H74.0.04.6C-06 Operating Manual GMH 3611 page 10 of 14
Air calibration without accessories ('Cal Air')
The electrode will be exposed to ambient air. To protect it
from draughts, wrap electrode in a paper towel. (before
calibration, expose sensor at least 15 minutes to the
ambient air, to let the temperature adjust and to dry
membrane)
Depending on the rel. atmospheric humidity [%] and the
temperature [°C] a small calibration error cannot be
avoided and will have to be accepted. The cooler the air is,
the smaller the deviation.
Recommended temperature: < 25°C.
For error compensation please refer to the opposite table.
Note: A correctly calibrated electrode shows 106 - 108% at air and at room temperature.
Air calibration for highly accurate measurements ('Cal Air')
The electrode is exposed to air with a relative atmospheric humidity of 100%.
Proceed as follows: Put some distilled water in a bottle. Close bottle and generate a water steam saturation
(100% rel. atmospheric humidity) in the remaining air by shaking it vigorously for approx. 3 minutes. Both
water and air temperatures should be identical. Open bottle and insert electrode so that the diaphragm is in
the air chamber.
Attention: By no means must the diaphragm get wet or be immersed in water. The bottle neck opening
should only be slightly larger than the electrode diameter; make sure to avoid over pressure in the container.
Note: A correctly calibrated electrode shows 106 - 108% at air and at room temperature.
Calibration in air saturated water (configuration: 'Cal AqUA')
This mode of calibration is more difficult than the ones already described. As water is easily oversaturated it
quickly results in calibration errors. In order to generate air saturated water, pure water (25-30°C) is poured
into another container from a height of 50 cm. This process is repeated 20 times. In order to let an
oversaturation evaporate, wait approx. 5 min. before carrying out the calibration (do not forget to stir).
Start calibration: press -key for 2 seconds
The display will show 'A,R , and as soon the values for oxygen and temperature are stable, the
calibration will be finished.
Then the electrode state resulting of the successful calibration will be shown for a short time (evaluation in
10% steps: xx% ELE[).
In case of error messages being displayed during the calibration process, please refer to our notes
at the end of this manual! If a calibration cannot be carried out after an extended period of time, at
least one of the measuring values is unstable (oxygen partial pressure, temperature).
Please check your measuring arrangements!
8.2 Evaluation of Sensor State (ELE[)
Watch sensor state: press key "CAL" shortly oncedisplay show for a short time xx% ELE[.
It will show the electrode state resulting of the last successful calibration carried out.
The valuation is displayed in 10 percent steps: 100% means optimal sensor condition. Lower values are
indicating that the sensor life time will be reached soon.
Remark: But also a erroneous pressure may be the cause of low valuation values.
8.3 Calibration Interval ([.INT)
You can enter the interval after which the device reminds you to recalibrate in the configuration.
The interval times should be chosen according to the application and the stability of the sensor.
“CAL” flashes on the display as soon as the interval has expired.
Deviations when carrying out an air calibration
without accessories,
O2-saturation= display value*corr. factor