
3
A few seconds after dispensing has ended, on the lower register,
the display switches from Reset Total to General Total: the word
RESET above the word TOTAL disappears, and the Reset Total is
replaced by the General Total.
This situation , where only “TOTAL” is displayed, is called
STANDBY mode. It remains stable until the user operates the
meter again.
B. MEASUREMENT UNITS CONFIGURATION
The METER comes with a menu through which the user can select
the main measurement unit, Quarts (Qts), Pints (Pts), Litres (Lit),
Gallons (Gal); The combination of the unit of measurement is
predefined according to the following table:
Reference No Unit of Measurement
Batch Total
Unit of Measurement
Total Register
1 Litres (L) Litres (L)
2 Gallon (Gal) Gallon (Gal)
3 Quarts (Qts) Gallon (Gal)
4 Pints (Pts) Gallon (Gal)
1. Wait for the METER to go to
Standby
2. Press the CAL and RESET keys
together. Keep these pressed until
the word “UNIT” appears on the
screen together with the unit of
measurement set at that time ( in
this example Quarts / Gallon )
3. Press RESET key to scroll among the four combinations of
units of measurement as shown :
12,345
123456
Unit QTS
GAL
QTS
GAL
SEQUENCE OF SETTING THE UNIT OF MEASUREMENT
L
L
R RR
TOTAL
C. NORMAL DISPENSING MODE
While the media is flowing through the meter, Batch Total and
Reset Total are displayed at the same time.
4. By pressing the CAL key at
length, the new settings will be
stored, the METER will pass
through the start cycle and
will then be ready to dispense in
the set units.
No new calibration is required
after changing the Unit of
Measurement.
L
L
12,345
2342,3
TOTAL
12.3 Reset
TOTAL
12,345
12.3 TOTAL
12,345
1. While in standby (i.e when the display shows TOTAL), press
the RESET button.
2. During reset, the display screen first of all shows all the lit-up
digits and then all the switched off digits.
3. At the end of the process, a display page is first of all shown
with the reset batch and the Reset TOTAL.
4. After a few moments, the Reset TOTAL is replaced by TOTAL.
E. RESETTING THE RESET TOTAL
The Reset Total can be reset by pressing the RESET key at length
while the display screen shows Reset TOTAL. The steps to be
taken are:
1. Wait untill the display shows Total only (standby mode)
2. Press the RESET key quickly.
3. The meter starts to reset the Batch Total.
4. While the display page showing the Reset Total is displayed,
press the Reset key again for at least 1 second.
5. The display screen again shows all the segments of the
display followed by all the switched-off segments and finally
shows the display page where the new Reset Total is shown.
D. RESETTING THE BATCH TOTAL
F. CALIBRATION
In standby mode, press the CAL key at length to see the current
calibration factor.
• Factory K Factor: Factory-set default factor.
It is equal to 1 (indicated as 1,000).
• User K Factor: Customized calibration factor,
meaning modified by calibration.
The meter has been calibrated at the factory under the following
operating conditions:
Fluid motor oil type : 10W40
Temperature : 20°C (68°F)
Flow rate : 5-25 litres/min
Calibration is needed to make the meter suitable for actual
conditions.