Henry Schein Maxima Turbo User manual

Maxima Turbo - 1
Turb o
Class B Sterilizer
Esterilizador Clase B
Stérilisateur de classe B
570-3362
INSTRUCTIONS FOR USE
INSTRUCCIONES DE USO
DIRECTIVES D’UTILISATION

2 - Maxima Turbo
INDICATIONS FOR USE
The Maxima Turbo is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by
means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Max-
ima Turbo has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of
such loads may result in incomplete sterilization and/or damage to the autoclave. Please refer to the table below for program name, load description,
sterilization temperature, exposure time, drying time and maximum load.
MAXIMA TURBO
Program
name
Load description Sterilization
temperature
Sterilization time Drying time maximum
load
Maximum load
134˚C solid objects, small
porous objects, sim-
ple objects recessed,
narrow-clearance items,
dental handpieces, and
textiles; wrapped and
unwrapped
134˚C
(273˚F)
4:00 minutes 3 minutes 0.5 Kg
/1.1 lbs
121°C solid objects, small
porous objects, sim-
ple objects recessed,
narrow-clearance items,
dental handpieces,
textiles, and plastics;
wrapped and unwrapped
121°C
(250˚F)
30 minutes 5:30 minutes 0.5 Kg
/1.1 lbs
134°C
FAST*
solid objects, non-po-
rous objects, simple
instruments (such as
scissors, handles, pliers,
chisels, probes, etc.),
and dental handpieces;
unwrapped
134˚C
(273˚F)
4:00 minutes N/A 0.5 Kg
/1.1 lbs
* - Immediate Use Steam Sterilization cycle
The sterilizer is suitable for use in the vicinity of other powered medical products.
The Maxima Turbo device may not be used to sterilize
liquids, biomedical waste or pharmaceutical products.
The device is intended for professional use by properly trained staff only.
SYMBOLS USED ON THE DEVICE
This symbol is located on the front of the appliance, on top of the drawer front, and advises extreme caution due to the high temperature of
the compartment and its immediate surroundings.
This symbol is located on the unit’s rating plate and identifies its individual serial number.
This symbol is located on the unit’s rating plate and identifies the year the unit was manufactured.
This symbol is located on the equipment rating plate and identifies the equipment manufacturer.
This symbol can be found on the unit’s rating plate and requires the user to read and follow the information in this manual.
Used electrical and electronic equipment must not be placed, thrown away or stored with other waste. Used equipment should be deliv-
ered to a local collection point for used electrical equipment, which is registered with the respective environmental protection office and
conducts selective waste collection.

Maxima Turbo - 3
INDICATIONS FOR USE
• Maxima Turbo complies with IEC 60601-1-2:2014 (Edition 4.0)
• Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify that they are operating normally. Max power cord length is 160 cm.
• Use of accessories and cables other than those specified or provided by Maxima Turbo could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
• Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the Maxima Turbo S, including cables specified by Maxima Turbo. Otherwise, degradation of the performance of this
equipment could result.
• The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
• The user is responsible for the installation, correct operation and maintenance of the device in accordance with instructions provided in this user
manual. If needed, contact the service or the supplier of the device.
• The sterilizer is not intended for sterilizing liquids, biomedical waste or pharmaceutical products.
• The sterilizer must not be used if explosive gases or vapours are present in the air.
• After the cycle is completed, the load is hot. Remove tools or packs from the chamber using appropriate thermal gloves or equipment that prevents
burns.
• Do not remove the rating plate or any other elements of labeling from the device.
• Follow guidelines for preparing tools for sterilization.
• Pouring water or other liquids on the device may cause a short-circuit.
• Prior to inspection, maintenance or servicing, turn off the device and disconnect it from the power source.
• Servicing may only be performed by trained service personnel and using original replacement parts.
• The device is intended for indoor use only.
• Max working high elevation of the device is 2,000 m above sea level.
• Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected.
• Overvoltage Category II.
• If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
• Use only detachable cord that rated equal or greater to the equipment electrical ratings.
Read this Operating Manual carefully before using this device. Install and operate the device strictly as specified herein. Comply with all safety
requirements for the device. This will ensure proper and safe operation of this device. Any other application, inconsistent with this manual, may lead
to dangerous accidents. Restrict unauthorised personnel access to the device and train the personnel handling the device. An operator of this device
is any person who, by training, experience and knowledge of applicable reference standards, manuals and occupational health and safety regulations
has been authorised for the essential operation with the device and who is capable of identifying and avoiding the hazards related to operation of this
product.
Always append this Operating Manual with the device if transferred to a new owner. The Operating Manual contains detailed information about
assembly, installation, initial start-up, use, repairs and maintenance of the device. If the device is used as intended, this Manual will provide sufficient
guidance to qualified personnel. Keep this Operating Manual close to the device and easily accessible at all times. As required by continuous improve-
ment of the product, the manufacturer has the right to amend this Manual or make changes to the device without prior notice. Henry Schein shall not
be liable for damage incurred during the wait for warranty service, any damage to the Customer’s property other than this device, or errors caused by
improper installation and/or improper operation of the device.
Detailed recommendations, coutnerindications and warnings are
described in the relevant sections.
PRODUCT STRUCTURE AND ACCESSORIES
22 in (561 mm) 9.9 in (252 mm)
6.4 in (162 mm)

4 - Maxima Turbo
Hepa filter
seat
Power socket
WATER IN feed
water inlet
USB socket
Condensate
outlet WATER OUT
Main switch
HEPA Filter Power Cord Water and condensate connection cables, rubber
plugs for water/condensate containers
Operating Manual (PDF, on USB drive) Data viewer software
* The electronic versions of the documents are stored on USB stick.
DEVICE SETUP
Please read these operating instructions carefully before using the Maxima Turbo. During operation, observe the guidelines in these instructions as
well as the applicable local health and safety rules. The Maxima Turbo unit is designed for self-assembly by the user and do not require specialized
installation on site. The user is responsible for proper installation of the unit on site in accordance with these instructions.
HEPA filter installation.
To protect the component during transport, a HEPA filter is not installed in the device. It was placed inside the chamber, on a tray. Take it out of the
chamber, take it out of the bag and by yourself, carefully screw it into the slot provided for this purpose (see fig.). The filter should be screwed in by
hand (approx. 7 turns) until resistance is felt - further screwing in will damage the filter and immobilize the sterilizer.

Maxima Turbo - 5
Feed water quality
It is unacceptable to use any liquids, solutions, chemicals, additives to the feed water - the device may be powered only with demineralized or distilled
water of appropriate quality.
No chemicals or additives must be added to the sterilization water, even if they are specifically intended for use in steam generators, steam production
or for use as additives in sterilization, disinfection, cleaning or corrosion protection.
The use of incorrect quality water or solutions other than water reduces the effectiveness of sterilization and leads to damage to the device and loss of
warranty.
The total mineral content of the sterilization water must be less than 10 ppm or, in the case of conductivity measurement, it must be less than 15 μS/
cm - therefore tap water cannot be used as feed water for the device.
The table below shows recommended water hardness and conductivity parameters. Permissible parameters of water used for sterilization
• Hardness <0.02 mmol / l
• Conductivity (at 20 ° C) <15 μS / cm
• PH - from 5 to 7.5

6 - Maxima Turbo
• Chemical additives - none
Water Filter & Water Quality Indicator
To ensure the proper feed water quality a water filter and water quality indicator has been attached to the inlet line. Any usage of improper water will
begin to discolor the indicator. Check the indicator frequently to ensure proper water is being used. The use of incorrect quality water or solutions
other than water reduces the effectiveness of sterilization and leads to damage to the device and a loss of warranty.
If the indicator is completely discolored the unit will need to be serviced and the inlet line will need to be replaced.
New Indicator Needs to be replaced
Connecting the WATER IN feed water
• Connect the blue connection hose to the blue feedwater connection, which is marked on the rear panel of the WATER IN device. The tubing should
be inserted into the connector, to the depth of the black line (placed on the tubing).
• The other end of the blue tube should be immersed in the tank with the feed water. The device is equipped with a water suction pump, so it is not
necessary to place the water tank above or on the same level as the device.
• To secure and immobilize the water supply hose to the machine, use the plug supplied with the hose and insert the plug into the opening of the
water supply tank. An unsecured tubing may jump out of the water and cause sterilization errors.
• To prevent the hose from coiling in the water tank, install the included weight with a rubber ring at 2 cm from the end of the hose (fig.)
• The minimum water load in the tank is 300 ml.
• Remember and check that the blue tube is always immersed in the water.
• Check the water level in the tank on a regular basis, depending on the frequency of the processes
WATER OUT Wastewater connection
• Connect the red wastewater hose to the orange wastewater connector that is marked on the rear panel of the WATER OUT device. The tubing
should be inserted into the connector, to the depth of the black line (placed on the tubing).
• Wastewater should be discharged directly to the water and sewage system or to a special wastewater tank. If a tank is used, the other end of the
red tube should be placed in the wastewater tank. To secure and immobilize the water drain hose from the machine, use the plug provided with the
hose and insert the plug into the opening of the wastewater tank. The unsecured tubing may jump out of the tank and flood the room.
• Make sure that the red tube is never immersed in water, otherwise the water will not drain properly, causing sterilization errors.
• The wastewater tank or drainage outlet must always be located below the unit.
• When using wastewater tanks, we recommend that you use tanks with the same capacity as the feed water tank. Their simultaneous replacement
will protect against the possibility of wastewater overflow.
• The water level in the tank should be checked on a regular basis, depending on the frequency of the processes.

Maxima Turbo - 7
0.4 in
2 in
Setting up the device
• The unit should be placed on a flat, level surface. Do not use the unit if it is at an angle.
• Only demineralized or distilled water should be used in the devices. The use of incorrectly specified water (see “Supply water quality”) reduces the
sterilization efficiency and leads to loss of warranty due to damage to the device.
• Leave a space of 5 cm (2 in) at the rear of the appliance and at least 1 cm (0.4 in) at the sides of the appliance from walls or other elements for
proper ventilation and fresh air access.
• The unit should be located to provide easy access to the main switch located on the rear panel of the unit.
• Do not place the unit near a sink with water or other places where the unit could be flooded - danger of short circuit.
• The device should be installed in a well-ventilated room, far from heat sources and rooms where mixtures of gases or liquids or other dangerous
factors may arise.
The following conditions must be ensured for safe and efficient operation of the device:
• Ambient temperature from + 5 ° C (41°F) to + 25 ° C(77°F),
• Relative humidity from 0 to 90%,
• Storage temperature from -20 ° C (-4°F) to + 60 ° C (140°F),
• 0 to 90% relative humidity.

8 - Maxima Turbo
Correct connection of the hoses to
the feed and wastewater connec-
tions
Correct installation of the weight
with the securing ring
Correct positioning of the
hoses in the water tank

Maxima Turbo - 9
PREPARATION AND LOADING OF INSTRUMENTS
The devices are suitable for sterilization of loads for which steam sterilization is stipulated. The instruments must be sterilized only if they are clean
and dry. Therefore they must be washed and disinfected before being loaded onto the tray in accordance with applicable regulations. Residues of used
chemicals or other solid particles may render the sterilization process impossible or even damage the device. In addition, sterilization of instruments
that have not been previously cleaned and disinfected is a biological hazard and can lead to damage (both to the instruments and to the sterilizer). For
instruments that must be lubricated, use steam sterilization lubricants. Excess lubricant must also always be removed.
• The 134 ° C (273°F) FAST program is intended only for sterilization of unwrapped instruments.
• After sterilization with 134°C (273°F) FAST the instruments are moist - it is recommended to leave the door open for a few minutes to evaporate
excess moisture.
• After sterilization of unpackaged loads, they are intended for immediate use.
Preparation of packages
It is recommended to use sterilization packaging that meets the requirements of EN ISO 11607-1: 2020 and EN 868-2: 2017. Appropriate packaging
should:
• Ensure good penetration of the sterilizing agent into the inside of the package.
• Provide resistance to damage during the sterilization process.
• Ensure tight and durable sealing of contents and safe removal for use.
• Provide a barrier to microorganisms and unwanted substances and contaminants.
• Use disposable sterilization sleeves (intended for steam sterilization) as recommended by their manufacturer.
• Sleeves should only be filled to about 3/4 full to allow for proper sealing and minimize the risk of damage to the package.
• A distance of at least 30 mm should be maintained between the seals and the sterilized load.
• The sharp edges of the load should be protected to avoid damaging the packaging.
• The packaging material must not be placed too loosely or be tightly stretched so as not to affect pressure variations during sterilization.
• a label with information about the contents of the package, the code of the packer, date of sterilization and use-by date as well as parameters of the
sterilization process should be placed on the package.
Arranging packages on the sterilizer tray (sterilization of packed items)
• Packets on the tray should be placed so that the paper side touches the paper side, because penetration of the sterilizing agent and air exchange
can only take place through the paper.
• The packages should be placed on the tray in such a way as to eliminate the contact of the package with the door seal and the phase of the steril-
ization chamber - failure to comply with this rule may cause the chamber to become unsealed and the sterilization cycle incorrectly carried out
• The edges of the packs must not protrude beyond the sterilizer tray, as this would cause the chamber to leak and the sterilization cycle to fail
• If the sterilizer chamber is heavily loaded, the first packs must face the foil side towards the bottom of the tray (this ensures faster and more effi-
cient drying of the packs).
Arranging tools on the sterilizer tray (sterilization of loads without packaging)
• In the case of sterilization without packages - the instruments must be placed in such a way that they do not have direct contact with each other, no
element of them falls into the openings of the tray, and does not rest against the edge of the tray or protrude beyond its outline.
• Failure to follow the above recommendations may cause permanent and irreversible damage to the sterilization chamber phase, which will result in
the lack of tightness of the sterilizer and loss of warranty.
Do not exceed the maximum load weight: 500 g for the Maxima Turbo device. In order to verify the correctness of the performed
sterilization, it is recommended to place a sterilization chemical indicator in the chamber for each process, which discoloration
during properly conducted sterilization.

10 - Maxima Turbo
Sample placing of sterilization packages.
Plastic
Paper
Plastic
Plastic
Paper
Plastic
Example of solid load.
OPERATING THE DEVICE
First run
Turn on the device using the main switch located on the rear panel of the device. Make sure that the supply and wastewater hoses are connected
correctly, and that there is water in the supply tank and the wastewater tank is empty. Place the load on the tray, slide the drawer in and lock it, i.e.
turn the knob clockwise. The sterilizer provides audible information about changes and the completion of the process.

Maxima Turbo - 11
Program selection
Depending on the type of load to be sterilized, the user must select the appropriate program - in accordance with the instructions of the sterilizer
manufacturer and the load manufacturer’s recommendations.
Maxima Turbo program 134°C (273°F) FAST 134°C (273°F) 121°C (250°F)
Type of load Unwrapped instruments only Wrapped or unwrapped instruments Wrapped or unwrapped instruments
Process temperature 134°C (273°F) 134°C (273°F) 121°C (250°F)
Pre-vacuum number 133
Sterilization duration 4:00 min 4:00 min 30 min
Drying duration -3 min 5:30 min
Total process duration* 100g: 7 min 100g: 13 min 100g: 45 min
Class S B B
Load Description Solid Objects, non-porous objects,
simple instruments (such as scissors,
handles pliers, chisels, probes, etc.)
and dental handpieces; unwrapped
Solid objects, small porous objects,
simple objects recessed, nar-
row-clearance items, dental hand-
pieces, and textiles; wrapped and
unwrapped
Solid objects, small porous objects,
simple objects recessed, nar-
row-clearance items, dental handpiec-
es, textiles, and plastics; wrapped and
unwrapped
After switching on the device, the start screen appears on the display. To go to the next screen, press the screen once with your finger (anywhere).
After switching on the device, the start screen appears on the display. To go to the next
screen, press the screen once with your finger (anywhere).
On the next screen, we can go to sterilization, test the device or open the information menu. The following sterilization programs can be selected from
the main menu: 121 ° C (250°F) , 134 ° C (273°F) , and 134 ° C (273°F) FAST.
Test programs Sterilization programs
Selection of sterilization programs
Settings Menus

12 - Maxima Turbo
When the chamber is opened, the DOOR OPEN symbol flashes.
When the chamber is closed by turning the locking knob clockwise, the START symbol
appears on the display, which informs that the chamber is properly closed.
If the USB memory is not inserted in the device, the USB disk symbol is not displayed in the
lower right corner of the screen.
A message appears that USB memory is missing. The program data will not be saved. You
can continue working without saving data to the USB memory stick by pressing the YES
box, or abort the work by selecting the NO box to place the stick in the port and start the
program steps from the beginning.
It is recommended to use a USB drive with each of the processes. The data stored on it will
allow you to print reports from selected processes.
If work continues or the START field has been selected, a symbolic pressure diagram of the entire process is displayed on the screen with the current
program progress highlighted against a background and information on the next process steps in the upper left corner of the screen. During the
program run, the screen displays the temperature of the selected sterilization program or the current temperature in the process chamber in the lower
left corner, the current pressure in the chamber in the lower right corner, and the time remaining until the end of the process in the upper right corner
of the screen. This is an estimated time, which may be increased due to the weight and type of the load.

Maxima Turbo - 13
Program stage
The current temperature
in the chamber
Temperature of the selected process
STOP button
Approximate time to the
end of the process
Current pressure in the
chamber
During the course of the program, in place of the START field, the STOP field is displayed, which allows the user to stop the process at any time. In
the upper left corner of the screen, the names of the individual stages of the program are displayed consecutively, e.g. chamber lock, heating of the
working chamber.
If the process was successful, the display alternately shows information screens about
the end of the process and the sterility of the load and about the possibility of opening the
chamber of the device. By pressing the FINISH field, you go to the welcome screen.
WARNING! After the process, the chamber, tray and load are hot. Be especially careful and use protective gloves to remove the
load or wait until it cools down. In the 134C FAST program, instruments are hot and wet after sterilization.
Test programs
The devices are equipped with special test programs to verify the correct operation of the sterilizer.
Maxima Turbo Bowie & Dick / Helix Vacuum test
Process temperature 134°C (273°F) -
Number of pre-vacuums 3 1
Sterilization time 3.5 min -
Drying time 3 min -
Total process time 15 min 16 min
By pressing the Test field you go to the test programs menu. From this level, we can choose
between the Vacuum test and Helix / B&D test programs. You select the appropriate pro-
gram by pressing the required field on the display.

14 - Maxima Turbo
When the device working chamber is closed, the word DOOR OPEN changes to START and
by pressing this field the selected test program is started.
If a USB flash drive has not been inserted into the unit, the USB symbol is not displayed
in the lower right corner of the screen and a message indicating no USB flash drive is
displayed. The test program data will not be saved. You can continue without writing data
to the USB flash drive by pressing the YES box or abort by selecting the NO box to place the
USB flash drive in the port and start the program from the beginning.
If work continues or the START field has been selected, a symbolic pressure diagram of the entire process is displayed on the screen with the current
program progress highlighted against a background and information on the next process steps in the upper left corner of the screen. During the
program run, the screen displays the temperature of the selected sterilization program or the current temperature in the process chamber in the lower
left corner, the current pressure in the chamber in the lower right corner, and the time remaining until the end of the process in the upper right corner
of the screen. This is an estimated time, which may be increased due to the weight and type of the load.
Vacuum test
The vacuum test should only be performed on a cold device before starting work. During the test, the device verifies:
• Vacuum pump capacity.
• Tightness of the pneumatic system.
After selecting a vacuum test program and starting it with the button, the vacuum test
program run screen is displayed, with Information about loss of pressure in the working
chamber and the duration of the test.
[Vacuum test PASSED]
When the test program ran successfully.

Maxima Turbo - 15
[Vacuum test FAILED]
When the test program did not run successfully.
After pressing the CONTINUE field, the welcome screen is displayed.
During the vacuum test the sterilizer chamber must be completely dry and cold. If not, the vacuum test may not be reliable even if the sterilizer is fully
operational. When the test is complete a message indicating the result will appear on the display. If the result is negative check, clean or replace the
seal, clean the front edge of the chamber and repeat the test. If the Vacuum test fails again, contact the supplier or manufacturer.
Bowie & Dick test
The Bowie & Dick test, also known as the steam penetration test, simulates a small, highly porous load. To perform the test it is necessary to have a
special test packet and placed inside the chamber. The package is not an accessory of the device, the user should purchase it on his own.
This test evaluates the device’s performance in sterilizing loads of porous objects:
• Pre-vacuum performance and steam penetration.
• Temperature and pressure of saturated steam reached for a certain period of time.
How to run the test:
• The test must be carried out with an empty chamber in accordance with EN 13060.
• Place the Bowie-Dick test package in the chamber in the center of the tray.
After selecting the Helix / B & D test program and starting it with the START button, the
program sequence screen is displayed.
Information about the process parameters is displayed.
The Helix / B & D test program can be stopped at any time by pressing the STOP field,
which is associated with incorrect termination of the test.
When the test program is complete, alternating screens are displayed:
[READY] Test complete / [READY] Chamber is safe to open. The process chamber of the sterilizer can be safely opened.
The control test should be removed.
WARNING! The package will be hot.
To correctly interpret the test, refer to the instructions provided by the test manufacturer.
Open the package and remove the indicator chemical from the center of the package.

16 - Maxima Turbo
Positive result
The chemical indicator turned a uniform dark color over the
entire surface.
Negative result
A bright field remained in the middle of the test because of the
remaining air inside the tested device.
Any color change, uneven coloring of the test, indicates the presence of air during the test cycle due to a malfunction of the sterilizer. If the test result
is abnormal, check the expiration date of the test pack and repeat the test.
Information menu
The info menu is accessible by pressing the Info. button.
This level displays information about the unit type, serial number, number of processes performed and available USB memory space for saving
process data, as well as the COUNTERS service menu - process counters for filter changes. It is also possible to change the date and time. To set the
date or time touch the digits on the display. When a certain field is selected, it starts flashing and the arrows for changing the value up or down are
displayed. This is how to set the date and time correctly. Pressing the number again confirms it and you can move on to changing the settings of the
next parameter. In the same way, you can select the language by clicking on its abbreviation.
The button labeled B turns off and on the blue backlight in the depth of the screen. Clicking the LED button launches the backlight control menu on the
sides of the device.
Clicking a button turns off the button sounds.
Setting the
date and
time
LED illumina-
tion settings
Counter
settings
Disable shaded
background
Autoclave
serial number
Process
count
Key sounds
activation
Language
choice
LED lighting
LED lighting has two modes:
• Free mode, where the user (by moving the sliders) freely sets the colors, intensity and brightness of the light to their preference.
• Continuous mode that indicates the stages of the entire sterilization process with colors.

Maxima Turbo - 17
Decorative illumi-
nation
Decorative illumination
color change
Illumination during a
process step
Turning the LED illumina-
tion on and off
Decorative illumination color
change
Counters
The Maxima Turbo sterilizer records the number of processes performed. This lets you know when you need to replace consumable parts.
Counter section with the counters reset to zero. The number of processes performed is on the left, while on the right is the number at which the com-
ponent should be replaced or a service inspection should be performed. After replacing the filter, the user can reset the values by pressing the RESET
button.
During the first process (Vacuum, Helix, FAST, 134/273, 121/250), the unit will record the current date as the date of the first process (bottom line on
the counter screen). The device will count processes and days since the first startup.
Displaying the counter values in yellow or red does not block the operation of the unit. However, exceeding the required replacement time may have a
significant impact on the operation of the unit and the sterility of the load. For replacement of individual components, please contact the manufacturer
or supplier.
No. Name Replacement frequency
(cycles)
Yellow
Early replacement call
(cycles)
Red
Latest replacement after
(cycles)
1HEPA filter and Door with
gasket 1000 980 1000
Restart
The restart of the process is forced when the user stops the process by pressing the STOP field, in the case of a power outage or an error during the
process, for example lack of supply water.
If the STOP field is selected, the following messages are displayed alternately:
•The user has stopped the process.
•Equalizing the pressure in the working chamber.
• Process is incorrect, which means that the load is not sterile.
When the pressure in the working chamber is equalized, the following messages appear alternately on the display. Now you can open the device
freely. After opening the chamber, the screen appears.

18 - Maxima Turbo
By selecting the field, we can return to the welcome screen. In the event of an error, we must additionally enter the 4-digit security code 0000. Enter-
ing this code is tantamount to the operator’s declaration that he is aware that the sterilization process has not been carried out properly and that the
batch is non-sterile.
DEVICE MAINTENANCE
In order to ensure the correct operation of the device, the user is obliged to perform the following maintenance actions.
Cleaning the tray
Keeping the tray clean helps to ensure proper operation of the appliance. The tray and its technical condition are a good indicator of using the correct
water. A stony, brown tray indicates the use of poor quality water. It is recommended to clean the inside of the tray on a weekly basis with a mild
detergent that does not add chlorine and does not react with aluminum. After cleaning, the tray requires thorough rinsing with water. In order to clean
the tray properly, it should be removed from the front of the device.
Lift the tray up gently and pull away from the front face. The mounting pins have notches
into which the drawer fits. Before reinstalling the tray into the unit, drain the tray and slide
it over the front face pins and press down gently to lock.
Cleaning the process chamber
Keeping the chamber clean helps to ensure proper operation of the device. It is recommended to clean the inside of the process chamber once a week
with a mild detergent without the addition of chlorine. After cleaning, the chamber should be wiped dry with a soft cloth. Cleaning should be per-
formed on a cold chamber.
Cleaning of external surfaces
Cleaning the external parts of the device should be done with a soft cloth moistened with water and a mild detergent (no chlorine added and not
reacting with plastics, varnish coatings, aluminum). Strong detergents should not be used. The use of mild detergents to maintain the device does not
affect the risk of toxic components coming into contact with device components.
Cleaning the gasket
It is recommended to clean the gasket each time after 100 processes. Use warm, clean water and microfiber to clean the gasket (microfiber with sil-
ver particles is allowed). Blunt and sharp cleaning tools are not allowed. Cleaning with chemicals is not allowed. Carry out cleaning on a cooled down
device, after opening the drawer. Be careful not to bend the drawer. After cleaning, leave the device open until the gasket is dry. During this time,
protect the device against damage.
Replacement of consumables
The sterilizer is equipped with a high performance sterilization system. A message on the screen informs the user when each component should be
replaced. If the sterilizer is in regular operation, alternating replacement screens will appear after the welcome screen has been pressed. The replace-

Maxima Turbo - 19
ment screens are described in detail in section “Warning Messages and Error Codes”.
Cleaning the water container
In order to ensure the correct parameters of the water supplying the device, it is recommended to check the water tank at least quarterly. If contami-
nation is found, the tank should be emptied, cleaned and refilled with new water.
Consumables
To ensure proper functioning of the sterilizer, it is recommended that the wearing parts be replaced according to the schedule below. And periodic
inspection of the individual components of the sterilizer in accordance with the following guidelines.
Name Replacement frequency
HEPA bacteriological filter Every 1000 cycles or every 12 months
Connection / drain hose If damage is observed or once a year
Plugs for water / condensate containers If damage is observed
Door with gasket Every 1000 cycles or every 12 months
Element to be controlled Frequency of inspections
Front gasket weekly or in the event of incorrect operation - performed by the user
Bacteriological filter every week - performed by the user
Connection / drain hose weekly or in the event of incorrect operation - performed by the user
Container stoppers weekly - performed by the user
Data archiving
The course of each performed sterilization is automatically saved on a data carrier (USB stick).
• The memory slot is located on the rear panel of the device.
• It is recommended to periodically archive the data on another medium e.g. computer, laptop.
• The USB flash drive should not be removed from the slot during the process.
Data Viewer software
DataViewer software allows you to view and archive sterilization programs on your computer and print them out.
Installing the software
• The software is delivered with the device and is on a portable drive - USB flash drive.
• To install the software, double click on the software installation file. After this operation an installation window will be displayed regarding language
selection.
• After confirming, you must accept the license terms for the software being installed.
• You will then be prompted to place a shortcut to the software on your computer desktop.
• After clicking “Next>”, the folder where the program should be installed is selected.
• After making your selection, click “Next”. By clicking the Install button, the Data Viewer software will be installed.
Program structure and main functionalities
The main window consists of three main areas.

20 - Maxima Turbo
Tree of all process-
es that have been
synchronized with the
flash drive - have been
sorted by execution
dates
Temperature and pressure graph with main autoclave and process data (date and
number).
Data on the duration and achieved
parameters of each process step.
The most important sterilization
parameters.
Possibility to save notes for each
process
The function buttons, eg “PDF report”, which will enable printing the protocol from the process, are marked in dark blue.
Drop-down menu:
By clicking on the File window, we have access to the following options:
• Loading the saved process flow from a flash drive or from another location.
• Print a saved process.
• Create a report to a PDF file.
• Exporting data to a file in order to send a package with data to the manufacturer in case of problems.
• Data export to CSV format.
By clicking on the Tools window, we have access to the following options:
• Synchronization of all files with saved processes after selecting memory location on pendrive (this process, depending on the number of steriliza-
tions and tests performed can take up to several seconds).
• Searching for any saved process from the database.
• Adding your own logo to PDF reports.
Search
The program allows you to search for processes by:
• Date range.
• Process number.
• Process type.
• Process outcome (success/fail).
PDF report
The program allows you to generate a report on each process performed by the autoclave. It contains all the necessary process data and the steriliza-
This manual suits for next models
1
Table of contents
Languages:
Other Henry Schein Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Agilent Technologies
Agilent Technologies MX3000P installation instructions

Finisar
Finisar WaveShaper A Series user manual

Agilent Technologies
Agilent Technologies 490 user manual

Vernier
Vernier STIR quick start guide

J.P. SELECTA
J.P. SELECTA UNITRONIC-ORBITAL manual

ChromaCon
ChromaCon Contichrom CUBE 30 System manual

Yamato
Yamato DL410 instruction manual

V&P Scientific
V&P Scientific VP 741ECE-3Z-KIT-96 Operation manual

Celexon
Celexon Microphone UV Sterilizer Professional Operation manual

Maxtec
Maxtec MaxFLO2 Operating manual & instructions for use

Bradley
Bradley S19-876 Installation

Milwaukee
Milwaukee MI456 PRO instruction manual