Follow these precautions to ensure safe operation and to obtain the full benefits
of the various functions.
Instrument Installation
General Specifications
Measurement functionality
Usage Notes
• Before using the device, make sure that the insulation on the
cable is undamaged and that no bare conductors are improp-
erly exposed. Using the device in such conditions could
cause an electric shock, so contact your dealer or Hioki repre-
sentative for repair.
• The maximum rated voltage between input terminals and
ground is 1000 VDC. Attempting to measure voltages exceed-
ing 450V with respect to ground could damage the device and
result in personal injury.
• To avoid shock and short circuits, turn off all power before
connecting the device.
• Do not allow the device to get wet, and do not take measure-
ments with wet hands. This may cause an electric shock.
• Do not use the device where it may be exposed to oil, chemi-
cals, or solvents. Contact with these substances may cause
cracking in the device, resulting in damage or electric shock.
• Do not slant the device or place it on top of an uneven sur-
face. Dropping or knocking down the device can cause injury
or damage to the device.
• This device is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it
might be splashed with liquid. This may cause damage.
• Do not store or use the device where it could be exposed to di-
rect sunlight, high temperature or humidity, or condensation.
Under such conditions, the device may be damaged and insu-
lation may deteriorate so that it no longer meets specifications.
• To avoid damage to the device, protect it from physical shock
when transporting and handling. Be especially careful to
avoid physical shock from dropping.
• To prevent an electric shock accident, confirm that the white or
red portion (insulation layer) inside the cable is not exposed. If
a color inside the cable is exposed, do not use the cable.
Avoid the following locations that could cause an accident or damage to the device.
Exposed to direct sunlight
Exposed to high tempera-
ture
In the presence of corro-
sive or explosive gases
Exposed to water, oil, oth-
er chemicals, or solvents
Exposed to high humidity
or condensation
Exposed to strong electro-
magnetic fields
Near electromagnetic
radiators
Exposed to high levels of
particulate dust
Near induction heating sys-
tems. (e.g., high-frequency
induction heating systems
and IH cooking utensils)
Subject to vibration
Specifications
Operating envi-
ronment
Indoors, altitude up to 2000 m (6562 feet),
Pollution degree 2
Operating tem-
perature and hu-
midity
0°C to 40°C (32°F to 104°F), 80%RH or less.
(non-condensing)
Storage tempera-
ture and humidity
-10°C to 50°C (14°F to 122°F), 80%RH or less.
(non-condensing)
Dielectric strength Between input terminals (together) and SM9001 enclosure:
7504 V DC, for 15 seconds
Electrode struc-
ture
Main electrode : Floating structure
(isolated from ring electrode)
Ring electrode : Fixed to enclosure
Dimensions
Approx. 100×223H mm (including handle and barrier)
Approx. 3.94” × 8.78”H
Connection Cable length: Approx. 1 m
Mass Approx. 2.5 kg (88.2 oz.)
Accessories
Instruction manual ..................................... 1
Counter-electrode with integrated stand.... 1
Dimensions: Approx. 154W×155D×25H mm
(6.06”W × 6.10”D × 0.98”H),
Mass: Approx. 1.2 kg (42.3 oz.)
Electrode dimensions: Approx. 100 mm×100 mm
Protective stand ......................................... 1
Carrying case (with lock) ........................... 1
Short bar .................................................... 1
Counter-electrode connection cable .......... 1
Cable length : Approx. 700 mm (27.56”)
Mass : Approx. 27.0 g (1.0 oz.)
Electrodes
Electrode rubber
Material: Conductive rubber
Profile : Main electrode ; 30.5 mm (D1),
Thickness: 3 mm
: Ring electrode ; 57.0 mm (inner diameter)(D2),
63.0 mm (outer diameter),
Thickness: 3 mm
Option
Model SM9002 Verification Fixture for Surface Resistance
Measurement
Temperature and humidity range during inspection
18°C to 28°C (64.4°F to 82.4°F),
60% RH or less. (non-condensing)
Product warranty
period
3 years (The electrode rubbers, which are consumables,
are not covered by the warranty.)
Reference stan-
dards
IEC61340-2-3:2000
JIS C2170:2004
Measurement
items
Surface resistance measurement [Rs]
The surface resistance of the sample is measured after placing
the electrode at roughly the center of the sample or at least 10
mm from the edge of the sample.
Volume resistance measurement [Rv]
The volume resistance of the sample is measured after placing
the sample on the counter-electrode with integrated stand and
then placing the electrode at roughly the center of the sample or
at least 10 mm from the edge of the sample.
*Switch measurement targets by switching terminal connections.
Measurement
ranges
Resistance measurement range: 1×103Ωto 1×1013 Ω
The minimum resistance measurement is determined by the speci-
fications of DSM series and SM series Super Megohm Meters.
Electrode internal
resistance
Between main electrode and ring electrode
: 1×1014 Ωor more
Maximum rated
voltage to earth 1000 V DC
Resistance when
shorted
Between main electrode and ring electrode
: 1×103Ωor less
Supported
models Super Megohm Meter SM series, DSM series
Conversion to resistivity (JIS C2170:2004/ IEC 61340-2-3:2000)
•(D2+D1)
Surface resistivity: s= × (Measured value)
D2D1
• D12(Measured value)
Volume resistivity: v = ×
4t 10
s: Surface resistivity (Ω)
v: Volume resistivity (Ω•mm)
: Circular constant = 3.14
D1 : Diameter of the main electrode (mm)
D2 : Diameter of the ring electrode (mm)
t: Thickness of sample (mm)
D2
D1
t
Ring electrode
Main electrode
Counter electrode
Object to be
measured