Introduction
Safetypette may be used with glass or plastic pipettes to facilitate the pipetting of a wide range of liq-
uids. Pipettes within the volume range of 1 ml to 100 ml can be used.
When correctly used, liquid only comes into contact with
the pipette.
Operating Instructions
1. Fitting the pipette
Carefully attach the pipette by inserting it into the
silicone holder inside the nose cone. Do not use force.
2. Filling the pipette
Press the suction bulb first whilst holding the
Safetypette in a vertical position with the pipette tip
down. Place the tip of the pipette into the liquid to be
aspirated. Press the aspirate lever slowly upwards
allowing the liquid to fill the pipette so that the
meniscus level is slightly above the required mark.
Note:
Do not allow liquid to enter the Safetypette. The
function of the filter will be severely impaired
reducing suction efficiency. Spare fiters are
available see over.
3. Meniscus Adjustment
Wipe the pipette tip with a suitable lint-free tissue.
Press the pipetting lever down slowly until the
required meniscus has been reached.
4. Dispensing
Hold the receiving vessel in an inclined position, plac-
ing the pipette tip against the receiving
vessel’s wall. Press the dispense lever down to empty the pipette.
5. ‘Blow-out’ Pipettes
As soon as the meniscus in the pipette comes to a standstill, press the discharge button to expel
the last few drops. To ensure a complete dispense procedure, wipe the pipette
tip upwards a few millimeters along the inside wall of the receiving vessel.
SAFE T YPETTE
Operator Instructions
Suction Bulb
Discharge
Button
Aspirate &
Dispense Control
lever
Nose Cone
6. ‘To Deliver’ Pipettes
As soon as the liquid in the pipette comes to a standstill, wait
for the period of time required to ensure the timed liquid has
been fully dispensed.
Note: Pipetting time is printed on the pipette e.g.’EX + 10s’
7. To Remove Pipette
Hold the pipette at its upper end and gently twist and remove
from the silicone pipette holder.
8. Cleaning
The Safetypette can be easily maintained and cleaned. A damp
cloth is all that is required.
9. Replacing Filter
When replacement of the filter is necessary, proceed as follows:-
a remove pipette
b unscrew the nose cone
c remove the filter and replace with a new one
(Cat. No. 266-003)
d re-assemble
10. Leak Test
Attach a pipette and aspirate liquid to a minimum of half the
capacity of the pipette. Hold the pipette over a suitable reser-
voir. If liquid drips from the pipette tip, dismantle the unit (see
9) and re-assemble correctly.
11. Returning For Repair
When returning Safetypette for repair, please ensure the
instrument has been cleaned and decontaminated. Please ask
us for a decontamination statement form.
SAFE T YPETTE
Operator Instructions (cont.)
Jencons Scientific Inc (USA)
800 Bursca Drive, Suite 801
Bridgeville, PA 15017
Phone: 412 257-8861
Toll Free: 800 846-9959
Fax: 412 257-8809
Email: [email protected]
Website: www.jenconsusa.com
Jencons Scientific Ltd
Cherrycourt Way Ind. Est.
Stanbridge Rd, Leighton Buzzard
LU7 4UA, ENGLAND
Phone: 01525 372010
Fax: 01525 379547
Int. Phone: +44 1525 372010
Int. Fax: +44 1525 851461
Email: [email protected]
Int. Email: [email protected]
Website: www.jencons.co.uk
Online sales: www.ecomcat.co.uk
ISO 9001:2000
FM 15114
Filter:
Cat No. 266-003
Silicone holder:
Cat No 266-004
Nose Cone:
White: Cat No. 475-156
Pink: Cat No. 475-157
Lilac: Cat No. 475-158
Mint: Cat No. 475-159
SAFETYPETTE

Other JENCONS Laboratory Equipment manuals

JENCONS Safetypette Pro Specifications

JENCONS

JENCONS Safetypette Pro Specifications

JENCONS Smoothie Specifications

JENCONS

JENCONS Smoothie Specifications

JENCONS Powerpette Pro Specifications

JENCONS

JENCONS Powerpette Pro Specifications

JENCONS Powerpette Specifications

JENCONS

JENCONS Powerpette Specifications

JENCONS Powerpette Pro Specifications

JENCONS

JENCONS Powerpette Pro Specifications

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.