4
4.0 TEMPERATURE MEASUREMENT
4.1 Ensure that the thermometer probe is inserted to a preferred depth of
25mm into the medium and that contact is maintained until readings
become stable.
5.0 MALFUNCTION INDICATORS
5.1 ‘HI’ display: temperature measured is above range.
5.2 ‘Lo’ display: temperature measured is below range.
5.3 ‘Er2’ display: The thermometer is exposed to rapid
changes in the ambient temperature.
5.4 ‘Er3’ display: when the ambient temperature exceeds
0 ºC (32ºF) or + 50ºC (122ºF). The thermometer
should be allowed plenty of time (minimum 30 minutes)
to stabalise to the working temperature.
5.5 ‘Er’ for all other error messages it is necessary to reset
the thermometer. To reset, turn the instrument off,
remove the battery and wait for a minimum of one minute,
reinsert the battery and turn on.
5.6Contact our technical helpline if diagnosis is not clear.
6.0 BATTERIES
6.1 The lithium battery has an expected lifespan of up to five years
allowing for two hours continuous use each day.
6.2 The battery symbols indicate power status.
Battery needs
to be replaced
6.3 For best results, return instrument to Klipspringer Service Centre for
battery replacement and service check. NB Battery is welded into
thermometer circuitry and is not suitable for ‘diy’ replacement.