HP 5308A Service manual

OPERATING
AND
SERVICE
MANUAL
15
MHz
TIMER/COUNTER
5308A
ET
PACKARD

CERTIFICATION
Hewlett-Packard
Company
certifies
that
this
instrument
met
its
published
specifications
at
the
time
of
shipment
from
the
factory.
Hewlett-Packard
Company
further
certifies
that
its
calibration
measurements
are
traceable
to
the
United
States
National
Bureau
of
Standards,
to
the
extent
allowed
by
the
Bureau's
calibration
facility,
and
to
the
calibration
facilities
of
other
International
Standards
Organization
members.
WARRANTY
AND
ASSISTANCE
This
Hewlett-Packard
product
is
warranted
against
defects
in
materials
and
workmanship
for
a
period
of
one
year
from
the
date
of
shipment.
Hewlett-Packard
will,
at
its
option,
repair
or
replace
products
which
prove
to
be
defective
during
the
warranty
period
pro-
vided
they
are
returned
to
Hewlett-Packard,
and
provided
the
preventive
maintenance
procedures
in
this
manual
are
followed.
Repairs
necessitated
by
misuse
of
the
product
are
not
covered
by
this
warranty.
NO
OTHER
WARRANTIES
ARE
EXPRESSED
OR
IMPLIED,
INCLUDING,
BUT
NOT
LIMITED
TO,
THE
IMPLIED
WARRANTIES
OF
MERCHANTABILITY
AND
FITNESS
FOR
A
PARTICULAR
PURPOSE.
HEWLETT-
PACKARD
IS
NOT
LIABLE
FOR
CONSEQUENTIAL
DAMAGES.
if
this
product
is
sold
as
part
of
a
Hewlett-Packard
integrated
instrument
system,
the
above
warranty
shall
not
be
applicable,
and
this
product
shall
be
covered
only
by
the
system
warranty.
Service
contracts
or
customer
assistance
agreements
are
available
for
Hewlett-Packard
products.
For
any
assistance,
contact
your
nearest
Hewlett-Packard
Sales
and
Service
Office.
Addresses
are
provided
at
the
back
of
this
manual.
019000-4-234

SECTION
IX
H
73
MHz
TIMER/COUNTER
5308A
OPERATING
AND
SERVICE
MANUAL
SERIAL
PREFIX:
1528A
This
section
applies
directly
to
Model
5308A
75
MHz
Timer/Counters
having
Serial
Prefix
1528A.
The
sec-
tion
is
provided
in
loose-ieaf
form
for
incorporation
into
the
5300B
Measuring
System
Manual.
NEWER
INSTRUMENTS
This
section
with
enclosed
“Manual
Changes”
sheet
applies
directly
to
HP
Model
5308A
75
MHz
Timer/
Counters
having
Serial
Prefix
numbers
above
1528A.
OLDER
INSTRUMENTS
Subsection
VII
of
this
document
contains
information
pertinent
to
all
older
instruments
with
Serial
Prefix
1440A
or
1524A.
Copyright
HEWLETT-PACKARD
COMPANY
1975
5301
STEVENS
CREEK
BLVD.,
SANTA
CLARA,
CALIF.
95050
Printed:
JAN
1976
MANUAL
PART
NO.
05308-90005
MICROFICHE
PART
NO.
05308-90006
PRINTED
IN
U.S.A.
HEWLETT
’
PACKARD


Model
5308A
General
Information
SECTION
IX
H
5308A
75
MHz
TIMER/COUNTER
SUBSECTION
|
GENERAL
INFORMATION
9H-1-1.
PURPOSE
AND
USE
OF
SECTION
IX
H
9H-1-2.
Section
IX
H
contains
the
information
neces-
sary
to
install,
operate,
and
maintain
the
HP
53084.
Timer/Counter.
Theory
of
operation,
parts
lists,
com-
ponent
locator
illustrations,
and
a
schematic
diagram
are
included.
Insert
this
document
in
the
HP
5300B
Measuring
System
manual
as
part
of
Section
IX.
9H-1-3.
INSTRUMENT
DESCRIPTION
9H-1-4.
When
plugged
onto
a
5300B
mainframe,
the
5308A
can
measure
frequency,
frequency
ratio,
period,
period
average,
time
interval,
and
time
interval
average.
A
single-channel
or
a
two-channel
totalize
function
is
provided.
The
two-channel
function
totalizes
pulses
on
Channel
A
during
pulses
on
Chan-
nel
B
or
between
pulses
on
Channel
B.
9H-1-5.
Front
panel
controls
are
provided
for
func-
tion
and
time
base
selection
and
for
signal
condition-
ing
(slope,
level,
and
attenuation).
Automatic
selection
of
maximum
resolution
is
provided
for
the
four
functions
of
frequency,
frequency
ratio,
period
average,
and
time
interval
average
measurements.
Trigger
indicator
lamps
are
provided
to
indicate
when
internal
amplifier
triggering
occurs.
9H-1-6.
Rear
panel
connectors
provide
a
gate
output
for
Z
axis
modulation
of
an
oscilloscope,
trig-
ger
level
outputs
for
monitoring
the
voltage
level
at
which
each
channel
triggers,
and
a
time
base cutput
for
use
in
totalizing,
period,
and
time
interval
measurements.
9H-1-7.
The
electrical
and
mechanical
specifications
are
listed
in
Table
9H-1-1.
9H-1-8.
INSTRUMENT
IDENTIFICATION
9H-1-9.
Hewlett-Packard
uses
a
two-section,
nine-
digit
serial
number
(0000A00000)
mounted
on
the
rear
panel
to
identify
the
instrument.
The
first
four
digits
are
the
serial
prefix
and
the
last five
digits
refer
to
the
specific
instrument.
If
the
serial
prefix
on
your
instrument
differs
from
that
listed
on
the
title
page
of
this
section,
there
are
differences
between
the
manual
and
your
instrument.
Lower
serial
prefixes
are
covered
by
a
manual
change
sheet
included
with
the
manual.
Figure
9H-1--1.
5308A
75
MHz
Timer/Counter
9H-1-1

Model
5308A
General
Information
9H-1-10.
The
printed
circuit
board
within
the
instru-
ment
is
identified
by
a
two-section,
10-digit
part
number
(e.g.,
05308-60001)
and
a
four-digit
series
number
(e.g.,
“SERIES
1440A”).
The
series
number
identifies
the
electrical
characteristics
of
the
com-
plete
printed-circuit
assembly.
A
replacement
circuit-
board
assembly
may
have
a
different
series
number
than
the
assembly
originally
supplied
with
the
instru-
ment.
Therefore,
when
troubleshooting
a
circuit-
board
assembly,
ensure
that
the
series
number
on
the
schematic
diagram
matches
the
series
number
on
the
board
assembly.
If
the
series
number
on
the
assembly
is
lower
than
the
number
on
the
schematic
diagram
in
Subsection
VIII,
refer
to
Subsection
VII
of
this
docu-
ment
for
change
information.
If
the
series
number
on
the
assembly
is
higher
than
the
number
on
the
sche-
9H-1-2
matic
diagram
in
Subsection
VIII,
the
change
infor-
mation
is
provided
in
a
manual
change
sheet
which
is.
`
available
from
the
nearest
Hewlett-Packard
Sales
and
Service
Office,
9H-1-11.
Manual
Changes
and
Options
9H-1-12.
The
title
page
lists
the
serial
prefix
number
to
which
this
manual
directly
applies.
If
the
serial
prefix
is
different
from
the
one
listed,
a
change
sheet
is
included
describing
the
required
changes.
If
this
change
sheet
is
missing,
the
information
can
be
sup-
plied
by
any
Hewlett-Packard
Sales
and
Service
office
listed
in
Section
VIII
of
the
5300B
Measuring
System
manual.
Options
are
listed
in
Section
IX
H,
Sub-
section
VII.

Model
5308A
General
information
Table
9H-1-1.
Specifications
INPUT
CHANNELS
A
AND
B
RANGE:
DC
Coupled;
0
to
75
MHz
AC
Coupled;
20
Hz
to
75
MHz
SENSITIVITY:
(Min)
25
mV
rms
sine
wave
to
10
MHz.
50
mV
rms
sine
wave
to
75
MHz.
150
mV
p-p
pulse
at
minimum
pulse
width
of 10
ns.
Input
must
be
less
than
1V
rms
from
50
to
75
MHz
with
Attenuator
switch
in
X1
position.
Sensitivity
can
be
decreased
by
10
using
Attenuator
switch.
IMPEDANCE:
1
MO
shunted
by
less
than
48
pF.
OVERLOAD
PROTECTION:
+
DC
Range:
Xl:
DC
to
400
kHz
125V
rms
400
kHz
to
5
MHz
Vrms
=
(5x107)+(Hz)
5
MHz
to
75
MHz
10V
rms
X10:
DC
to
4
MHz
250V
rms
4
MHz
to
75
MHz
Vrms
=
(1x10°)+(Hz)
|
AC
Ranges:
DC
to
10
Hz
200V
peak
in
addition
to
above
ratings.
TRIGGER
LEVEL:
Set
position
centers
triggering
at
O
volts,
or
continuously
variable
over
the
range
of
22.0
V
with
Attenuator
in
X1
position
or
+20V
with
Attenuator
in
X10
position.
Trig-
ger
level
available
on
rear
panel
BNC
con-
nectors
for
DVM
monitoring.
SLOPE:
Independent
selection
of
triggering
on
positive
or
negative
slope.
CHANNEL
INPUTS:
Separate
or
Common
B
GATE
OUTPUT:
Rear
panel
BNC.
TTL
low
level
while
gate
is
open
may
be
used
to
intensity
modulate
an
HP
oscilloscope.
TIME
BASE/SCALING
OUTPUT:
Available
at
rear
panel
BNC.
FREQUENCY
RANGE:
0
to
75
MHz,
Channel
A
or
Channel
B.
GATE
TIMES:
8
manually
selectable
times
from
1
us
to
10
seconds.
AUTO
position
selects
gate
time
for
maximum
resolution
within
a
1.1
second
measurement
time.
ACCURACY:
+1
count
+
time
base
accuracy.
DISPLAY:
Hz,
kHz,
and
MHz
with
positioned
deci-
mal
point.
FREQUENCY
RATIO
DISPLAY:
Fa/Fb,
measured
over
N
periods
of
Fb.
N=1
to
10%,
selectable
in
decade
steps
with
automatic
decimal
position
and
annunciators.
AUTO
position
selects
N
automatically
for
maximum
resolution
within
a
1.1
second
measurement
time.
RANGE:
Channel
A;
0
to
75
MHz.
Channel
B;
0
to
5
MHz.
ACCURACY:
+1
count
of
Fa
+
trigger
error
of
Fb*.
`
PERIOD
RANGE:
0
Hz
to
5
MHz,
Channel
B-
RESOLUTION:
100
ns
to
10
s
in
decade
steps.
ACCURACY:
+1
count
+
time
base
accuracy
+
trig-
ger
error*.
DISPLAY:
us,
ors
with
positioned
decimal
point.
PERIOD
AVERAGE
RANGE:
0
to
5
MHz;
(100
ns
to
10
s),
Channel
B.
PERIODS
AVERAGED:
1
to
10%
manually
selectable
in
decade
steps.
AUTO
position
automati-
cally
selects
number
of
periods
for
maximum
resolution
within
a
1.1
second
measurement
time.
FREQUENCY
COUNTED:
10
MHz.
ACCURACY:
+1
count
+
time
base
accuracy
+
trig-
ger
error*,
DISPLAY:
mus
(ns),
us,
with
positioned
decimal
points.
(In
the
108
position
of
the
TIME
BASE
switch,
the
display
will
be
ps
with
no
annunciator
shown.)
TIME
INTERVAL
RANGE:
200
ns
to
10°
s.
25
ms
minimum
pulse
width.
Separate
or
Common
B.
RESOLUTION:
100
ns
to
10
s
in
decade
steps.
ACCURACY:
+1
count-+
time
base
accuracy
+
trig-
ger
error’,
DISPLAY:
us,
s,
or
ks
with
positioned
decimal
point.
TIME
INTERVAL
AVERAGE
RANGE:
1
ns
to
10
s.
200
ns
dead
time
between
intervals.
Channels
A
and
B
separate
or
Com-
mon
B.
INTERVALS
AVERAGED:
1
to
108
selectable
in
de-
cade
steps.
AUTO
position
selects
number
of
intervals
to
give
maximum
resolution
with-
in
a
measurement
time
of
1.1
second.
DISPLAY:
mus
(ns),
us,
or
s
with
positioned
deci-
mal
point.
(In
the
108
position
of
the
TIME
BASE
switch,
the
display
will
be
in
ps
with
no
annunciator
shown.)
ACCURACY:
+
time
base
accuracy
£5
ns
+
[Trigger
Error*
+
100
ns]
Y
Intervals
Averaged
TOTALIZE
Totalizes
Channel
A
during
pulses
on
Channel
B.
Totalizes
Channel
A
between
pulses
on
Channel
B.
RANGE:
75
MHz
on
Channel
A
in
position
1
of
TIME
BASE
switch,
5
MHz
in
other
positions
of
TIME
BASE
switch.
ACCURACY:
+1
count
+
trigger
error
on
Channel
B*.
DISPLAY:
Displays
count.
Can
scale
display
with
annunciator
by
use
of
TIME
BASE
switch
to
increase
count
capacity.
GENERAL
NOTE:
5308A
is
compatible
with
8
digit
5300B
mainframe
only.
CHECK:
Inserts
internal
10
MHz
reference
fre-
quency
into
counting
decades.
Displays
10
counts
for
TIME
BASE
switch
positions
10
through
108
with
proper
decimal
position
and
annunciator.
OPERATING
TEMPERATURE:
0°
to
50°C.
POWER
REQUIREMENTS:
Including
5300B
main-
frame,
nominally
15
watts.
WEIGHT:
Net,
0.9
kg
(2
Ib).
Shipping,
1.5
kg
(3%
Ib),
DIMENSIONS:
With
mainframe
89
mm
H
3%”)
x
760
mm
W
(6547)
x
248
mm
L
(9%”).
+SEE
WARNING
ON
PAGE
9H-3-11.
*For
any
waveshape,
trigger
error
(us)
is
less
than
+
0.005
For
period
average,
this
is
less
than
+0.3%
of
one
period
+
periods
averaged
for
signals
with
40
dB
or
better
signal-to-noise
ratio.
9H-1-3

TEE

Model
5308A
Installation
SECTION
IX
H
5308A
75
MHz
TIMER/COUNTER
SUBSECTION
Il
INSTALLATION
9H-2-1.
INTRODUCTION
9H-2-2.
This
section
contains
information
on
unpack-
ing
and
inspection,
storage
and
shipment,
installation
and
removal
of
plug-on,
and
portable
operation.
9H-2-3.
UNPACKING
AND
INSPECTION
9H-2-4.
If
the
shipping
carton
is
damaged,
ask
that
the
carrier’s
agent
be
present
when
the
instrument
is
unpacked.
Inpsect
the
instrument
for
damage
such
as
scratches,
dents,
broken
knobs,
etc.
If
the
instrument
is
damaged
or
fails
to
meet
performance
tests
when
used
with
the
5300B
mainframe,
notify
the
carrier
and
the
nearest
Hewlett-Packard
Sales
and
Service
Office
immediately.
Performance
check
procedures
are
located
in
Subsection
V,
and
Sales
and
Service
Offices
are
listed
in
Section
VI
of
the
5300B
portion
of
the
manual.
Retain
the
shipping
carton
and
the
pad-
ding
material
for
the
carrier’s
inspection.
The
Sales
and
Service
Office
will
arrange
for
the
repair
or
re-
placement
of
the
instrument
without
waiting
for
the
claim
against
the
carrier
to
be
settled.
9H-2-5.
STORAGE
AND
SHIPMENT
9H-2-6.
PACKAGING.
To
protect
valuable
elec-
tronic
equipment
during
storage
or
shipment,
always
use
the
best
packaging
methods
available.
Your
Hewlett-Packard
Sales
and
Service
Office
can
pro-
vide
packaging
material
such
as
that
used
for
original
factory
packaging.
Contract
packaging
companies
in
many
cities
can
provide
dependable
custom
pack-
aging
on
short
notice.
The
unit
was
originally
pack-
aged
as
follows:
The
original
container
is
a
corrugated
cardboard
box
with
200
Ibs.
burst
test
(HP
9211-1620).
The
instrument
is
secured
and
protected
in
the
box
by
a
top
and
bottom
molded
frame
of
polystyrene
foam
(HP
No.
9220-1545).
9H-2-7,
ENVIRONMENT.
Conditions
during
stor-
age
and
shipment
should
normally
be
limited
as
follows:
a.
Maximum
altitude:
25,000
feet.
b.
Minimum
temperature:
-40°F
(-40°C).
c.
Maximum
temperature:
+167°F
(+75°C).
9H-2-8.
INSTALLATION
AND
REMOVAL
OF
PLUG-ON
9H-2-9.
The
counter
must
be
used
with
a
5300B
main-
frame
to
make
measurements.
Mate
the
counter
and
the
mainframe
according
to
the
instructions
given
in
Paragraph
2-11
and
Figure
2-1
in
the
5300B
main-
frame
documentation.
9H-2-10.
PORTABLE
OPERATION
9H-2-11.
Use
of
the
HP
Model
5310A
Battery
Pack
enables
the
5300B
mainframe
and
the
counter
to
be
used
in
areas
removed
from
ac
power
sources.
The
battery
pack
provides
a
minimum
of
3
hours
oper-
ation
(typically
5
hours)
at
20°
to
30°C
operating
and
charging
temperature.
Tables
1-2
and
1-4
of
the
5300B
portion
of
the
manual
lists
the
battery
pack
as
an
available
accessory.
Documentation
for
the
battery
pack
is
also
included
in
Section
IV
through
VIII
of
the
5300B
portion
of
the
manual.
To
prepare
the
5300B/5308A
for
portable
operation,
refer
to
Paragraph
2-13
and
Figure
2-2,
of
the
5300B
portion
of
the
manual.
9H-2-1

ere
EEEREN
NAON
e
reee,
‘20400938903
tees
green

Model
5308A
Operation
SECTION
IX
H
5308A
75
MHz
TIMER/COUNTER
SUBSECTION
Ili
OPERATION
9H-3-1.
INTRODUCTION
9H-3-2.
This
section
provides
operating
information
and
describes
modes
of
operation
and
operating
proce-
dures.
The
5308A
plug-on
may
be
used
to
measure
frequency
or
frequency
ratio
up
to
75
MHz,
period
to
5
MHz,
period
average
to
5
MHz,
time
interval
from
200
nanoseconds
to
10
seconds
and
time
interval
aver-
age
from
1
nanosecond
to
10°
seconds.
Signals
up
to
75
MHz
may
be
totalized
on
Channel
A
in
the
single-
channel
totalize
function.
In
the
two-channel
totalize
function,
signals
up
to
75
MHz
may
be
totalized
on
Channel
A
during
pulses
on
Channel
B
or
between
pulses
on
Channel
B.
9H-3-3.
OPERATING
INFORMATION
9H-3-4,
The
front
panel
contains
a
FUNCTION
selec-
tor,
a
TIME
BASE
selector
and
signal
conditioning
controls
for
use
with
the
Channel
A
and
Channel
B
input
connectors.
The
signal
conditioning
controls
are
labeled
LEVEL
A
and
LEVEL
B,
ATTENUATORS
and
AC-DC.
An
O/C
(open/close)
switch
is
provided
for
Channel
B
and
a
SEP-CHK-COM
B
switch
is
pro-
vided
for
selecting
channel
inputs
and
self
check
func-
tion.
The
purpose
and
use
of
these
controls
is
described
in
the
following
paragraphs.
9H-3-5.
Function
Selector
9H-3-6.
The
FUNCTION
selector
selects
one
of
the
eight
operating
modes
listed
in
paragraph
9H-3-2.
9H-3-7.
Time
Base
Selector
9H-3-8.
The
TIME
BASE
selector
provides
gate
time
control
for
frequency
measurements,
resolution
settings
for
period
or
time
interval
mode,
selection
of
number
of
cycles
averaged
during
frequency
ratio,
period
average,
or
time
interval
average
modes,
and
provides
a
scaling
factor
of
1
to
10%
in
the
totalize
mode.
The
AUTO
position
automatically
sets
the
TIME
BASE
selection
(for
maximum
resolution
within
1.1
second
measurement
time)
for
frequency
gate
time,
and
for
number
of
cycles
averaged
in
ratio,
period
average
and
time
interval
average
measurements.
9H-3-9.
Signal
Conditioning
Controls
9H-3-10.
Four
types
of
signal
conditioning
controls
are
provided
for
each
channel
(A
and
B)
input.
These
controls
allow
adjustment
of
trigger
level
and
selection
of
attenuation,
slope
and
ac
or
de
coupling.
The
effect
of
these
controls
is
to
minimize
inaccuracies
in
measurements
(refer
to
Figure
9H-3-1).
9H-3-11.
ATTENUATORS.
X1
and
X10
positions
(both
channels).
The
X10
position
of
the
attenuator
switch:
is
used
to
attenuate
large
amplitude
signals.
The
XI
position
represents
no
attenuation.
The
switch
setting
X10
increases
the
trigger
levels
and
hysteresis
voltages
by
a
factor
of
10.
Ifinput
signal
levels
are
unknown,
the
initial
measurements
should
be
made
with
the
attenua-
tor
switches
set
to
X10.
The
setting
may
then
be
changed
to
X1
if
necessary
to
obtain
a
stable,
usable
measurement.
9H-3-12.
SLOPE
SWITCHES
(both
channels).
The
slope
switches
allow
triggering
on
the
positive
slope
or
the
negative
slope
of
the
input
signals
(see
Figure
9H-3-1).
9H-3-13.
LEVEL
A
and
LEVEL
B.
The
level
controls
allow
adjustment
of
the
triggering
point
on
the
input
signal
waveform.
With
the
LEVEL
controls
set
to
the
“0”
position
(or
to
the
SET
position
when
the
rear
panel
“SET”
TRIG
LEVELS
switch
is
in
the
.0V
position
as
described
in
Figure
9H-3-3)
triggering
is
centered
around
zero
volts.
The
voltage
range
over
which
triggering
may
be
set
is
+2
volts
with
the
attenuators
in
the
X1
position
or
+20
volts
with
the
attenuator
switch
in
the
X10
position.
9H-3-14.
AC-DC
SWITCHES
(both
channels).
The
AC-DC
switches
select
ac
or
de
coupling.
In
AC
posi-
tion
with
the
LEVEL
control
set
for
triggering
centered
around
zero
ac
volts,
the
de
components
on
the
signal
have
no
effect.
In
DC
position
with
the
LEVEL
control
in
SET
(as
described
in
the
preceding
paragraph)
triggering
is
centered
around
zero
volts
de.
9H-3-15.
0/C
(open/close)
Switch
9H-3-16.
This
pushbutton
switch
controls
totalizing
of
Channel
A
input
signals
as
described
in
paragraph
9H-3-34.
This
switch
also
controls
the
stopwatch
function
as
described
in
paragraph
9H-3-40.
9H-3-17.
SEP-CHK-COM
B
Switch
9H-3-18.
This
switch
provides
selection
of
separate
Channel
A
and
Channel
B
inputs
in
the
SEP
position.
The
CHK
position
enables
the
self
check
function
as
described
in
paragraph
9H-3-39
and
the
stopwatch
function
as
described
in
paragraph
9H-3-41.
The
COM
B
position
connects
the
CHAN
B
connector
on
the
front
panel
to
the
Channel
A
and
Channel
B
ampli-
fier
circuits
in
parallel.
9H-3-1

Model
53084.
Operation
TRIGGER
VOLTAGE
\
\
RECOVERY
VOLTAGE
+
SLOPE
=
O
LEVEL
HYSTERESIS
PSET
(NOTE
E)
nn
ha
—
SLOPE
HYSTERESIS
(NOISE
REJECTION}
VOLTAGE
IS
CONSTANT
FOR
EACH
ATTENUATOR
SETTING
AND
DOES
NOT
VARY
WITH
TRIGGER
LEVEL
THIS
ALLOWS
ZERO
VOLTS
TRIGGERING
WITH
GOOD
NOISE
REJECTION
ATTENUATOR:
X1,
X10
CHANGES
TRIGGER
LEVEL
AND
HYSTERESIS
VOLTAGES
BY
FACTOR
OF
1,
10,
Figure
9H-3-1.
Signal
Conditioning
Using
Attenuator,
Level
and
Slope
Controls
SH-3-1S.
MODES
OF
OPERATION
9H-3-20.
The
5308A
provides
seven
modes
of
operation
in
addition
to
a
self-check
function
and
a
stopwatch
function.
Some
modes
of
operation
are
available
on
both
channels
(A
or
B)
while
other
modes
are
available
on
only
one
channel.
In
all
cases,
the
modes
are
selected
by
the
FUNCTION
switch.
The
modes
and
functions
are:
a.
Frequency
Mode
(Channel
A
or
B).
b.
Frequency
Ratio
Mode
(Channel
A/B).
c.
Period
Mode
(Channel
B
only).
d.
Period
Average
Mode
(Channel
B
only).
e.
Time
Interval
Mode
(Channel
A
to
B
or
Channel
B
only).
f.
Time
Interval
Average
Mode
(Channel
A
to
B
or
Channel
B
only).
g.
Totalize
Mode
(Single
or
two
channel).
h.
Self
Check
Function.
i.
Stopwatch
Function.
9H-3-2
9H-3-21,
Frequency
Mode
(Channel
A
or
B)
9H-3-22,
Frequency
measurements
may
be
made
on
|
Channel
A
or
B
from
0
to
75
MHz.
Front
panel
controls
are
set
to
the
same
relative
positions
for
measurements
on
either
channel
with
the
exception
that
the
SEP-
CHK-COM
B
switch
is
set
to
SEP
for
Channel
A
and
to
COM
B
for
Channel
B
measurements.
The
LEVEL
A
control
adjusts
the
triggering
point
on
the
input
signal
for
Channel
A
or
Channel
B
in
the
frequency
mode.
Accuracy
is
+1
count
+
time base
accuracy.
9H-3-23.
Frequency
Ratio
Mode
(Channel
A/B)
9H-3-24.
The
frequency
ratio
of
Channel
A
signals
(from
0
to
75
MHz)
to
Channel
B
signals
(0
to
5
MHz)
can
be
measured.
The
TIME
BASE
switch
setting
(1
to
10°)
selects
the
number
of
cycles
of
the
Channel
B
signal
over
which
the
ratio
A/B
is
averaged.
Accuracy
is
+1
count
of
Frequency
A
+
trigger
error
of
Frequency
B.
Increasing
the
number
of
cycles
of
Frequency
B
(higher
setting
of
TIME
BASE
switch)
results
in
increased
accuracy.
9H-3-25.
Period
Mode
(Channel
B)
9H-3-26.
The
period
mode
allows
single
period
mea-
surements
to
be
made
with
frequencies
of
0
to
5
MHz
into
Channel
B.
The
resolution
is
selectable
from
100
nanoseconds
to
10
seconds
by
the
TIME
BASE
switch.
Accuracy
is
+1
count
+
time
base
accuracy
+
trigger
error.
9H-3-27.
Period
Average
Mode
(Channel
B)
9H-3-28.
The
period
average
mode
allows
multiple
period
averages
to
be
made
with
frequencies
of
0
to
5
MHz
into
Channel
B.
The
number
of
periods
to
be
averaged
is
selected
by
the
TIME
BASE
switch
(1
to
105).
Accuracy
is
#1
count
+
time
base
accuracy
+
trigger
error.
The
+
1
count
and
trigger
error
is
reduced
in
proportion
to
the
number
of
periods
averaged.
9Fi-3-29.
Time
Interval
Mode
(Channel
A
to
B
or
Channel
B
only)
9H-3-30.
The
time
interval
mode
allows
time
measure-
ments
to
be
made
between
points
on
one
waveform
or
between
two
waveforms.
The
range
of
measurements
is
200
nanoseconds
to
10%
seconds
(minimum
pulse
width
of
25
nanoseconds)
with
resolution
of
100
nano-
seconds
to
10
seconds
in
decade
steps.
A
Start
signal
(Channel
A)
opens
the
main
gate
and
a
Stop
signal
(Channel
B)
closes
it.
(For
Start
and
Stop
signals
from
a
common
source,
the
CHAN
B
input
is
used
and
the
SEP-CHK-COM
B
switch
is
set
to
COM
B.)
A
TTL
low
output
is
available
at
the
GATE
OUT
connector
on
the
rear
panel
while
the
gate
is
open.
This
output
is

+
base
accuracy
+
used
to
intensity
modulate
an
oscilloscope
for
display
of
the
time
interval
measured.
Accuracy
is
+
1
count
+
time
trigger
error.
For
more
precise
mea-
surement
of
trigger
level,
connectors
are
provided
on
the
rear
panel
for
DVM
monitoring
of
the
trigger
level
voltage.
9H-3-31.
Time
Interval
Average
Mode
(Channel
A
to
B
or
Channel
B
Only)
9H-3-32.
The
time
interval
average
mode
provides
greater
resolution
of
time
interval
measurements
than
the
time
interval
mode
provides.
The
+
1
count
error
can
be
reduced
by
averaging
a
number
of
measure-
ments.
In
the
time
interval
average
mode,
the
main
gate
is
open
for
the
number
of
time
intervals
selected
by
the
TIME
BASE
switch.
Up
to
108
time
intervals
may
be
averaged
in
this
mode.
Since
the
measurement
accuracy
is
increased
by
Fiz
(instead
of
=
as
shown
on
the
display)
the
displayed
measurement
should
be
evaluated
by
computing
the
true
resolution
which
is
2100
ns
where
N
is
equal
to
the
number
of
time
inter-
vals
selected
by
the
TIME
BASE
switch.
For
detailed
information
on
time
interval
averaging,
refer
to
HP
Application
Note
162-1.
9H-3-33.
Totalize
Mode
(Single
or
Two
Channel)
H-3-34.
Input
signals
up
to
75
MHz
applied
to
Chan-
nel
A
can
be
totalized
with
the
TIME
BASE
switch
in
position
1.
In
all
other
positions
of
the
TIME
BASE
switch
input
signals
up
to
5
MHz
can
be
totalized.
The
input
frequency
is
scaled
by
a
E
equal
to
the
setting
of
the
TIME
BASE
switch
CLON).
9H-3-35.
For
single-che
anel
totalizing,
the
Channel
A
signal
is
totalized
under
control
of
the
O/C
switch
when
the
FUNCTION
switch
is
in
the
TOT
A,
AA.
B
position.
Totalizing
is
initiated
by
pressing
the
O/C
switch
and
terminated
by
pressing
the
O/C
switch
a
second
time.
9H-3-36.
For
two-channel
totalizing,
the
Channel
A
signal
is
totalized
under
control
of
the
Channel
B
signal.
When
the
FUNCTION
switch
is
in
the
TOT
A,
“LIB
position
(totalize
Channel
A
during
pulses
on
Channel
B)
the
Start
and
Stop
signals
on
Channel
B
occur
between
two
successive
trigger
level
crossings
in
the
opposite
direction,
as
follows:
Channel
B
Slope
Switch
_/
Channel
B
Slope
Switch
A.
Start
Stop
Start
Stop
AR
PS
FUNCTION
`
TOT
A
TOT
ALR
Model 5308A
Operation
9H-3-37.
When
the
FUNCTION
switch
is
in
the
TOT
A,
AA
B
position
(totalize
Channel
A
between
pulses
on
Channel
B)
the
Start
and
Stop
signals
on
Channel
B
occur
between
two
successive
trigger
level
crossings
in
the
same
direction
as
follows:
Channel
B
Slope
Switch
_/
Channel
B
Slope
Switch
ML
È
Stop
A
Ai
SC
FUNCTION
TOT
A,AAB
\
\
9H-3-38.
Self
Check
Function
9H-3-39.
The
CHK
position
of
the
SEP-CHK-COM
B
switch
on
the
front
panel
provides
a
position
where
proper
operation
of
the
majority
of
the
logic
circuits
in
the
5308A
and
5300B
is
verified.
When
circuits
are
functioning
properly,
the
10N
counts
are
displayed
for
positions
10
through
10%
of
the
TIME
BASE
switch
with
the
proper
decimal
position
and
annunciator.
9H-3-40.
Stopwatch
Function
9H-3-41.
The
CHK
position
of
the
SEP-CHK-COM
B
switch
also
provides
an
electronic
stopwatch
function
when
the
FUNCTION
switch
is
setto
TOTA,
AA
B.
After
the
RESET
switch
is
pressed,
the
stopwatch
count
is
started
by
pressing
the
O/C
switch.
The
count
is
terminated
by
pressing
the
O/C
switch
a
second
time.
To
initiate
a
new
measurement,
the
RESET
switch
must
be
pressed.
9H-3-42.
CONTROLS
AND
CONNECTORS
9H-3-43.
The
front
panel
controls
and
connectors
are
described
in
Figure
9H-3-2
and
the
rear
panel
switch
and
connectors
are
described
in
Figure
9H-3-3.
9H-3-44,
OPERATING
PROCEDURES
9H-3-45,
The
operating
procedures
for
making
mea-
surements
in
the
various
modes
are
listed
in
Figures
9H-3-4
through
9H-3-12.
9H-3-3

Model
5308A
Operation
3
3
2
È.
9H-3-4
FUNCTION.
Eight
position
switch
selects
mode
of
operation,
as
follows
(blue
lettering
matches
corresponding
blue
lettering
on
TIME
BASE
switch):
a.
FREQ
A.
Enables
frequency
measure-
ments
of
0
to
75
MHz
to
be
made
using
Channel
A.
FREQ
A/B.
Enables
the
frequency
ratio
of
Channel
A
signals
to
Channel
B
signals
to
be
taken.
The
ratio
of
Channel
A
input
signals
(from
0
to
75
MHz)
to
the
Chan-
nel
B
input
signals
(0
to
5
MHz)
can
be
measured.
The
ratio
is
averaged
over
N
cycles
of
the
Channel
B
signal.
N
is
selectable
from
1
to
108
by
the
TIME
BASE
switch.
Accuracy
is
+1
count
of
frequency
A,
+
trigger
error
of
Fre-
quency
B.
TIME
AVG
A-B
Sets
counter
to
measure
average
time
interval,
A
to
B.
Channel
A
starts
the
interval
and
Channel
B
stops
the
interval.
The
TIME
BASE
switch
(blue
lettering)
sets
the
number
of
time
intervals
to
be
averaged.
Time
interval
input
range
is
1
ns
to
10
s;
there
must
be
a
200
ns
deadtime
between
intervals.
d.
TIME
A—B.
Sets
counter
to
measure
time
interval
A
to
B.
Channel
A
starts
the
mea-
surement
and
Channel
B
stops
the
mea-
surement.
Time
interval
input
range
is
200ns
to
10%s
(25ns
minimum
pulse
width).
The
TIME
BASE
switch
(black
lettering)
selects
resolution
of
100ns
to
10
s.
PER
AVG
B.
Enables
multiple
period
averages
to
be
made
from
Channel
B
with
frequencies
of
0
to
5
MHz
(range
of
200
ns
to
10
s).
Use
TIME
BASE
switch
(blue
lettering)
to
select
number
of
periods
to
be
averaged
(1
to
108).
PER
B.
Enables
single
period
measure-
ments
to
be
made
from
Channel
B
with
frequencies
of
0
to
5
MHz.
The
TIME
BASE
switch
(black
lettering)
selects
resolution
of
100
ns
to
10s.
TOT
A,
LI
B.
Totalizes
signals
on
Channel
A
during
pulses
on
Channel
B.
(Refer
to
paragraph
9H-3-33.)
Totalizes
signals
up
to
75
MHz
with
TIME
BASE
switch
in
position
1
(blue
lettering),
up
to
5
MHz
with
TIME
BASE
switch
in
any
other
position
(10N),
Figure
9H-3-2.
Front
Panel
Controls
and
Connectors

b.
TOT
A,
AAR
Totalizes
Channel
A
be-
tween
pulses
on
Channel
B.
(Refer
to
para-
graph
9H-3-33.)
Totalizes
signals
up
to
75
MHz
with
TIME
BASE
switch
in
position
1
(blue
lettering),
up
to
5
MHz
with
TIME
BASE
switch
in
any
other
position
(10N).
2.
TIME
BASE.
Ten
position
switch
performs
the
following
functions:
a.
lps,
10
us,
.Ims,
1
ms,
10
ms,
.1s,1s,10s
(black
lettering).
Provides
eight
selectable
times
as
gate
time
controls
for
frequency
(FREQ
A)
measurements.
b.
.1
us,
1
us,
10
us,
.1
ms,
1
ms,
10
ms,
Ja
1
s,
10
s
(black
lettering).
Provides
resolu-
tion
of
100
ns
to
10
s
when
FUNCTION
switch
selects
period
(PER
B)
or
time
inter-
val
(TIME
A
—
B)
mode
of
operation.
c.
1,
10,
102, 103,
104,
105,
108,
107, 108,
(blue
lettering).
Provides
selection
of
number
of
eycles
or
events
averaged
when
FUNC-
TION
switch
selects
frequency
ratio
(FREQ
A/B),
period
average
(PER
AVG
B),
or
time
interval
average
(TIME
AVG
A-B)
d.
1,
10,
102,
10%,
104,
105,
108,
107, 108,
(blue
lettering).
Provides
scaling
factor
when
FUNCTION
switch
selects
totalize
(TOT
A,
“LI
B/TOT
A,
AA
B)
modes
of
opera-
tion.
e.
AUTO/.1
us
1.
The
AUTO
position
pro-
vides
four
function
auto-ranging
(pertains
to
gray-bordered
FUNCTION
switch
posi-
tions)
including
frequency
(FREQ
A),
fre-
quency
ratio
(FREQ
A/B),
period
average
(PER
AVG
B)
and
time
interval
average
(TIME
AVG
A-B).
Provides
maximum
resolution
that
can
be
obtained
in
each
function
within
a
maximum
of
1.1
second
measurement
time.
f.
AUTO/.1
us
1.
The
.1
us
1
position
per-
tains
to
the
FUNCTION
switch
lower-half
positions
(corresponding
letter
colors).
The
.1
us
position
(black
lettering)
pertains
to
TIME
A—B
and
to
PER
B.
The
1
position
(blue
lettering)
pertains
to
TOT
A,
LI
B
and
to
TOT
A,
-AALB.
O/C
(open/close).
Pushbutton
switch
enables
input
signals
to
Channel
A
to
be
totalized
when
the
FUNCTION
switch
selects
TOT
A,
AA
D
Totalizing
is
initiated
by
pressing
the
O/C
switch
and
is
terminated
by
pressing
the
switch
a
second
time.
To
reinitiate
the
Model
5308A
Operation
totalize
function
the
reset
switch
on
the
front
panel
of
the
5300B
mainframe
must
be
pressed.
An
additional
function
of
the
O/C
switch
is
to
operate
the
display
as
a
stop-
~
watch
(see
Figure
9H-3-12),
LEVEL
A
and
LEVEL
B.
Used
in
conjunction
with
ATTENUATORS
switch
to
determine
voltage
at
which
triggering
occurs
for
Chan-
nel
A
or
Channel
B.
In
SET
position,
trigger-
ing
is
centered
around
0
volts
when
the
rear
panel
“SET”
TRIG
LEVELS
switch
is
in
the
OV
position.
Trigger
level
is
continuously
variable
over
the
range
of
+2V
(ATTENUA-
TORS
IN
X1)
or
+20V
(ATTENUATORS
in
X10).
Adjacent
lamp
indicates
when
ampli-
fier
triggering
occurs.
CHAN
A
and
CHAN
B.
Input
connectors
for
Channel
A
and
Channel
B.
Input
impedance
is
1MQ
shunted
by
less
than
48
pF.
By
using
a
10
to
1
divider
probe,
input
impedance
can
be
increased
to
10
MO.
Slope
Switch.
Selects
the
triggering
on
posi-
tive
or
negative
slope
of
the
input
signals
to
Channel
A
(switch
on
left
side)
or
Channel
B
(switch
on
right
side).
SEP-CHK-COM
B.
mon
B).
(Separate-Check-Com-
a.
SEP.
Channel
A
and
Channel
B
are
con-
nected
for
separate
source
inputs
(1
MO
input
impedance).
b.
CHK.
Self
check
verifies
that
the
major-
ity
of
the
5300B
mainframe
logic
circuits
and
the
5308A
logic
circuits
are
function-
ing
properly.
c.
COM
B.
Disconnects
front
panel
input
connector
CHAN
A
and
connects
CHAN
B
to
the
amplifiers
for
Channel
A
and
Channel
B
in
parallel
(IMQ
input
impedance).
ATTENUATORS.
Selects
attenuation
of
the
input
signals
to
Channel
A
(switch
on
left
side)
and
to
Channel
B
(switch
on
right
side).
a.
X1
connects
input
signals
directly
to
input
amplifiers.
b.
X10
attenuates
input
signals
by
a
factor
of
10.
.
AC-DC.
Selects
ac
coupling
or
de
BE
of
the
input
signals
to
Channel
A
(switch
on
left
side)
or
Channel
B
(switch
on
right
side).
Figure
9H-3-2.
Front Panel
Controls
and
Connectors
(Cont'd)
9H-3-5

Model
5308A
Operation
T;
9H-3-6
‘SET’
TRIG
LEVELS.
Three
position
switch
presets
the
trigger
level
to
the
indicated
vol-
tage
(-.13V,
.0V,
+.15V)
when
the
LEVEL
control
on
the
front
panel
is
in
the
SET
posi-
tion.
With
a
10:1
divider
probe
connected
to
the
CHAN
A
or
CHAN
B
connector
on
the
front
panel,
the
-.13V
position
is
used
when
checking
ECL
logie
circuits
and
the
+.15V
position
is
used
when
checking
TTL
logic
cir-
cuits.
The
.OV
position
is
used
when
measur-
ing
a
symmetrical
waveform
such
as
a
sine
Wave.
TRIG
LEVELS
A
and
B.
Connectors
provide
for
measurement
of
Channel
A
or
Channel
B
trigger
levels
by
a
monitoring
DVM.
The
trigger
level
voltage
is
measured
at
the
mid-
point
of
the
hysteresis
band.
GATE
OUT.
Provides
a
TTL
level
signal
(LOW)
while
the
5300B
main
gate
is
open.
Used
for
time
interval
measurements
to
intensity
modulate
an
oscilloscope
to
indi-
cate
the
interval
being
measured.
4.
TIME
BASE
OUT.
This
connector
provides
two
different
outputs,
selectable
by
the
posi-
tion
of
the
FUNCTION
switch,
as
follows:
a.
In
the
TOT
A,
LI
Band
TOT
A,
AA
B
positions
of
the
FUNCTION
switch,
the
TIME
BASE
OUT
connector
provides
a
scaled
(divided)
output
of
the
input
signal
frequency
applied
to
Channel
A.
The
output
is
scaled
by
a
factor
of
from
10
to
10%
as
set
on
the
TIME
BASE
control.
b.
In
the
PER
B
and
TIME
A-
B
positions
of
the
FUNCTION
switch,
the
TIME
BASE
OUT
connector
provides
a
scaled
output
of
the
internal
10
MHz
oscillator
fre-
quency.
The
output
is
scaled
by
a
factor
of
from
10
to
10%
as
set
on
the
TIME
BASE
control.
The
period
of
this
output
is
indi-
cated
by
the
black
lettering
on
the
TIME
BASE
control.
The
same
output
is
pro-
vided
when
the
FUNCTION
switch
is
in
TOT
A,
LF
B
or
TOT
A,
AA
B
posi-
tion
and
the
SEP-CHK-COM
B
switch
is
in
the
CHK
position.
F
igure
9H-3-3.
Rear
Panel
Control
and
Connectors

Model
5308A
Operation
1
(REAR)
5,6,7,8,9,10
NOTE
4.
Set
the
FUNCTION
switch
to
FREQ
A.
If
proper
indications
are
not
received
during
the
self
check,
refer
to
the
trouble-
Set the
TIME
BASE
switch
to
the
positions
SE
procedures
in
Sec-
listed
below
and
observe
the
display.
ion
V.
Connect
ac
power
to
5300B
ac
receptacle.
Frequency
A
Self
Check
Turn
ac
power
on
with
53008
SAMPLE
RATE
control.
Adjust
SAMPLE
RATE
for
desired
TIME
BASE
DISPLAY
ANNUNCIATOR
display
time.
AUTO
10000.000
kHz
Set
the
SEP-CHK-COM
B
switch
to
CHK
lus
10000.0090
kHz
|
lus
00000010
MHz
|
NOTE
10
us
0000010.0
MHz
1
ms
000010.00
MHz
All
display
indications
are
+1
ims
-
00010.000
MHz
count.
In
the
10%
position
of
10
ms
0010000.0
kHz
the
TIME
BASE
switch,
wait
1s
010000.00
kHz
10
seconds
for
the
count
ls
10000.000
kHz
display.
10s
°0000000.0
Hz
e
=
OVERFLOW
Figure
9H-3-4.
Making
Self
Check
Measurements
9H-3-7

Model
5308A
Operation
Ratio
Seli
Check
TIME
BASE
DISPLAY
ANNUNCIATOR
6.
Set
the
FUNCTION
switch
to
FREQ
A/B.
AUTO
1000000.0
us
Set
the
TIME
BASE
switch
as
shown
below
lus
1000000.0
ES
and
observe
the
display.
lus
00000010
us
10
us
000.00100
-
s
TIME
BASE
DISPLAY
ANNUNCIATOR
1
ms
0000.1000
8
ims
00010.000
$
AUTO
1000000.0
u
10
ms
001000.00
s
1
10000000
‚is
0100000.0
S
10
0000001.0
is
10000000
S
102
000001.00
10s
e000000.00
ks
108
00001
.000
104
0001.0000
e
=
Overflow
10°
001.00000
108
01000000
Uu
Time
Interval
Self
Check
sn
“i
S
9.
Set the
FUNCTION
switch
to
TIME
A-B.
Set
the
TIME
BASE
switch
as
shown
below
s=
Ovidio
and
observe
the
display.
TIME
BASE
DISPLAY
ANNUNCIATOR
AUTO
1000000.5
MS
A
ps
1000000.5
us
l
us
00000015
us
i
|
10
us
000.00105
s
Period
Average
Self
Check
Ta
0000.1005
>
7.
Set the
FUNCTION
switch
to
PER
AVG
B.
Da
GE
e
Set
the
TIME
BASE
switch
as
shown
below
10
ms
001000.05
E
and
observe
the
display.
at
bo
See
j
TIME
BASE
DISPLAY
ANNUNCIATOR
SC
RE
Ke
AUTO
100.00000
Mus
e
=
Overflow
ei
nu.
i
Time
interval
Average
Self
Check
10?
00000.100
Us
10.
Set the
FUNCTION
switch
to
TIME
AVG
105
0000.1000
Jus
A-B.
Set
the
TIME
BASE
switch
as
shown
10°
000.10000
MS
below
and
observe
the
display.
105
00100.000
Mus
105
0100.0000
Mus
TIME
BASE
DISPLAY
ANNUNCIATOR
10°
100.00000
Mus
108
200000.000
AUTO
100.00000
Mus
1
1000000.0
HS
e
=
Overflow
10
000000.10
us
102
-00000.100
HS
Mus
=
Nanoseconds
103
0000.1000
us
104
000,10000
MS
10°
00100.000
Mus
106
0100.0000
Mus
107
100.00000
Mus
Period
Self
Check
103
*00000.000
8.
Set the
FUNCTION
switch
to
PER
B.
Set the
e
=
Overflow
TIME
BASE
switch
as
shown
below
and
observe
the
display.
Mus
=
Nanoseconds
Figure
9H-3-4.
Making
Self
Check
Measurements
(Cont’d)
9H-3-8

Model
5308A
Operation
1
(REAR)
2
5,11
3,9
4,10
8,14
5,7,11,13
6
5,11
12
SEE
WARNING
ON
PAGE
9H-3-11
Connect
ac
power
to
5300B
ac
receptacle.
10.
Set
SEP-CHK-COM
B
to
COM
B.
Turn
ac
power
on
with
the
5300B
SAMPLE
NOTE
RATE
control.
In
the
Frequency
mode
of
operation,
the
Channel
A
CHAN
B
input
is
processed
by
the
Set
FUNCTION
to
FREQ
A
and
adjust
Channel
A
amplifier
circuit.
Conse-
SETE
Se
e
quently,
the
slope
and
LEVEL
controls
SM
RALE
for
este
display
time:
for
Channel
A
are
effective
as
listed
Set
SEP-CHK-COM
B
to
SEP.
below.
The
Channel
B
ATTENUATORS
and
AC-DC
switches
are
effective
(see
Set
ATTENUATORS
to
X10;
AC-DC
to
AC
Figure
9H-8-1).
and
Channel
A
slope
switch
to
polarity
desired.
Set
LEVEL
A
to
SET.
NOTE
11.
Set
ATTENUATORS
(Channel
B)
to
X10;
sia
ge
AC-DC
(Channel
B)
to
AC
and
Channel
A
Ensure
that
the
‘SET
TRIG
LEVELS
slope
switch
to
polarity
desired.
Set
LEVEL
A
switch
on
the
rear
panel
is
in
the
.0V
to
SET.
eo
position
when
the
LEVEL
control
is
in
NOTE
the
SET
position;
Ensure
that
the
‘SET’
TRIG
LEVELS
-
switch
on
the
rear
panel
is
in
the
.0V
position
when
the
LEVEL
control
is
in
the
SET
position.
Connect
input
signal
(up
to
75
MHz)
to
be
measured
to
CHAN
A
connector.
Observe
display.
If
display
is
not
stable,
n
rotate
LEVEL
A
control
until
trigger
light
12.
Connect
input
signal
(up
to
75
MHz)
to
be
illuminates.
If
necessary,
set
ATTENUA-
measured
to
CHAN
B
connector.
TORS
switch
for
Channel
A
to
X1.
13.
Observe
display.
If
display
is
not
stable,
Set
TIME
BASE
to
desired
gate
time
or
to
rotate
LEVEL
A
control
until
trigger
light
AUTO.
illuminates.
If
necessary,
set
ATTENUA-
Channel
B
TORS
switch
for
Channel
A
to
X1.
Set
FUNCTION
to
FREQ
A
and
adjust
14.
Set
TIME
BASE
to
desired
gate
time
or
to
SAMPLE
RATE
for
desired
time.
AUTO.
Figure
9H-3-5.
Making
Channel
A
and
Channel
B
Frequency
Measurements
9H-3-9

2
m
SE
F
he
Z
MO
E
EZ
Ei
>
45
ope
È
Q
S
we
E
3
=
EX.
A
Weg
ES
A
N
a
m
Sp
9
dv
+
oe
n]
pe
KÉ
O
Zo
Za
en
ri
mo
Mm
d
Da
Gi
E
È
ZELDA
y
Ii,
5
5
3
E
as
Gi
opg
=
>.
ala
E
pra
w
©
bai
S
A
E
Seto
als
de
.
E
=
z
ADA
g
23°
3
2
2
Ë
Sas
y
EE
E
2
2
SR
6
5
RER
E
a
Ex
TRES
=
328
E
z
n
zu
"EI
=
eg
om]
q
Gi
D
E
W
Gx
NES
ef,
ovo
2
9
ES
22
Sein
3
aa
È
<
SÈ
a
Fo
a
go
88
BEBO
SE
È
p
det
08
OLA
zz
3
emi
D
D
D
D
E
5
o
e
vi
z
2
eb]
+
S
=
«ri
o
s
a
o
e
Ge
a
i
ES
ERSTE
=
2
>
ce
=
.
2
di
E
SES
a
a
D
EZ
S
a
È
<
E
T
=
2
=
3
o
lia
KS:
i
u
v
S
:
2354
ES
2
me
N,
A
isa
ES
p
2
M
a
E
228
SÉ
a
S
©
Qu
ang
dò
Ha
“~~
y
Ka
=
Q
=
ka
E
B
F
S
Rest
i
YN
10
Ka
=
Ù
SS
=
ni
og
o
E
To
o
ee
oc
"=
S
M
È
es
ys
Bä
Seier
DW
Gel
=
O
SLA
Lo
S
=
a
O
£
e
J
o7
©
©
Le
O
ems
E
+,
$
i
Na
oh
Qu
Ex
Ze
O
yx
“Ra
moa
2
CH
:
EN
H
sc
E
&
2
o
=
Si
zaog
SE
a,
O
S
So
Lis
Sts
B
oz
A
A
sI
Bas
ba
D
BS
=
El
Moa,
2.4
MOL
>
S
cë
&
Q
Ad
oO
=
mS
+
+2
42
=
Sq
D
e
>
5
Do
>
Doo
2.28
A
O
E
3
DI
Re
NS
oO
pitt
D
D
D
D
+
ur
35
rc
N
oD
xt
uf)
A
Li
20
D
Table of contents
Other HP Cash Counter manuals