HP 333A Service manual

OPERATING AND SERVICE MANUAL
HEWLETT
PACKARD
L
@CopyrightHewlett-PackardCompany 1966
P.O.
Box
301,
Loveland, Colorado, 80537
U.S.A.
Reproduced with Permission, Courtesy
of
Agilent Technologies, Inc.
MODEL 333A/334A
DISTORTION
ANALYZER
Serial Numbers:
333A: 1137A03146and greater
334A: 1140A05641
and
greater
Appendix
C,
Manual Backdating Changes, adapts this manual
to
lower serial numbers.
WARNING
I
To
help minimize thepossibility of electricalfire
or
shock hazards,
do
not expose thisinstrument
to rain
or
excessive moisture.
Manual
Part
NO.
00333-90008
Microfiche
Part
No.
00333-90058
prlnttd:
October
1975,1985

HEWLETT
Pa
PACKARD
CERTIFICATION
Hewlett-Packard Company certifies that
this
product met itspublishedspecifcations at thetime
of
shipment
from
the
factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States
Na-
tional Bureau
of
Standards, to the extent allowed by the Bureau’s calibrationfacility, and to thecalibrationfacilities
of
other International Standards Organization members.
WARRANTY
This Hewlett-Packard product is warranted against defects in material and workmanship for a period of one year
from date of shipment [,except that in the case of certain components listed in Section
I
of this manual, the warranty
shall be
for
the specified period] .During the warranty period, Hewlett-Packard Company will, at
its
option, either
repair
or
replace products which prove to be defective.
For
warranty service
or
repair, this product must be returned to a service facility designated by -hp. Buyer shall
prepay shipping charges
to
-hp- and -hp- shall pay shipping charges to return the product to Buyer. However, Buyer
shall
pay
all shipping charges, duties, and taxes for products returned to -hp- from another country.
Hewlett-Packard warrants that its software and firmwaredesignated by -hp- for use with an instrument will execute its
programming instructions when properly installed on that instrument. Hewlett-Packard does not warrant that the
operation
of
the instrument,
or
software,
or
firmware will be uninterrupted
or
error free.
LIMITATION
OF
WARRANTY
The foregoing warranty shall not apply to defects resulting from improper
or
inadequate maintenance by Buyer,
Buyer-supplied software
or
interfacing, unauthorized modification
or
misuse, operation outside of the environmental
specifications for the product,
or
improper site preparation
or
maintenance.
DISCLAIMSTHE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE.
NO OTHER WARRANTY
IS
EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. HEWLETT-
PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSE-
QUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
Product maintenance agreements and other customer assistance agreements are available
for
Hewlett-Packad products.
For
any assistance, contact
your
nearest Hewlett-Packard Sales and Service Office.
Addnsscr
areprovided at the back
of
this manual.

HEWLETT
PACKARD
SAFETY
SUMMARY
Tho following
ganorol
ufotypneoutionrmurt
bo
oburvodduring
all
phrmof oporotior,
urviw,
and nprir
of
thir
inrtrumont friluro
to
comply
with
thaw
procautions
or
with rprcific
warnings
oluwhon
in
thir
manual
violotw
rrfoty rtandrrdrof dosign, mrnufrcturo, and intondod
uu
of tho inrtrumontHowlott-PieLordCompany
assume8
no
liabilityfor tho curtomar'r foiluro
to
comply
with thou roquinmont8.
This
is
I
Srfoty
Class
1
inrtrumont
GROUND
THE
INSTRUMENT
To minimize shock hazard, the instrument chassis and cabinet must be connected to an elec-
trical ground. The instrument is equipped with a three-conductor ac power cable. The power
cable must either be plugged into an approved three-contact electrical outlet
or
used with a
three-contact to two-contact adapter with the grounding wire (green) firmly connected to an
electrical ground (safetyground) at the power outlet. The power jack and mating
plug
of the
power cable meet International Electrotechnical Commission (IEC) safety standards.
DO
NOT
OPERATE
IN
AN
EXPLOSIVE
ATMOSPHERE
Do
not operate the instrument
in
the presence of flammable gases
or
fumes. Operation of any
electrical instrument
in
such an environment constitutesa definite safety hazard.
KEEP
AWAY
FROM
LIVE
CIRCUITS
Operating personnel must not remove instrument covers. Component replacement and internal
adjustments must be made by qualified maintenance personnel.
Do
not replace components
with powercable connected. Under certain conditions, dangerous voltages mayexisteven with
the power cable removed. To avoid injuries, always disconnect power and discharge circuits
before touching them.
DO
NOT
SERVICE
OR
ADJUST
ALONE
Do
notattempt internal service
or
adjustment unless another person, capable of rendering first
aid
and resuscitation, is present.
DO
NOT
SUBSTITUTE
PARTS
OR
MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install substitute parts
or
per-
form any unauthorized modification to the instrument. Return the instrument to a Hewlett-
Packard Sales
and
Service Office for service and repair to ensure that safety features are main-
tained.
DANGEROUS
PROCEDURE
WARNINGS
Warnings, such as the example below, precede potentially dangerous procedures throughout
this manual. Instructions containe#
in
the warnings must be followed.
WARNING
I
Dongarour
voltagu,
wprbla of wuring drth,
on
prwont
in
thir inrtrumont Uso
ox.
tnmo
wution whon handling, twting,
and
rdjurting.
A

Instruction manual symbol: the product
will
be marked with this
symbol when
it
is necessary for the user to refer
to
the instruction
manual in order
to
protect against damage
to
the instrument.
A
Indicates dangerous voltage (terminals fed from the interior by
voltage exceeding
lo00
volts must be
so
marked).
/s
Protective conductor terminal. For protection against electrical
shock in case
of
a
fault.
Used with field wiring terminals
to
in-
dicate the terminal which must be connected
to
ground before
4
-
OR@
operating equipment.
Low-noise
or
noiseless, clean ground (earth) terminal. Used for
a
signal common, as well as providing protection against electrical
shock in case of a
fault.
A terminal marked with this symbol must
.beconnected to ground in the manner described in the installation
(operating) manual, and before operating the equipment.
D
,
Frame or chassis terminal. A connection to the frame (chassis) of
/f;r
OR
1,
the equipment which normally includes
all
exposed metal struc-
tures.
Alternating current (power line).
-
--
-
Direct current (power line).
Alternating or direct current (power line).
The WARNING sign denotes
a
hazard.
It
calls attention
to
a
pro-
cedure, practice, condition
or
the like, which,
if
not correctly per-
0
formed or adhered to, could result in injury or death
to
personnel.
WARNING
SAFETY
SYMBOLS
General Definitions of Safety Symbols
Used
On Equipment
or
In Manuals.
The CAUTION sign denotes
a
hazard.
It
calls attention
to
an
operating procedure, practice, condition
or
the like, which,if not
correctly performed or adhered to, could result in damage
to
or
destruction of part or all
of
the product.
CAUTION
The NOTE sign denotes important information.
It
calls attention
to
procedure, practice, condition or the like, which is essential
to
highlight.
E
'

Model 333A/334A
TABLE
OF
CONTENTS
Table
of
Contents
Section Page
I
GENERAL INFORMATION
........
1-3
1.1
. Description
............
1-3
1.6
.
Accessory Features
........
1-3
1.8
.
Option
...............
1-3
1.10
.
Instrument Identification
.....
1-3
Section
11
INSTALLATION
..............
2
.
1
. Introduction
............
2.3
.
Inspection
.............
2.5
.
Power Requirements
.......
2.7
.
Three-Conductor Power Cable
..
2
.
10
.
Installation
............
2.12
.
Bench Installation
.......
2.14 . Rack Installation
........
2.16
.
Repackaging For Shipment
....
page
2-1
2-1
2-1
2-1
2-1
2-1
2-1
2-1
2-1
Section
-ge
III
OPERATINGINSTRUCTIONS
......
3-1
3-
1
.
Introduction
............
3
.
1
3.4
.
Controls and Indicators
......
3-1
3.6
.
General Operating Information
.
. 3-1
3.7 . Input Connections
.......
3-1
3.9 . Voltmeter Characteristic$
.
.
3-1
3.12
.
Output Terminals
.......
3-2
3.14
.
Operating Procedures.
......
3-2
3.15 . Instrument Turn.On
......
3-2
3.16
.
Adjustment of Meter
Mechanical Zero
.......
3-2
3.18
.
DistortionMeasurement
...
3-2
3.21
.
RF Carriers (334A only)
...
3-3
3.22
.
Voltage Measurement
.....
3-3
3.23
.
Meter
Indication
.......
3-4
3.27
.
Use
of
Output
Terminals ... 3-4
3.30 . 333A1334A with Option 01 .. 3-4
3.32
.
Manual Nulling
........
3-4
Section
Page
IV THEORY OF OPERATION
........
4-1
4.1
.
Overall Description
........
4-1
4.3
.
Block Diagram Description
....
4-1
4.4
.
Distortion Measuring
4.8 . Carriers.
..........
4-1
4.10 . Voltmeter Operation
.....
4-1
4.12
.
Detailed Circuit Description
...
4-1
4.13
.
Impedance Converter Circuit
.
4-1
4.17
.
Rejection Amplifier Circuit
.
.
4-2
4.39
.
High Pass
Filter
.......
4-5
4.41
.
Meter Amplifier
.......
4-5
4.49
.
Power Supply Circuit
.....
4-7
4.54
.
RF Detector Circuit(334Aonly) 4-7
Section Page
5.1
.
Introduction
............
5-1
5.3
.
Required Test Equipment
.....
5-1
Distortion Measurement
of
AM
Operation
..........
4
.
1
Distortion Measurement in
AM
V
MAINTENANCE
.............
5-1
Section Page
V
MAINTENANCE (Cont’d)
.........
5-1
5.5 . Performance Checks
.......
5-1
5.9 . FundamentalRejection Check . 5-1
5.10
.
Second Harmonic Accuracy .
.
5-2
5.11
.
Instrument Induced
Distortion Check
.......
5-2
5.12 . Frequency CalibrationAccuracy
Check
.............
5-3
5.13
.
Automatic Nulling Mode Check 5-4
5.14
.
Input Impedance Check
....
5-4
5.17 . Minimum Input Level Check.
.
5-5
5.18 . DC Isolation Check
......
5-5
5.19 . Voltmeter Accuracy and
Frequency Response Check 5-6
5.20
.
High
Pass
Filter
Check
....
5-7
5.21
.
Residual Noise Check ..... 5-7
5.22 . AM Detector Check (Model
5.23 .Adjustment and Calibration
Procedure
............
5-8
5.26
.
Adjustments
.........
5-8
5.27
.
Voltmeter Gain Adjustments . 5-8
5.28
.
Voltmeter Frequency Response
Adjustment
.........
5-10
5.29
.
Bridge Balance Adjustment
.
. 5-10
5.30 . Sensitivity SwitchCalibration
.
5-11
5.31
.
Components
.........
5-11
334Aonly)
..........
5-7
Power Supply and Bias
Adjustment of Factory Selected
5.33 . Troubleshooting Procedure
....
5-11
5
.38
.
Troubleshooting Voltmeter
Function
...........
5
.
12
5.40
.
Troubleshooting Distortion
Function
...........
5-14
5.45
.
Photoresistor Check and
Adjustment (A6V1thruA6V5)
.
5-15
5.49
.
Servicing Etched Circuit
Boards
...........
5-16
5.52 . Servicing Rotary Switches
...
5-16
5-54
*
Converter
........
5-16
Section Page
VI REPLACEABLE PARTS
.........
6-1
6.1
.
Introduction
............
6-1
6.4
.
Ordering Information
.......
6-1
6.6
.
Non-Listed
Parts
.........
6-1
Isolating the Impedance
Section Page
M
CIRCUITDIAGRAMS
...........
7-1
7-
1
.
Introduction
............
7-1
APpen&
A
CODE
LIST
OF
MANUFACTURERS
B SALES AND SERVICE OFFICES
C MANUAL BACKDATING CHANGES
iii

Table of Contents
LIST
OF
TABLES
Model
333A1334A
Page
1-
1
.
Specifications
..............
1
.
1
5.5
.
Power Supply and
Bias
Adjustments
..
5-8
Measurements
............
3-1 5.7
.
Factory Selected Components
.....
5-12
5.2
.
Second Harmonic Accuracy Check
...
5-2 5.8
.
Troubleshooting Aid
..........
5-13
Number Page Number
3.1
. Effect of Harmonics on Voltage
5.6
.
Sensitivity Switch Calibration
.....
5-12
5.1
. RequiredTestEquipment
........
5-0
5.3
. Frequency Calibration Accuracy Check
5-4 5.9
.
Photoresistor Check
..........
5-15
5.4
.
Voltmeter Frequency Response Check .
5-6 6.1
.
Replaceable Parts
...........
6-6
LIST
OF
ILLUSTRATIONS
Number Page Number Page
1.1
.
3.1
.
3.2
.
3.3
.
4.1
.
4.2
.
4.3
.
4.4
.
4.5
.
4.6
.
4.7
.
4.8
.
5.1
.
5.2
.
5.3
.
Model
333A
Distortion Analyzer
....
Front and Rear Panel Description
...
Manual Nulling Test Setup
Impedance Correction Graph
......
Block
Diagram
.............
Bridge Waveforms
...........
Wien Bridge Circuit and Rejection
Characteristics
...........
Auto Control Loop Detector
.......
Reference and Error Phase
Relationship
.............
Rejection Amplifier Block Diagram and
Typical Frequency Rejection
Characteristic
............
Bandwidth Versus
Null
Depth
......
Simplified Metering Circuit
.......
Fundamental Rejection Check
.....
Instrument Induced Distortion Check
..
Frequency Calibration Accuracy Check
111
5.4
.
3
-0
3-
5 5.5
.
3-5 5.6
.
4-0
4-2 5.7
.
5.8
.
4-3 6.1
.
4-3 6.2
.
6.3
.
4-4 7.1
.
7.2
.
4-5
7.3
.
4-6
4-7 7.4
.
5-1 7.5
.
5-2
5-3
7.6
.
Capacitance Check
...........
5-5
DC
Isolation Check
...........
5-6
Voltmeter Accuracy and Frequency
Response Check
...........
5-7
Shielded Load Assembly
........
5-8
Adjustment and Chassis Mounted
Component Location
..........
5-9
Mechanical
Parts
...........
6-2
Modular Cabinet
Parts
.........
6-4
Frequency Tuning Assembly
......
6-5
Impedance Converter (P/O
A2)
and RF
RejectionAmplifier (A3). Bridge Balance
(P/OA6) andH h
Pass
Filter
(A7)
7-7/7-8
Meter AmplifiervP/O
A2)
......
7-9/7-10
Control Circuit
(A5)
andBridge
Power Supply
(Al)
.........
7-13/7-14
Internal Wiring
Data
.........
74/74
Detector
(A4)
...........
7-5/74
Balance (P/O A6)
........
7-11/7-12
lv

Model 333A/334A
RANGE
100%-0.3%
0.1%
Section
I
*3%
*ts%
*12%
1MHz
3MHz
10
HZ
10H~
30
Hz
20
Hz
10 Hz
300
kHz
500
Wz
1.2
MHz
SECTION
I
GENERAL INFORMATION
1-1.
DESCRIPTION.
1-2.
The Hewlett-Packard Models 333A
and
334A
Distortion
Analyzers axe solid
state
instrumentsfor
measuring distortion on ac voltages. The
Models
333A and 334A include
two
control loops
that
auto-
maticdy
tune
both
legs of
a
bridge
circuitwhich
re-
jects the fundamental when the rejection circuit
is
initially
set
within the
range
of
the loops. The 334A
has
a
high
impedance detector which operates from
550
Mz
to
greater
than
65
MHz and provides the ca-
pability
of
monitoring the distortion of the amplitude
modulation on an
rf
carrier.
1-3.
Distortion levels
of
0.1% to 100%fullscale
are
measured in seven rangesforanyfundamentalfrequen-
cy of
5
Hz
to
600
Wz.
Harmonics are indicated upto
3
MHz.
The high sensitivity of these instruments re-
quires only 0.3V rms for the 100%set level reference.
The distortion characteristics can
be
monitored atthe
OUTPUT connectors with an oscilloscope,
a
true rms
voltmeter, or
a
wave analyzer. The instruments are
capable of an isolation voltage of 400 volts above
chassis ground.
1-4. The voltmeter can
be
used separatelyforgeneral
purpose voltage and gain measurements.
It
has
a
frequency range of
5
Hz
to 3
MHz
(20
Hz
to
500
Mz
for 300 pV
range)
and
a
voltage range of 300pV to
300
V
rms full scale.
1-5. The
AM
detector included in the Model 334A
is
a
broadband dc restoring peak detector consisting of
a
semiconductor diode and
filter
circuit.
AM
distor-
tion levels
as
low
as
0.3% can be measured on
a
3
V
to
8
V
rms
carrier modulated 30% in the standard
broadcast band. Distortion less than
1%
can
be
measured
at
the 8amelevelof the carrier up to
65
Mc.
1-6. ACCESSORY FEATURES.
1-7. The accessory available with the 333Aand334A
Distortion Analyzers
is
a
voltage divider probe, -hp-
Model No. 10001A. The features
of
the probe are:
a.
10 megohms shunted
by
10 pF, giving
1O:l
b.
c.
2%
division accuracy.
d.
600Vpeakinput.
e.
5
ns rise-time.
attenuation.
DC
to 30 MHz bandwidth.
1-8.
OPTION.
1-9. Option 01
is
a
standard -hp- Model333A
or
334A
with
a
special meter and meter amplifier, compen-
sated to permit response to
W
(volume
units)
characteristic
8.
1-10
INSTRUMENT IDENTICATION.
1-11.
Hewlett-Packard uses
a
two-section
serial
number. The first section (prefix) identifies
a
series
of instruments. The
last
section (suffjx) identifies
a
particular instrument within the series.
If
a letter
is
included with the serial number, it identifies the coun-
try in which the instrument
was
manufactured.
If
the
serial prefix of
your
instrument differs from the one
on the title page of this manual,
a
change sheet
will
be
supplied to make this manual cornpatable
with
new-
erinstruments
or
the backdating information in Appen-
dix
C will
adapt this manual to earlier instruments.
All correspondence with Hewlett-Packard should in-
clude the complete serial number.
Table
1-1.
Specifications
MODEL 333A Fundamental
Input
Greater Than 30
V
DISTORTION MEASUREMENT RANGE
Any fundamental frequency,
5
Hz to
600
kHz.
Distortion levels of
0.
1%-100%are measured
full scale
in
7
ranges.
Harmonic measurement accuracy
(full
scale)
Fundamental Input Less Than 30
V
DISTORTION MEASUREMENT
ACCURACY
RANGE
I
*3%
1
*6%
I
*12%
50 Hz
Elimination Characteristics:
Fundamental Rejection
>
80
dB
Second Harmonic Accuracy for
a
fundamental
of:
5
Hz
to 20 Hz: better than
+1
dB
20 Hz to 20
Wz:
better than
4.6
dB
20
kHz
to 100
kHz:
better
than
-1
dB
100
Wz
to 300
kHz:
better than -2 dB
300
kHz
to
600
kHz:
better
than -3 dB
1-1

Model
333A/334A
Section
I
300
pv
1
mV-30 V
100
V-300 V
Table
1-1.
Specifications (Cont'd)
30
HZ-300
IrHz
20
HE-~OO~HZ
10
He-1
MHe
5
Hz-3
MHe
10 HE-300
~HZ
5
He-500
kHz
Distortion Introduced
by
Instrument:
>
-70
dB
from
5
He
to 200
kHz
>
-64
dB
from 200 kHz to
600
kHz
Meter
indication
is
proportional to the aver-
age
value
of
a
waveform.
FREQUENCY CALIBRATION ACCURACY
Better
than
*5%
from
Better
than
f
10%
from
300
kHz
to
600
kHz
5
Hz
to 300 kHz
INPUT IMPEDANCE
Distortion
Mode:
1
MS1
*5%
shunted
by
<70
pF.
Voltmeter Mode:
1
Mi2
15%
shunted
by
40
pF (333A
only),
1
Mi2
*5%
shunted by
45
F
(334A
only),
1
to
300
V
ranges;
1
MS1
*5
4l
shunted
by<70
pF,
300
pV to
0.3
V
ranges.
INPUT LEVEL
FORDISl?ORTIONMEASOaEMENTS
0.3
V
rrns for
100%
set
level (up to
300
V
may
be
attenuated to
set
level reference). The
minimum measurable distortion
for
floating
operation on the
X1
frequency
range
L
50dB
below the fundamental.
DC ISOLATION
Signal
ground may
be
rl00
Vdc from external
chassis.
VOLTMETER RANGE
300
pV
to
300
V
rrns
full scale
(13
ranges),
10
dB
per
range.
5
Hz
to
3
MHz
(300
pV
range:
20
Hz-500
Hie).
VOLTMETER FREQUENCY RANGE
VOLTMETER ACCURACY:
RANGE
I
*2%
I
*5%
I
NOISE MEASUREMENTS
Voltmeter
residual
noise on
the
300
pV
range:
<
25
pV
rms terminated
in
shielded
60OQ
<30
pV
rrns
terminated in shielded
100
kS1.
OUTPUT
For input frequencies from
20
Hz
to 600
kRz,
0.1
V
rma
*
0.01
V
open circuit
for
full
scale
meter deflection;
0.05
V
rms
f
0.005
V
into
2
kS2
for
full scale meter deflection.
AUTOMATIC
NULLWG
MODE
Set
Level:
At
least
0.2
V
rms.
Frequency Ranges:
X1,
manual null tuned to
less
than
3%
of
set
level; total frequency hold-in
io.
5%
about
true manual
null.
X10
through
X10
K,
manual
null
tuned
to
less
than
10%
of
set
level;
total
frequency
hold-in
1
1%
about
true manuaX null.
AUTOMATIC NULL ACCURACY
5
Hz
to
100
Iiz:
Meter reading within
0
to
+3
dB
of manual null.
100 Hz
to
600
kHz:
Meter reading within0 to
+l.
5
dB
of manual null.
HIGH-PASS FILTER
3
dB point
at
400
Hz
with
18
dB
per
octave
roll
off.
60
Hz
rejection
>
40
dB.
Normally used
only
with fundamental frequencies greater than
1
BHZ.
POWER SUPPLY
100
V/120 V/220 V/240
V
+
5%
-
108,
48
-
66
Hz,
approximately 4
watts.
-
MODEL 334A
1.6
MHz
to
65
MHz:
Distortion introduced
by
detector
is
c
1%
for
3
to
8
volts rms carriers
modulated
30%.
Same
as
Model 333A except
as
indicated below:
AM
DETECTOR
High
impedance dc restoring
peak
detector with
semi-conductor
diode
operates from
550kHz
to
greater than
65
MHz.
Broadband input.
Maximum
input;
40
V
p-p
ac
or
40
V
peak
transient.
CARRIER
FREQUENCY
550kHzto
1.6
MHz:
Distortion introduced
by
detector
is
c
0.3%
for
3
to
8
volt carriers
modulated
30%.
NOTE
Distortion measurement
at
carrier levels
as
low
as
1
volt may
be
made with reduced
accuracy.
OPTION
01
Indicating meter
has
W
characteristics con-
forming to FCC Requirements for
AM,
FM,
and TV broadcasting.
1-2

Model 333A/334A
StCTION
II
INSTALLATION
Section
XI
2-1.
I
NTR ODUCTION.
2-2.
This section contains information
and
instruc-
tions necessaryforthe installation
and
shipping
of
the
Models 333A/334A Distortion Analyzers. Included
areinitial inspectionprocedures, power and grounding
requirements, installation information,
and
instruc-
tions for repackaging for shipment.
2-3.
INSPECTION.
2-4.
This instrument
was
carefully inspected both
mechanically
and
electrically before shipment.
It
shouldbe physically
free
of mms
or
scratches
and
in
perfect electrical order upon receipt. To confirm
this,
the instrument should
be
inspected for physical
damage
in
transit. Also checkforsuppliedaccessories,
and
test
the electricalperformance of
the
instrument
using the procedure outlined
in
Paragraph 5-5.
If
thereisdamageordeficiency,
see
the warranty on
the
inside front cover
of
this manual.
2-5.
POWER REQUIREMENTS.
2-6. The
Model
333A/334A can
be
operated from any
ac source
of
100 V/120 V/220 V/240
V
+
5%
-
10%
48
-
66
Hz. With the instrument disconnected from
the ac power source, move the slide (located on the
rear panel) until the desired line voltage appears.
Power dissipation
ie
approximately 4
watts.
2-7.
THREE-CONDUCTOR POWER
CABLE.
2-8. To protect operating personnel, the National
ElectricalManufacturers'
Association (NEMA) recom-
mendsthat
the
instrumentpanelandcabinet
be
grounded.
All
Hewlett-Packard instruments are equipped with
a
three-conductor power cable, which, when plugged in-
toanappropriate receptacle, grounds the instrument.
Theoffsetpinon the power cable three-prong connec-
tor
is
the
ground
wire.
2-9
.INSTALLATION.
2-10. The Model 333A/334A
is
fully transistorized;
therefore, no special cooling
is
required. However,
the instrument should not
be
PerateBwhere the am-
bient
temperature exceeds 55
C
(131
F).
2-11.
BENCH mSTALLATION,
2-12.
The
Model
333A/334A
is
shipped with plastic
feet and tilt
stand
in place, ready for use
as
a
bench
instrument
.
2-13.
RACK
INSTALLATION,
2-14. The Model 333A/334A may
be
rack mounted
by
usingthe 5"RackMount Kit (-hp- Part No. 5060-0775).
Instructions for the conversion are included with the
kit. The rack mount for
the
Model 333A/334A
ie
an
EIA standard width of
19
inches. When mounted in
a
rack using
the
rack mount
kit,
additional support
at
the rearof theinstrument should be provided
if
vibra-
tion
or
similar
stress
is
likely.
2-15.
REPACKAGING FOR SHIPMENT.
2-16. The following paragraphs contain
a
general
guide
for
repackaging
of
the instrument
for
shipment.
Refer
to Paragraph 2-17
if
the
original
container
is
to
be
used; 2-18
if
it
ia
not.
If
you
have any questions, con-
tact your
local
-
hp
-
Sales and Service Office. (See
Appendix
B
for
&ice locations.)
NOTE
If
the instrument
is
to
be
shippedsto
Hewlett-Packardfor service
or
repair,
attachatag to the instrument identify-
ing the owner and indicate the service
or
repairto be accomplished. Include
themodelnumber and
full
serial
num-
ber
of
the instrument.
In
any corres-
pondence, identify the instrument
by
model number, serial number, and
serial number prefix.
2-17.
a
able.
If
original
container
is
to
be
used, proceed
as
follows:
Place instrument in
original
container
if
avail-
If
original
container
is
not available,
a
suitable
one can
be
-purchased from your nearest -hp-
Sales
and Service Office.
b.
Ensure that container
is
well sealed with strong
tape
or
metal
bands.
2-18.
If
original
container
is
not to
be
used, proceed
as follows:
a.
Wrap
instrument in heavy paper
or
plastic
be-
fore placing in an inner container.
b.
Place packing material around
all
sides of in-
strumentandprotectpanel facewith cardboard strips.
c. Place instrument andinner container in
a
heavy
carton or wooden box
and
seal
with strong
tape
or
metal bands.
d. Mark shipping container with "DELICATE
INSTRUMENT,
I'
"FRAGILE,
I'
etc.
2-1

Section
TII
Model
333A/334A
@LINE switch turns instrument ac power on.
Pilot lamp glows when instrument
is
turned
ON.
@Meter indicates distortion
or
voltage level of
input.
@MODE switch selects MAIUUAL
or
AUTOMATIC
bridge tuning operation.
@FREQUENCY RANGE switch selects frequency
range which corresponds
to
fundamental of input
@COARSE BALANCE control provides coarse
ad-
justment for balancing the
Wien
bridge circuit.
6
FINE BALANCE control provides
a
vernier
ad-
'justment forbalancing the Wien bridge circuit.
7
Frequency
dial
tunes
the
Wien
bridge
circuit to
'fundamental
of
input signal.
@HJGH PASS FILTER switch inserts
or
bypasses
HIGH
PASS
FILTER
in
SET LEVEL and
DIS-
TORTION function. When inserted, filter pro-
vides
>
40
dB attenuation
to
50
-
60
Hz hum
components but no attenuation
to
frequencies
over
1
kHz.
@OUTPUT
connectors provide means of monitor-
ing output of meter circuit.
@Frequency vernier provides
fine
adjustment of
frequency dial.
@METER RANGE switch selects
full
scale range
of meter
in
percent,
dB
and rms volts.
Signal.
OSENSITMTY selector provides
r)
-
50
dB
atteau-
@SENSITIVITY VERNIER control provides fine
@Mechanical zero adjust provides mechanical
OFUNCTION switch selects type of operation
of
@Shorting
bar
connects circuit ground
to
chassis
@INPUT terminalsprovide connections for input
ation of input signal
in
10
dB
steps
in
SET LEVEL
and
DISTORTION functions.
adjustment of sensitivity setting.
zero adjustment of meter.
instrument.
ground.
stgnal.
@RF
INPUT
connector (Model
334A
only) provides
input connection for
AM
RF carrier input signal.
20
FUSE provides current overloadprotection for
0
instrument circuits.
21
Line voltage switch sets inrrtrument to
'operate
from
100
V/120
V/220
V/240
V.
@)
Ac power connector provides input connectton
for ac power.
~~~
Figure
3-1.
Front and Rear Panel Description
3-
0

Model 333A/334A
100
100.5
102
112
100.5
102
112
Section
III
I
SECTION
111
OPERATING INSTRUCTIONS
3-1.
INTRODUCTION.
3-2. The Models 333A and 334A DistortionAnalyzers
measure
total
harmonic distortion
of
fundamentalfre-
quencies from 5
Hz
to
600
kHz;
harmonics upto3
MHc
are included. The sharp elimination characteristics
(
>80
dB),
the low level of instrument induced
dtstor-
tion, and the meter accuracy
of
theseinstrumentsre-
sult
in
highly accurate measurement of low level har-
monic distortion.
3-3.
An
rms
calibrated voltmeter
is
inherent in the
333A and
334A.
The voltmeter provides
a
full
scale
sensitivity
of
300
p
voltsrms(residualnoise
<25p
volts).
The voltmeter frequency range
is
from 5
HZ
to 3 MHz
except on the 0.0003 volt range, which
is
from 20
Hz
to 500
kHz.
3-4.
CONTROLS AND INDICATORS.
3-5. Figure 3-1
illustrates
anddescribesthefunction
of
all
front
and
rear
panel
controls, connectors, and
indicators.
keyed tothe drawing included within the
figure,
The description
of
each component
is
3-6.GENERAL OPERATING INFORMATION.
3-7.
INPUT
CONNECTIONS.
3-8. The
input
signal
can
be
connected to the 333A
and 334A through twisted pair
leads
or
a
shieldedcable
with banana plug connectors. Keep
all
test
leads
as
short
as
possible to avoid extraneous pickup from
stray
ac
fields. When measuring Low-level
signals,
ground loope may occur causing erroneous readings.
Ground loops may
be
avoided
by
connecting the 333A/
334A Distortion Analyzer toan appropriate isolation
transformer to break the chassis ground from power
supply ground. Connect
all
other
instruments
to one
power
strip
with the three-prong connectors
as
close
as
possible.
3-9. VOLTMETER CHARACTERISTICS.
3-10. The RMS VOLTS markmgs on the meter face
arebased ontheratiobetweentheaverageandeffective
(rms)values
of
a
pure
sine
wave, Theratioof average
to effective values in a true sine wave
is
approximately
0.9
to
1.
When the meter
is
used to measure complex
waves,
the voltage indicated may not
be
the rms value
of the
signal
applied. This deviation
of
meter indica-
tion exists because the ratios of average to effective
values are usually not the same in
a
complex wave
as
in
a
sine wave. The amount of deviation depends on
the magnitude and phase relation betweenthe harmonics
and fundamental frequency of the signal applied.
Table 3-1
lists
the deviation of the meter indication
of
a
sine wave
partly
distorted
by
harmonics.
As
in-
dicated in the table, harmonic content of less than
108
results
in
very small errors.
Table
3-1
Effect of Harmonics on Voltage Measurements
Input Voltage
Characteristics
Fundamental
=
100
Fundamental
+
10%
2nd harmonic
Fundamental
+
20%
2nd harmonic
Fundamental
+
50%
2nd harmonic
Fundamental
+
10%
3rd harmonic
Fundamental
+
208
3rd harmonic
Fundamental
+
50%
3rd harmonic
Meter
100
100
100
-
102
100
-
110
96
-
104
94
-
108
90
-
116
NOTE
This chart
is
universal in application
since these errors are inherent in
all
average-responding voltage-measuring
instruments.
3
-11.
When making distortion measurements where
tne fundamental frequency
is
suppressed and the re
-
mainder of the
signal
is
measured, the reading ob-
tained on an average responding meter may deviate
from the true total rms value. When the residual
wave contains many inharmonically related sinusoids,
the maximum error
in
the distortion reading
is
about
11
%
(11
%
of the measured distortion) low for dis-
tortion levels below 10
%.
EXAMPLE
:
Measured Maximum Error Total
Distortion In Meter Indication Distortion
2.5%
+O.
llx0.025= 0.025+0.0027=
0.00027 0.0277
or
2.8%
This example representsthe maximumpossible
error,
and in most cases the error
is
less. In distortion
measurements, the reading
of
an
average-responding
meter
is
sufficiently close to the rms value to
be
satisfactory for most applications.
3-12.
OUTPUT
TERMINALS.
3-13. The
OUTPUT
terminals provide
a
0.1
V
rms
open circuit
output
for
full
scale meter deflection.
These terminals can
be
used to monitor the output
signal
with
an
oscilloscope,
a
true rms voltmeter,
or
a wave analyzer. The combination
of
the distortion
meter and oscilloscope provides more significant in-
3-1

Section
IIf
Model
333A/334A
formation about the device under test than the
expression of distortion magnitude alone. Information
obtained from the oscilloscope pattern
is
specific
and
reveals
the
nature
of distortion which sometimes
occurs
at
such low levels that
it
is
difficult to detect
in
the presence
of
hum and
noise.
The impedance
at
the
OUTPUT terminals
is
2000
ohms,
therefore, ca-
pacitive loads greater than
50
pF should
be
avoided
to
maintain the accuracy
of
meter readings.
3-14.
OPERATING PROCEDURES.
3-15.
INSTRUMEN"
TURN-ON.
a
Set the
115-230
VAC switch
to
coincide with the
line
voltage in
we.
Turn switch to ON position.
Pilot lamp
will
glow, indicating application
d
primary power.
3-16.
ADJUSTMENT OF METER MECHANICAL
ZERO.
3-17.
The meter
ia
properly zero-set when the
pointer rests over the zero calibration mark and the
instrument
ie
in
its
normal operating enivronment
and
is
turned
off,
Zero-set the meter
as
follows
to
obtain
maximum accuracy and mechanical stability:,
Turn instrument on and allow
it
to
operate
for
at
least
20
minutes, to let meter move-
ment reach normal operating temperature.
b.
Turn
instrument off and allow
30
seconde
for
all
capacitorsto discharge.
c. Rotate zero adjustment screw clockwise
un-
til
pointer
is
left
of
zero and moving upscale.
d.
Continue rotating screw clockwise; stopwhen
pointer
ia
exactly
at
zero.
e.
When
pointer
is
exactly over zero, rotate
adjustment screw
slightly
counterclockwise
to
relieve
tension on pointer suspension.
If
pointer moves
aff
zero, repeatstepsc
through
e,
but make counterclockwise
rotation
lese.
a.
3-18.
DISTORTION
MEASUREMENT.
e
DO
NOT EXCEED THE INPUT VOLT-
DAMAGING COMPONENTS
ON
A2
BOARD.
AGES
LISTED
BELOW
ro
PREVENT
VOLTMETER FUNCTION
-lV
RANGE AND BELOW, AND
DISTORTIONANALYZER
FUNCTION
-
MAXIMUM
SENSITIVITY.
1.
300
VABOVE
100
HE
2.
50
VABOVE
1
kHz
3-19.
MANUAL MODE.
a.
Turn
instrument on and mechanically
zero
meter according to procedure in
Paragraphs
3-15
and
3-16.
b.
C.
d.
e.
f.
g.
h.
k.
1.
m.
Set NORM-R.
F.
DET. switch to NORM.
Set FUNCTION switch to SET
LEVEL.
Set MODE switch to MANUAL.
If
fundamentalfrequency
is
1
kHz
or
greater,
set
HIGH
PASS FILTER switch to
IN.
Set SENSITIVITYselector to
MIN.
position,
and rotate VERNIER control maximum
counterclockwise.
NOTE
The bandwidth of the SENSI-
TMTY selector
is
reduced
in
the
two
extreme
CCW
positions (positions used with
an input
signal
greaterthan
30
V).
Set METER RANGE switch to SET
LEVEL,
and
set
BALANCE
COARSE
and
FINE
con-
trols to center position.
Connect
signal
to
be
measuredto333A/334A
INPUT terminals.
I
WARNING
1
REMOVE SHORTING
STRAP
BETWEEN
FRAME GROUND
(A
)
AND
CHASSIS
GROUND (
)
TERMINALSON
FRONT
PANEL INPUT TERMINALS WHEN
MEASURING DISTORTION BETWEEN
TWOPOINTSWHICHAFEDCOFFSET
FROM GROUND POTENTIAL.
Set
SENSITIVITY selector to
obtain
meter
indication greater than
1/3
full
scale.
Adjust SENSITIVITY VERNIERfor
full
scale
meterindication
if
making
distortion
measure-
ment in percent;
if
makingdistortionmeasure-
ment
in
dB
adjust SENSITMTY
VERNIER
for
0
dB
meter indication.
NOTE
If
unable to adjust for full scale
or
OdB
indication, (whichindicates
input
signal
is
below
0.3
volts),
set METER RANGE selectordown-
scale.
Use
this
new
setting
as
the
100'KDor
OdB
SET LEVEL position,
thus making the next range
30%
or
-10
dB,
etc.
Set FREQUENCY RANGE switchandfrequency
dial
to fundamental frequency of input
signal.
Set FUNCTION switch to DISTORTION.
Adjust frequency
dial
vernier and BALANCE
COARSE and FINE controls for minimum
3-2

Model
333A/334A
Section
III
meter indication.
Set
METERRANGE
switch
down-scale
as
necessary to keep meter
in-
dication on scale.
n.
Repeat step
m
until
no further reduction
in
meter indication can
be
obtained.
0.
Observe distortion
either
in
percentage or
dB,
as
indicated
by
meter deflection and
METER RANGE
switchsetting.
For
example,
if
meter indicates
0.4
and
METER
RANGE
setting
is
18,
distortion measured
is
0.4%
of
fundamental. Similarly,
if
meter
indi-
cates
-6
dB
and
METER RANGE
setting
ie
-40
dB, distortion measured
is
-46
dB from
fundamental.
NOTE
In
MANUAL
mode
the
accuracy
of
distortion measurements
is
affected
by
frequency stability
of
the input
signal.
An
inaccuracy
in
distortion indications occurs
when
the
frequency
drift
of
the
input
signal
exceeds
the
bandwidth
of
the rejection curve.
p.
If
desired, rms voltage of
input
signalcan
be
measured
by
setting
FUNCTION
switch to
VOLTMETER,
and
setting
METER
RANGE
switch
to
obtain
an on-scale indication.
3-20. AUTOMATIC
MODE.
a
b.
C.
d.
e.
f.
Perform steps athroughlof Paragraph
3-19.
Adjust frequency
dial
vernier and
BALANCE
COARSE
and
FINE
control8for minimum
meter indication.
When meter indication
is
less than
10%
of
SET LEVEL
indication,
set
MODE
switch to
AUTOMATIC.
(If
fundamental cannot
be
manually
nulled below
10%
of
SET LEVEL
indication, automatic mode cannot
be
used).
Set
METER RANGE
switch down-scale to
obtain on-scale meter indication.
Observe distortion either in percentage
or
dB,
as
indicated
by
meter deflection and
METER RANGE
switch
setting.
Forexample,
if meter indicates
0.4
and
METER RANGE
setting
is
18,
distortion measured
is
0.4%
of
fundamental. Similarly,
if
meter indicates
-6
dB
and
METER RANGE
setting
is
-40
dB,
distortion measured
is
-46
dB from funda-
mental.
If
desired, rms voltage of input
signal
canbe
measured by setting
FUNCTION
switch to
VOLTMETER,
and
setting
METER RANGE
switch to obtain
an
on-scale indication.
3-21. DISTORTION MEASUREMENT OF
AM
RF
CARRIERS
(334A
only).
DO
NOT EXCEED
MAXIMUM
INPUT
VOLTAGES LISTEDON
REAR
PANEL.
a.
Turn instrument
on
and
mechanically zero
meter according toprocedure
in
Paragraphs
3-15
and
3-16.
b.
Set
NORM.-R
F.
DET.
switch to
R.
F.
DET.
c. Connectinput
signal
to
R
F.
INPUT
terminal
on rear panel.
Refer
to Paragraph
3-19
formanual distort-
ion measurement;
refer
to
Paragraph
3-20
for
automatic distortion measurement.
d.
NOTE
If
no
meter deflection can
be
obtained with
an
RF
input,
diode
A4CR1
should
be
checked.
A
spare diode
is
located on
the
outside
of
the
A4
shield.
3-22. VOLTAGE MEASUREMENT.
a.
Turn instrument on and mechanically zero
meteraccordingto procedure in Paragraphs
3-15
and
3-16.
b.
Set
NORM.-R.
F.
DET.
switch to
NORM.
c.
Set
FUNCTION
switch to
VOLTMETER.
6
Set
METER RANGE
switch to
a
range
exceeding amplitude
of
signal to
be
mea-
sured.
e. Connect
signal
to
be
measured to
INPUT
terminals.
f.
Set
METER RANGE
switch to give a reading
as
close to
full
scale
as
possible, and
ob-
serve meter indication.
g. The dB scale
of
the
333A/334A
is
calibrated
in
dBm, such
that
0
dBm =
1
milliwatt
dissi-
pated
by
600
ohms.
Therefore,
a
dBm mea-
surement must
be
made across
600
ohms.
However,
dB
measurements
across
other
impedances can
be
converted to dBm
by
use
of
the Impedance Correction Graph of
Fig-
ure
3-3.
For example: to convert
a
-30
dB
reading
across
200
ohms
to dBm, locate the
200
ohm
impedance line
at
the bottom
of
the
graph.
heavy
black
line, and
read
the
meter correc-
tion
at
that
point. The correction for
200
ohms
is
+5
dBm;
thus
the
corrected reading
is
-25
dBm.
3-3
Follow the impedance
line
to
the

Section
III
3-23. METER INDICATION.
3-24. The 333A/334A meter
is
calibrated to indicate
in
both dB and volts.
It
is
interesting to notethat the
METER RANGE markings differ from most ac volt-
meter range markings.
On
most
ac
voltmeters
(600
ohms)
0
dB
corresponds
to
the
1
volt
range.
This
is
not
true
inthecase
of
the333A/334A. Sincethe
instru-
ment
is
primarily
a
distortionanalyzer, measurements
are
in
dB (relative measurement) rather than
in
dBm
(absolute measurement). Zero dB on the 333A/334A
corresponds to0.3 volt range rather than
the
1
volt
range. This allows a10
dB
greater
dynamic
range
of
distortion measurements.
Model 333AI334A
3-30. 333A/334A
WT”H
OPTION 01.
3-25.
E
measurements
are
to
be
made in dBm, 10
dB
must
be
subtracted from the METER RANGE
setting.
Thus
0
dB
becomes the -10 dBm
range
for absolute
power measurements. Zero dBm
is
equal
to
1
milli-
watt dissipatedbyanyimpedance andinthis particular
case
is
800
ohms.
The +10 DECIBELS marking on the
meter
face
indicates
thatwhenvoltmetermeasurements
are
being made, the indication (METER RANGE plus
meter indication)
is
10
dB
greater
than when power
(dBm) measurements
are
being made.
3-26.
In
short, when distortion and voltage measure-
ments
are
being made, utilizethe instrumentMETER
RANGE and meter scale
as
they
exist.
For absolute
power measurements in dBm, simply subtract 10
dB
from the METER RANGE
setting.
3-27. USE OF OUTPUT TERMINALS.
3-28.
InVOLTMETERand SET LEVEL functions, the
333A/334A can
be
used
as
a
low distortion, wide-band
amplifier. A portion of the meter
input
(0.1
V
rms
open
circuit for
full
scale meterdeflection
is
provided
at
the
OUTPUT terminals.
3-29.
In
DISTORTION function, the distortion (0.1
V
rmsopen circuit forfull
scale
deflection)
is
provided
at
the OUTPUT terminals for monitoring purposes.
NOTE
The INPUT terminal and the
OUTPUT terminalshouldnot be
connected directly together when
making low level measurements.
These terminals
are
isolated from
each other
by
1
ohm which reduces
the
effects
of
commonmodevoltages.
3-31. Operating procedures for the 333A/334A with
Option01
are
the same asforthe standard instrument.
The onlydifference between the standard and optional
instrument
is
that the Option 01has
a
special
meter
and meter amplifier which
is
compensated to respond
to
W
(volume
unit)
characteristics.
3-32. MANUAL NULLING.
3-33. Since the frequency and balance controls
are
rather
sensitive in
the
MANUAL mode,
the
following
information
is
suppliedto
simplify
nulling the
333A/334A
in
the MANUAL mode. When nulling
the
333A/334A
in
the MANUAL mode, connect
the
equip-
ment
as
shown
below
and
adjust
the
333A/334A
fre-
quency
and
balance controls for the waveform shown
in
step
a
below. Additional waveforms
are
provided
to simplify nulling.
No harmonic
dietortion.
Frequency
and
balance adjustment correct.
Frequency and balance control
im-
properly adjusted.
Frequency approximately
correct;
balance
incorrect.
Balance
approximately correct;
fre-
quency incorrect.
Second harmonic predominant;
fre-
quency and balance adjusted.
Second harmonic predominant;
fre-
quency and balance
adjusted;
phase
changed.
Second harmonic predominant;
fre-
quency and balance
adjusted;
phase
changed.
Third harmonic predominant.
Balance incorrect; meter reading
&
scale.
Frequency incorrect; meter
readin(
off
scale.
3-4


Section
IV
Model
333A/334A
I
I
I
I
I
I
I
I
I
I
I
I
I
I3
If
I
I
I
I
I
I
I
I
13
jP
I
I
L
PI
I
I
I
I
I
-----
I
I
I
I
I
I
I
I
I
I
---A-
Y
Y
m
0
W
W
lb
I
I
-
.-
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
1
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
1

Model 333A/334A Section IV
SECTION
IV
THEORY
OF
OPERATION
4-1.
OVERALL DESCRIPTION.
4-2. Models 333A and 334A Distortion Analyzers in-
clude
an
impedance converter,
a
rejection amplifier,
a
metering circuit, and
a
power supply. The Model
334A also
contains
an
AM
detector. The impedance
converter provides a
low
noise input circuit with
a
highinput impedance independent
of
source impedance
placed
at
the INPUT terminals. The rejection ampli-
fier
rejects the fundamental frequency of the
input
signal
and
passes
the remaining frequency components
on to the metering circuit for measuring distortion.
The metering circuit provides visual indications of
distortion
and
voltage
levels
on the front panel meter,
M1. The
AM
detector (Model 334A only) detects the
modulating
signal
from the
RF
carrier
and
filters
all
RF
components from the modulating
signal
before
it
is
applied
to
the impedance converter circuit,
(Refer to Figure 4-1)
4-3.
BLOCK DIAGRAM DESCRIPTION.
level
applied to
the
metering circuit to
1
mV for full
scale deflection. Sensitivity
of
the metering circuit
is
increasedto 300
pV
for full scale deflection onthe
.0003
Vrange.
Themeteringcircuitprovides a visual
indication of the distortion level
of
the input signal.
In
addition totheindication provided
by
the
meter, the
OUTPUT terminals provide
a
means
of
monitoring
the distortion components.
4-8. DISTORTION MEASUREMENT
IN
AM
CARRIERS.
4-9. The
Model
334A DistortionAnalyzer contains an
AM
detector circuitformeasuringenvelope distortion
in
AM carriers. The input
signal
is appliedto the in-
pt
of
the AM detector circuit where the modulating
signal
is
recovered from
the
RF carrier. The sipal
is
then applied to the impedance converter circuit
through the one megohm attenuator and then through
the same circuits previously described in the distor-
tion measuring mode operation.
4-4. DISTORTION MEASURING OPERATION. 4-10. VOLTMETER OPERATION.
4-5. For distortion measurement, the input signalis
appliedto
the
impedance converter (Assembly
A2)
through the FUNCTION selector, S1, and the one
megohm attenuator. The one megohm attenuator
is
a
voltage
divider
network which provides 50 dB atten-
uation
in
10
dB
steps. The desiredlevelof attenuation
is
selected
by
the SENSITIVITY selector,
S2.
The
impedance
converterprovides impedance matching and
unity
gain between the INPUT terminals and the input
of the rejection amplifier.
4-6. Therejectionamplifierconsistsof
a
preamplifier,
a
Wien
bridge, andabridgeamplifier. TheSENSITIVITY
VERNER control,
at
the input
of
the preamplifier,
provides
a
set
level
signal
to
obtain
a
full scale read-
ing on the meter for any voltage level at the input
of
the
instrument.
With the FUNCTION switchintheSET
LEVEL position,
a
ground
is
applied to theWienbridge
circuit to allow
a
signal
reference level to be set up
on the meter.
With
the FUNCTION switch
in
theDIS-
TORTION position, the Wien bridge
is
used asan in-
terstage coupling network between the preamplifier
and bridge amplifier. The Wien bridge
is
then tuned
and balanced to reject the fundamental frequency
of
the
applied input
signal.
Two automatic control loops con-
sisting of
two
phase detectors, lamp drivers, lamps,
and photocells provide fine
tuning
and balance in the
AUTOMATIC MODE. The remaining frequency com-
ponents are appliedto the bridge amplifier and
are
measured
as
distortion by the metering circuit.
Nega-
tive feedback from the bridge amplifier to the preamp-
lifiernarrowstherejectionresponse of theWienbridge.
4-7. The output of the rejection amplifier
is
applied
to the metering circuit through the post-attenuator.
The post-attenuator
is
used to limit the input signal
4-11. In the voltmeter mode of operation, the input
signal
is
applied to the impedance converter circuit
through the
1:l
and
1OOO:l
attenuator. The
1:l
atten-
uation ratioisusedonthe0.0003 too.
3
VOLTSrangcs,
and the
1OOO:l
attenuationratiois used in the
1
to 300
VOLTS ranges.
With
the FUNCTION switch
in
the
VOLTMETER position, the output of
the
impedance
converter bypasses the rejection amplifier and
is
applied to the metering circuit through the
post-
attenuator (METER RANGE switch). Meteringcircuit
sensitivity
is
increased from
1
mV for full scale
deflection to
300
pV on the ,0003V range,
as
it was in
the distortion measuring operation. The function
of
the post-attenuator andmeteringcircuit
is
thesame
for
voltmeter operation as for the distortion measuring
operation.
4-12.
DETAILED CIRCUIT DESCRIPTION.
4-13. IMPEDANCE CONVERTER CIRCUIT.
;
4-14. The input
signal
is
applied to the impedance
converter circuitthroughthe
1:1
and 1000:
1
attenuator
S3R12
in voltmeter operation, and through the one
megohm attenuator S2R1through
S2R6
in distortion
operation. Capacitive dividers SC1through SC10
in
the attenuator keep
the
frequency response flat.
The impedance converter
is
a
low distortion, high
input impedance amplifier circuit with gain indepen-
dent of the source impedance placed at the INPUT
terminals.
4-15. Instrument induced distortion of the signal
being measured
is
minimized
hy
keeping the input
impedance and the gain of
the
impedance converter
4-1

Section IV Model 333AI334,
linear. The input impedance
is
kept linear
by
use of
local positive feedback from the source of A2Ql to
the gate of A2Ql and to the protective diodes A2CR2
and A2CR3. Thus signals with
a
large sourceimped-
ance can
be
measured accurately. Overall induced
distortion
is
further minimized
by
a high open loop
gain and 100%negative feedback. The high open loop
gain
is
achieved
by
local
positive feedback from the
emitter
of
A2Q3 to the collector
of
A2Q2. Overall
negative feedback from the emitter circuit
of
A2Q4
to the source of A2Ql results in unity gain from the
impedance converter.
4-16. Thebias
points
of thetransistors
in
the imped-
ance converter
are
selected to minimize instrument
induced distortion. A2Q1, an extremely low noise,
high impedance field
effect
transistor,
is
the major
component that makes linearity of the impedance con-
verter independent
of
the
signal
source impedance.
4-17. REJECTION AMPLIFIER CIRCUIT.
(Refer to
Figures
7-3
and
7-5)
4-18. The rejection amplifier circuit consists of the
preamplifier (A3Q1
thru
A3Q3), the Wien bridge
resistive leg and auto control loop (A5Q1 thru A5Q9
with associated lamp and photocell), the reactiveleg
and auto control loop (A5Q10 thru A5Q18 with associ-
ated lamp and photocell), and the bridge amplifier
(A3Q4 thru A3Q6).
4-19. PREAMPLIFIER CIRCUIT.
4-20. The signal from the impedance converter
is
applied tothe preamplifier, which
is
used during SET
LEVEL
and DISTORTION measuring operations.
Negative feedback from the junction of A3R10 and
A3Rll
is
applied to the junction of A3R2 and A3C2 to
establish the operating point for A3Q1. Negative
feedback from the emitter
of
A3Q3
is
applied to the
emitter
d
A3Q1 to stabilize the preamplifier. The
preamplifier, like the impedance converter,
is
designed for high open loop
gain
and low closed loop
gain
to minimize
instrument
induced distortion.
4-21. WIEN BRIDGE CIRCUIT.
4-22. In the distortion measuringoperationthe Wien
bridge
circuit
is
used as
a
rejection
filter
for the
fundamental frequency of the input
signal.
With the
FUNCTION selector, S1, inthe DISTORTION position,
the Wien brldqe
is
connected
as
an
interstagecoupling
network between the preamplifier circuit and the
bridge amplifier circuit. The bridge
is
tuned to the
fundamental frequency
of
the input signal by setting
the FREQUENCY RANGE selector,
S4,
for the appli-
cable frequency
range,
andtuning the capacitors C4A
through C4D. The bridge circuit
is
balanced by ad-
justing
the
COARSE balance control, R4, andtheFINE
balance control, R5.
In
the AUTOMATIC MODE fine
tuningandbalancing
are
accomplished
by
photoelectric
cells which
are
in the resistive and reactive legs of
the Wien bridge. The error signals for driving the
photocells
are
derived by detecting the bridge output
using the input signal
as
a
reference,
4-23. When the Wien bridge
is
not tuned exactly to
the frequencytobe nulled, aportion of thefundamental
4
-2
INPUT TO BRIDGE
OUTPUT OF BRIDGE
IFRESISTIVE LEG
IS
UNBALANCED
OUTPUT
OF
BRIDGE
IF
REACTIVE LEG
IS DETUNEO
Figure
4-2. Bridge Waveforms
frequency will appear
at
the bridge output. The phase
of this
signal
depends
on
which leg
of
the bridge
is
nc
tuned,
or
on the relative errors in tuning
if
neither
if
set
correctly. The magnitude of the signal
is
propor
tional to the magnitude of the tuning error
of
either
0.
both legs of the bridge.
4-24.
Figure
4-2a shows
a
sinusoidinputtothe Wien
bridge.
If
the resistive leg of the bridge
is
slightly
unbalanced, the output of the bridge
is
very small,
but has the waveform shown
in
Figure 4-2b and
is
in
phase with the input.
As
the resistive leg
is
tuned,
the signal approaches zero amplitude at null and then
becomes larger, but 180° out of phase, if the
null
position
is
passed.
Whentheresistive leg
is
correctly
tuned and the reactive leg
is
tuned through null,
a
similar waveform
is
produced, Figure 4-2c. The
only difference
is
that the reactive signal
is
90°
out
of phase with the resistive
signal.
4-25. When
the
bridge output
is
detected using the
input signal
as
the
reference, the error
signals
in
phase
or
1800 out of phase with the reference develop
a voltage that is used to vary the resistance
in
the
resistive leg
of
the bridge, to tune it to the correct
null position. Signalsof theformin Figure4-2c donot
develop any voltage,
as
the resistive detector is in-
sensitivetoinputs differingfromthereferenceby
90°.

Model
333A/334A
Section IV
(5
E IN
0
~
TYPICAL WlEN
BRIDGE
I
Figure
4-3.
Wien Bridge Circuit and Rejection Characteristics
4-26.
In
an
independent, but similar, control loop,
the bridge input
signal
is
shifted
900
and used
as
the
reference
signal
for the detector. This detector
develops control voltages to null the reactive leg of
the bridge, but
is
insensitive to signals of the form
in
Figure
4-2b,
which
are
caused by small tuning
errors
of
the
resistive branch.
4-27.
The
result
is
that the two control loops derive
information from
a
common source and develop two
independent control signalsfornulling the two legs
of
the bridge. These control voltages are used to vary
the brilliance of lamps, which
in
turn cause resistance
changes
in
photocells
that
form part of
the
Wien bridge.
4-28.
When the bridge circuit
is
tuned and balanced,
the voltage and phase
of
the fundamental, which
appears
at
junction of the
series
reactive leg
(SQR1,
3, 5,
7,
or
9 and
C4A/B)
and the shunt reactive leg
(SQRll, 13, 15, 17,
or
19,
and
C4C/D),
is
the
same
as
at
the midpoint of the resistive leg
(A3R12
and
A3R14).
When
these
two
voltages
are
equal
and
in
phase, the fundamentalfrequency will not appear
at
the drain of the field effect transistor
A3Q4.
For
frequenciesother than the fundamental, the reactive
leg of the Wien bridge offers various degrees
of
attenuation and phase shift which cause a voltage at
the output points of the bridge. This difference volt-
age
betweenthe reactive leg and resistive leg
is
amp-
lified by
A3Q4, A3Q5,
and
A3Q6.
Figure
4-3
illus-
trates a typical Wien bridge circuit and the rejection
characteristics for
it.
4-29.
The Wien bridge circuit
is
designed
to
cover
a
continuous frequency range of over a decade for each
position of the
FREQUENCY
RANGE
selector
S4.
S4
provides coarse tuning
of
the
reactive leg by changing
the bridge circuit constants in five steps at
1
decade
per
step.
For
the automatic control loop, the refer-
ence voltage
is
taken from
R6
at the input to the re-
jection amplifier and applied to
the
buffer amplifier
A5Q7.
The reference voltage
is
amplified andclipped
by
A5Q8
and
A5Q9,
and coupled to the detector
A5Q4.
The output of the metering circuit, which contains the
fundamentalfrequency
if
either leg of the bridge
is
untuned,
is
applied to the
buffer
amplifier
A5Q1.
It
is
amplified by
A5Q2
and
A5Q3
and coupled to the
detector
A5Q4.
4-30.
RefertoFigure
4-4,
partial scheiiiatic
for
detector operation. The discussion
is
applicable
to both resistive and reactive detector circuits.
4-31.
The
signals
from the error amplifier,
(A5Q2
andA5Q3)
will
be
equal and of opposite phase, and
will
cancel out each other when
the
detector,
A5Q4,
is
off. However, when the positive half of the referencc
square wave gates
A5Q4
on,
the
signal from thc coll-
ector of
A5Q3
will
be shorted to ground. Thus the
signal from the collector
of
A5Q2
will
be
couplcd
through the filter network to the base of
A5Q5.
If
thc
signal from
A5Q2
is
in phase with the referencc, thc
positive half of the signal
will
be
passcd, and
if
it
is
out of phase, the negative half will
be
passed.
4-32.
The normal working voltage at
A5TP3
is
be-
tween0and
-1
volt. Thedcoutput of thefilter nctwork
causes the voltage
at
A5TP3
to go in
a
positive dircc-
tion (toward zero) for in phase error signals, and
in
a
negative direction (toward
-1
V)
for
out of phase
error
signals. The change in base voltage
is
then ampIificd
by
A5Q5
and lamp driver
A5Q6.
This
will
change thc
brilliance of lamp
A6DS1,
which
will
vary
theresist-
anceof
A6V1
inthe directionnecessaryto balance the
resistive leg of the bridge.
I
I
I I
I-1
I
Figure
4-4.
Auto
Control
Loop
Detector
4-3

Section
IV
Model
333A/334
4-33.
Refer
to
Figure
4-5
for the phase relationship
of
the
bridge
error
signal
andreferencevoltage
at
the
base
of
A5Q4.
The
shaded
portions of
the
error
sig-
nals
(bandc) indicate that
part
of the error
signal
which contributesto
the
dc
lamp control voltage.
As
indicatedind, any error
signal
that
ie
90°
outofphase
with
the
reference
does
not affect
the
dc lamp control
voltage, because
equal
amounts
of
the
positive and
negative
portions
are
passed.
4-34.
The operation
d
the reactive branch control
loop
ie
similar to
that
of
the
resistive branch. The
phase
delay
circuit,
A5Q15, A5Q16, S4AF
and
S4C1
through
S4C5,
shifts
the
reference voltagc
90°,
aa
shown
in
Figure
4-5f.
This makes the detector
A5Ql2
sensitive to components
of
the bridge error
signal
that are
900
out
of
phase
(g
and
h).
The outpui
of
the
lamp driver,
A5Q14,
controls the brilliance
of
A6DS2,
which varies the resietance
of
A6V2
through
A6V5
to
tune
the
branches
of
the
reactive leg. Deck
AR
of
the
FREQUENCY RANGE switch,
54,
switches
A5R56
in
parallel
with
A5R55
on
the
top
three
fre-
quency
ranges.
A6DS2
will
becomebrighter, andlower
the resistance
of
A6V2
through
A6V5,
makingvariatio
45'
135'
225'
315'
a
Ref. signalforResistive
Detector
(A564)
b
Resistive
Error
signal
in
phase
with
reference
at
bridge
Resptive
Error
signal
180
out
of
phase
d
Reactive
Error
signal
30
out
of
phase
(Results
in
no
output
from
Re-
sistive Detector)
e
Error
signal,
reactive
leg detuned, resistive
.eg unbalanced
f
Ref.
signal
for Reactive
Detector
(A5612
b
Reactive
Error
signal
in
phase with reference
at
bridge
Reactive
Error
out
of
phase
h
signal
i
Reptive
Error
signal
90
out of phase(Results
in
no output from
Re-
active Detector)
Error
signal,
Reactive
leg detuned, Resistive
leg unbalanced
45'
13s'
225'
316'
Figure
4-5.
Reference and
Error
Phase Relationship
4-4
Other manuals for 333A
3
This manual suits for next models
1
Table of contents
Other HP Measuring Instrument manuals