Keysight PXIe Series Use and care manual

Keysight PXIe
Chassis Family
Firmware Guide
Notices
© Keysight Technologies, Inc. 2018
No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or
translation into a foreign language) without
prior agreement and written consent from
Keysight Technologies, Inc. as governed by
United States and international copyright
laws.
Manual Part Number
M9019-90025
Edition
First Edition, May 2018
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED "AS IS," AND IS
SUBJECT TO BEING CHANGED, WITHOUT
NOTICE, IN FUTURE EDITIONS. FURTHER,
TO THE MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS
ALL WARRANTIES, EITHER EXPRESS OR
IMPLIED WITH REGARD TO THIS MANUAL
AND ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIMITED TO
THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT SHALL
NOT BE LIABLE FOR ERRORS OR FOR
INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE
FURNISHING, USE, OR PERFORMANCE OF
THIS DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY
TERMS COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH THESE
TERMS, THE WARRANTY TERMS IN THE
SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of
a U.S. Government prime contract or
subcontract, Software is delivered and
licensed as “Commercial computer
software” as defined in DFAR 252.227-
7014 (June 1995), or as a “commercial
item” as defined in FAR 2.101(a) or as
“Restricted computer software” as defined
in FAR 52.227-19 (June 1987) or any
equivalent agency regulation or contract
clause. Use, duplication or disclosure of
Software is subject to Keysight
Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights
as defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls
attention to an operating procedure,
practice, or the like that, if not correctly
performed or adhered to, could result in
damage to the product or loss of important
data. Do not proceed beyond a CAUTION
notice until the indicated conditions are
fully understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure,
practice, or the like that, if not correctly
performed or adhered to, could result in
personal injury or death. Do not proceed
beyond a WARNING notice until the
indicated conditions are fully understood
and met.
PXIe Chassis Firmware Guide 3
Safety Summary
The following general safety precautions must be observed during all
phases of operation of this instrument. Failure to comply with these
precautions or with specific warnings or operating instructions in the
product manuals violates safety standards of design, manufacture, and
intended use of the instrument. Keysight Technologies assumes no liability
for the customer's failure to comply with these requirements. Product
manuals are provided with your instrument on CD-ROM and/or in printed
form. Printed manuals are an option for many products. Manuals may also
be available on the Web. Go to www.keysight.com and type in your product
number in the Search field at the top of the page.
General This product is a Safety Class 1 instrument (provided with a protective earth
terminal). The protective features of this product may be impaired if it is
used in a manner not specified in the operation instructions.
All Light Emitting Diodes (LEDs) used in this product are Class 1 LEDs as
per IEC 60825-1.
Environment Conditions This instrument is intended for indoor use in an installation category II,
pollution degree 2 environment. It is designed to operate at a maximum
relative humidity of 95% and at altitudes of up to 2000 meters.
Refer to the specifications tables for the ac mains voltage requirements and
ambient operating temperature range.
Before Applying Power Verify that all safety precautions are taken. The power cable inlet of the
instrument serves as a device to disconnect from the mains in case of
hazard. The instrument must be positioned so that the operator can easily
access the power cable inlet. When the instrument is rack mounted the
rack must be provided with an easily accessible mains switch.
Ground the Instrument To minimize shock hazard, the instrument chassis and cover must be
connected to an electrical protective earth ground. The instrument must be
connected to the ac power mains through a grounded power cable, with
the ground wire firmly connected to an electrical ground (safety ground) at
the power outlet. Any interruption of the protective (grounding) conductor
or disconnection of the protective earth terminal will cause a potential
shock hazard that could result in personal injury.
Do Not Operate in an
Explosive Atmosphere
Do not operate the instrument in the presence of flammable gases or
fumes.
Do Not Remove the
Instrument Cover
Operating personnel must not remove instrument covers. Component
replacement and internal adjustments must be made only by qualified
personnel.
Instruments that appear damaged or defective should be made inoperative
and secured against unintended operation until they can be repaired by
qualified service personnel.
Contents
1
Firmware Overview .................................................................................8
2
Determining the Installed Version...........................................................9
3
Firmware Version Contents...................................................................10
4
Firmware Version Detail........................................................................11
5
Upgrading or Downgrading Firmware Process .....................................13
2
Determining the Installed Version
Firmware version contents are displayed by the Soft Front Panel (SFP)
Help About screen of the PXIe Chassis Family driver. However, you can
only see the Trigger Bridge firmware version when using PXIe Chassis
Driver 1.7.82.1 or later. You cannot observe the Trigger Bridge firmware
version using the PXIe Chassis Family driver prior to 1.7.82.1. Please
download and install the latest driver from www.Keysight.com. Search
on your chassis model number. Click on “Visit Technical Support” and
then click on the tab for “Drivers, Firmware & Software”. After installing
version 1.7.82.1 or later, run the SFP for your chassis and click on
Help>About. Refer to the next page of this document to identify which
firmware version is installed.
The figure below shows an example of the Help>About for an M9019A
chassis. In this example, the version of the Trigger Bridge version
number matches that of the 2018 firmware. If the 2017 firmware were
installed, it would show version 0 for the Trigger Bridge.
Figure 1.M9019A SFP Help About
PXIe Chassis Firmware Guide 9
4
Firmware Version Detail
Select the version of firmware based on your need. It is recommended
that you always use the latest version of the firmware. Please check the
firmware download webpage for the latest version available. Below is
the detail showing each firmware capabilities.
2017 Firmware Version Detail
If you do not need the enhanced capability of the front panel trigger
ports and you are using the 2017 firmware version, then you can use
any version of the PXIe Chassis Family driver, including 1.7.82.1. The
1.7.82.1 PXIe Chassis Family driver automatically detects the firmware
version installed and performs accordingly.
The only difference between the 2017 Firmware version and the 2018
Firmware version is in the behavior of the front panel trigger ports. The
following paragraph describes how they work in the 2017 Firmware
version.
With the 2017 firmware, the front panel trigger port, when configured as
an Input, can connect to only a single PXI_TRIG line on Trigger Bus
Segment 2 (Segment 1 for the M9010A). The front panel trigger port,
when configured as an Output, can only output the trigger signal from a
single PXI_TRIG line on Trigger Bus Segment 2 (Segment 1 for the
M9010A). In both these situations, only one PXI_TRIG trigger signal line
can be used and it must be on Trigger Bus Segment 2 (Segment 1 for
the M9010A). If you want to connect a front panel trigger port to a
PXI_TRIG trigger signal line on a different Trigger Bus Segment, you
must add a route to the chassis configuration. See the PXIe Chassis
Family User Guide for more information.
2018 Firmware Version Detail
The PXIe Chassis 2018 firmware adds enhanced capability to the front
panel trigger ports, but it can only be used with PXIe Chassis Family
driver 1.7.82.1 or later. If you want the enhanced capability for the front
panel trigger ports, then you must use both the 2018 firmware and the
1.7.82.1 driver or later.
PXIe Chassis Firmware Guide 11
The 2018 firmware can only be used with the 1.7.82.1 or later PXIe Chassis
Family driver. It will not work with driver version 1.6.493.1 or older versions. If
the 2018 Firmware version is installed in your chassis, you must use the
1.7.82.1 or later driver. If you cannot use 1.7.82.1, then you must downgrade
your firmware.
When the 2018 firmware is installed in your chassis, the PXIe chassis
Driver 1.7.82.1 offers the enhanced front panel trigger port capability in
both the SFP Trigger Ports tab and in the IVI software API.
The front panel trigger port, when configured as an Output, can connect
to multiple PXI_TRIG lines on Trigger Bus Segment 2 or 3 (or both). For
the M9010A chassis, the front panel trigger ports can connect to
Segments 1 or 2 (or both). You have a great deal of flexibility in
configuring the destination of the incoming trigger signal. You can
configure the incoming trigger signal to go to a single PXI_TRIG line on
a single segment or to multiple PXI_TRIG lines on two segments.
The front panel trigger port, when configured as an Output, can only
output the trigger signal from a single PXI_TRIG line, but you can
connect it to either Trigger Bus Segment 2 or 3 (Segments 1 or 2 for the
M9010A). Unlike the 2017 firmware, which is limited to a single Trigger
Bus Segment, the 2018 firmware can output from either of two Trigger
Bus Segments.
12 PXIe Chassis Firmware Guide
5
Upgrading or Downgrading
Firmware Process
This section describes how to upgrade or downgrade your chassis
firmware. PXIe Chassis Family driver version 1.7.82.1 or later is required
to perform the firmware change process.
Step 1
- Determine the model number of the chassis (i.e. M9010A,
M9018B, or M9019A) you want to upgrade or downgrade.
Step 2
– Determine whether you are upgrading or downgrading. It is
recommended that you always use the latest version of the firmware.
Step 3
– Download the firmware zip file from Keysight.com. Click on the
link appropriate for your chassis model number:
www Keysight com/find/M9010A
www.Keysight.com/find/M9018B
www Keysight com/find/M9019A
•Click on “Visit Technical Support”.
•Click on the “Drivers, Firmware & Software” tab.
•Find the version you want and download it.
Step 4
– Unzip your firmware zip file.
Step 5
– Start the SFP.
Step 6
– Click on Utilities > Firmware Update.
Step 7
– Click on “Yes” if the below Alert screen pops up.
Step 8
– The Firmware Update Utility will start as shown in the screen
shot below. Click on “Browse” then navigate to the location where you
unzipped the Firmware zip file. Select the Manifest*.* file and click
”Open”.
PXIe Chassis Firmware Guide 13
Step 9
– With your browse path filled in, click “Next”.
Step 10
– The Firmware Update Utility asks you to verify that you want
to update the “Current version” to the “New version”. In the example
below, we are downgrading from 2018 firmware to 2017 firmware. To
start the process, click “Update”.
Step 11
– When the Firmware Update Utility finishes, it displays this
message. Please follow the instructions to complete the firmware
change process.
14 PXIe Chassis Firmware Guide
This information is subject to change
without notice.
© Keysight Technologies 2018
Edition 2, Aug 2018
Published in U.S.A.
*M9019-90025*
M9019-90025
http://www.keysight.com
PXIe Chassis Firmware Guide 15

This manual suits for next models

3

Other Keysight Chassis manuals

Keysight Ixia XGS2 User manual

Keysight

Keysight Ixia XGS2 User manual

Keysight M9018A User guide

Keysight

Keysight M9018A User guide

Keysight Ixia XG12 User manual

Keysight

Keysight Ixia XG12 User manual

Keysight M9502A User manual

Keysight

Keysight M9502A User manual

Keysight Ixia IxOS XGS2 Series User manual

Keysight

Keysight Ixia IxOS XGS2 Series User manual

Keysight M9018A User manual

Keysight

Keysight M9018A User manual

Keysight PXIe M9018B User guide

Keysight

Keysight PXIe M9018B User guide

Keysight M9506A User manual

Keysight

Keysight M9506A User manual

Keysight ixia XGS12 User manual

Keysight

Keysight ixia XGS12 User manual

Keysight M9514A Operating instructions

Keysight

Keysight M9514A Operating instructions

Keysight M9514A User guide

Keysight

Keysight M9514A User guide

Keysight PXIe M9010A User guide

Keysight

Keysight PXIe M9010A User guide

Keysight E8401A Troubleshooting guide

Keysight

Keysight E8401A Troubleshooting guide

Popular Chassis manuals by other brands

VESTEL 11AK37 Service manual

VESTEL

VESTEL 11AK37 Service manual

HP A10504 Product End-of-Life Disassembly Instructions

HP

HP A10504 Product End-of-Life Disassembly Instructions

AOpen H800 user guide

AOpen

AOpen H800 user guide

EK Quantum Vector FTW3 RTX 3080 user guide

EK

EK Quantum Vector FTW3 RTX 3080 user guide

BURL B16-BMB5 MOTHERSHIP quick start guide

BURL

BURL B16-BMB5 MOTHERSHIP quick start guide

Rosewill RSV-X4000 user manual

Rosewill

Rosewill RSV-X4000 user manual

Supermicro SC815B user manual

Supermicro

Supermicro SC815B user manual

iStarUSA D-206 Product Quick Guide

iStarUSA

iStarUSA D-206 Product Quick Guide

Cobra 427 Instruction book

Cobra

Cobra 427 Instruction book

Advantech IPC-623 Series user manual

Advantech

Advantech IPC-623 Series user manual

Advantech HPC-7480 user manual

Advantech

Advantech HPC-7480 user manual

Pickering LXI 60-103 user manual

Pickering

Pickering LXI 60-103 user manual

Hisense MT5659AUHT Service manual

Hisense

Hisense MT5659AUHT Service manual

National Instruments PXI EXPRESS PXIe-1095 user manual

National Instruments

National Instruments PXI EXPRESS PXIe-1095 user manual

Supermicro SC116B user manual

Supermicro

Supermicro SC116B user manual

Advantech HPC-8316 Series user manual

Advantech

Advantech HPC-8316 Series user manual

iStarUSA D-400-7 Product Quick Guide

iStarUSA

iStarUSA D-400-7 Product Quick Guide

TP-Link TL-MC1400 installation guide

TP-Link

TP-Link TL-MC1400 installation guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.