Page 18
Note: The accuracy of the Calibrator’s Frequency function must be appropriate for
the specified Accuracy of the Meter.
Frequency Test: Model 61-480
Range Source Level Reading
3KHz 3.000KHz 0.5 Vp-p 2.999 to 3.001
30KHz 30.00KHz 0.5 Vp-p 29.99 to 30.01
300KHz 300.0KHz 0.5 Vp-p 299.9 to 300.1
3MHz 3.000MHz 0.5 Vp-p 2.999 to 3.001
10MHz 10.00MHz 0.5 Vp-p 9.99 to 10.01
Note: The accuracy of the Calibrator’s Frequency function must be appropriate for
the specified Accuracy of the Meter.
Frequency Test: Model 61-481 and 61-482
Range Source Level Reading
4KHz 3.600KHz 0.5 Vp-p 3.599 to 3.601
40KHz 36.00KHz 0.5 Vp-p 35.99 to 36.01
400KHz 360.0KHz 0.5 Vp-p 359.9 to 360.1
4MHz 3.600MHz 0.5 Vp-p 3.599 to 3.601
10MHz 10.00MHz 0.5 Vp-p 9.99 to 10.01
HZ Test (push button) (61-481 only)
Select ACV at 4V range. Push the HZ key and apply 500Hz at 100mv.
Reading should be 500.0 Hz +/- 1½ count.
Note: Sensitivity of the HZ mode is 1/100 of full scale.
The measurement frequency is from 40 Hz to 1 kHz.
Testing the Capacitance Function: Model 61-481 and 61-482
The meter measures capacitance by charging the capacitor with a known direct current, measuring the
resultant voltage, and calculating the capacitance. If the same capacitance is measured on an impedance
bridge, a different reading may result. This variance is likely to be greater at higher frequencies.
Note: Lead capacitance can effect accuracy on the lower ranges [4nF, 40nF], This error is not included.
Note: The meter selects the proper range automatically. Each measurement takes about one second per
range, 4mF and 40mF range take approx. 15 seconds.
Capacitance Test: Model 61-481 and 61-482
Range Source Reading
4nF 3.700nF 3.579 to 3.821
40nF 37.00nF 35.79 to 38.21
400nF 370.0nF 361.8 to 378.2
4µF 3.700µF 3.618 to 3.783
40µF 37.00µF 36.18 to 37.82
400µF 360.0µF 352.0 to 368.0
4mF (4,000µF) 3.600mF 3.400 to 3.800
40mF (40,000µF) 36.00mF 34.00 to 38.00
Form number TM61480-1-2 Rev 5 July 2002