Battery Replacement
Priority of function
The LCD display shows
voltage mode when the
input Terminal is satisfied
by the following conditions:
ACV 1.3V ~ 600.0V
DCV 2.1V ~ 600.0V
-0.7V ~ -600.0V
The LCD display shows
Ωmode when the input
terminal is satisfied by the
following conditions:
0 ~ ∞Ω
ACV 0 ~ 0.9V
DCV -0.4 ~ -0.2V
DCV 1.0 ~ 2.0V
The LCD display shows
diode mode when the
Input Terminal is satisfied
by the following conditions:
DCV 0.4 ~ 0.8V
The LCD display shows
AC mode when the
Current Input Jaw is
satisfied by the following
conditions :
ACA 0.9A ~ 400.0A
Caution :
The calibration seal can only be
punctured by certified distributor for
annual calibration otherwise the meter’s
performance can not be warranted.
Maintenance
Do not attempt to repair this Tester.
It contains no user-serviceable parts.
Repair or serving should only be
performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry
cloth and detergent do not use abrasives
or solvents.
TEST EQUIPMENT RISK ASSESSMENT
Users of this equipment and or their
employers are reminded that Health and
Safety Legislation require them to carry
out valid risk assessments of all electri-
cal work so as to identify potential
sources of electrical danger and risk of
electrical injury such as from inadvertent
short circuits. Where the assessments
show that the risk is significant then the
use of fused test leads constructed in
accordance with HSE guidance note
GS38 “ Electrical Test Equipment for
use by Electricians” should be used.
Specifications
1-1 General Specifications
LCD display digits :
6000 digit large scale LCD readout.
Measuring rate :
3 times / sec of ACV/DCV diode.
3 times / sec of resistance.
2 times / sec of ACA.
Overrange display :
“OL” is displayed for “Ω” function,
shows the real value for “A” and “V”
function.
Automatic power off time :
Approximately 20 minutes after
power on without operations.
Low battery indicator :
is displayed. Replace the battery
when the indicator appears in the
display.
Power requirement : 9V battery.
Battery type and life :
ALKALINE PP3, 9V 250 hours.
ⅠThe circuits not connected to mains.
ⅡThe circuits directly connected to
Low-voltage installation.
ⅢThe building installation.
ⅣThe source of the Low-voltage
installation.
Conductor Size : 27mm diameter.
Pollution degree : 2
EMC : EN 61326-1
Shock vibration :
Sinusoidal vibration per MIL-T-
28800E (5 ~ 55 Hz, 3g maximum).
Function Range Accuracy
V1.3V ~ 600.0V ±(1.5%+ 3 dgt)
50Hz ~ 500Hz
V2.1V ~ 600.0V
-0.7V ~ -600.0V ±(1% + 2 dgt)
±(1% + 4 dgt)
Overload protection : 600 Vrms
Max Operation time :
DT=30s for ≧30V
Input impedance : ≧4K for input
voltage up to 30V. Impedance
increases with input voltage to
approximate 200KΩ at 600V.
Resistance & Continuity
Function Range Accuracy
Ω;0.0Ω ~ 9999Ω±(2% + 2 dgt)
90.4V ~ 0.8V ±(1% + 3 dgt)
Overload protection : 600V rms
Max. open circuit voltage : 1.5V
Continuity check : Internal sound
activates if the resistance of the
circuit under test is less than 25Ω.
It will then turn off if the resistance
is increased beyond 400Ω.
AC Current
Function Range Accuracy
A0.9A ~ 400.0A ±(1.8% + 3 dgt)
50Hz ~ 60Hz
Overload protection : 400A
1-2 Environmental Conditions
Indoor Use.
Calibration :
One year calibration cycle.
Operating temperature :
0°C ~ 30°C (≦80% RH)
30°C ~ 40°C (≦75% RH)
40°C ~ 50°C (≦45%RH)
Storage temperature :
-20 to +60°C, 0 to 80% RH (batteries
not fitted).
Operating altitude : 2000m (6562 ft)
Overvoltage category :
IEC 61010-1 600V CAT.Ⅲ.
CAT Application field
1-3 Electrical Specifications
Accuracy is ±(% reading + number
of digits) at 23°C ±5°C < 80%RH.
Temperature coefficient :
0.2 x (Specified accuracy) / °C,
< 18°C, > 28°C .
Voltage