human care Diana II Comfort 4MSB User manual

Service manual / Servicemanual / Servicehandbuch /
Servicehandleiding
Diana II Comfort 4MSB 90559

2www.humancaregroup.com
InnehållContent
Inhalt Inhoud
SVENSKA
DEUTSCH NEDERLANDS
ENGLISH
Service Manual 3
Compliances and Standards 4
Useful informaon 4
Service manual descripon 5
Tools & Maintenance 5
Troubleshoong & Solving 7
Spare parts list 14
Servicemanual 15
Eerlevnad och standarder 16
Användbar informaon 16
Beskrivning av servicemanual 17
Verktyg och underhåll 17
Felsökning och lösning 19
Reservdelslista 26
Servicehandbuch 27
Konformität und Standards 28
Nützliche Informaonen 28
Beschreibung des Servicehandbuchs 29
Werkzeuge & Wartung 29
Fehlersuche und -behebung 31
Ersatzteilliste 38
Servicehandleiding 39
In overeenstemming met en normen 40
Nuge informae 40
Beschrijving van de servicehandleiding 41
Gereedschappen & Onderhoud 41
Probleemoplossing 43
Reserveonderdelenlijst 50

www.humancaregroup.com 3
ENGLISH
Symbols
WARNING symbol indicates a poten-
ally hazardous situaon which, if not
avoided, could result in serious injury
to the user or damage to property
and/or the device itself.
INFORMATION symbol indicates
recommendaons and informaon for
proper, trouble-free usage.
WEIGHT CAPACITY symbol indicates
maximum user weight for the product.
Do not exceed this weight in any
circumstance.
CE MARKING symbol indicates product
meets requirements of EU Direcve
93/42/EEC (MDD) and/or EU
Regulaon 2017/745 (MDR).
RECYCLE according to local regulaons.
MEDICAL DEVICE symbol indicates
that the product is a medical device
according to the denion of
medical device in EU Direcve 93/42/
EEC (MDD) and/or EU Regulaon
2017/745 (MDR).
READ INSTRUCTIONS and ensure you
fully understand them before using
this product.
Important!
You must read the user manual for your device prior to usage. Keep this
booklet and informaon handy for future use.
Product modicaon is not permied.
Before use, always check that the product
is not damaged or worn.
Human Care’s products are connuously
being developed and updated to provide our
customers with the highest quality. We reserve
the right to make design changes without prior
noce.
Always make sure that you have the most
recent version of the manual which is available
for downloading from our website at
www.humancaregroup.com.
Contact your local distributor if you have any
quesons about the product and its use.
Service Manual

4www.humancaregroup.com
Diana II Comfort 4MSB Service Manual
Human Care is an ISO 13485:2016 cered
Swedish medical device company. The Quality
Management System is in compliance with US
21 CFR part 820.
The product is CE marked in accordance with EU
Direcve 93/42/EEC (MDD) and/or EU Regula-
on 2017/745 (MDR), as class I medical device.
The CE mark is on the product.
This sling has been tested in accordance with
ISO 10535.
Compliances and Standards
Useful informaon
a) Service manual
Service manual was created for technicians
involved in servicing of Diana II Comfort 4MSB
mobile li. Safe performance of mobile li ser-
vice procedures requires experience, technical
skills and general knowledge of safety rules.
This manual uses warnings and cauons to pay
special aenon to service procedures.
Warning – this informaon tells us
about the possibility of safety risk of
technician and user.
Cauon – this informaon tells us about
acons that can cause potenal damage
of the li.
b) Diana II Comfort 4MSB installaon and
operang
For installaon and operaon of the li, see the
user manual.
c) Service of Diana II Comfort 4MSB
The service can be only made by Human Care’s
authorized service team or authorized units by
Human Care.
Human Care shall bear no responsibility for the
consequences of repairs or modicaons that
have been done by unauthorized personnel and
for the use of non-original replacement parts
dedicated by Human Care for mobile lis.
d) Service documentaon
Service documentaon and service repair do-
cument are available on a website accessible for
authorized service.
e) Safety precauons
The mobile li design provides maximum of sa-
fety. Safety of the device are ensured by safety
standards IEC60601-1.
Read these cauons carefully before
start servicing!
Note - do not perform service repairs
and funconal tesng when the paent
is on the device.
f) Addional informaon
In case of any problems with mobile li, contact
the intermediary or manufacturer’s service.
Service contact details:
HUMAN CARE SWEDEN (HQ)
Årstaängsvägen 21B
117 43 Stockholm
Phone: +46 8 665 35 00
Fax: +46 8 665 35 10
Email: info.se@humancaregroup.com

www.humancaregroup.com 5
ENGLISH
Service manual descripon
Tools & Maintenance
The service manual contains informaon about
repairing, maintaining and idencaon of
mobile li. The li can be idened by using
the idencaon code which is placed on the
name plate.
Note – if the plug from baery charger
does not match a socket in your country,
you can use an adapter.
Tools
Standard set of tools is required for servicing of
the mobile li:
Allen key set Ordinary set of various keys
Flat screwdriver
Hammer Volt meter

6www.humancaregroup.com
Diana II Comfort 4MSB Service Manual
Device maintaining
All movable joints must be lubricated every six
months or when experiencing squeaks or other
noise. These nodes are:
• axle transport wheels,
• axle actuator mounng,
• jib arm bearing sleeve.
We recommend using any available penetrang
lubricants.
Report all loose connecons that are supposed
to be xed to the manufacturer’s authorized
service agent and stop using the product unl
the problem is resolved.
This device is equipped with electric actuator
which allows to adjust the height of the jib
arm. In order to ensure proper performance of
the actuators check the actuators for unusual
sounds every 3 months. These unusual sounds
contain rales, squeaks, creaks. Any perceived
performance deciency must be reported to
Human Care technical service immediately cea-
sing device operaon unl remove their causes.
Not less frequently than every 1 years and aer
each failure/repair of actuator technical service
must check the li for electric safety.
The minimal range of the review should include:
• check if there were no mechanical wiring
damages of the device,
• check if there were no mechanical damages
of the controller and the actuators.
See checklist in user manual.
Basic periodic review
One year aer producon date or every 8000
cycles of using the li, the indicator lamp light
up.
If this LED is on, operator must check condion
of the li (cables connecons, looseness on the
frame etc.).
Press and hold “up” and “down” buon for 5
sec., to turn o the light aer review.
Transport wheels
Actuator mounng
Jib arm bearing sleeve
LED indicator

www.humancaregroup.com 7
ENGLISH
Troubleshoong
Symptom Possible cause - soluon
Device doesn’t work properly. Diagnose the mobile li.
Height adjustment doesn’t work. Replace main actuator.
Spreading legs emits sounds. Lubricaon legs spread mechanism.
Problem with moving device. Check/replace castors.
Troubleshoong & Solving
Diagnose the mobile li
In case of problems with height adjustment,
please complete following steps:
Step 1:
• Check the baery level on hand control. If
any led is lighng, please check baery with
voltmeter,
• If the baery is empty, please charge it,
• If aer a few hours the baery is sll empty,
please check if the charger is working
correctly by using the voltmeter. If charger
is ok, you have to replace the baery for a
new one.
Step 2:
If the baery level is correct, check that the
wires are connected properly to the control box
in the boom part of the device.
Checking voltage on baery
Checking voltage on charger
Checking wires connecon

8www.humancaregroup.com
Diana II Comfort 4MSB Service Manual
Step 3:
If everything looks good so far but device sll
doesn’t work, for further diagnose, spare parts
will be necessary (hand control, control box and
actuator).
• At rst, replace the hand control for a new
one (or switch with another device if it’s
possible) and check,
• Next, prepare a new control box and connect
it with slots from device,
• Now test what is wrong, control box, remote
or actuator.
Note – in case of situaon that one of
actuators is not working, make sure that
problem is with actuator, by swapping
the wires 1 and 2 for actuators in the
control box and use the remote control
to test (a working actuator means a
faulty controller).
Step 4:
To remove the control box, just loosen two
main screws and li up whole holder with
control box.
Loosening two screws Removing main holder

www.humancaregroup.com 9
ENGLISH
Replace main actuator
Cauon - Following procedures require turn
down device on aside. To protect it from any
damage or scratches we strongly recommend to
prepare some kind of protecon (card boards,
sponges, etc.) on the oor.
Step 1:
Take out blockade from plug with small
screwdriver and take out the plug.
View of a plug
Unplugged
Removing the blockage
Sponges under device

10 www.humancaregroup.com
Diana II Comfort 4MSB Service Manual
Step 2:
Use a at screwdriver to remove the star lock
cap and circlips from the upper rod holding the
actuator (from only one side of the pin) and
remove the pin. Do the same operaon for
second connecon actuator with frame.
Step 3:
Remove old and install the new actuator in the
reverse order to dismounng, carefully ghten
the gland.
Removing star lock cap
Removing pin
Removing circlip (second side)
Removing pin (second side)
Removing pin (second side)
Removing circlip

www.humancaregroup.com 11
ENGLISH
Replace boom actuator
Cauon - Following procedures require
turn down device on aside. To protect
it from any damage or scratches we
strongly recommend to prepare some
kind of protecon (card boards, spong-
es, etc.) on the oor.
Step 1:
Remove boom metal cover. With remote
control try to spread legs and check with
actuator does not work.
Step 2:
With hand control try to spread legs and check
with actuator does not work.
Removing the screws
Internal view
Removing the screws
Sponges under device

12 www.humancaregroup.com
Diana II Comfort 4MSB Service Manual
Step 3:
Remove only two screws from failure actuator’s
side (two top screws in that occasion), pull
up actuator and remove boom pins holding
actuator.
Step 4:
Disconnect the cable
Removing rst screw
Take out actuator from the frame
Removing safety ring
Take out blockade
Removing second screw
Use at screw driver
View aer disconnecng actuator
Removing plug

www.humancaregroup.com 13
ENGLISH
Step 5:
Remove old and install the new actuator in the
reverse order to dismounng, carefully ghten
the gland.
Wheel replacement
To replace rear wheels, use a socked wrench
size 19”. To replace front wheels, use an allen
key and socked wrench size 17”.
Rear wheels replacement Front wheels replacement

14 www.humancaregroup.com
Diana II Comfort 4MSB Service Manual
Spare parts list
INDEX Price list for Diana II Comfort 4MSB
04-N0101-0012 LINAK Baery BAJ
04-N0107-0041 LINAK 2 funcons hand control HB3X
04-N0109-0040 LINAK CBJ-CARE control box
04-N0301-0072 LINAK main actuator LA34
04-N0301-0080 LINAK secondary actuator LA23
04-N0901-0035 LINAK micro-t grey wire 750mm
04-N0901-0043 LINAK control box holder + screw M5x8
04-N0901-0018 LINAK external charger
04-N0901-0094 LINAK safety ring for micro-t wire
04-N0901-0096 LINAK safety lock for secondary actuator’s plug
04-N0901-0112 LINAK Jack mini-t grey wire 1050mm
04-N0901-0123 EU LINAK power supply cord
04-M2010-0064 TENTE double wheel set D125 White with red brake
04-M2010-0066 TENTE double wheel set D100 white
04-M0501-0024 Round, at D10 PE white cap
04-E4043-0047 Grommet rubber D10x7
04-M3010-0013 Spring plate DIN 6799- 8
04-M3005-0030 Outer snap ring steel 10 with a hook
04-E4041-0071 Gray gland with metric thread M12x1.5
04-M3010-0014 Retaining plate (pin on sha) 10 steel
04-M0530-0017 M12 bolt / nut cover black
04-M0520-0054 Round overlay 8x12 black
04-M0505-0115 Oval end cap 30x15 black
04-M0501-0066 Round lencular D22 PE black cap
04-K8400-0051 Roller cover
04-K8400-0050 Roller
02-TAP-0337 4MSB adjustment belt
02-OBR-1203 Hook sling

www.humancaregroup.com 15
SVENSKA
Symboler
VARNING Symbolen indikerar en
potenellt farlig situaon som om den
inte kan undvikas kan medföra allvarlig
skada på brukaren eller egendomsska-
da och/eller skada på själva enheten.
INFORMATION Symbolen indikerar re-
kommendaoner för och informaon
om korrekt, problemfri användning.
VIKTKAPACITET Symbolen indikerar
maximal brukarvikt för produkten.
Denna vikt får aldrig överskridas.
CE-MÄRKNING Symbolen indikerar a
produkten uppfyller kraven i EU-di-
rekv 93/42/EEG (MDD) och/eller
EU-förordning 2017/745 (MDR).
ÅTERVINN enligt lokala förordningar.
MEDICINTEKNISK PRODUKT Symbolen
indikerar a produkten är en medicin-
teknisk produkt enligt denionen av
en medicinteknisk produkt i EU-di-
rekv 93/42/EEG (MDD) och/eller
EU-förordning 2017/745 (MDR).
LÄS INSTRUKTIONERNA och se ll a
du förstår dem fullt ut innan du börjar
använda denna produkt.
Vikgt!
Du måste läsa bruksanvisningen för din enhet före användning. Håll denna
manual och informaonen i den llgänglig för framda bruk.
Det är inte llåtet a ändra produkten.
Före användning ska du alld kontrollera
a produkten inte är skadad eller sliten.
Human Cares produkter utvecklas och uppda-
teras konnuerligt för a erbjuda våra kunder
högsta kvalitet. Vi förbehåller oss räen a göra
förändringar i konstrukonen utan a dea
meddelas på förhand.
Se alld ll a du har den senaste versionen
av manualen som är llgänglig för nedladdning
från vår webbplats www.humancaregroup.com.
Kontakta din lokala distributör om du har frågor
om produkten och dess användning.
Servicemanual

16 www.humancaregroup.com
Diana II Comfort 4MSB Servicemanual
Human Care är e ISO 13485:2016-cererat
svenskt medicintekniskt företag. Kvalitetsled-
ningssystemet uppfyller kraven i US 21 CFR del
820.
Produkten är CE-märkt i enlighet med EU-di-
rekv 93/42/EEG (MDD) och/eller EU-förord-
ning 2017/745 (MDR), som en medicinteknisk
produkt i klass I.
CE-märkningen nns på produkten.
Denna sele har testats i enlighet med ISO
10535.
Eerlevnad och standarder
Användbar informaon
a) Servicemanual
Servicemanualen skapades för tekniker som
arbetar med service på mobillyen Diana II
Comfort 4MSB. För a mobillyen ska fungera
på e säkert sä förutsäer serviceprocedurer-
na a man har erfarenhet, tekniska kunskaper
och allmänna kunskaper om säkerhetsregler.
I denna manual används varningar och säker-
hetsinstrukoner för a lya fram olika service-
procedurer.
Varning – denna informaon beskriver
eventuella säkerhetsrisker för tekniker
och brukare.
Se upp – denna informaon beskriver
åtgärder som kan orsaka potenell
skada på lyen.
b) Installaon och dri av Diana II Com-
fort 4MSB
För montering och dri av lyen, se bruksanvis-
ningen.
c) Service på Diana II Comfort 4MSB
Servicen får endast uöras av Human Cares
auktoriserade serviceteam eller enheter som
auktoriserats av Human Care.
Human Care bär inte risken för konsekvenserna
av reparaoner eller ändringar som har gjorts
av obehörig personal eller för användning av
reservdelar som inte är originaldelar från Human
Care för mobillyar.
d) Servicedokumentaon
Servicedokumentaon och reparaonsdoku-
ment vid service nns på en webbplats som är
llgänglig för auktoriserad servicepersonal.
e) Försikghetsåtgärder
Mobillyens design ger maximal säkerhet.
Enhetens säkerhet garanteras genom säkerhets-
standard IEC60601-1.
Läs dessa försikghetsåtgärder noggrant
innan du påbörjar servicen!
Obs! – uör inte servicereparaoner och
funkonstest när paenten benner sig
i enheten.
f) Övrig informaon
Vid problem med mobillyen ska du kontakta
försäljarens eller llverkarens serviceavdelning.
Kontaktuppgier service:
HUMAN CARE SWEDEN (HQ)
Årstaängsvägen 21B
117 43 Telefon: +46 8 665 35 00
Fax: +46 8 665 35 10
E-post: info.se@humancaregroup.com

www.humancaregroup.com 17
SVENSKA
Beskrivning av servicemanual
Verktyg och underhåll
Servicemanualen innehåller informaon om
reparaon, underhåll och idenering av mobil-
ly. Lyen kan ideneras genom a använda
den ideneringskod som nns på namnplåten.
Obs! – om kontakten på baeriladda-
ren inte passar i uag i di land kan du
använda en adapter.
Verktyg
En standarduppsäning verktyg krävs vid servi-
ce på mobillyen:
Insexnyckelsats Vanlig sats med olika nycklar
Spårskruvmejsel
Hammare Voltmätare

18 www.humancaregroup.com
Diana II Comfort 4MSB Servicemanual
Underhåll av enhet
Alla rörliga leder måste smörjas var sjäe må-
nad eller när de avger gnissel eller annat ljud.
Dessa noder är:
• axel för transporthjul
• axel för ställdonsfäste
• lagerhylsa motviktsarm
Vi rekommenderar a du använder e llgäng-
ligt penetrerande smörjmedel.
Rapportera alla lösa delar som skall vara xe-
rade ll llverkares utsedda serviceagent och
använd ej produkten lls problemet är åtgärdat.
Denna enhet är utrustad med elektriska ställ-
don som gör det möjligt a justera höjden på
motviktsarmen. För a säkerställa a ställdonen
fungerar korrekt ska du kontrollera ställdonen
var tredje månad för a se om det förekommer
ovanliga ljud. Ovanliga ljud kan vara skrammel,
gnissel och knarrande. Alla upplevda prestanda-
brister ska rapporteras ll Human Cares teknis-
ka service och du ska omedelbart sluta använda
enheten ll dess a orsakerna åtgärdats.
Som minst varje år och eer fel i/reparaon
av ställdonet måste teknisk service kontrollera
lyen och se över elsäkerheten.
Översynen ska som minst omfaa:
• kontroll av a det inte nns några mekaniska
kabelskador på enheten
• kontroll av a det inte nns några mekanis-
ka skador på styrenheten och ställdonen
Se checklista i bruksanvisningen.
Återkommande grundgenomgång
E år eer llverkningsdatum eller eer 8 000
cyklers användning av lyen tänds indikator-
lampan.
Om denna LED-lampa lyser ska operatören
kontrollera lyens skick (kabelanslutningar,
glapp i ramen etc.).
Tryck och håll ned ”upp”- och ”ner”-knappen i 5
sek. för a stänga av lampan eer kontroll.
Transporthjul
Montering av ställdon
Lagerhylsa motviktsarm
LED-indikator

www.humancaregroup.com 19
SVENSKA
Symptom Möjlig orsak – lösning
Enheten fungerar inte korrekt. Ställ diagnos på mobillyen.
Höjdjusteringen fungerar inte. Byt ut huvudställdonet.
Ljud avges när bredden mellan benen justeras. Smörj benens justeringsmekanism.
Problem med a ya enheten. Kontrollera/byt ut hjulen.
Felsökning och lösning
Felsökning på mobillyen
Vid problem med höjdjustering ska du genom-
föra följande steg:
Steg 1:
• Kontrollera baerinivån på handkontrollen.
Om någon LED-lampa lyser ska du kontroll-
era baeriet med voltmätare
• Om baeriet är tomt ska du ladda det
• Om baeriet eer några mmar forarande
är tomt ska du kontrollera om laddaren fung-
erar korrekt genom a använda voltmätaren.
Om laddaren fungerar måste du byta ut
baeriet mot e ny.
Steg 2:
Om baerinivån är korrekt ska du kontrollera
a kablarna är ordentligt anslutna ll styrboxen
i den nedre delen av enheten.
Kontrollera spänningen hos baeriet
Kontrollera spänningen hos laddaren
Kontrollera kabelanslutningarna

20 www.humancaregroup.com
Diana II Comfort 4MSB Servicemanual
Steg 3:
Om allt ser bra ut så långt, men enheten
forarande inte fungerar, krävs reservdelar
för fortsa diagnos (handkontroll, styrbox och
ställdon).
• Först byter du ut handkontrollen mot en ny
(eller byter med en annan enhet om så är
möjligt) och kontrollerar
• Däreer förbereder du en ny styrbox och
ansluter den med uag från enheten
• Testa sedan vad som är fel: styrbox, ärrkon-
troll eller ställdon.
Obs! – i en situaon där e av ställdonen
inte fungerar ska du kontrollera om pro-
blemet nns i ställdonet genom a byta
ut kabel 1 och 2 för ställdonen i styrbox-
en och använda ärrkontrollen för a
testa (om ställdonet fungerar innebär det
a det är fel på styrenheten).
Steg 4:
Du tar bort styrboxen genom a lossa de två
huvudskruvarna och lyer upp hela hållaren
med styrboxen.
Lossa två skruvar Ta bort huvudhållare
Other manuals for Diana II Comfort 4MSB
1
This manual suits for next models
1
Table of contents
Languages:
Other human care Fitness Equipment manuals