King Innovation KING CM User manual

KING CM
CLAMP METER
INSTRUCTION MANUAL
True RMS
Milliamp Clamp Meter
#42095
31318 84

2
I
KING INNOVATION
TABLE OF CONTENTS
1. SAFETY INFORMATION........................................3
2. GENERAL SPECIFICATION...................................4
3. ELECTRICAL SPECIFICATION...............................5
3-1 ACmA Measurement.............................................5
3-2 ACA Measurement................................................5
3-3 ACV Measurement ................................................6
3-4 DCV Measurement................................................6
3-5 Continuity ( ) ....................................................6
3-6 Resistance (Ω)........................................................6
4. DESCRIPTION..............................................................7
4-1 Description of the display .....................................7
4-2 Description of front and rear ................................8
5. BUTTON INSTRUCTION ........................................9
5-1 HOLD & LPF Function............................................9
5-2 PEAK HOLD Function............................................9
5-3 ZERO Function.......................................................9
5-4 BACKLIGHT Function..........................................10
6. MEASURING INSTRUCTION............................10
6-1 ACA & ACmA Measurement..............................10
6-2 ACV Measurement ..............................................12
6-3 DCV Measurement..............................................13
6-4 Continuity Measurement.....................................13
6-5 Resistance Measurement ....................................14
7. CHANGING THE BATTERY........................... 14,15
8. MAINTENANCE........................................................15
8-1 Cleaning...............................................................15

2
I
KING INNOVATION KING INNOVATION
I
3
MILLIAMP CLAMP METER
1. SAFETY INFORMATION
• Never attempt to make any measurement if any
abnormal conditions are noted, such as broken case,
cracked test leads and exposed metal part.
• Set the Function Switch to an appropriate
position before starting measurement.
• Do not perform resistance and continuity test on a live
power system.
• Do not apply voltage between the test terminals and test
terminal to ground that exceed the maximum limits stated
in this manual.
• Keep your fingers behind the protection ring on the test
probes when using the test leads.
• Change the battery when the symbol appears to
avoid incorrect reading.
Environmental Conditions
Operation Temperature: 0°C to 40°C (32°F to 104°F);
<80 % RH
Storage Temperature: -10°C to 60°C (14°F to 140°F);
<80 % RH
Altitude: Up to 2000 meters
Pollution Degree: 2
Explanation of Symbols
Attention refer to operation instructions.
Approvals
1. SAFETY INFORMATION........................................3
2. GENERAL SPECIFICATION...................................4
3. ELECTRICAL SPECIFICATION...............................5
3-1 ACmA Measurement.............................................5
3-2 ACA Measurement................................................5
3-3 ACV Measurement ................................................6
3-4 DCV Measurement................................................6
3-5 Continuity ( ) ....................................................6
3-6 Resistance (Ω)........................................................6
4. DESCRIPTION..............................................................7
4-1 Description of the display .....................................7
4-2 Description of front and rear ................................8
5. BUTTON INSTRUCTION ........................................9
5-1 HOLD & LPF Function............................................9
5-2 PEAK HOLD Function............................................9
5-3 ZERO Function.......................................................9
5-4 BACKLIGHT Function..........................................10
6. MEASURING INSTRUCTION............................10
6-1 ACA & ACmA Measurement..............................10
6-2 ACV Measurement ..............................................12
6-3 DCV Measurement..............................................13
6-4 Continuity Measurement.....................................13
6-5 Resistance Measurement ....................................14
7. CHANGING THE BATTERY........................... 14,15
8. MAINTENANCE........................................................15
8-1 Cleaning...............................................................15
CONFORMS TO UL STD. 61010-1,
61010-2-032, 61010-2-033
CERTIFIED TO CSA STD. C22.2 #61010-1-12,
61010-2-032, 61010-2-033
3131884

4
I
KING INNOVATION
INSTRUCTION MANUAL
2. GENERAL SPECIFICATION
The King Innovation KING CM Digital Milliamp Clamp Meter
is ideal for use in troubleshooting two-wire irrigation,
landscape lighting and electrical systems. The KING CM
is for use in indoor and outdoor dry locations.
Digital Display:
6000 count backlit LCD display
Polarity:
Automatic, + Implied, Negative Polarity Indicator
Low Battery Indication:
The meter has a low battery indicator that will alert the
user when the voltage is too low to provide accurate readings.
Sample Rate:
2 times/sec
Power Source:
(2) 1.5V AAA batteries
Typical battery life: (without buzzer, backlight)
40 hours at ACmA and ACA function;
60 hours at DCV and Ohm function. (alkaline battery)
Auto Power Off:
The meter will shut off after 30 minutes of non-use to preserve
battery life. This function can be disabled by pressing and
holding the “HOLD” button while powering on the unit.
Over Load:
When the applied signal is larger than the maximum for the
range in use the display will show OL.
Maximum Jaw Opening:
.79” (20mm)
Dimensions:
8.27” x 2.87” x 1.50” (210 x 73 x 38mm)
Accessories:
Carrying case, batteries, test leads and instruction manual.

4
I
KING INNOVATION KING INNOVATION
I
5
MILLIAMP CLAMP METER
3. ELECTRICAL SPECIFICATION
The accuracy specification is defined
as ± (percent of reading + digit) at 23±5°C, ≤80 %RH.
Safety Class - CAT III 600V
3-1 ACmA (True RMS)
Zero correction: Fractions smaller than approximately
0.006 mA are calibrated to zero.
Low Pass Filter
3-2 ACA (True RMS)
Low Pass Filter
Range Resolution Accuracy
50~60Hz 60~500Hz
6.000mA 0.001mA (1μA) 1.0% + 8dgts 2.0% + 8dgts
60.00mA 0.01mA (10μA) 1.0% + 5dgts 2.0% + 5dgts
600.0mA 0.1mA (100μA)
Range Resolution Accuracy
6.000mA 0.001mA 2.0% + 8dgts
60.00mA 0.01mA 2.0% + 5dgts
600.0mA 0.1mA
Range Resolution Accuracy
6.000A 0.001A 2.0% + 5dgts
60.00A 0.01A
Range Resolution Accuracy
50~60Hz 60~500Hz
6.000A 0.001A 1.0% + 5dgts 2.0% + 5dgts
60.00A 0.01A

6
I
KING INNOVATION
INSTRUCTION MANUAL
3-3 ACV (True RMS)
Input impedance: 2MΩ
3-4 DCV
Input impedance: 2MΩ
3-5 Continuity
3-6 Resistance (Ω)
Range Resolution Accuracy
50~500Hz
60.00V 0.01V 1.0% + 3dgts
600.0V 0.1V
Range Resolution Accuracy
60.00V 0.01V 1.0% + 2dgts
600.0V 0.1V
Range Resolution Accuracy
600.0Ω0.1Ω
1.0% + 2dgts
6.000KΩ0.001KΩ
60.00KΩ0.01KΩ
600.0KΩ0.1KΩ
Range Buzzer Function
Ohm < 45Ω

6
I
KING INNOVATION KING INNOVATION
I
7
MILLIAMP CLAMP METER
4. DESCRIPTION
4-1 Description of the display
Auto power off indication
Polarity indication
Low battery indication
~
AC measurement indication
DC measurement indication
ACurrent measurement indication
VVoltage measurement indication
ZERO ZERO indication
HOLD Data HOLD indication
PEAK PEAK hold indication
Low pass filter indication
Continuity test indication
K
Measurement unit
Ω Resistance measurement indication
mMeasurement unit

8
I
KING INNOVATION
INSTRUCTION MANUAL
4-2 Description of front and rear
1 Current Sensing Clamp
2 Safety Protection Ring
3 Backlight Button
4 Clamp Opening Handle
5 Function Select Dial
6 ZERO Button
7 PEAK and /ΩButtons
8 Data HOLD & LPF Button
9 LCD Display
10 COM Input Terminal
11 Positive Input Terminal
12 Battery Cabinet
12

8
I
KING INNOVATION KING INNOVATION
I
9
MILLIAMP CLAMP METER
5. BUTTON INSTRUCTION
5-1 HOLD & LPF Function
The meter has a data hold feature that will retain the
measured value on the LCD screen. Press the “HOLD”
button momentarily to retain the measured value. Press the
“HOLD” button again or power the meter off and back on
to resume normal operation.
To activate the Low Pass Filter (LPF) feature, press and hold
the “LPF” button until the symbol displays. Press and hold
the “LPF” button for 2 seconds to deactivate the “LPF”
feature and resume normal operation.
The Cut-Off frequency of the low pass filter is about 160Hz
with attenuation characteristic of approx. -24dB/octave.
Note: the LPF function operates on the AC A
and AC mA ranges.
5-2 PEAK Hold and /ΩFunction Selection
Press “PEAK” button to enter the “PEAK” mode, the meter
records the peak maximum value. Press the “PEAK” button
again, to exit the peak mode.
When you turn the rotary switch to the Ohms/Continuity
Test position, press the “PEAK” button to select continuity
test with buzzer.
5-3 ZERO Function
Press “ZERO” button to enter the “ZERO” mode, ZERO
Indicator will appear and the display will ZERO. The
reading is stored as reference value for the subsequent
measurement.
Press the “ZERO” button again, to exit the ZERO mode.

10
I
KING INNOVATION
INSTRUCTION MANUAL
5-4 BACKLIGHT Function
When the “☼“ button is pressed, the LCD display backlight
will be turned on. Press the button again to turn the
backlight off. If the button is not pressed again then the
backlight will be automatically turned off in about 30
seconds after it’s turned on.
6. MEASURING INSTRUCTION
6-1 ACA, ACmA Measurement:
(1) Set the range selector switch to the desired position.
The current to be measured should be within the
measuring range.
(2) Normal measurement (see Fig. 1):
Using the trigger, open the clamp meter jaw and clamp
around the conductor. Make sure that the conductor
is positioned in the center and that the flow of current
matches the arrows located on the jaws of the clamp meter.
The display will show the value of the measurement being
taken. Earth leakage current or any small AC current that
flows through a ground wire can also be measured
by this method.
NOTE: Keeping the cable in the center of the clamp will
return more accurate test results when making current
measurements. Zero meter to accurately measure low
milliamp circuits.
(3) Measuring out of balance leakage current (see Fig. 2):
Disconnect the test leads from the meter. Clamp the
jaw around all conductors except the ground wire. The
measured imbalance current value is shown on the display.
If the display is not giving a steady and consistent reading,
press the HOLD button and read the display after the
clamp is removed from the conductors.

10
I
KING INNOVATION KING INNOVATION
I
11
MILLIAMP CLAMP METER
Fig. 1 Normal Measurement:
INCORRECT CORRECT
Fig. 2 Measuring out of balance leakage current:
Single-Phase 2-Wire System
LOADLOAD
3-Phase 3-Wire System

12
I
KING INNOVATION
INSTRUCTION MANUAL
6-2 ACV Measurement:
Turn the dial to the ṽrange. Connect the red test lead to
the “+” terminal and the black lead to the “COM” terminal.
Place the red and black test leads at both the positive and
negative points to be measured. The display will show the
value of the measurement being taken.
WARNING!
Maximum Input Voltage is 600V AC/DC. Do not
attempt to take any voltage measurement that may
exceed this maximum to avoid Electrical Shock hazard
and/or damage to this instrument.

12
I
KING INNOVATION KING INNOVATION
I
13
MILLIAMP CLAMP METER
6-3 DCV Measurement:
Turn the dial to the Vrange. Connect the red test lead to
the “+” terminal and the black lead to the “COM” terminal.
Place the red and black test leads at both the positive and
negative points to be measured. The display will show the
value of the measurement being taken.
6-4 Continuity Test with Buzzer:
Turn the dial to the Ωrange. Connect the red test lead to
the “+” terminal and the black lead to the “COM” terminal.
Place the red and black test leads at both the positive and
negative points to be measured. The buzzer will sound when
the resistance is less than 45 Ω.
NOTE:
When measuring continuity, all power to the circuit or cable
being tested MUST be turned off to prevent damage to the
user or the clamp meter.
Short Circuit Open Circuit

14
I
KING INNOVATION
INSTRUCTION MANUAL
6-5 Resistance Measurement:
Turn the dial to the Ωrange. Connect the red test
lead to the “+” terminal and the black lead to the “COM”
terminal. Place the red and black test leads at both the
positive and negative points to be measured. The display
will show the resistance value of the measurement being
taken on the display.
NOTE:
When measuring resistance, the circuit should be powered
off and all capacitors should be completely discharged
prior to testing.

14
I
KING INNOVATION KING INNOVATION
I
15
MILLIAMP CLAMP METER
7. CHANGING THE BATTERY
Replace the batteries when a low battery symbol is
displayed on the LCD screen. When the battery is
completely exhausted, the display will appear blank
and no symbol will be shown. Rotate the dial to the “OFF”
position and remove the test leads. Remove the cover using a
screwdriver. Replace the batteries. Use two new 1.5V (AAA)
batteries. Reinstall the door and tighten the screw
to secure the cover.
NOTE: Remove the batteries if the mter is not going to be
used for an extended period of time. The meter should be
stored in a cool, dry location.
8. MAINTENANCE
8-1 Cleaning:
Use only a soft cloth dampened with water or
neutral detergent for cleaning the meter. Do not use
abrasives, solvents or harsh chemicals. Allow to dry
thoroughly before use.
WARNING!
Before opening the meter, disconnect both test leads.
Never use the meter unless the battery cover is in place
and the screw is tightened.

©2020 King Innovation
42 N. Central Drive
O'Fallon, MO 63366
www.kinginnovation.com
800.624.4320
MADE IN CHINA
REV H 11/21
For technical support, service inquiries or
warranty requests, call (877) 771-8885 or email
kingser[email protected]

©2020 King Innovation
O'Fallon, MO 63366
42 N. Central Drive
www.kinginnovation.com
800.624.4320
HECHO EN CHINA
REV H 11/21
Para soporte técnico, consultas de servicio
o solicitudes degarantía, llame (877) 771-8885
o correo electrónico
kingser[email protected]

KING INNOVATION
I
15
AMPERÍMETRO DE PINZA EN MILIAMPERIOS
7. CAMBIO DE BATERÍA
Reemplace las baterías cuando en la pantalla LCD aparece
el símbolo de batería baja . Una vez que la batería se
agota completamente, la pantalla aparece en blanco y no
se muestra ningún símbolo. Gire el dial a la posición “OFF”
(apagado) y saque los cables de prueba. Saque la cubierta
con un destornillador. Reemplace las baterías. Use dos
baterías de 1.5 V (AAA) nuevas. Vuelva a colocar la puerta
y ajuste el tornillo para asegurar la cubierta.
NOTA: Saque las baterías si no usará el medidor durante
un período prolongado. Deberá guardar el medidor en un
lugar fresco y seco.
8. MANTENIMIENTO
8-1 Limpieza:
Use solo una tela suave humedecida con agua o detergente
neutro para limpiar el medidor. No use abrasivos,
disolventes o productos químicos agresivos. Deje secar
completamente antes del uso.
¡ADVERTENCIA!
Antes de abrir el amperímetro, desconecte ambos
conductores de prueba. Nunca use el amperímetro sin la
cubierta de la batería en su lugar y el tornillo apretado.

14
I
KING INNOVATION
MANUAL DE INSTRUCCIONES
6-5 Medición de resistencia:
Gire el dial al rango Ω. Conecte el conductor de
prueba rojo al terminal “+” y el conductor de prueba negra
al terminal “COM”. Coloque los cables de prueba de color
rojo y negro en los puntos positivo y negativo que se desean
medir. En la pantalla aparecerá el valor de resistencia de la
medición que se realiza en pantalla.
NOTA:
Al medir la resistencia, deberá apagar el circuito y descargar
completamente todos los capacitores antes de la prueba.

KING INNOVATION
I
13
AMPERÍMETRO DE PINZA EN MILIAMPERIOS
6-3 Medición de voltaje de corriente directa (VCD):
Gire el dial hasta el rango V. Conecte el conductor de prueba
rojo al terminal “+” y el conductor de prueba negro al terminal
“COM”. Coloque los cables de prueba de color rojo y negro
en los puntos positivo y negativo que se desean medir. En la
pantalla aparece el valor de la medición que se toma.
6-4 Pruebe la continuidad con el indicador acústico
(Buzzer):
Gire el dial al rango Ω. Conecte la punta de prueba
roja al terminal “+” y la punta de prueba negra al terminal
“COM”. Coloque los cables de prueba de color rojo y
negro en los puntos positivo y negativo que se desean medir.
El indicador acústico sonará cuando la resistencia es inferior
a 45Ω.
NOTA:
Al medir la continuidad, se DEBE apagar toda la energía al
circuito o al cable que se desea probar para evitar que se
lesione el usuario o se dañe la pinza amperimétrica.
Corto circuito Circuito abierto
Table of contents
Other King Innovation Measuring Instrument manuals