HP 8719D Installation and operation manual

Installation Note
HP
8719DU/872ODU/8722DU
Option 000 Performance Upgrade Kit
for the HP
8719D/872OD/8722D
Network Analyzers
IMPORTANT! Be sure to read the enclosed section, “Installation
Options” before starting.
HP Part Number 08720-90372
Printed in USA April 1998

Notice.
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this material.
@
Copyright Hewlett-Packard Company 1998
All Rights Reserved. Reproduction, adaptation, or translation without prior written permission
is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA

P
RODUCT
A
FFECTED
:
HP
8719D/8720D/8722D
Network Analyzer
S
ERIAL
N
UMBERS
:
To B
E
P
ERFORMED
B
Y
:
E
STIMATED
I
NSTALLATION
T
IME
:
All serials
[X] HP Service Center
[X] HP personnel on-site
[X] Customer
See
“Installation
Options”
below
1 hour (add the time needed to back up
instrument states, if desired.)
E
STIMATED
V
ERIFICATION
T
IME
:
30 minutes (not included in the kit price)
Description
The HP 8720DU Option 000 Performance Upgrade Kit is used to replace the A7 CPU board in
the analyzer with a new board designed with a faster microprocessor and enhanced firmware.
Faster measurement speeds, data transfers, and instrument state recalls are made possible
with this upgrade. Loading new firmware revisions can now be accomplished with a 3.5 inch
floppy disk rather than by replacing
EEPROMs.
This note contains the information required to
upgrade the A7 CPU board in the HP
8719DU/8720DU/8722DU
network analyzer.
Installation Options
1.
2.
3.
Installation at HP Service Center
included in the kit price
call l-800-403-0801 to schedule with a service center
send the analyzer and upgrade kit to the designated service center
does not include repair
-
requires an operational analyzer
Installation by HP personnel at customer’s site
n
travel time and labor (including backup of instrument states, if desired) in excess of the 1
hour estimated installation time to be paid by customer.
Installation by customer
Installation Note 08720-90372

Warranty
This Hewlett-Packard instrument product is warranted against defects in material and
workmanship for a period of 90 days from date of shipment. During the warranty period,
Hewlett-Packard Company will, at its option, either repair or replace products which prove to
be defective.
For warranty service or repair, this product must be returned to a service facility designated by
Hewlett-Packard. Buyer shall prepay shipping charges to Hewlett-Packard and Hewlett-Packard
shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all
shipping charges, duties, and taxes for products returned to Hewlett-Packard from another
country.
Hewlett-Packard warrants that its software and firmware designated by Hewlett-Packard for
use with an instrument will execute its programming instructions when properly installed on
that instrument. Hewlett-Packard does not warrant that the operation of the instrument, or
software, or firmware will be uninterrupted or error-free.
L
IMITATION OF
W
ARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or
misuse, operation outside of the environmental specifications for the product, or improper
site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY
DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
E
XCLUSIVE
R
EMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT,
OR ANY OTHER LEGAL THEORY.
4
Installation Note 08720-90372

Performance Upgrade Kit: Parts List
1
CPU Repair kit: 08720-69253
Required Equipment and Tools
Required Equipment and Tools
Item
IIP
Fart/Model Number
3.5~inch
floppy disk HP 92192A (box of 10)
T-10 TORX screwdriver
N/A
T- 15 TORX screwdriver
N/A
Antistatic Wrist Strap
9300-1367
Antistatic Wrist Strap Cord
9300-0980
Static-Control Table Mat and Earth Ground Wire
9300-0797
Safety Considerations
Warning
Before you disassemble the instrument, turn the power switch OFF and
unplug the instrument. Failure to unplug the instrument can result in
personal injury.
Caution
Electrostatic discharge (ESD) can damage or destroy electronic components. All
work on electronic assemblies should be performed at a static-safe workstation.
Refer to the documentation that pertains to your instrument for information
about static-safe workstations and ordering static-safe accessories.
Installation
Note 08720.90372
5

Record the Analyzer Information
1. Turn on the analyzer.
2. Record the firmware revision, serial number, and option numbers which appear on the
display.
Perform the Operator’s Check
1.
2.
3.
4.
5.
6.
Press
(i%ZQ
(System)
SERVICE
%NU
TESTS
121)
a
EXECUTE TEST
.
Follow the displayed prompts for making the necessary connections.
Press CONTINUE
.
When the analyzer passes the test, it will display: Op Ck Port
1
DONE.
Press RETURN TESTS 122)
Ixl)
EXECUTE TEST
.
Follow the displayed prompts for making the necessary connections.
Press CONTINUE
.
When the analyzer passes the test, it will display: Op Ck Port 2 DONE.
Note
If either check did not pass, the instrument must be adjusted or repaired before
being upgraded. Repair costs are
not
included in the cost of the upgrade
installation.
Back Up the Correction Constants
The correction constants, which are unique to the instrument, are stored in EEPROM on the A7
CPU board. By creating an EEPROM backup disk, you will have a copy of all the correction
constants data that you need to replace the A7 CPU board.
7. Insert a 3.5-inch disk into the analyzer disk drive.
8. If the disk is not formatted, follow these steps:
a.
Press
(-1
FILE UTILITIES FORMAT DISK.
b. Select the format type:
n
to format a LIF disk, select FORMAT:LIF
.
’
to format a DOS disk, select
FORMAT:DOS
.
c.
Press FORMAT INT DISK and answer YES at the query.
g.
Press
[s)
SERVICE MENU SERVICE MODES MORE STORE EEPR ON [Save/Recall)
SELECT DISK INTERNAL DISK RETURN SAVE STATE.
6
Installation Note 08720-90372

Note
The analyzer creates a default file: “FILEnn”(nn= 00 or the next highest
available number). The filename appears in the upper-left corner of the display.
The file type:
“ISTATE(E
indicates that the file is an instrument-state with
EEPROM backup.
10.
11.
Press FILE
UTI$?%IES
RENtikE
‘FsL&
ERASE,
T?‘I@;.
Use the front panel knob and the
,
.1^,
.,ir.**a*lr,x
>
..,,
,
.+.
“**,
,
rx*x
>
,a”_,,,
SE$EC$~~J~I$‘%$<,
softkey
(or an external keyboard) to rename the file “FILEO” to
“N12345”
where
I2345
represents the last 5 digits of the instrument’s serial number. (The first
character in the filename must be a letter.) When you are finished renaming the file, press
DoYE,.
Write the following information on the disk label:
n
analyzer serial number
H
today’s date
n
“EEPROM Backup Disk”
Back Up the Instrument States
All instrument states that you have saved to the analyzer’s internal memory will be lost when
the A7 CPU board is replaced. If you want to save these instrument states, you must back
them up on a disk before replacing the board. For an explanation of this procedure, refer
to “Printing, Plotting, and Saving Measurement Results,” located in the User’s
Gbide.
This
procedure may be very time consuming, depending on the number of instrument states that
you choose to back up. Each instrument state file will have to be renamed.
Remove the Rear Panel Assembly
Note
Although not mandatory, it is recommended that the rear panel assembly be
removed before replacing the A7 CPU board. It is easier to work with the
ribbon cable connectors with the rear panel assembly removed.
Refer to Figure l-l.
1.
2.
3.
4.
Disconnect the power cord.
5.
Remove the four rear standoffs (item 2).
Loosen the top cover screw and slide the top cover off.
Remove the instrument bottom cover and record the keyword label(s) that are located
on the exposed sheet metal next to the A7 CPU assembly. Note that each keyword is for
EACH option installed in the instrument.
If the analyzer has option lD5, remove the BNC jumper from the high stability frequency
reference (item 3).
6.
If the analyzer has option 085, remove the RF cable (item 4) and the connectors’
attachiw
hardware (item 5).
7.
Remove the hardware (item 6) that attaches the RS-232 connector to the rear panel.
Installation Note 09720.90372

8. Remove the four screws (item 7) that attach the interface bracket to the rear panel.
9. Remove the six screws (item
8)
and (item
9),
that attach the preregulator to the rear panel.
10. Remove the eight screws (item 10) from the rear frame: four from the top edge and four
from the bottom edge.
11. Remove the screw from the pc board stabilizer and remove the stabilizer.
12. Lift the reference board (A12) from its motherboard connector and disconnect the flexible
RF cable (item 11).
13. Pull the rear panel away from the frame. Disconnect the ribbon cable (item 12) from the
motherboard connector, pressing down and out on the connector locks. Disconnect the
wiring harness (item 13) from the motherboard.
8
Figure
l-l.
Installation Note
08720-9OBpP

Remove the A7 CPU Board
Refer to Figure l-2.
1.
Turn the analyzer over and remove the screw (item 2) that secures the CPU board to the
deck.
2.
Disconnect the ribbon cable (item
3),
sliding your finger nail between the cable and the
connector.
3.
4.
Disconnect the ribbon cable (item 4) from the CPU board.
Slide the board towards the front of the instrument so that it disconnects from the three
standoffs (item 5).
5.
Disconnect the ribbon attached at the rear of the CPU board.
6.
Lift the board off of the standoffs and place it in an antistatic bag.
Install
the New A7 CPU Board
1. Reverse the order of the removal procedure.
2. On the CPU board, set the A7 Switch to the Alter Mode configuration (see Figure l-3).
installation Note
08720-80372

.
n-
-.--.n”-“n
10
sb6108d
Figure
l-2.
Installation Note 08720-90372

I
CPU Assembly
dg62d
Normal Mode Alter Mode
Figure 1-3.
Rocker
Slide
Load the New Firmware
1.
2.
3.
Place the firmware disk into the disk drive.
Turn the instrument on. The firmware will be loaded automatically during power-on. The
front panel
LEDs
should step through a sequence for approximately one minute as firmware
is loaded. The display will be blank during this time.
At the end of a successful loading, the
LEDs
for Channel 1 and Testport 1 will remain on
and the display will turn on indicating the version of firmware that was loaded.
The display may indicate No IF or No Phase Lock. This is normal for this procedure.
Remove the firmware disk from the disk drive.
In Case of Difficulty
If the firmware did not load successfully, LED patterns on the front panel can help you isolate
the problem.
n
If the following LED pattern is present, an acceptable firmware filename was not found
on the disk. (The desired format for firmware filenames is 8720D_07.
-02.)
Check that the
firmware disk used was for the
HP
8719D/20D/22D.
Installation Note 08720-90372
11

n
If any of the following LED patterns are present, the firmware disk may be defective.
LED
Pattern
XlCH2
RL Ts
.
.
.
.
.
.
.
.
a
. .
.
.
.
. .
.
.
.
.
.
.
.
.
.
.
.
.
.
n
If any other LED pattern is present, the CPU board is defective.
Note
If firmware did not load, a red LED on the CPU board will be flashing.
n
If the following LED pattern is present on the CPU board, suspect the disk drive or
associated cabling:
.
.
.
0
.
.
0
cl
(front of instrument
lJ)
Store the Analyzer Serial Number
This procedure stores the analyzer serial number in the A7 CPU assembly
EEPROMs.
1.
2.
3.
Press @GZ) DISPLAY MORE TITLE ERASE TITLE to erase the HP logo from the display.
Enter the analyzer serial number (recorded earlier) by using an external keyboard, or by
rotating the front panel knob to position the arrow below each character of the instrument
serial number, and then pressing SELECT LETTER to enter each character. Enter a total of
ten characters.
Press BACKSPACE if you made a mistake.
Press DONE when you have finished entering the title.
Caution
You CAA?.OT
correct mistakes after you perform step 4, unless you contact
the factory for a keyword to clear the serial number. If needed, this
keyword should be used in the next procedure, “Store the Analyzer Option(s)
Information,” for clearing the incorrect serial number. After this is done,
repeat the procedure “Store the Analyzer Serial Number.”
4. Press
System_]
SERVICE MENU TESTS
(Z’J
@
EXECUTE TEST YES
o
5. Observe the analyzer for the results of the routine:
12
Installation Note 88720-90372

If the analyzer displays the message Serial Cor DONE, you have completed this
procedure.
If the analyzer does not display DONE, then either the serial number that you entered in
steps 3 and 4 did not match the required format, or a serial number was already stored.
Check the serial number recognized by the analyzer:
b.
Look for the serial number displayed on the analyzer screen.
c. Rerun this adjustment test.
If the analyzer continues to fail this adjustment routine, contact your nearest HP Sales and
Service Office, listed at
the end of this document.
Store the Analyzer Option(s) Information
This procedure stores instrument option(s) information in CPU assembly
EEPROMs.
You can
also use this procedure to remove a serial number by using the factory supplied keyword, as
referred to in “Store the Analyzer Serial Number.”
1.
2.
3.
Press
IPreset)
IDisplay)
MORE TITLE ERASE TITLE
.
Enter an option keyword (recorded earlier in the section “Remove the Analyzer Rear Panel”)
by using an external keyboard, or by rotating the front panel knob to position the arrow
below each character of the keyword, and then pressing SELECT LETTER to enter each
letter.
Press BACKSPACE if you made a mistake.
Press DONE when you have finished entering the title.
Caution
Do not
confuse
“I”
with
“1”
or “0” with
“0”
(zero).
4. Press
(s?)
SERVICE MENU TESTS
150)
a
EXECUTE TEST YES.
5.
Observe the analyzer for the results of the adjustment routine:
w
If the analyzer displays Option Cor DONE, you have completed this procedure.
n
If the analyzer has more than one option, repeat steps 2-5 to enter the remaining
keyword(s) for options.
n
If the analyzer displays Option Cor FAIL, check the keyword used in step
3
and make
sure it is correct. Pay special attention to the letters “I” or “O”, the numbers
“1”
or “0”
(zero). Repeat this entire adjustment test.
n
If the analyzer continues to fail the adjustment routine, contact
your nearest
HP Sales and
Service Office, listed at the end of this document.
Installation Note 08720.90372
13

Re-load the Correction Constants
1.
2.
3.
4.
Insert the “EEPROM Backup Disk” (created earlier) into the disk drive.
Press
(-1
‘Sl$&?T?%SK’~
;TNT%KEDISF.
Use the front panel knob to highlight the
&
a
*a
,“,,
*A,
x.~~*-d/.
*.;
,,A_,.~
,es:*-~~
file
“N12345”
where N12345 represents the file name of the EEPROM data for the analyzer.
Press
Rl?I’URN
RECALL
‘STiTK
to down load the correction constants data into the
*i
=,,,A_
,,,
,,
instrument
EEPROMs.
Press
IPreset)
to load the EEPROM Constants. Verify that good data was transferred to
EEPROM by performing a simple measurement.
Re-load the Instrument States
If you want to re-load the analyzer’s instrument states from your backup disk (created earlier),
refer to “Printing, Plotting, and Saving Measurement Results,” located in the User’s Guide. This
procedure may be very time consuming, depending on the number of instrument states that
you choose to re-load. Each instrument state file will have to be renamed.
Configure the Analyzer to Produce a Time Stamp
You
can set a clock, and then activate it, if you want the time and date to appear on your
hardcopies.
1.
Press (System) SET CLOCK
.
2.
Press
3.
Press
4.
Press
5.
Press
6.
Press
SET YEAR and enter the current year (four digits), followed by
Ixl).
SET
MONTH
and enter the current month of the year, followed (xl.
SET DAY and enter the current day of the month, followed by
Ixl).
SET HOUR
and enter the current hour of the day (O-23), followed by
a.
SET MINUTES
and enter the next immediate minute, followed by
1x11.
7.
Press ROUND SECONDS when the current time is exactly as you have set it.
*.
Press TIME STAMP until TIME STAMP’ ON appears on the
softkey
label.
14
Installation Note
09720-90372

Verify Successful Installation
1. Turn the analyzer off and then on.
2. Verify the following information on the display:
a. The new firmware revision number matches the number printed on the firmware disk.
b. All other options have been re-established.
3. Repeat the section titled “Perform the Operator’s Check,”located earlier in this document,
4. Turn the analyzer off.
5. Move the A7 switch back to its NORMAL position. Refer to Figure l-3.
6. Reinstall the rear panel assembly by reversing the order of the removal procedure.
Troubleshooting
1. If the display information is not observed, as listed previously in the section titled “Verify
Successful Installation,” examine all connections to the CPU board. Improper connections
may be diagnosed as follows:
a. If the Channel 1 and Testport 1
LEDs
are on but the display is blank, examine the item 4
ribbon cable connector. Refer to Figure l-2.
b. If the Testport 2 LED is on but the display is blank, examine the item 3 ribbon cable
connector.
Installation Note 09720.90372
15

Firmware History
Firmware Revision 7.12 (includes all features of revisions 6.14 and 6.12)
Firmware Enhancements and Features
1.
2.
3.
A new IF bandwidth filter setting of 6000 Hz can now be selected in addition to the
previously available settings of 10, 30, 100, 300, 1000, 3000, and 3700 Hz.
Swept list mode, a new list frequency sweep mode, takes data while sweeping through the
defined frequency points, increasing throughput by up to 6 times over a stepped sweep. In
addition, this mode allows the test port power and IF bandwidth to be set independently for
each segment that is defined.
New HP-IB commands:
HP-IB Cmd Description
LISIFBW<ONIOFF> Enables/disables the IFBW setting for a list-frequency table in swept
list mode.
LISPWRM<ONIOFF> Enables/disables the power setting for a list-frequency table in
swept list mode.
LIS’FIYFE<LSTF~LSWP> Selects either the stepped list mode or the swept list mode to use
with a list-frequency table.
SEGIFBW[D]
Sets the IFBW for the active segment of a list-frequency table in
swept list mode.
SEGPOWER[D]
Sets the power for the active segment of a list-frequency table in
swept list mode.
Firmware Revision 6.14 (includes all features of revision 6.12)
Firmware Revision 6.14 is the revised code for both the HP 8753D (including Option 011) and
the HP
8719D/8720D/8722D
Network Analyzers.
Firmware Enhancements and Features
1.
2.
3.
4.
5.
6.
7.
8.
A major improvement has been made in disk writes for the
CITIfile
and
S2P
(Touchstone)
ASCII formats.
Error numbers (error table) are now synchronized with all revisions.
Data files with
S2P
now have suffixes dependent on the channel number (Sl or S2).
In operating parameters pages, the specific model number has been replaced with the word
ANALYZER.
Calibration kit labels under the SELECT CAL KIT menu now include the HP calibration kit
most commonly associated with that connector geometry.
Single parameter, linear magnitude marker display format now shows in engineering units
(autoscales to U,
mu,
pU,
etc.). HP 8753D (including Option 011) only.
Changed the list table default frequency from MHz to
GHz.
HP
8719D/8720D/8722D
only.
Updated the 7mm calibration constants based on the HP 85050 series calibration kits.
HP
8719D/8720D/8722D
only.
16
Installation Note 08720-90372

9. The DEFINE STANDARD menu for the selection of
c$bration,standard
“type”
has been
changed.
‘RI
access this menu, press
a
CAL
KiT
MODIFY
DEFINE STANDARD
.
The
choices of open, short, load, delay/thru, and arbitrary impedance now work the same as
the choices in other “one-of-n” menus. (For a “one-of-n” menu example, press the
(FORMAT)
key.) The DEFINE STANDARD menu remains displayed after pressing
UPEti
,
or
SHURT,
,
or
whatever “type” the user has selected, thus showing the latest selection for a particular
calibration standard number. Further, a new message box appears above the active entry for
calibration standard number. The message reads, “Select
#
with arrow keys, then press
MODIFY STD”. Pressing the new
softkey
MOI$FY
S’I’U
DEFiNI’I?ON
then brings up the menu
appropriate for the “type.” For example, if an open was selected, the CO
,
Cl’, etc. menu
appears. If a load was selected, the FIXED
,
SLIDIIVC
,
etc. menu appears.
These changes help the front panel user only. HP-IB operation remains unchanged.
F’irmware
Problems Fixed
1. Fixed the problem experienced with downloading cal coefficients using INPUCALC when the
number of coefficients in the current I-state does not match the download number of points.
2. Fixed the slow HP-IB access to the internal disk.
3. Fixed the inadvertent test set switching by programming the test port on each sweep.
HP 8753D (including Option 011) only.
4. Fixed the sampler correction phase error experienced when measuring B/R with an IF
bandwidth setting of 3700 Hz. HP 8753D (including Option 011) only.
5. Option 011, HP 85047A Test Set: fixed the unwanted activation of the doubler during
attenuator changes. HP 8753D (including Option 011) only.
6. Fixed the POSSIBLE FALSE LOCK error message from appearing at incorrect times.
HP
8719D/8720D/8722D
only.
7. Option 400: fixed the loss of phase lock experienced under the following
conditions-
during the first sweep after test set switching, with power meter calibration on one port.
HP
8719D/8720D/8722D
only.
8. Fixed R channel drop-out when sweeping under the following conditions- CF= 2.55
GHz,
Span= 100 MHz. HP
8719D/8720D/8722D
only.
Firmware Revision 6.12
Firmware Enhancements and Features
Firmware Revision 6.12 merges the codes of the HP 8753D (including Option 011) and the
HP
8719D/8720D/8722D
Network Analyzers.
Carry-Over From the HP 8719D/8720D/8722D Network Analyzer. The following
enhancements and features are a carry-over from the HP
8719D/8720D/8722D
but are
new in firmware revision
6.12 to the
EIP
8753D (including Option 011):
1. There are now 5 markers per channel (previously 4 per channel). The MARKER FCTN menu
has been revamped accordingly.
2.
Calibration features, such as the
softkey
DO BOTH FWD
&
REV in Full 2-Port or TRL
calibration, are improved and more user friendly. It is easier to modify or create a
TRL
cal
kit because separate TRL class entries are now allowed. HP-IB commands are like
thme for
the HP 8510 Network Analyzer.
Installation Note 08720-90372
17

3. Minor improvements have been made to the user interface for disk operations. The file last
selected on the disk directory page is remembered when the user selects SAVE/RECALL. If
the user tries to save a file to the internal disk drive when the disk has been removed, the
internal disk drive remains the selected drive.
4. Waveguide delay can now be selected by pressing
[SCALE)
ELECTRICAL DELAY
.
5.
The dump-graphics feature, previously added as a carry-over from the
872OC,
can be
<,,,l,.,X
:
,I
,,,
I,,
,.
selected by pressing
I-J
,~ERVJCE‘MENU:
,TEST^
.OPT$NS
DUMP
GRAPH
.
This feature can
L
*I(__
i
be toggled between on and off, which enables/disables the dumping of the graphical output
during the System Verification service tests (SYS VER TESTS).
6. The SELL (set learn string revision) command has been updated to work with post 5.00
revisions (HP 8753D including Option 011). The current HP 875x can now generate and
receive instrument learn strings from all the previous shipment releases. Disk compatibility
between the various HP 8753 revisions and the current revision is now supported.
7. The following are
the
choices from the new CONFIGURE menu, selected by pressing
Cm)
CONFIGURE
:
’
TESTSET
SW Toggles
testset
switching between continuous and hold.
. RAW OFFSET Toggles attenuator and sampler cal offsets on or off.
. SPUR AVOID Toggles spur avoidance on or off. HP 8753D (including Option 011)
only.
New Enhancements
and
Features.
The following enhancements and features are new in firmware revision 6.12:
1. The capability to store S-parameter data to component data files using the “Touchstone”
format
(S2P)
has been added.
2. A new IF bandwidth filter setting of 3700 Hz has been added.
3. The individual power ranges will now allow 3
dl3
over and under the values shown with
each power range selection.
4. Enhancements have been made to the file naming capability for files generated by the
instrument during execution of a sequence. The new menu can be found by pressing
(
SAVE
/
RECALL
_)
FILE UTILITIES SEQUENCE FILENAMING
.
The TITLE FILE menu also gained
a new
softkey
to support the inclusion of the sequence loop counter in the filenames.
5. A method of external calibration,
‘Ihke
4 mode, provides a remote-only command that
directs the network analyzer to take a single “group” or sweep (consisting of a forward
and reverse sweep) and to collect the raw data for all 4 S-parameters. The user can then
extract the raw data for the S-parameters (or the pre-raw data, which has not had sampler
correction nor attenuator offsets applied) and perform the error correction in an external
computer (using previously extracted calibration arrays).
6. A new feature, Adapter Removal, is designed for cases where the device under test is
“non-insertable” and thus an adapter is required. This feature requires two full 2-port
calibrations, one for each port, and subsequently provides a calibration for the test device
with the effects of the adapter removed. The menu is found by pressing
m
MORE
ADAPTER REMOVAL.
18
Installation Note 08720.90372

7. New HP-IB commands:
HP-IB Cmd
IFBW3700
SWPSTART
RAWOFFS<ONIOFF>
TAKE4<0NIOFF>
SMS<ONIOFF>
OUTPPREX
TITP
TITFO
ADAPl[D]
ADPTCOAX
ADP’IWAVE
MODS
CALSPORTl
CALSPORT
Description
Set IF Bandwidth to 3700 Hz.
Sweep start (Zrke4).
Raw Offsets
(Thke4).
Take4 mode On/off.
Spur Avoidance On/off. HP 8753D (including Option 011) only.
Output “pre-raw” arrays,
x=
l-4
(l%ke4).
Title Plot to Disk file.
Title the save state filename, only in sequence mode.
Set adapter electrical delay.
Set adapter to
COAXial.
Set adapter to
WAVEguide.
Compute new cal set using adapter removal.
Recall cal set associated with Port 1 for adapter removal.
Recall cal set associated with Port 2 for adapter removal.
8. Additional Limit Line and Data Point Functions:
These HPIB-only functions were previously available only with option K96, but have now
been added to the standard firmware set. Example programs using these functions are
available in the Programmer’s Guides with a print date of July 1997 or later.
BP-IB Cmd
MINMAX
OUTPAMAX
OUTPAMIN
OUTPSEGAM
OUTPSEGM[n]
SELSEG[n]
SELMAXPT
SELMINPT
SELPT[n]
OUTPDATP
Installation Note 08720.90372
Description
Enable/disable min/max recording per segment. Min and max values
are recorded per limit segment.
Output max values for all limit line segments.
Output min values for all limit line segments.
Output limit test min/max all segments. Output the segment number,
max stimulus, max value, min stimulus, min value for all active
segments.
Output limit test
min/max
for a specified segment.
Select segment number for the OUTPSEGF and OUTPSEGM
commands to report on.
Select the last point number in the range of points that the
OUTPDATR command will report.
Select the first point number in the range of points that the
OUTPDATR command will report.
Select point number that OUTPDATP will report.
Output trace data indexed by point (see SELPT).
19

OUTPDATR
OUTPLIMl
OUTPLIMZ
SELSEG[n]
OUTPSEGAF
OUTPSEGF
OUTPFAIP
Output trace data for range of points (see SELMINPT, SELMAXPT).
Output status of limit test for channel 1.
Output status of limit test for channel 2.
Select segment number for the OUTPSEGF and OUTPSEGM
commands to report on.
Output the segment number and its status for all active segments.
Output limit test status per segment.
Output Fail Point: this command is similar to OUTPLIMF except that
it reports the number of failures first, followed by the stimulus and
trace values for each failed point in the test (note: use command
LIMITEST<ON>
to function properly).
20
Installation Note 08720-90372
This manual suits for next models
2
Table of contents
Other HP Measuring Instrument manuals