
DANGER
Ensure the rig is not doing any
operations and that the travelling
block is STATIONARY and others
are aware you are working
Do not continue until you are
certain the rig is no longer
operating, and that the rig manager
is aware of what you are doing.
Normal Usage: (parts 13 through 21)
13) The unit requires minimal to no supervision to operate
normally if installation is done correctly. Particularly
ensuring careful clearance of the cable outside the unit.
14) During operation the pneumatic motor will keep tension
on the cable and you will be able to monitor the distance
the travelling block moves up/down.
15) The drum has many cable layers, so each revolution will
have a slightly different linear length. Calibration is
required and is set up per end user systems.
16) The equivalent of this revolution distance to your
travelling block movement depends on the installation
and alignment of cable with the drilling mast.
iCalibration should be done to
ensure correct readings.
DANGER
Improper calibration can cause
incorrect readings relative to the
position of the travelling block,
resulting in possible injury.
17) The pressure gauge can be seen directly on the FRL. It
is good practice to check that the air pressure is at about
4 bar (52 psi) periodically
18) If you would like to use a different maximum speed on
your unit, please adjust the air pressure accordingly as
per the graphs and table shown on the previous page.
19) Also regularly check for any accumulation of dirt and
grime on the unit, if excessive, it is a good idea to keep
it relatively clean and to keep the telescope area clean of
any debris.
CAUTION
Cable moving at speed can cause
serious injury and / or death.
Ensure sufficient clearance
between it and other machinery and
personnel during normal
operations.
Ensure cable paying out from
device is well clear of any obstacles
and that bends, sharp objects or
turns via pulleys or other methods
are minimized.
20) The unit is designed to operate with minimal to no
maintenance, but we are aware that a drilling rig
environment differs between locations, as does the
environment around it. Therefore keeping it clean and
the cable clean helps with the life time. The unit is
designed to minimize tangling and damage inside the
unit, any damage to the cable outside will result in the
cable reeling in at a safe speed and causing no damage
to the internal mechanism.
21) Cleaning the cable as it comes in and out is not required,
though a good wipe with a thick cloth while using
gloves is good practice once a week. Do this by gently
applying a ‘damp’ cloth to the cable (right by the
telescope) while the cable is paying in or out. Be sure to
wear safety gloves and glasses during this operation. It
is advised that restricted access is in place in normal
operation (especially to the top telescope section, where
the cable enters the geolograph via two pulleys).
CAUTION
Take extra precautions when
applying the damp cloth to the cable
during operation as the cable can
have sharp or worn edges and to
avoid an
potential static char
es.
Failure Mode: (parts 22 through 31)
22) If during operation the cable snaps or brakes, the cable
will automatically reel into the device at a maximum
controlled speed as per table 1 (when set at 4Bar). This
is a safe speed which will minimize the damage and
injury to machines and personnel while it reels in.
23) Most of the failures will be a result of the cable
snapping or getting damaged outside of the unit.
Therefore it is important to have a clear path for the
cable to track the target.
24) If the cable does snap, the drum will continue to turn
until the air shutoff valve is turned off.
CAUTION
It is recommended to wait for the
cable to fully reel back in before
approaching the Geolograph in
such a case, and only then shutting
the air off at the main supply or the
manual shut off valve.
25) Once the drum comes to a complete stop, ensure the
manual valve is truly set to OFF. Turn the regulator in
the direction of the minus sign to fully shut off the air to
the unit. Now the air is ‘twice’ shut off from the device,
and it’s now safe to open the device.
26) Open the device by removing the 4x M12 hex bolts on
corners of the device. Keep them in a safe place.
DANGER
Ensure wear and tear to the cable is
minimized as any fraying or kinks
might cause the cable to snap and
possibly cause damage to the unit
itself and injuries to others. It is
important to monitor the condition
of the cable regularly and replace it
if necessar
.