Introduction
Thank you for purchasing the HIOKI Model SM9001 Surface/Volume Resis-
tance Measurement Electrode and the HIOKI Model SM9002 Verification Fix-
ture for Surface Resistance Measurement . To obtain maximum performance
from the device, please read this manual first, and keep it handy for future ref-
erence.
Initial Inspection
When you receive the device, inspect it carefully to ensure that no damage oc-
curred during shipping. If damage is evident, or if it fails to operate according to
the specifications, contact your dealer or Hioki representative.
The SM9001, an electrode for measuring the surface resistance and volume re-
sistance of flat samples, is an optional product designed for use with Hioki Super
Megohm Meters and Digital Super Megohm Meters/Micro-ammeters. It com-
plies with JIS C 2170 (IEC 61340-2-3) and is capable of reliable measurement
in applications ranging from evaluation of electrostatic countermeasures to
high-resistance measurement in materials development and other fields.
This manual contains information and warnings essential for safe operation of
the device and for maintaining it in safe operating condition. Before using it, be
sure to carefully read the following safety precautions.
Safety Symbols
The following symbols in this manual indicate the relative importance of cautions
and warnings.
SM9001
SURFACE/VOLUME RESISTANCE
MEASUREMENT ELECTRODE
SM9002
VERIFICATION FIXTURE FOR
SURFACE RESISTANCE MEASUREMENT
Instruction Manual
May 2019 Revised edition 3
SM9001A981-03 19-05H
EN
Overview
Safety Information
Mishandling this device during use could result in injury or
death, as well as damage to the device. Be certain that you
understand the instructions and precautions in the manual
before use. We disclaim any responsibility for accidents or
injuries not resulting directly from device defects.
Indicates cautions and hazards. When the symbol is printed on the instru-
ment, refer to a corresponding topic in the Instruction Manual.
Indicates DC (Direct Current).
Indicates an imminently hazardous situation that will result in death or
serious injury to the operator.
Indicates a potentially hazardous situation that may result in death or
serious injury to the operator.
Indicates a potentially hazardous situation that may result in minor or
moderate injury to the operator or damage to the device or malfunction.
Indicates advisory items related to performance or correct operation of
the device.
Stowing the electrode in its
carrying case
Protective
stand
SM9001
Counter-elec-
trode with inte-
grated stand
SM9002
Verification Fixture
for Surface Resis-
tance Measure-
ment
(Option)
Cables
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Follow these precautions to ensure safe operation and to obtain the full benefits
of the various functions.
Instrument Installation
General Specifications
Measurement functionality
Usage Notes
• Before using the device, make sure that the insulation on the
cable is undamaged and that no bare conductors are improp-
erly exposed. Using the device in such conditions could
cause an electric shock, so contact your dealer or Hioki repre-
sentative for repair.
• The maximum rated voltage between input terminals and
ground is 1000 VDC. Attempting to measure voltages exceed-
ing 450V with respect to ground could damage the device and
result in personal injury.
• To avoid shock and short circuits, turn off all power before
connecting the device.
• Do not allow the device to get wet, and do not take measure-
ments with wet hands. This may cause an electric shock.
• Do not use the device where it may be exposed to oil, chemi-
cals, or solvents. Contact with these substances may cause
cracking in the device, resulting in damage or electric shock.
• Do not slant the device or place it on top of an uneven sur-
face. Dropping or knocking down the device can cause injury
or damage to the device.
• This device is not designed to be entirely water- or dust-proof.
Do not use it in an especially dusty environment, nor where it
might be splashed with liquid. This may cause damage.
• Do not store or use the device where it could be exposed to di-
rect sunlight, high temperature or humidity, or condensation.
Under such conditions, the device may be damaged and insu-
lation may deteriorate so that it no longer meets specifications.
• To avoid damage to the device, protect it from physical shock
when transporting and handling. Be especially careful to
avoid physical shock from dropping.
• To prevent an electric shock accident, confirm that the white or
red portion (insulation layer) inside the cable is not exposed. If
a color inside the cable is exposed, do not use the cable.
Avoid the following locations that could cause an accident or damage to the device.
Exposed to direct sunlight
Exposed to high tempera-
ture
In the presence of corro-
sive or explosive gases
Exposed to water, oil, oth-
er chemicals, or solvents
Exposed to high humidity
or condensation
Exposed to strong electro-
magnetic fields
Near electromagnetic
radiators
Exposed to high levels of
particulate dust
Near induction heating sys-
tems. (e.g., high-frequency
induction heating systems
and IH cooking utensils)
Subject to vibration
Specifications
Operating envi-
ronment
Indoors, altitude up to 2000 m (6562 feet),
Pollution degree 2
Operating tem-
perature and hu-
midity
0°C to 40°C (32°F to 104°F), 80%RH or less.
(non-condensing)
Storage tempera-
ture and humidity
-10°C to 50°C (14°F to 122°F), 80%RH or less.
(non-condensing)
Dielectric strength Between input terminals (together) and SM9001 enclosure:
7504 V DC, for 15 seconds
Electrode struc-
ture
Main electrode : Floating structure
(isolated from ring electrode)
Ring electrode : Fixed to enclosure
Dimensions
Approx. 100×223H mm (including handle and barrier)
Approx. 3.94” × 8.78”H
Connection Cable length: Approx. 1 m
Mass Approx. 2.5 kg (88.2 oz.)
Accessories
Instruction manual ..................................... 1
Counter-electrode with integrated stand.... 1
Dimensions: Approx. 154W×155D×25H mm
(6.06”W × 6.10”D × 0.98”H),
Mass: Approx. 1.2 kg (42.3 oz.)
Electrode dimensions: Approx. 100 mm×100 mm
Protective stand ......................................... 1
Carrying case (with lock) ........................... 1
Short bar .................................................... 1
Counter-electrode connection cable .......... 1
Cable length : Approx. 700 mm (27.56”)
Mass : Approx. 27.0 g (1.0 oz.)
Electrodes
Electrode rubber
Material: Conductive rubber
Profile : Main electrode ; 30.5 mm (D1),
Thickness: 3 mm
: Ring electrode ; 57.0 mm (inner diameter)(D2),
63.0 mm (outer diameter),
Thickness: 3 mm
Option
Model SM9002 Verification Fixture for Surface Resistance
Measurement
Temperature and humidity range during inspection
18°C to 28°C (64.4°F to 82.4°F),
60% RH or less. (non-condensing)
Product warranty
period
3 years (The electrode rubbers, which are consumables,
are not covered by the warranty.)
Reference stan-
dards
IEC61340-2-3:2000
JIS C2170:2004
Measurement
items
Surface resistance measurement [Rs]
The surface resistance of the sample is measured after placing
the electrode at roughly the center of the sample or at least 10
mm from the edge of the sample.
Volume resistance measurement [Rv]
The volume resistance of the sample is measured after placing
the sample on the counter-electrode with integrated stand and
then placing the electrode at roughly the center of the sample or
at least 10 mm from the edge of the sample.
*Switch measurement targets by switching terminal connections.
Measurement
ranges
Resistance measurement range: 1×103Ωto 1×1013 Ω
The minimum resistance measurement is determined by the speci-
fications of DSM series and SM series Super Megohm Meters.
Electrode internal
resistance
Between main electrode and ring electrode
: 1×1014 Ωor more
Maximum rated
voltage to earth 1000 V DC
Resistance when
shorted
Between main electrode and ring electrode
: 1×103Ωor less
Supported
models Super Megohm Meter SM series, DSM series
Conversion to resistivity (JIS C2170:2004/ IEC 61340-2-3:2000)
•(D2+D1)
Surface resistivity: s= × (Measured value)
D2D1
 • D12(Measured value)
Volume resistivity: v = ×
4t 10
s: Surface resistivity (Ω)
v: Volume resistivity (Ω•mm)
: Circular constant = 3.14
D1 : Diameter of the main electrode (mm)
D2 : Diameter of the ring electrode (mm)
t: Thickness of sample (mm)
D2
D1
t
Ring electrode
Main electrode
Counter electrode
Object to be
measured
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Before using the device, check it and verify that it operates properly to make
sure that it suffered no damage during storage or transportation.
Check the following before using the device.
Names of Parts
Pre-Operation Inspection
Check Items Diagnose and Solution
Is the insulation of the cable to be used damaged,
or is bare metal exposed?
Do not use a damaged
cable doing so may
result in electric shock.
Is there a break in a cable?
• Between the main electrode rubber and the input
cable's inner wire
• Between the guard terminal (blue) and the input
cable's outer wire
• Between the output terminal (red) and the output
cable
Have the SM9001
repaired if there is a
break in a cable.
Is the electrode's internal resistance at least
1×1014 Ω?
• Place the SM9001 in a horizontal orientation and
take a measurement while the main electrode
and ring electrode are not in contact with any-
thing.
(Applied voltage: 100 V/1 min.)
Remove dirt and other
foreign matter from
between the main elec-
trode and ring elec-
trode.
Measurement Procedures
When measuring insulation resistance, a dangerous voltage will
be applied to the electrodes in the START, MEASURE, and
CHARGE states. To avoid electric shock, do not touch the elec-
trodes.
Short bar
Handle
Input cable
Output cable
Protective stand
Counter-electrode terminal
SM9001
Counter-elec-
trode with inte-
grated stand
Counter-electrode
surface
Stand surface
Main electrode rubber
Ring electrode
rubber
Bottom of SM9001
Top of SM9001
Guard terminal
(Blue)
Common terminal (gray)
Output terminal
(Red)
Barrier
• To avoid breaking the cable, do not bend or pull it.
• Avoid stepping on or pinching cables, which could damage
the cable insulation.
• The cable is hardened under the 0 degree or colder environ-
ment. Do not bend or pull it to avoid tearing its shield or cut-
ting cable.
• Keep the cables well away from heat sources, as bare con-
ductors could be exposed if the insulation melts.
• When placing the SM9001 in contact with the sample, do so gently and from
directly above the sample.
• Check the surface of the electrode rubber for dirt, foreign matter, scratches,
or other impediments to proper operation.
• Place the measurement target (or when measuring volume resistance, the
counter-electrode) on a level surface.
• While performing measurement, avoid contact with the SM9001. Avoid sub-
jecting the SM9001 to excessive force by pulling on or twisting cables.
• Do not subject the electrode rubber to twisting (friction) or mechanical shock.
• If unable to obtain stable measurement results, clean the surface of the
SM9001's electrode rubber with industrial-use ethanol (use a concentration
of at least 95%). If cleaning fails to yield improvement, remove the SM9001's
electrode rubber and clean the entire unit with industrial-strength ethanol
(use a concentration of at least 95%). For more information, see "Removing
and reattaching the electrode rubber."
• To protect the electrode rubber, place the SM9001 on its protective stand
when not using it to measure samples and avoid sitting it directly on the bot-
tom surface. When placing the SM9001 on the protective stand, do so care-
fully and avoid subjecting the stand to excessive force.
Measuring surface resistance
Measuring volume resistance
1. Connect the short bar to the surface resistance measurement [Rs] terminals
(common terminal [gray] and output terminal [red]).
2. Place the sample on the stand surface of the counter-electrode with integrat-
ed stand and gently place the SM9001 on top of the sample at least 10 mm
from the edge, lowering it from directly above.
3. Verify that no voltage is being output from the Super Megohm Meter. Con-
nect the input cable to the Super Megohm Meter's input terminal and the out-
put cable to the Super Megohm Meter's output terminal.
4. Operate the Super Megohm Meter to start measurement.
For more information about measurement, see the Super Megohm Meter's
instruction manual.
1. Connect the short bar to the volume re-
sistance measurement [Rv] terminals
(common terminal [gray] and guard ter-
minal [blue]).
2. Place the sample on the counter-elec-
trode surface of the counter-electrode
with integrated stand and gently place
the SM9001 on top of the sample at least
10 mm from the edge, lowering it from di-
rectly above so that it does not extend
beyond the counter-electrode surface.
3. Connect the included counter-electrode
connection cable to the output terminal
(red) and the counter-electrode terminal.
4. Verify that no voltage is being output
from the Super Megohm Meter. Connect
the input cable to the Super Megohm
Meter's input terminal and the output ca-
ble to the Super Megohm Meter's output
terminal.
5. Operate the Super Megohm Meter to
start measurement.
For more information about measure-
ment, see the Super Megohm Meter's in-
struction manual.
3
Counter-electrode
connection cable
1
4
2
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
• If damage is suspected, check the "Before Returning for Repair" section
before contacting your dealer or Hioki representative.
• Use the original packing materials when transporting the device, if possible.
Cleaning
To clean the device, wipe it gently with a soft cloth moistened with water or mild
detergent. Never use solvents such as benzene, acetone, ether, ketones, thin-
ners or gasoline, as they can deform and discolor the case. Clean the electrode
rubber with industrial-use ethanol (use a concentration of at least 95%).
About electrode rubber
Removing and reattaching the electrode rubber
• The electrode rubber can be removed by gripping it carefully so as not to
scratch it and pulling it off.
• The electrode rubber has a top and a bottom. Check for proper orientation
before reattaching it. The surface that makes contact with the measurement
target is thinner, while the surface that makes contact with the metal portion
of the SM9001 is thicker.
• Gently insert the electrode rubber and then fit it tightly against the metallic
surface by pushing it against a level base such as the counter-electrode.
Contact your dealer or Hioki representative if:
• The electrode rubber no longer covers the adjacent metal
• The electrode rubber exhibits visible damage
• The electrode rubber is excessively dirty and efforts at cleaning prove to be
ineffective
Before Returning for Repair
If abnormal operation occurs, check the following items.
The SM9002 Verification Fixture for Surface Resistance Measurement is a sys-
tem verification fixture for use in measuring surface resistance as required by
Section 8.4 of JIS C 2170:2004 (IEC 61340-2-3:2000). The SM9002 is an option
designed specifically for use with the SM9001 and cannot be used for system
verification of other devices.
General Specifications
Measurement functionality
The following describes how the fixture is used in a typical application. Custom-
ers needing to perform a rigorous verification process that complies with JIS C
2170:2004 should be sure to check the actual JIS document.
Low-resistance measurement
If the acquired resistance values are both within 5% of 500 kΩ, then the system
is operating properly.
High-resistance measurement
Maintenance and Service
Symptom Check Items Solution
The resistance value is
abnormal.
Are the terminals con-
nected properly?
Connect the terminals
securely.
Has the short bar been
connected properly?
Connect the short bar to
the proper pair of termi-
nals for the type of mea-
surement you intend to
perform.
Is the electrode rubber
dirty?
Clean the electrode rub-
ber.
SM9002
Verification Fixture for Surface
Resistance Measurement
Specifications
Operating environ-
ment
Indoors, altitude up to 2000 m (6562 feet),
Pollution degree 2
Operating tempera-
ture and humidity
18°C to 28°C (64.4°F to 82.4°F), 60% RH or less.
(non-condensing)
Storage tempera-
ture and humidity
-10°C to 50°C (14°F to 122°F), 80% RH or less.
(non-condensing)
Dielectric strength Between electrodes (together) and SM9002 enclosure:
1120 V DC, for 15 seconds
Electrode structure
High-resistance : Chrome-plated metal (brass)
Low-resistance : Gold-plated copper cladding formed on
PC board
Dimensions Approx. 100×56H mm (3.94” × 2.20”H)
Mass Approx. 300 g (10.6 oz.)
Product warranty
period 3 years
Measurement items
Low-resistance measurement
Resistance measurement electrodes connected to a measuring
instrument are secured to the inspection fixture, and a judgment
is made concerning whether the system is functioning properly.
(Test voltage: 10 V DC)
Verification consists of a series of repeated measurements
while varying the phase position of the fixture and electrodes by
90°.
High-resistance measurement
Resistance measurement electrodes connected to a measuring
instrument are secured to the inspection fixture, and a judgment
is made concerning whether the system is functioning properly.
(Test voltage: 100 V DC)
Circuit design
Low-resistance measurement
The circuit consists of contact electrodes arranged in a radial
pattern at the center and ring measurement electrode positions
and twenty 10 MΩ±1% resistors placed at an even interval.
The circuit forms a 500 kΩ±1% parallel connection.
High-resistance measurement
The circuit consists of contact electrodes placed at the center
and ring measurement electrode positions and one 1 TΩ±5%
resistor.
Maximum rated volt-
age to earth 100 V DC
Appearance
Using the fixture
• Some Super Megohm Meter models are capable of generating a terminal-
to-ground voltage of 1,000 V, but the SM9002 is rated for a maximum ter-
minal-to-ground voltage of 100 V.
• Check the SM9002 test surface and the SM9001 electrode rubber for dirt,
foreign matter, scratches, or other impediments to proper operation.
• While performing measurement, avoid contact with the SM9001. Avoid
subjecting the SM9001 to excessive force by pulling on or twisting cables.
• Due to the delicate nature of the contact with the SM9001, you may obtain
out-of-range measured values even if the product is operating properly if
the SM9001 has not been placed on the surface from directly above.
Remove and then reposition the SM9001 if the value is out of range.
• If unable to obtain stable measurement results, clean the surface of the
SM9001's electrode rubber and the SM9002 test surface with industrial-
use ethanol (use a concentration of at least 95%). If cleaning fails to yield
improvement, remove the SM9001's electrode rubber and clean the entire
unit with industrial-strength ethanol (use a concentration of at least 95%).
For more information, see "Removing and reattaching the electrode rub-
ber."
• Place the SM9002 on a level surface.
• The SM9002 should be stored in a pouch inside the SM9001's included
carrying case.
• Do not touch the test surface or get it dirty.
High-resistance test surface Low-resistance test surface
Flip over
1. Connect the short bar to the surface resistance measurement [Rs] terminals
(common terminal [gray] and output terminal [red]).
2. Gently place the SM9001 on the SM9002's low-resistance test surface, low-
ering it from directly above.
3. Apply 10 V and read the resistance value after 15 sec.
4. Stop applying the voltage. Lift up the SM9001 and rotate the SM9002 90°.
5. Repeat steps (3) and (4) until the SM9002 has been rotated through 360°.
1. Connect the short bar to the surface resistance measurement [Rs] terminals
(common terminal [gray] and output terminal [red]).
2. Gently place the SM9001 on the SM9002's high-resistance test surface, low-
ering it from directly above.
3. Apply 100 V and verify that a stable resistance value of 1 TΩ±5% is obtained.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

This manual suits for next models

1

Other Hioki Measuring Instrument manuals

Hioki CM4371 User manual

Hioki

Hioki CM4371 User manual

Hioki BT3562-01 User manual

Hioki

Hioki BT3562-01 User manual

Hioki 7016 User manual

Hioki

Hioki 7016 User manual

Hioki 3169-20 User manual

Hioki

Hioki 3169-20 User manual

Hioki CT6700 User manual

Hioki

Hioki CT6700 User manual

Hioki CM4375 User manual

Hioki

Hioki CM4375 User manual

Hioki CM4141 User manual

Hioki

Hioki CM4141 User manual

Hioki MR8740T User manual

Hioki

Hioki MR8740T User manual

Hioki FT6380 User manual

Hioki

Hioki FT6380 User manual

Hioki 9695-02 User manual

Hioki

Hioki 9695-02 User manual

Hioki SM-8213 User manual

Hioki

Hioki SM-8213 User manual

Hioki 3273 User manual

Hioki

Hioki 3273 User manual

Hioki LCR HiTESTER 3535 Instruction manual

Hioki

Hioki LCR HiTESTER 3535 Instruction manual

Hioki lr5001 User manual

Hioki

Hioki lr5001 User manual

Hioki 3390-10 Use and care manual

Hioki

Hioki 3390-10 Use and care manual

Hioki IM3523A User guide

Hioki

Hioki IM3523A User guide

Hioki MR8740T Instruction manual

Hioki

Hioki MR8740T Instruction manual

Hioki CKT3548 User manual

Hioki

Hioki CKT3548 User manual

Hioki MR8875 User manual

Hioki

Hioki MR8875 User manual

Hioki FT3424 User manual

Hioki

Hioki FT3424 User manual

Hioki 9010 User manual

Hioki

Hioki 9010 User manual

Hioki CM4375 User manual

Hioki

Hioki CM4375 User manual

Hioki RM9010-01 User manual

Hioki

Hioki RM9010-01 User manual

Hioki CM4003 User manual

Hioki

Hioki CM4003 User manual

Popular Measuring Instrument manuals by other brands

Delmhorst Instrument Co TechCheck PLUS owner's manual

Delmhorst Instrument Co

Delmhorst Instrument Co TechCheck PLUS owner's manual

Electro Industries/GaugeTech Shark 100T quick start guide

Electro Industries/GaugeTech

Electro Industries/GaugeTech Shark 100T quick start guide

Kolida KTS440 Series manual

Kolida

Kolida KTS440 Series manual

American Aerospace Intelligent Series Operator's manual

American Aerospace

American Aerospace Intelligent Series Operator's manual

Hitachi L44K instruction manual

Hitachi

Hitachi L44K instruction manual

REED ST-128M instruction manual

REED

REED ST-128M instruction manual

DAKOTA ULTRASONICS MAX II manual

DAKOTA ULTRASONICS

DAKOTA ULTRASONICS MAX II manual

FlowLine LI1 Series instruction manual

FlowLine

FlowLine LI1 Series instruction manual

Diamond Antenna sx-1100 Operation instructions

Diamond Antenna

Diamond Antenna sx-1100 Operation instructions

Shaw SADP operating instructions

Shaw

Shaw SADP operating instructions

KSV Surface Potential Meter user manual

KSV

KSV Surface Potential Meter user manual

Renishaw PH10M-iQ PLUS Installation and user guide

Renishaw

Renishaw PH10M-iQ PLUS Installation and user guide

PCB Piezotronics 3711F11100G Installation and operating manual

PCB Piezotronics

PCB Piezotronics 3711F11100G Installation and operating manual

Vega VEGACAL 63 Specification sheet

Vega

Vega VEGACAL 63 Specification sheet

DV Power RMO40TD manual

DV Power

DV Power RMO40TD manual

Fuji Electric ZSU-5 instruction manual

Fuji Electric

Fuji Electric ZSU-5 instruction manual

Flexijet 3D 710100 quick start guide

Flexijet

Flexijet 3D 710100 quick start guide

ENDA EPV241A user manual

ENDA

ENDA EPV241A user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.