hadewe Medeas User manual

Art. 0980, 0981
FB 04-288 Rev. 2
Date: 23.01.2019
Manual
Medeas

2
Directory
Foreword 3
Safety Instructions 4
Cleaning & Disinfection 9
Discover Device 10
Before First Use 11
Area of Application 11
Operation 11
Usage 13
Operating Steps 13
Working Positioning 14
Suction 16
Exchange Micro Filter 16
Exchange Control Unit Filter 16
Clamping the Tool/Bur (Medeas Plus) 17
Clamping the Tool/Bur (Medeas Chrome) 17
Adjusting the Speed 17
Memory Keys 18
Foot Switch 18
Switsch on the Handpiece (Medeas Plus) 18
Select Menu 19
Help Menu 19
Settings Menu 20
Confirm Filter Change 20
Turn Light ON/OFF (Medeas Plus) 20
Pictures & Videos 21
Messages 24
Maintenance & Care 25
Cleaning the Chuck (Medeas Plus) 25
Cleaning the Clamping System (Medeas Chrome) 26
Upload Firmware 27
Maintenance Overview 27
Handling Instructions 28
Troubleshooting 29
Technical Details 30

3
Foreword
The Medeas footcare unit was produced according to strict quality criteria and
uses the latest technology in its design. The images in this manual may vary
slightly from the product you have bought as we are constantly developing our
products.
This user manual includes a detailed description and explanation of how to
use the Medeas unit. If you have any questions or suggestions, you can con-
tact us by telephone or email.
This user manual belongs with the Medeas unit. Keep it somewhere safe. If
you give this product to someone else, please include this document as it
includes important information on how to correctly use the product.
Short description of functions
The Medeas is a device for the treatment of callus and nails. The handpiece is
compatible with rotating tools such as diamond polishers, steel cutters and
ceramic cutters. These can be worked with up to a speed of 30,000 rpm
(Medeas Chrome) respectively 40,000 rpm (Medeas Plus) thus providing an
efficient way of removing tissue. The direction of rotation can be changed by
pressing a button. The integrated filtered suction function effectively sucks up
any dust particles. The rotation speed and suction function can be adjusted
using the keyboard. All other functions can be accessed with the touch display.
The device display will notify you when the filter needs to be changed and
when the device is due for inspection. Functions and maintenance steps will
be described with pictures and videos shown in the display. You can further
use an USB stick featuring your own pictures. For example, you may want to
show your customers the separate successful steps from treatment to treat-
ment.

4
Safety Instructions
The installation of the working area has to cor-
respond to the relevant regulations. Set up the
unit in such a way, that the air slits are not
blocked and the extracted air can escape well.
This device shall only be used by persons who
are familiar with feature and function of the unit.
This appliance can be used by children aged
from 8 years and above and persons with re-
duced physical, sensory or mental capabilities or
lack of experience and knowledge if they have
been given supervision or instruction concerning
use of the appliance in a safe way and under-
stand the hazards involved. Children should be
supervised not to play with the appliance.
Cleaning and user maintenance shall not be
made by children without supervision.
To avoid the risk of electrocution only connect
the device to the main power supply. If you pull
the power plug, it will stop the connection to
the main power supply. Set up the unit so that
the power plug is always accessible. The power
switch should also always be easily accessible.
Only switch on the unit, if the handpiece rests
secured in the handpiece holder, or if you hold it
safely in your hand.

5
Safety Instructions
Prevent any liquids from entering into the control
unit, since this could damage the electronic el-
ements. Clean the outside of the box only with a
damped cloth. Make sure, that you have
unplugged the unit before.
Only when pulling the mains cable, you can en-
sure that the device will be completely discon-
nected from mains. Check the mains cable
regularly for damages and in case of such, have
it changed or repaired by an authorized service
center.
Prevent your hair from wrapping itself around
moving parts. If applicable, wear a hair net.
When working with materials, which might cre-
ate dust or moisture, use a drilling unit with suc-
tion or spray technique. Wear a nose and mouth
protection. Always follow the safety instructions,
provided by the manufacturer of the materials
being formed and used.
For safety reasons, you must always wear dis-
posable gloves and protective glasses when
operating this unit. In addition, a face mask con-
forming to EN 14683 type IIR or EN 149 FFP 2
must also be worn.
When changing the filter, it is advisable to wear
a face mask conforming to EN 14683 type IIR
or EN 149 FFP 2.

6
Safety Instructions
Ensure that the patient’s foot and the handpiece
are held firmly in place during the operation. To
prevent injury, work with extreme caution on the
foot of the patient, and use only light pressure.
To ensure that the bur/tool sits firmly in the
chuck, do not work with tools which
•have an oily shaft.
•are worn.
•are bent.
Before using the equipment, assure that
•the tool can be pushed far enough into the
chuck
(test this with a dipstick and clean if neces-
sary),
•the inserted tool sits properly in the chuck
and will not cant/jam while clamping it,
•the tool is approved for using it at the max
speed (see manufacturer’s instructions), or
only use the speed required by the instru-
ment,
•the tool is absolutely dry.
For safety reasons, do not change the tool while
the motor is running.
Switch off the unit immediately if
•the bur comes off during the treatment itself,
•the motor blocks during the treatment.

7
Safety Instructions
While doing any footcare, only operate the unit
with the suction turned on.
Only use the suction function if you have in-
stalled the dust filter. Check filter bag every 10
operating hours. If it is full, please exchange it. A
filter bag should be exchanged at least after 50
operating hours.
Medeas Plus: The light on the handpiece is for
lighting the area being treated. Do not look di-
rectly into the light.
Never use the suction to vacuum any liquids (or
air/water fog which contains particles of liquids).
Always follow the operating instructions. Do not
carry out any repairs, modifications or mainte-
nance work yourself. This is only to be accom-
plished by an authorized specialist. The mains
cable may also only be changed by the manu-
facturer, or an authorized specialist. Mainte-
nance records may be supplied to repair work-
shops and trained technicians upon request.
Never remove the housing, especially not while
the equipment is plugged into the socket, since
some components are under electric voltage.
Any unauthorized or improper handling (e.g. try-
ing to repair the unit) of the equipment will im-
mediately nullify the guarantee/warranty, and
any other liabilities of the manufacturer.

8
Safety Instructions
Attention! Only use burs/tools which are permit-
ted for speed level saved under (12) to (14)!
When the device is brought out of standby
mode, it will run with the rotation speed and suc-
tion power used last. Take care to not start the
mill at an overly high rotation speed. Select the
required rotation speed before changing the
milling tool.
In case the touch display shows a damage or
does not work properly the device may not be
used any longer. Please send it to a qualified
service workshop for repair.

9
Cleaning & Disinfection
The device is designed to be compatible with rotating in-
struments which do not penetrate the skin or mucosal or
inadvertently may still cause injuries. For this application please
adhere to the cleaning and disinfection instructions provided be-
low. After every treatment please carry out the following steps:
•Remove the rotating tool (e.g. bur, capper etc.) first. Then
handle the tool according to the instructions of its manufac-
turer.
•Remove tissue remains by wiping off the unit with a mildly
moistened paper cloth. Afterwards clean all surfaces with dis-
infection cloth (e.g. Mikrozid AF Tücher of Schülke and Mayr).
There is no limit to the number of times this device can be
cleaned and disinfected during its recommended lifespan.

10
Discover Device
1 main power switch
2 increase speed
3 reduce speed
4 press briefly: switch unit on
press again briefly: turn on the handpiece
press long (keep pressed): switch unit off (standby)
5 speed display
6 time
7 date
8 changing the direction of rotation (left/right switch)
9 maintenance required
10 USB stick plugged in (green light)
11 Medeas Plus: light (yellow when switched on)
12,13,14 memory keys for speed/suction
15 suction stage set
16 bringing up menu with additional functions
1 2 3 4
5 6 7 8 9 10 11
12 13 14 15 16

11
Before First Use
Please read through the instructions
carefully before using the unit and
carefully note the safety advice.
Always keep this manual accessible
to any users of the drill.
Area of Application
This device should only be used by
people who have completed a foot-
care qualification. Before use the
operator should make themselves
familiar with the device by reading
this manual or being instructed by
another.
The device must only be used in
closed spaces. It may be used in
hospitals, in a commercial environ-
ment or in patient’s homes. The
device is not intended for use in
operating rooms.
The control unit can be set up in a
cabinet, or can be used as a porta-
ble unit (e.g. in a carry case). It is
suitable for the following scopes:
•foot care (milling, trimming and
polishing),
•manicure (grinding, milling and
polishing works).
The device is determined of
operating rotating instru-
ments, which will not
penetrate the skin or murcous
membrane if used appropriate, but
could evoke damages accidently.
Caution: In the area of
pedicure you have to work
with suction techniques.
Operation
The working area must conform to
the conditions described. Set up the
unit in such a way, that the air slits
are not blocked and the extracted
air can escape well. To guarantee
unhindered air flow there must a
gap of at least 12cm to the right of
the device. Position the device so
that the operating buttons are easy
to reach and the outlet of the hose
is a maximum distance of 60cm
from the patient’s foot. The unit
should be placed 30 - 90cm above
the floor.
Image: Example of positioning of footcare unit
in treatment room
Image: Bird's-eye view of work area

12
Before First Use
If you have a foot switch, insert its
plug into the port on the right of the
device which displays this symbol
.
Check if the filter bag and control
unit filter are correctly fitted. Ensure
that the tool is fixed.
On the left side of the device you
will find the main power switch.
Press this switch to turn the unit on.
The display prompts you to set the
device language:
This is followed by 5 other pages
with important clues. With (con-
tinue) you proceed to the next page.
On the last page press to start
the device. The counter for calculat-
ing the maintenance proposal will
automatically be set to zero.
The display will shut down now and
the unit is in standby.
Take the handpiece in hand. If the
hose is twisting the handpiece, the
exit point at the control unit can be
turned around to adjust:
To start the handpiece and the
suction, press button (4) on the
front page.

13
Usage
In the following section the order in
which you should use the device is
described. Additionally, the individ-
ual steps will be explained in detail.
Read the complete manual before
use.
Operating Steps
Once you have clamped in the
instrument, press the standby
button (4) to turn on the screen.
Select with the keys (2) and (3) for
the desired speed. The instrument
manufacturers often give out
recommendations for the optimum
speed. In any case, it must be
ensured that the maximum permis-
sible speed is not exceeded. Now
press the standby button (4) again
to rotate the handpiece.
By means of the icon (8) you can
choose the direction of rotation. As
a rule, it works clockwise. Also,
most rotational instruments (steel
instruments with cross cutting) are
optimized for the clockwise rotation.
In diamond cutters or grinders, the
direction of rotation does not affect
the removal rate. Working in re-
verse may be useful if the resulting
milling dust is swirled in a clockwise
direction in the field of view.
Now switch to suction. By means of
the icon (15) you can adjust the
strength of suction.
Frequently used speed and suction
capacity can be saved. First set the
desired speed and suction. Save
this by pressing one of the three
memory keys (12), (13) or (14) for 4
seconds. To recall the stored speed
and suction power quickly, briefly
press the corresponding memory
key.
After speed and suction power are
set you can start working.
If you want to interrupt or finish the
job, press the standby button (4).
Handpiece and suction switch off
then. You can put the handpiece on
the handpiece holder present at the
right side of the device. Alternative-
ly, by holding down the switch on
the handpiece or by pressing the
button on the footswitch ether can
be turned off.
By long pressing the standby button
(4) the device turns off and is in
standby mode.
We recommend you to disconnect
the device from the mains power
supply if it’s not going to be used for
a while. This will save electricity.
The mains plug must also be re-
moved when it comes to retiring the
device.

14
Usage
Work Positioning
The following images are examples
of how to work on various parts of
the foot. Always make sure that the
foot is firmly gripped and that you
can work freely and without strain.

15
Usage

16
Usage
Suction
Set the suction by using the icon
(15). The suction power can be
regulated in 5 steps. The suction
stage selected is displayed.
In case of overheating of the
suction motor, this switches off au-
tomatically. Allow the unit to cool
down, to put it into operation again.
Send the device for inspection if the
error occurs again.
Exchange Micro Filter
When changing the filter, it
is advisable to wear a face
mask conforming to EN
14683 type IIR or EN 149 FFP 2.
Only operate the unit with a filter
bag assembled. Control the bag
after every 10 operating hours. Ex-
change a full filter bag immediately.
A filter bag should be exchanged at
least after 50 operating hours.
At first, unfold the paper of the filter
bag (art. 5179) and push your finger
into the opening in the front.
Pull out the lid of the dust chamber
and remove the old filter. Hold the
filter in the middle to prevent it from
bending and sliding off.
Therefore, you avoid it from coming
off again.
It is important to use micro
filter bag (art. 5179). You
may not use paper filter
bag (art. 5115).
After the filter has been changed,
this must be entered in the device.
See chapter Confirm Filter
Change on page 20.
Exchange Control Unit Filter
This filter (Art. 3752) has to be
exchanged after every 200 operat-
ing hours. Remove the old filter and
place the new one in its.

17
Usage
Clamping the Tool / Bur
(Medeas Plus)
The handpiece is equipped with a
chuck, which can hold all burs with
a diameter of 2.35 mm (this is the
diameter usually used for tools in
podiatry). To open the chuck, push
the knob forward with your thumb
and leave it there.
Now insert the tool as far as possi-
ble. Pull back the knob and the tool
will be firmly clamped.
For security reasons, only
change the tool, if the
handpiece is turned off.
Clamping the Tool / Bur
(Medeas Chrome)
Clamp the tools/bur only when the
handpiece is not working. Insert the
tool/bur until it stops. Ones can use
all tools/burs which has a shaft
according to DIN EN ISO 1797-1
with a diameter of 2,35 mm.
For security reasons, only
change the tool, if the
handpiece is turned off.
Adjusting the Speed
Take the handpiece (with tools) in
the hand or put it into the handpiece
holder.
Select button (4) of the device. Now
you can set the desired speed with
the buttons (3) and (4) to adjust and
select with the icon (8) the direction
of rotation.

18
Usage
Memory Keys
Frequently used speed/suction
capacity can be saved. But there
are only 3 memory spaces availa-
ble. First set the desired
speed/suction. Save this by press-
ing one of the three memory keys
(12), (13) or (14) for 4 seconds. The
saved speed/suction is now directly
available on pressing the memory
key. To recall the stored
speed/suction quickly, briefly press
the corresponding memory key
(12), (13) or (14).
Foot Switch
Operation of the footswitch
Only use the original hadewe foot-
switch (Art. No. 0970) to avoid any
damages. Insert the plug of the foot
switch into the socket with the
symbol .
Function of footswitch
By holding down the footswitch, the
power can be switched ON/OFF.
Briefly press the left memory key
(12) to recall the speed/suction.
Another quick press of the central
lock key (13) will recall the
speed/suction. Another quick press
of the first memory key (12) can
also be used.
This only work with tools
that are approved for the
(12) to (14) stored speeds!
When the display is
switched off, you can turn it
on by keeping the foot
switched pressed (keep pressed
long).
Switch on the Handpiece
(Medeas Plus)
By holding down the switch, the
power can be switched ON/OFF.
Briefly press the left memory key
(12) to recall the speed/suction.
Another quick press of the central
lock key (13) will recall the speed/
suction. Another quick press of the
first memory key (12) can also be
used.
This only work with tools
that are approved for the
stored speeds!
When the display is off, you
can activate it by pressing
the footswitch.

19
Usage
Select Menu
In the select menu, you can use the
following functions:
Picture/Video: viewing pictures or
videos from your USB stick
Help: instructions, help tutorials,
troubleshooting
Light (Medeas Plus): turn light
ON/OFF
Maintenance: shows the last main-
enance and error messages
Change Filter: enter date of change
of the filter bag and display the date
of last change
Settings: for setting the time, date
and language
By means of this button you can
return to the previous menu.
Help Menu
Manual: Here is a short form of the
manual. This does not replace the
original manual in paper form, as
the latter contains more detailed
instructions and safety instructions.
Insert Burs: this clip demonstrates
the clamping of rotary tools
Clean Chuck: video clip that
demonstrates the cleaning of the
chuck
Filter Change: video clip that shows
how to change the filter bag
Troubleshooting: solutions to known
problems
By means of this button you can
return to the previous menu.

20
Usage
Settings Menu
Time: set the system time
Date: set the system date
Language: set the language for the
device
Safety Precaution: Set whether a
safety message is to appear from a
speed increase of 10,000 rpm, which
has been caused by the memory keys.
If you only use instruments that are
approved for a speed of 40,000rpm, the
safety message can be deactivated.
Confirm Filter Change
Every 50 hours of operation you will
get the message to "Check filters".
After the filter change, the following
must be entered into the instrument.
On the start screen, click the icon
for the select menu:
Then click “Change Filter”. Confirm
the filter change with YES.
Turn Light ON / OFF
(Medeas Plus)
To turn the light on or off, click the
top of the bar on the light icon:
Alternatively, you can switch the
light on or off in the menu. Just click
the icon for the Select menu on the
start screen:
Then click “Light”. Now confirm with
YES or NO.
Table of contents
Other hadewe Pedicure Spa manuals

hadewe
hadewe Argos 2 User manual

hadewe
hadewe Azuras PLUS User manual

hadewe
hadewe Veloria MD User manual

hadewe
hadewe HELIUS-40 User manual

hadewe
hadewe Argos2 User manual

hadewe
hadewe Veloria Chrome User manual

hadewe
hadewe Auroria MD User manual

hadewe
hadewe Auroria User manual

hadewe
hadewe Xantos-BT User manual

hadewe
hadewe Azuras CHROME User manual