Phosphate High Range Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 717 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• One bottle of Phosphate HR Reagent A
• Twenty packets of Phosphate HR Reagent B
• 1 x 1.5V AAA Battery
• Instruction Manual
HI 717
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.0” indicates that the sample absorbs
less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Cuvette cap.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button.
For more details about spare parts and accessories
see “Accessories”.
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at [email protected]
or view our worldwide contact list at www.hannainst.com.
Technical specifications:
Range 0.0 to 30.0 ppm
Resolution 0.1 ppm
Accuracy ±1.0 ppm ±5% of reading @ 25 °C / 77 °F
Light Source Light Emitting Diode @ 525 nm
Light Detector Silicon Photocell
Method Adaptation of the
Standard Methods for the Examination
of Water and Wastewater, 18Th edition
, Heteropoly-
molybdenum Blue method. The reaction between
orthophosphate (reactive phosphorus) and the reagent
causes a blue tint in the sample.
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 10 minutes of non-use
Dimensions 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
•Turn the meter on by pressing the button.
all segments will be displayed. When the
display shows “Add”, “C.1” with “Press”
blinking, the meter is ready.
•Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
•Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
•Remove the cuvette from the meter and
unscrew the cap. Add 10 drops of HI 717AS
reagent and the content of one packet of
HI 717B-25 reagent. Replace the cap and
shake gently until the reagent is completely
dissolved. Place the cuvette back into the
meter.
•Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 5 minutes and press the
button.
•The instrument displays the results in mg/L of
phosphate. The meter automatically turns
off after 10 minutes.
INTERFERENCE
Interference may be caused by:
Sulfide (S2¯)
Fluoride (F¯)
Arsenate (AsO4
3¯)
Ferrous iron (Fe2+) above 100 mg/L
10 mL
•It is important that the sample does not contain any debris.
•Whenever the cuvette is placed into the measurement cell, it must be dry
outside, and completely free of fingerprints, oil and dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
•Shaking the cuvette can generate bubbles, causing higher readings. To obtain
accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.
•Do not let the reacted sample stand for too long after reagent is added, as
accuracy will be affected.
•After the reading it is important to immediately discard the sample, otherwise
the glass might become permanently stained.
REAGENT SETS
HI 717-25 Reagent set for 40 Phosphate High Range tests
OTHER ACCESSORIES
HI 717-11 Phosphate HR Certified Standard Kit
HI 731318 Cloth for wiping cuvettes (4 pcs.)
HI 731321 Glass cuvettes (4 pcs.)
HI 731225 Cuvette black cap for checker HC (4 pcs.)
HI 731353 Cuvette seal cap for checker HC (4 pcs.)
HI 740028 1.5V AAA batteries (4 pcs.)
HI 93703-50 Cuvette cleaning solution (230 mL)
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
www.hannainst.com
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
Battery management
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•Turn the instrument off by holding the button until the meter shuts off.
•Turn the instrument upside down and remove the battery cover with a screwdriver.
•Remove the battery from its location and replace it with a new one, inserting the
negative end first.
•Insert the battery cover and replace the screw with a screwdriver.
IST717 03/13
Measurement procedure: Accessories:Tips for an accurate measurement

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HALO 2 HI9810332 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810332 User manual

Hanna Instruments HI 9026 User manual

Hanna Instruments

Hanna Instruments HI 9026 User manual

Hanna Instruments HI759 User manual

Hanna Instruments

Hanna Instruments HI759 User manual

Hanna Instruments HI 9835 User manual

Hanna Instruments

Hanna Instruments HI 9835 User manual

Hanna Instruments GroLine HI98168 User manual

Hanna Instruments

Hanna Instruments GroLine HI98168 User manual

Hanna Instruments HI 713 User manual

Hanna Instruments

Hanna Instruments HI 713 User manual

Hanna Instruments HI 83206 User manual

Hanna Instruments

Hanna Instruments HI 83206 User manual

Hanna Instruments HI88713-01 User manual

Hanna Instruments

Hanna Instruments HI88713-01 User manual

Hanna Instruments OHI 98140 User manual

Hanna Instruments

Hanna Instruments OHI 98140 User manual

Hanna Instruments HI 8014 User manual

Hanna Instruments

Hanna Instruments HI 8014 User manual

Hanna Instruments HALO 2 HI9810352 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810352 User manual

Hanna Instruments HI 120 User manual

Hanna Instruments

Hanna Instruments HI 120 User manual

Hanna Instruments HI96709C User manual

Hanna Instruments

Hanna Instruments HI96709C User manual

Hanna Instruments HI 96804 User manual

Hanna Instruments

Hanna Instruments HI 96804 User manual

Hanna Instruments HI 746 User manual

Hanna Instruments

Hanna Instruments HI 746 User manual

Hanna Instruments HI 93737 User manual

Hanna Instruments

Hanna Instruments HI 93737 User manual

Hanna Instruments HI 749 User manual

Hanna Instruments

Hanna Instruments HI 749 User manual

Hanna Instruments HI 93739 User manual

Hanna Instruments

Hanna Instruments HI 93739 User manual

Hanna Instruments HI 991401 User manual

Hanna Instruments

Hanna Instruments HI 991401 User manual

Hanna Instruments HI 96104C User manual

Hanna Instruments

Hanna Instruments HI 96104C User manual

Hanna Instruments HI 981402 User manual

Hanna Instruments

Hanna Instruments HI 981402 User manual

Hanna Instruments HI774 User manual

Hanna Instruments

Hanna Instruments HI774 User manual

Hanna Instruments HI 991001 User manual

Hanna Instruments

Hanna Instruments HI 991001 User manual

Hanna Instruments HI772 User manual

Hanna Instruments

Hanna Instruments HI772 User manual

Popular Measuring Instrument manuals by other brands

wtw Cond 3110 quick start guide

wtw

wtw Cond 3110 quick start guide

Benning MM 5-2 operating manual

Benning

Benning MM 5-2 operating manual

BIRD 7027 Series Operation manual

BIRD

BIRD 7027 Series Operation manual

Colibri SMARTHUB user manual

Colibri

Colibri SMARTHUB user manual

Ametek WDG-V user manual

Ametek

Ametek WDG-V user manual

Citec LOOP420 instruction manual

Citec

Citec LOOP420 instruction manual

Mezon DSTM500 user manual

Mezon

Mezon DSTM500 user manual

Flexim FLUXUS F704**-NN Series user manual

Flexim

Flexim FLUXUS F704**-NN Series user manual

pH ELECTRODE PE-03 Operation manual

pH ELECTRODE

pH ELECTRODE PE-03 Operation manual

Honeywell DVC100 (M) MK2 installation guide

Honeywell

Honeywell DVC100 (M) MK2 installation guide

Milwaukee MW101 user manual

Milwaukee

Milwaukee MW101 user manual

Decagon Devices KD2 Pro Operator's manual

Decagon Devices

Decagon Devices KD2 Pro Operator's manual

Doepke DCTR B-X Installation and operating instructions

Doepke

Doepke DCTR B-X Installation and operating instructions

Smart Storm Universal Smart Meter instruction manual

Smart Storm

Smart Storm Universal Smart Meter instruction manual

M-system AS4M instruction manual

M-system

M-system AS4M instruction manual

Rixen SR-496 instruction manual

Rixen

Rixen SR-496 instruction manual

Vetek MO.DTR.526.R6 operating manual

Vetek

Vetek MO.DTR.526.R6 operating manual

ThermoWorks ThermaQ 2 operating instructions

ThermoWorks

ThermoWorks ThermaQ 2 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.