on the LED matches the set number, the buzzer will start to beep, indicating that the time
is up. At this time, the “Timer” switch is turned off to turn off the buzzer.
3.The pour point is tested in accordance with GB/T3535-2006 "Peak Point
Determination Method for Petroleum Products". For detailed methods, see
GB/T3535-2006 "Pour Point Determination Method for Petroleum Products".
4. If the actual value detected by the glass thermometer does not match the
displayed value of the temperature control table, correction is required. For example, if
the meter display value is -30.0 °C and the glass thermometer detection value is -29.7 °C,
press the " " button for more than 4 seconds, the PV window displays rE, and then
press the " " button. The PV window displays rt to make the SP value. It is 0.3 (if the
glass thermometer has a detection value of -30.3 °C, the SP value is -0.3). After the
correction is completed, press the " " button for 4 seconds or more to exit (the key is
automatically returned to the standard mode after 1 minute without pressing the button.
(See the specification of the temperature control table for details).
Warning: when the instrument fails, the power should be cut off immediately. After
the cold groove of the instrument is restored to room temperature, please ask
professional personnel to repair and remove the fault before continuing to use, to prevent
accidents!V.Common Troubleshooting
Power switch
does not work
(1) Fuse blown
(2) The power cord is loose
or the plug is not in good
contact.
(1) Replace the fuse
(2) Exclude after checking the power
supply and plug
Temperature
display is
confusing or
not displayed
(1) Sensor plug contact is
not good
(2) Sensor damage
(3) Internal circuit failure
(1) Re-inserted
(2) change sensor
(3) Ask the professional to repair
Cooling rate is
slow or not
cooling
(1) too much dust in the
condenser
(2) Refrigerant leakage or
circuit failure
(1) Open the front door and side
door of the refrigerator to remove
dust
(2) Ask the professional to repair
VI.Matters Needing Attention
1.The instrument should be placed in a ventilated and dry place and kept clean.
2.The instrument power supply must be well grounded.
3.The tilt of the instrument should not be greater than 45° during handling and
installation.
4.Do not dry heat heater without medium.
5.If you need to start again after turning off the refrigerator, you must do so at an
interval of 5 minutes, otherwise it will damage the compressor.