
14
11. Do not apply excessive pressure on cable connection. If the cable does not connect, check to
see if the plug is appropriate for socket. Damaged cable plug and socket shall be repaired by
certified service technician.
12. Do not pull the cable. Always hold and pull plug when removing the cable.
13. This equipment can be used as other international purposes for general examination of
lenses in accordance with this instruction.
14. Always conduct a visual examination of the equipment and inspection for proper functioning
before operation.
15. Do not block the air grafting or holes.
16. Immediately turn off the power and remove plug when smoke, flame, unusual noise or smell
is detected from the equipment.
17. This equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to other
devices in the vicinity. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to other devices,
which can be determined by turning the equipment off and on, the user is encouraged to try
to correct the interference by one or more of the following measures:
A. Reorient or relocate the receiving device
B. Increase the separation between the equipment
C. Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected
D. Consult the manufacturer or field service technician for help
18. "External equipment intended for connection to signal input, signal output or other connectors,
shall comply with relevant IEC standard (e.g., IEC 60950 for IT equipment and IEC 60601
series for medical electrical equipment). In addition, all such combinations - systems - shall
comply with the standard IEC 60601-1-1, Safety requirements for medical electrical systems.
Any person who connects external equipment to signal input, signal output or other
connectors has formed at system and is therefore responsible for the system to comply with
the requirements of IEC60601-1-1. If, in doubt, contact qualified technician or your local
representative." (Or some cases, for example LCD Monitor, "This device is intended to
connect to the medical device only, which complies with standards of IEC 60601 series.")
19. To avoid risk of electric shock, this equipment must only be connected to a supply mains with
protective earth
20. Do not position the equipment so that it is difficult to operate the disconnecting device
(disconnecting device: power cord, appliance inlet, etc.)