H31.0.11.6C-11 User’s Manual GMH3110 page 5
_______________________________________________________________________________________________________________________
5 Pressure Connection To The Sensors
The device is designed to be connected to the sensors of the GMSD...-series without a new calibration being
necessary. Therefore a great variety of replaceable sensors of e.g. –1.999...2.500 mbar relative up to
0...400.0 bar absolute pressure can be connected to the device (p.r.t. chapter 9)
Relative Pressure Sensors (Types: GMSD...MR. GMSD...BR)
•For measurements of over- or under pressure:
Connect plastic tube with internal dia of 4 mm to pressure port "B". Port "A" will not be used!
Pressure sensors GMSD 2,5 MR, GMSD 25 MR and GMSD 350 MR allow for measurements of under
pressure up to the entire over pressure measuring range by re-plugging the tube to pressure port "A".
Please note that all values are displayed as positive values. No minus sign will be shown. (Example for
GMSD 25 MR: For tube connection "B" the measuring range covers -19.99 to 25.00 mbar. If you replug
to port "A" under pressure measurements down to -25.00 mbar could be carried out with the display
showing the value 25.00 (no minus sign).
•For measurements of pressure differences:
Connect both plastic tubes with an internal dia of 4 mm to pressure port "B" and "A"; make sure to apply
higher pressure to port "B".
Absolute pressure sensors: (types: GMSD...BA)
Connect plastic tube with an internal dia of 4 mm to pressure port "A". (Port "B" is not used.)
Stainless steel pressure sensors: (types: GMSD...MRE, GMSD...BRE, GMSD...BAE)
For measurements of over-, under- or absolute pressure screw sensor to G1/4" pressure terminal or plug
plastic tube to a suitable adapter.
6 The Serial Interface
By means of the serial interface and a suitable electrically isolated interface adapter (GRS3100, GRS3105 or
USB3100) the device can be connected to a computer for data transfer. With the GRS3105 up to 5 devices
of the GMH3xxx- series can be connected to one interface (see also manual of GRS3100, GRS3105 or
USB3100). To avoid transmission errors, there are several security checks implemented e.g. CRC.
The following standard software packages are available:
!EBS9M: 9-channel software to display the measuring values
!EASYCONTROL: Universal multi channel software (EASYBUS-, RS485-, or GMH3000- operation
possible) for real-time recording and presentation of measuring data of one GMH3xxx
device in the ACCESS®-data base format
In case you want to develop your own software we offer a GMH3000-development package including:
•a universally applicable Windows functions library ('GMH3000.DLL') with documentation
•Programming examples Visual Basic 4.0, Testpoint (Keithley Windows measuring software)
Note: The measuring and display range values read back from the interface are always in the
selected measurement unit (mbar, bar...)!
Supported functions:
Code Name/Function Code Name/Function
0 Read measurement value 180 Read kind of measuring of sensor
3 Read system state 194 Set display unit
6 Read min memory 199 Read kind of measuring of display
7 Read max memory 200 Read min display range
12 Read ID number 201 Read max display range
202 Read display range - unit32 Read configuration flag
BitCorrectToSealevel:32 (only abs. press.
Sensors) 204 Read display range – decimal point
160 Set configuration flag (see above) 208 Read # of channels
174 Clear min memory 220 Read altitude (only abs. press sensors)
175 Clear max memory 221 Set altitude (only abs. press sensors)
176 Read min measuring range 222 Read power off time (Conf-P.oFF)
177 Read max measuring range 223 Set power off time (Conf-P.oFF)
178 Read measuring range – measuring unit 240 Reset
179 Read measuring range – decimal point 254 Program version