below)
The display may be powered by battery, externally powered
or both. When both are used (external power terminal as
primary with the battery terminal as back up) the batteries
should last up to 5 years. Connection and disconnection of
either power input, while the other is active will not interrupt
operation of the display.
Determine maximum power supply voltage after de-
termining maximum allowable voltage of all electronic
devices in the system.
WIRING
T E R M I N A L L O C A T I O N S
9V_BAT+
Battery, 6.5V to 20V
EXT_V(+)
External Power, 7V to 30V
COIL_B /
COIL_A
Any magnetic (variable-reluctance) pickup coil can
drive this input. Minimum signal amplitude is 10 mV
P-P. Maximum recommended signal amplitude is
1V P-P. Shielded cable may be used to increase
distance.
HI-F_IN
Either open-collector or active-drive signals are
accepted. If active-drive, signals of 5V to 12V P-P
amplitude are acceptable. Maximum frequency is
approximately 1000 Hz. Note: Optional to route a
reed-switch signal directly to this input.
LO-F_IN
A reed switch may be connected directly between
this terminal and any
COM
terminal. Maximum
frequency is approximately 150 Hz.
OC_OUT
•Thisisanopen-collectororcurrent-sinkingout-
putwhenNOJUMPERisattachedatOUTPUT
P.U.R. location. It can sustain closed-circuit
current of up to 200 mA and open-circuit voltage
of up to 60V.
•Thisisanactive-driveoutputwhentheJUMPER
(supplied)isattachedatlocationOUTPUTP.U.R.
TheJUMPERappliesaninternal10KOhmresis-
tor as a “pull-up” to the regulated power supply
of 5.5V.
•Theoutputgroundcircuitmaybeconnectedtoany
COM
terminal.
•DonotuseJUMPERifoperatingfromabattery;
it will cause significantly increased system
current consumption.
•CAUTION: Only use JUMPER when the receiving
equipment has a active-drive input with 5V or less.
COM
All COM terminals are internally connected and may
beusedeitherasNEGATIVEpowersupplytermi-
nals or return terminals for any inputs or outputs.
3
CAUTION

P O W E R
Connect battery (included) red wire (+) to
9V_BAT+
terminal.
Connect battery (included) black wire (–) to any
COM
terminal.
9V Battery (included)
S I G N A L I N P U T
Connect either coil wire to
COIL_A
terminal.
Connect other coil wire to
COIL_B
terminal.
Connect shield cable, if any, to any
COM
terminal.
NOTE:Leavemetersideofshieldwireunconnected.
Sine Wave (coil)
continued on next page
4
OUTPUTP.U.R.
OUTPUTP.U.R.
Connect either wire to
LO-F_IN
terminal.
Connect other wire to any
COM
terminal.
Contact Closure (reed switch)
OUTPUTP.U.R.
GM
METER
Connect signal wire to
HI-F_IN
terminal.
Connect ground wire to any
COM
terminal.
Square Wave (active drive)
OUTPUTP.U.R.
G SERIES
OR
G2
METERS
Connectpowerwire(7-30VDC)to
EXT_V(+)
terminal.
Connect ground wire to any
COM
terminal.
External Power
OUTPUTP.U.R.
Optional to connect
to HI-F_IN.

5
S I G N A L I N P U T
continued on next column
Connect signal wire to
HI-F_IN
terminal.
Connect ground wire to any
COM
terminal.
Connect power wire (+) to
EXT_V (+)
terminal.
Connect1Kpullupresistorbetween(+)powerwireandsignal
wire.
Open Collector with External Power to Display
S I G N A L O U T P U T
Connect signal wire to
OC_OUT
terminal.
Connect ground wire to any
COM
terminal.
NOTE:Jumpershouldnot connect pins marked Output P.U.R.
Open Collector
Connect signal wire to
HI-F_IN
terminal.
Connect ground wire to any
COM
terminal.
Connect1Kpullupresistorbetween(+)powerwireandsignal
wire.
Open-Collector
OUTPUTP.U.R.
OUTPUTP.U.R.
OUTPUTP.U.R.
Connect signal wire to
OC_OUT
terminal.
Connect ground wire to
COM
terminal.
PlaceJUMPER(supplied)overbothpinsmarkedOUTPUTP.U.R.
Active-Drive Output
OUTPUTP.U.R.
NOTE: Multiple COM terminals are provided for convenience
in making connections.

The calibration and display features may have been pro-
grammed at the factory, indicated by the word “PRESET”
on the display (see Display 1), and further user programming
is not required. If the unit was purchased as an accessory,
remote or as a spare part, the calibration features will not be
appropriate for the user installation and must be calibrated
before use. If desired, the display features can be changed
in the field.
DISPLAY CALIBRATE
ELECTRONIC
DIGITAL
METER
INDUSTRIAL
GRADE
D I S P L A Y 1
OPERATIONS
6
Configuration determines what information is present on the
LCD display. For instance, total, flowrate, number of decimal
places, number of calibration curves, etc.
The Display has been programmed with many features, which
can be enabled by the end user through the configuration
process. By disabling unnecessary features, day-to-day
flowmeter operation can be greatly simplified, making the
unit easier to use. Alternately, there are several features not
found in the default configuration.
Available features include:
• 0to3totals,eitherresettable(Batch)ornon-resettable
(Accumulative).
• Flowrateinunitsperminute,hourordayornoowrate.
• 1to3differentcalibrationcurves.
• K-factorentryorDispense-Displaycalibrationproce-
dure.
• 0to3decimalresolution(0000;000.0;00.00)
Changing Configuration Settings
Access to the configuration settings require a specific pro-
cedure and a pin code available through the GPI website at
www.gpi.net or call GPI Customer Serivce at 888-996-3837.
All non-PRESET displays must be calibrated before
use!
CONFIGURATION
About Calibration
Calibration determines that the information presented
matches your particular meter. For instance, the “S20N” 2
inch turbine has approximately 100 pulses per US gallon
whereas the “GM020” 2 inch oval gear meter has 25.3 pulses
per US gallon. The display must be calibrated for correct and
accurate display information.
To determine if the Pulse Out with Display has been configured
and calibrated at the factory to your specific meter, view the
display. If “GAL” is followed by the word “PRESET,” it has
been configured and calibrated. If “GAL” is not followed by
the word “PRESET,” the display must be calibrated.
WARNING
All operations are reflected in the LCD readout. The top line
identifies the calibration curve. The middle line reflects flow
information. The bottom line shows information from the
totalizer.Wordsor“ags”displayonthetopandbottomline
to further identify specific information.
NOTE: Operations can be practiced prior to installation. To
simulate flow conditions, blow gently through the meter.
Turn On
The unit is on when any display is present. It turns on auto-
matically when liquid flows through the meter. It can be turned
on manually by pressing and releasing the DISPLAY button.
Turn Off
Whenever no flow has been sensed for one minute, the unit
automatically switches to a power-saving “sleep” mode with
a blank display. The unit will automatically “wake up” the
moment any flow is sensed and will remain awake as long
as fluid is flowing. Totals are never lost during sleep periods.
Batch and Cumulative Totals
Total flags are displayed on the bottom line. The Cumula-
tive Total (labeled TOTAL 1 LOCKED) is the total of all fluid
measured since the meter’s power was connected. (At your
rstuse,theCumulativeTotalmaynotreadzerobecauseof
calibration at the factory.) The Batch Total (labeled TOTAL 2)
indicates flow during a single use.
Clearing a Totalizer
TheBatchtotalizerregister(TOTAL2)maybeindependently
clearedto0.00atanytime.Toclearabatchtotalizer,with
thedesiredtotalizerdisplayed,pressandholdtheDISPLAY
button. At about 3 seconds, the displayed total will be cleared
to “0.00.” You can do this even while fluid is flowing, in which
case counting will resume after you release the DISPLAY
button.
The Cumulative totalizer register is labeled as TOTAL 1
LOCKEDindicatingthatitcannotbemanuallyzeroed(See
Display2).TheCumulativetotalizercanbeclearedonlywhen
the batteries are removed, go dead or when the Cumulative
Total reaches the maximum value of 999,999.

7
DISPLAY CALIBRATE
ELECTRONIC
DIGITAL
METER
INDUSTRIAL
GRADE
D I S P L A Y 2
DISPLAY CALIBRATE
ELECTRONIC
DIGITAL
METER
INDUSTRIAL
GRADE
D I S P L A Y 3
Flowrate Mode
The Rate of Flow feature is accessed by briefly pressing and
releasing the DISPLAY button as described above. When this
feature is activated, the word “FLOWRATE” displays to the
left on the bottom line (See Display 3) and the numbers in the
middle of the display reflect the rate of flow (instead of total).
Units are set to update the display every 5 seconds, so the
first reading after flow starts or changes and the last reading
after flow stops or changes will not be correct. This is normal.
Changing Display Registers
TochangetoanothertotalizerregisterortoFLOWRATEmode
during normal operation, watch the bottom line display flags
while you briefly press and release the DISPLAY button. When
you press and release the display buttons, the mode will ad-
vance as follows: TOTAL 1 LOCKED (Cumulative Total), TOTAL
2 (Batch Total), FLOWRATE, TOTAL 1 LOCKED (etc.). You can
change registers at any time, even during flow. Non-visible
totalizerregisterswillcontinuetoaccumulate.
NOTE: Generally, display registers change when the buttons
are released.
Calibration Curves
The GPI electronics have enhanced calibration features. All
calibration information is visible to the user as words in the
upper part of the display, above the numeric digits.
The calibration curve(s) may be set by the user, and can be
changed or modified at any time using the calibration pro-
cedure described below in the CALIBRATION section. Totals
or flowrate derived from the field calibration are visible when
the field calibration setting is selected.
Selecting a Different Calibration Setting
You can switch between GAL and LTR modes at will without
“corrupting”totalizercontents.Forexample,thecomputer
cantotalize10.00gallons.IftheuserswitchestoLTRmode,
the display will immediately change to “37.85” (the same
amount in units of litres). GAL / LTR switching also works in
FLOWRATE mode.
To select a different calibration setting, first press and hold
the CALIBRATE button. Continue to hold it while also briefly
pressing and releasing the DISPLAY button (you may then
also release the CALIBRATE button). The flag indicators in
the upper area of the display will change to show the newly
selected calibration setting. Calibration settings change in this
order: GAL, LTR, CAL B, CAL C, GAL, etc. While fluid is flow-
ing only the GAL and LTR selections may be made, however,
when NO fluid flow is occurring, any setting may be selected.
Propeller
A small propeller displays to indicate liquid is flowing through
the meter.

8
Your Actions Notes
1. First select a Calibration Curve. Remember that Factory
To do so, hold down CALIBRATE Calibration curves are
while pressing and releasing displayed with the word
DISPLAY until the desired Field PRESET. If LTR is present,
Calibration curve appears (GAL, it will always show PRESET.
CAL B or CAL C message will be
displayed). Release both buttons. Local Mount: The Field
Calibration curves are
shown as CAL B or CAL C
with the PRESET message
absent.
Remote Mount: The Field
Calibration curves are
shown as GAL, CAL B or
CAL C with the PRESET
message absent.
2. To calibrate, press and hold both This step puts the unit in
the CALIBRATE button and the K-factor input mode
DISPLAY button. Hold for about (KF CAL).
3 seconds until the KF CAL mes-
sage appears. Then release both
buttons. Your are now in K-factor
input mode.
0000.00
GAL
KF CAL
GAL
CALIBRATION Your Actions Notes
3. Once you release the buttons, The computer is waiting
the display will show the blinking for you to make a decision
message KF 01 (or KF 02, KF 03, to either enter a K-factor
etc., if you are repeating this step). or exit calibration. If you
want to exit the calibration
now, go to Step 11.
4. If you want to continue with the It is possible to set up to
calibration, press either button 15 cal-curve points, and
briefly and release it. the KF ## and Pr ##
messages will increase
increments each time you
repeat the calibration pro-
cess (KF 01, Pr 01, KF 02,
Pr 02, KF03, Pr 03, etc.,
up to KF 15 Pr 15).
5. The display will change to a nu- NOTE: K-factors are
meric representation of the current entered as number of
K-factor for this point (for example pulses per unit measure.
00564.0) with the left hand digit When calibrating GAL, use
blinking. Use the CALIBRATE but- only pulses per gallon.
ton to change the numeric value When calibrating CAL B or
and the DISPLAY button to change CAL C, most unit
the digit location, enter the K-fac- measures are acceptable
tor of the new cal point. When you (for example, per litre,
are finished, press both buttons
briefly and release.
KF 01
GAL
GAL
00564.0
6. Once you release the buttons, The computer is waiting
the display will show the blinking for you to enter a pulse
message “Pr 01” (or Pr 02, Pr 03, rate. Pulse rates are
etc., if you are repeating this entered as pulses per
step). Press either button briefly second. NOTE: If you
and release it. elected to enter a single
K-factor, you must still
enter a pulse rate (any
rate will work, such as
100), otherwise you will
receive an Error message.
7. After you have released one of To enter numbers, use the
the buttons, the display will CALIBRATE button to
change to a numeric representa- change the value of the
tion of the old pulse rate for this digit that is blinking and
point (for example 00047.1) with use the DISPLAY button
the left hand digit blinking. Enter to shift the blink to the
the pulse rate of the new cal next digit.
point.
8. Once the correct number has You have installed the new
been entered, press both but- cal-curve point. You are
tons, and the display will show a ready to end calibration
“KF CAL” message. When you (Step 10) or enter another
release the buttons, the display new calibration point
will change to a blinking “KF 02” (Step 9).
(or greater) message. At this
point the computer is reminding
you that the next value to be
entered will be a K-factor.
GAL
Pr 01
r
The GPI electronics allows for “field” calibration, that is,
user entry of custom calibration parameters. A “single point”
calibration may be used on the GPI GM series, or any other
positive displacement meter, with excellent results. Any pulse
generating device that changes K-factor with varying flowrate
(turbine meter, paddlewheel, etc.) should be calibrated to at
least three points throughout the anticipated flow range for
best accuracy. Up to 15 points can now be entered. The use
of dependable, accurate calibration equipment is highly rec-
ommended for the most accurate results. Additionally, each
curve can be calibrated to a different fluid.
The display will have either a K-factor Entry or a Dispense/
Display Field Calibration. If you need to switch from one kind of
calibration to another, please see CONFIGURATION section.
The K-factor of your meter is usually located on the housing
or nameplate of the meter body. It will also appear on the
Calibration report. Or call GPI Customer Service at 888-996-
3837 for assistance.
The litre calibration is determined using the gallon calibration
information. Enter the K-factor as pulse per gallon in the GAL
curve. This will correctly calibrate the LTR PRESET curve. You
may still remove the GAL indication on the display using the
computer electronics configuration procedure.
K-FACTOR CALIBRATION PROCEDURES

Your Actions Notes
9. To enter another calibration point,
go back and repeat Steps 3
through 8.
10. To end calibration, press and After you release the
hold both buttons for about 3 buttons, the computer will
seconds until you see the resume normal operations
“CAL End” message. with the new cal point(s)
active.
11. If you HAVE NOT entered a After you release the but-
K-factor, you can exit calibration tons, the computer will
without changing the cal curve. resume normal operations
If the message KF 01 is showing and the old curve (if you
and you have not entered a have entered one in the
K-factor, hold both buttons for past) is still intact.
about 3 seconds until you see a
“CAL End” message.
GAL
CAL EA
A
Your Actions Notes
1. Hold down CALIBRATE while Remember that Field
pressing and releasing DISPLAY Calibration curves are not
until the Field Calibration curve preset.
appears (“CAL B” or “CAL C”
message will be displayed).
Release both buttons.
2. To calibrate, press and hold the This step puts the unit
CALIBRATE button. While con- in dispense-display field
tinuing to hold CALIBRATE, also calibration mode
press and hold the DISPLAY (“dd-CAL”).
button. Hold both buttons for
about 3 seconds until you see a
blinking “dd-CAL” message.
Once the “dd-CAL” message
appears, release both buttons.
You are now in field calibration
mode.
DISPENSE / DISPLAY
FIELD CALIBRATION PROCEDURES
Your Actions Notes
3. Once the buttons have been The computer is waiting
released from Step 2, the display for you to make a decision
will show the blinking message to either exit from field
“run 01.” calibration mode or to
begin a dispense run. If
you want to exit the cali-
bration now, go to Step 11.
4. If you want to continue with the
calibration, but have not dispens-
ed any fluid yet, make your final
preparations to your pumping
system, but don’t start pumping
yet.
5. Start your pumping system so When the computer dis-
that fluid flows through the meter. plays a non-blinking “run
The display will stop blinking and 01” message, it is sensing
show the “run 01” message. fluid flow. For the most
Dispense into a container that accurate results, dispense
allows you to judge the amount of at a flowrate which best
fluid pumped. When you have simulates your actual
pumped the desired amount operating conditions.
(for example, 10 gallons), stop the Avoid “dribbling” more
fluid flow quickly. fluid or repeatedly starting
and stopping the flow -
these actions will result in
less accurate calibrations.
6. Once the flow has stopped, briefly When the display shows
press and release both buttons. “0000.00” the computer
At this point the computer display has stopped “watching”
will change to “0000.00” with the for fluid flow and is now
left-hand digit blinking. waiting for you to enter
some numbers.
7. Enter the volume (amount) of fluid
that you dispensed (for example,
if your 10-gallon container is full,
enter “10.00” for gallons or “37.85”
for litres). To enter numbers use
the CALIBRATE button to change
the value of the digit that is blink-
ing and use the DISPLAY button
to shift the “blink” to the next digit.
8. Once the correct number has been You have installed the
entered, briefly press and release new cal-curve point. You
both buttons. The display will now are ready to end calibra-
change to a blinking “run 02” tion (Step 10) or enter
message. another new calibration
point (Step 9).
9

10
Your Actions Notes
9.To enter another calibration point, It is possible to set up to
go back and repeat Steps 3 15 cal-curve points, and
through 8. the “run ##” message will
increment each time you
repeat the calibration
process (run 01, run 02,
run 03, etc., up to run 15).
10. To end calibration, press and hold After you release the
both buttons for about 3 seconds buttons, the computer will
until you see the “CAL End” resume normal operations
message. with the new cal point(s)
active.
11.If you HAVE NOT dispensed any After you release the
fluid, you can exit calibration with- buttons, the computer will
out changing the cal curve. If the resume normal operation
message “run 01” is showing and and the old curve (if you
you have not dispensed any fluid, have entered one in the
hold both buttons for about past) is still intact.
3 seconds until you see a
“CAL End” message.
SYMPTOM PROBABLE CAUSE CORRECTIVE ACTION
A. LCD REGISTER 1. Battery not connected properly Check battery connections.
NOT WORKING
2. Battery flat Replace battery.
3. Faulty wiring connections Check wiring for loose or faulty connections.
4. Faulty LCD Replace computer assembly.
5. Faulty connection from computer Check wiring connections.
to Pulse PCB
TROUBLESHOOTING
MAINTENANCE
Replacing the Battery
Replace the battery when the readout becomes dim or blank.
Replace the battery with a 9-volt lithium battery. Order GPI
part number 902006-44.
To replace the battery:
1. Remove the two large screws and two small screws from
the battery coverplate.
2. Remove the battery coverplate and gasket.
3. Remove the battery and clean any corrosion from the
terminals.
NOTE: Coat the terminals with petroleum jelly to protect
against corrosion.
4. Install the new battery.
5. Check the gasket for damage and replace as needed.
Position gasket and coverplate to align, insert screws
and tighten.
NOTE:BatchandCumulativeTotalreturntozerowhenthe
battery fails or is disconnected. The factory calibration
is retained in the computer permanently and will display
when the battery is replaced.
Storeattemperaturesbetween14°Fto140°F(–10°Cto+60°C).

1
2
3
2
4
5
6
7
15
9
10
11
12
13
14
8
ILLUSTRATED PARTS DRAWING – GG510 & GM SERIES GM005-GM020
Item No.
No. Part No. Description Req’d.
1 120009-2 Battery Cover ......................................................... 1
2 120028-1 Battery Gasket........................................................ 1
3 12051801 Computer Kit (Hours).............................................. 1
12051802 Computer Kit (Minutes) .......................................... 1
4 12051701 Adapter Kit (GG510) ............................................... 1
120509-02 Adapter Kit (GM005 & GM007) - Local .................. 1
120509-03 Adapter Kit (GM010, GM015 & GM020) - Local..... 1
5 12009601 Circuit Assembly (Core).......................................... 1
6 12009801 Circuit Assembly (Pigtail)........................................ 1
7 42100117 Jumper (2-Circuit)................................................... 1
8 901002-82 O-Ring .................................................................... 1
9 902004-97 Battery Terminal...................................................... 1
10 902006-44 Battery, 9-volt Lithium ............................................ 1
11 904005-27 Sems Screw, 6-32 x 3/8 in. & Plain Washer ........... 2
12 904005-28 Sealing Screw, 1/4-20 x 5/8 in., m SS, Viton.......... 4
13 904005-63 Screw, 4-40 x 3/16 in., X-Recess ........................... 2
14 906005-47 Threaded Plug ........................................................ 2
15 906005-48 Seal......................................................................... 2
11

12
ILLUSTRATED PARTS DRAWING – GG500, G2 SERIES & GM SERIES GM001-GM003
Item No.
No. Part No. Description Req’d.
1 120009-2 Battery Cover ......................................................... 1
2 120028-1 Battery Gasket........................................................ 1
3 902006-44 Battery, 9-volt Lithium ............................................ 1
4 12051801 Computer Kit (Hours).............................................. 1
12051802 Computer Kit (Minutes) .......................................... 1
5 12009801 Circuit Assembly (Pigtail)........................................ 1
6 901002-82 O-Ring .................................................................... 1
7 120509-01 Adapter Kit - Remote Display & Local (GM001,
GM002 & GM003).............................................. 1
8 120058-01 Bracket ................................................................... 1
9 904005-13 Screw, 6-32 x 1/2 in. (Remote Model).................... 4
10 904002-44 Screw, 8-32 x 5/16 in. ............................................ 2
11 125066-20 Cable, 20 ft. ............................................................ 1
125066-3 Cable, 100 ft. .......................................................... 1
Item No.
No. Part No. Description Req’d.
12 902005-9 Strain Relief ............................................................ 1
13 901002-87 O-Ring .................................................................... 1
14 906005-47 Threaded Plug ........................................................ 1
15 906005-48 Seal......................................................................... 1
16 904005-63 Screw, 4-40 x 3/16 in. ............................................ 2
17 42100117 Jumper (2-Circuit)................................................... 1
18 904005-28 Sealing Screw, 1/4-20 x 5/8 in. .............................. 4
19 904005-27 Sems Screw, 6-32 x 3/8 in. .................................... 2
20 12009601 Circuit Assembly (Core) .......................................... 1
21 902004-97 Battery Terminal...................................................... 1
22 904006-94 Screw, 10-16 x 5/8 in. (GM001, GM002
& GM003)........................................................... 2
12051901 Battery Conversion Kit (not shown) - Kit includes
Items 3, 14, 15 and 21....................................... 1

SPECIFICATIONS – LOCAL MODEL
Materials:
Acetal, Amorphous Nylon, PET Polyester, Polyester (de-
cals), Viton (gasket & seals), Stainless Steel (fasteners)
Power Source:
Battery (9V): 6.5V to 20V acceptable range.
Quiescent current (over and above current drawn by
display module) at the battery input is typically less
than 15uA.
External Power: 7V to 30V acceptable range. Quiescent
current (over and above current drawn by display
module) at the external power input is typically less than
90uA.
Battery Life:
1,800 hours
Configuration:
2-Totals (1 cumulative and 1 batch), Rate, 2 Cals
(Factorycalibrationingallonsorlitres;2eldcalibrations)
Input Signal:
Hall Effect, Reed Switch, NPN, Open Collector or
Sine Wave
Time Base:
Hours or minutes
Unit of Measure:
U.S. gallons or litres
Accuracy:
No additional error over coupled flow meter’s accuracy
Frequency Range:
0to1,000hertz
Batch Total:
Up to 999,999
Cumulative Total:
Up to 999,999
Temperature:
+14°Fto+140°F(–10°Cto+60°C)
Cable:
N/A
Mechanical Connections:
Display is mounted directly to flow meter body
Electrical Connections:
GG500 - One strain relief port: one threaded plug
GG510 - Two threaded plugs
G2 Series - One strain relief port: one threaded plug
GM Local Models - One strain relief port: one threaded
plug (GM001, GM002 & GM003)
GM 1/2 in. and Larger - Two threaded plugs (GM005 -
GM020)
Shipping Weight:
1 lb. (.45 kg)
SPECIFICATIONS – REMOTE MODEL
Materials:
Acetal, Amorphous Nylon, PET Polyester, Polyester (de-
cals), Viton (gasket & seals), Stainless Steel (fasteners),
PVC (cable jacket)
Power Source:
Battery (9V): 6.5V to 20V acceptable range.
Quiescent current (over and above current drawn by
display module) at the battery input is typically less
than 15uA.
External Power: 7V to 30V acceptable range. Quiescent
current (over and above current drawn by display
module) at the external power input is typically less than
90uA.
Battery Life:
1,800 hours
Configuration:
2-Totals (1 cumulative and 1 batch), Rate, 2 Cals
(Factorycalibrationingallonsorlitres;2eldcalibrations)
Input Signal:
Hall Effect, Reed Switch, NPN, Open Collector or
Sine Wave
Time Base:
Hours or minutes
Unit of Measure:
U.S. gallons or litres
Accuracy:
No additional error over coupled flow meter’s accuracy
Frequency Range:
0to1,000hertz
Batch Total:
Up to 999,999
Cumulative Total:
Up to 999,999
Temperature:
+14°Fto+140°F(–10°Cto+60°C)
Cable:
20 feet, 3-conductor (red, black & white), tinned drain
wire, 22 AWG, PVC jacket .212 dia., (Reference Belden
9363 or equivalent cable)
Mechanical Connections:
Wall or pipe mountable with standard U-bolts
Electrical Connections:
GG500 - One strain relief port: one threaded plug
GG510 - Two threaded plugs
G2 Series - One strain relief port: one threaded plug
GM Local Models - One strain relief port: one threaded
plug (GM001, GM002 & GM003)
GM 1/2 in. and Larger - Two threaded plugs (GM005 -
GM020)
Shipping Weight:
1 lb. (.45 kg)
13

14
DIMENSIONAL DRAWINGS

SERVICE
For warranty consideration, parts, or other service informa-
tion, please contact your local distributor. If you need further
assistance, contact the GPI Customer Service Department
in Wichita, Kansas, Monday-Friday, 8:00 a.m. to 5:00 p.m.
Central time.
Tel: 316-686-7361
Fax: 316-686-6746
Toll free: 1-888-996-3837
To obtain prompt, efficient service, always be prepared with
the manufacturing date code, found behind the coverplate.
For warranty work, always be prepared with your original sales
slip or other evidence of purchase date.
Please contact GPI before returning any part. It may be
possible to diagnose the trouble and find a solution with a
telephone call. GPI can also inform you of any special require-
ments you will need to follow for shipping.
15
The Waste Electrical and Electronic Equipment (WEEE) directive
(2002/96/EC) was approved by the European Parliament
and the Council of the European Union in 2003. This symbol
indicates that this product contains electrical and electronic
equipment that may include batteries, printed circuit boards,
liquid crystal displays or other components that may be
subject to local disposal regulations at your location. Please
understand those regulations and dispose of this product in
a responsible manner.
WEEE DIRECTIVE

Other manuals for GM Series
1
This manual suits for next models
2
Table of contents
Other GPI Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

TSI Instruments
TSI Instruments ALNOR AVM440 Operation and service manual

Novus
Novus N1540-485 instruction manual

Kusam-meco
Kusam-meco KM-CAL-906 user manual

IME
IME Nemo 96 HDLe Mounting instructions

Gossen MetraWatt
Gossen MetraWatt SECULIFE ITBASE operating instructions

Beta
Beta 601C Instructions for use

Agilent Technologies
Agilent Technologies 8510C Service quick reference guide

McCrometer
McCrometer ULTRA MAG UM06 manual

HACH LANGE
HACH LANGE POLYMETRON 8350 Series user manual

Landis+Gyr
Landis+Gyr T450 installation instructions

Malmbergs
Malmbergs 99 090 00 manual

Kamstrup
Kamstrup READy Collector installation guide