Replacing the Source and Detector in
the LI-8x0 Analyzers and LI-8100/A
This document describes how to replace the optical source
and detector on the LI-830, LI-850, and LI-870 gas analyzers,
legacy LI-800, LI-820, LI-840, and LI-840A gas analyzers, and
the LI-8100/A Automated Soil CO2Flux System.
Gas
Analyzer
Part Numbers
Source Detector 14 cm
Bench
5 cm
Bench
LI-850/870 800-902 800-918 800-904 n/a
LI-830 800-902 800-916 800-904 n/a
LI-840/A 800-902 800-906 800-904 n/a
LI-820 800-902 800-903 800-904 800-905
LI-800 800-902 800-903 800-904 800-905
LI-8100/A 8100-902 800-906 800-904 n/a
Contents
LI-800, LI-820, LI-840/A, LI-830, LI-850, and LI-870 1
LI-8100/A Source and Detector Replacement 3
LI-800, LI-820, LI-840/A, LI-830, LI-850, and
LI-870
The LI-800, LI-820, LI-840/A, LI-830, LI-850, and LI-870
optical source and detector may need to be replaced if one
fails. The life expectancy of the source is typically >20,000
hours, or >2 year of continuous operation, but this will vary
depending upon the number or warm-up cycles, vibration,
and other factors. Typically, detectors need to be replaced
rarely, if ever.
Disassembly
You'll need #1 or #2 Phillips screwdriver and a clippers or
utility knife. The installation, including calibration, may
take an experienced technician 30 minutes to 1 hour. If the
instrument has been running recently, the bench may be
hot to the touch. Let it cool for 5 or 10 minutes after power-
ing off if needed.
Note: Be sure that you are properly grounded to avoid elec-
trostatic discharge that can damage the electronics. Use an
anti-static wrist strap, electrostatic discharge grounding mat,
or occasionally touch bare metal that has a clear path to
ground, such as an unpainted computer case.
1Power off the gas analyzer and remove the top cover
(LI-800, LI-820, LI-830, LI-840/A and LI-850) or open
the case (LI-870 only).
The LI-800, LI-820, LI-830, LI-840/A and LI-850 top cover
is held in place by 6 screws. Loosen the screws and lift
off the cover. If there is a ribbon cable connecting the top
panel, be careful not to strain it. The LI-870 is opened by
removing two screws that hold the case closed; then
remove four screws that secure the optical bench cover.
2Unplug the source and detector ribbon cables from
the source and detector.
3Lift the optical bench out of the foam insulation.
Replacing the Optical Source
With the optical bench free of the case (except the tubing),
you can remove the source from the optical bench.
1Remove the tube from the source.
The simplest way is to cut the tubing as close to the hose
barb as possible.
1
Installation Guide
2Remove the four larger screws from the optical
source.
3Separate the source from the optical path.
Be sure that the inside of the optical path is clean—free
of dust and other contaminants. Clean it if needed, as
described in the instruction manual.
4Install the new O-ring that is included with the source.
5Install the replacement optical source.
Tighten the screws firmly.
6Press the air-in tube over the hosebarb—be sure it
covers both barbs on the hosebarb.
Removing a section of tubing will reduce the total
volume of the analyzer, and this may affect some char-
acteristics of the analyzer, such as the T90 specification,
but the reduction in volume is generally negligible.
There is enough slack in the original tubing for a single
replacement of the source, but if you replace it more
than once, you may need to replace the entire segment of
tubing. Should you choose to recalculate system volume
after removing a segment of tubing (which is not
required), the Bev-A-Line® has an inside diameter of 1/8"
(0.3175 cm). If you removed 1 cm (0.39 inches) of tube,
the volume removed is about 0.079 cm3(π × r2× length;
3.14 × (0.3175 cm / 2)2× 1 cm = 0.079 cm3).
7Work the optical bench into the foam.
Be sure that the bench is secured tightly in the foam insu-
lation and that no parts are in direct contact with the sur-
rounding metal cradle.
8Connect the ribbon cables that were removed pre-
viously.
9Reassemble the instrument.
Make sure that the foam insulation on the inside top
cover is positioned over the optical bench; it is required
for thermal stability. Reconnect the top cover ribbon
cable and then install the top cover.
10 In the LI-830, LI-850, and LI-870, reset the source
hours to "0".
The LI-830, LI-850, and LI-870 record the number or
operating hours for the optical source. You should reset
this count after replacing the source. Connect using the
software and then click Settings >Diagnostics and click the
gear button near Source, and then reset the hours.
11 Zero and span the analyzer as described in the instruc-
tion manual.
A simple zero and span will be adequate if the span gas is
close to the concentrations you will measure. A dual
span, however, will bring the instrument performance
within specifications across the full measurement range.
If you do not have zero and span gases, send the instru-
ment back to LI-COR for a factory recalibration.
Replacing the Optical Detector
You can also replace the optical detector.
1Remove the tube from the detector.
When replacing the detector on the LI-830, LI-850, or LI-
870, you will probably need to replace the entire segment
of urethane tubing, because cutting a portion to remove
it from the hose barb will not leave enough tubing to re-
connect using the original segment. A replacement seg-
ment of urethane tubing is included in the detector
replacement kit for this purpose. On the LI-800, LI-820,
or LI-840/A, the tubing is long enough to reconnect.
2
2Remove the four screws that secure the optical
detector and shield.
On the LI-840/A, there are four spacers between the
shield and detector. Collect the four spacers.
3Separate the detector from the optical path.
Be sure that the inside of the optical path is clean. It
should be free of dust and other contaminants. Clean it if
needed, as described in the instruction manual.
4Install the new O-ring that is included with the
detector.
5Install the replacement optical detector.
On the LI-840/A, place the four spacers on the screws so
they fit between the shield and detector circuit board.
Tighten the screws firmly.
6Press the tube over the hosebarb—be sure it covers
both barbs on the hosebarb.
7Work the optical bench into the foam.
8Connect the wiring harnesses to the detector and
source.
9Reassemble the instrument.
Make sure that the foam insulation on the inside top
cover is positioned over the optical bench; it is required
for thermal stability. Reconnect the top cover ribbon
cable and then install the top cover.
Important: LI-840's that are running older software
must be updated to the latest LI-840A embedded soft-
ware, which is available for free from
licor.com/env/support/LI-840A/software.html.
10 Zero and span the analyzer as described in the instruc-
tion manual.
A simple zero and span will be adequate if the span gas is
close to the concentrations you will measure. A dual
span, however, will bring the instrument performance
within specifications across the full measurement range.
If you do not have zero and span gases, send the instru-
ment back to LI-COR for a factory recalibration.
LI-8100/A Source and Detector Replacement
The LI-8100A Soil Gas Flux System the source and detector
on this optical bench can be replaced if they fail. These
instructions describe how to change the source, detector
and/or optical bench. A Phillips head screwdriver and a 1/2”
open end wrench are the only tools needed to replace the
source and/or detector in all instruments. To replace the
source or detector:
Note: Be sure that you are properly grounded to avoid elec-
trostatic discharge that can damage the electronics. Use an
anti-static wrist strap, electrostatic discharge grounding
mat, or occasionally touch bare metal that has a clear path
to ground, such as an unpainted computer case.
1Power the instrument off and remove the battery or
other external power source.
Allow the instrument optical bench to cool for about 10
minutes.
2Open the Analyzer Control Unit and remove the
access panel by loosening the 4 thumbscrews.
3Remove both air filters by pressing the orange part of
the quick-connect fittings toward the white part of
the connector and pulling the filter out.
Discard the filters. There are replacement filters included
that can be used during reassembly.
4Disconnect the 3 electrical connectors.
3
Two of them are on a single wiring harness, just to the
right of filter connectors; the third is at the left end of the
bundle of blue cables. Pull straight out on the connectors
to remove them.
5Remove the 2 screws from the top of the air manifold.
The top half of the manifold rests on two pins; lift the
top half until it is completely separated from the bottom
half. It is not necessary to remove any of the hoses.
6Remove the 12 screws around the outer edge of the
white panel.
7Lift the assembly out of the yellow case and turn it
upside down.
8Remove the 6 screws from the optical bench cover.
9Leaving the hoses connected to the optical bench, lift
the bench out of the foam casing.
10 There are four screws on the source and detector
housings that must be removed.
Remove only the screws from the assembly you are repla-
cing. Remove the four screws in the corners of the
source housing circuit board. Do not remove the remain-
ing four screws. Remove the four screws from the
detector housing cover. Note that there are some small
standoffs inserted over the screws behind the cover; tilt
the housing down when removing the cover so these
standoffs don’t get lost.
4
11 Disconnect the ribbon cable connectors attached to
the source and/or detector housing circuit board.
12 Remove the Bev-A-Line® tubing connected to the
source and/ or detector housing at the quick-connect
fitting.
Press down on the orange part of the fitting to remove
the tubing.
13 The source and/or detector housing, with the
attached circuit board and tubing can now be
removed and discarded.
Note that there are O-rings on each end of the optical
bench. It is a good practice to replace the O-ring when
replacing the source. Make sure the optical path tube is
clean (no dust, dirt, etc.) before proceeding. Clean it if
needed, as described in the instruction manual.
14 Re-assemble the bench using the replacement source
and/or detector.
Be sure the O-rings are in place on both ends of the
bench. Note that the orientation of the optical bench cyl-
inder is not important; either end can be inserted into
the source or detector housing.
15 Replace the tubing.
Any tubing that you removed should be replaced with a
piece of equal length, which can be cut from the tubing
that is included in the instrument spares kit.
16 Re-assemble the LI-8100/A case.
Make sure that the foam insulation on the inside top
cover is positioned over the optical bench; it is required
for thermal stability. The remaining assembly procedures
are the opposite of the disassembly.
17 Perform zero and span calibrations as described in the
LI-8100/A Instruction Manual.
A simple zero and span will be adequate if the span gas is
close to the concentrations you will measure. A dual
span, however, will bring the instrument performance
within specifications across the full measurement range.
If you do not have zero and span gases, send the instru-
ment back to LI-COR for a factory recalibration.
5
Copyright © 2019 LI-COR, Inc. All rights reserved.
984-17308•10/19
LI-COR Biosciences
4647 Superior Street
Lincoln, Nebraska 68504
Phone: +1-402-467-3576
Toll free: 800-447-3576 (U.S. and Canada)
envsales@licor.com
LI-COR Distributor Network:
www.licor.com/env/distributors
Regional Offices
LI-COR Biosciences GmbH
Siemensstraße 25A
61352 Bad Homburg
Germany
Phone: +49 (0) 6172 17 17 771
envsales-gmbh@licor.com
LI-COR Biosciences UK Ltd.
St. John’s Innovation Centre
Cowley Road
Cambridge
CB4 0WS
United Kingdom
Phone: +44 (0) 1223 422102
envsales-UK@licor.com

This manual suits for next models

6

Other LI-COR Measuring Instrument manuals

LI-COR LI-6800 User manual

LI-COR

LI-COR LI-6800 User manual

LI-COR LI-830 User manual

LI-COR

LI-COR LI-830 User manual

LI-COR LI-200R User manual

LI-COR

LI-COR LI-200R User manual

LI-COR LI-3100C User manual

LI-COR

LI-COR LI-3100C User manual

LI-COR LI-600 User manual

LI-COR

LI-COR LI-600 User manual

LI-COR LI-200R User manual

LI-COR

LI-COR LI-200R User manual

LI-COR LI-600 User manual

LI-COR

LI-COR LI-600 User manual

LI-COR li-6262 User manual

LI-COR

LI-COR li-6262 User manual

LI-COR LI-800 Guide

LI-COR

LI-COR LI-800 Guide

LI-COR LI-250A User manual

LI-COR

LI-COR LI-250A User manual

LI-COR Li-189 User manual

LI-COR

LI-COR Li-189 User manual

LI-COR LI-800 User manual

LI-COR

LI-COR LI-800 User manual

LI-COR LI-180 User manual

LI-COR

LI-COR LI-180 User manual

LI-COR LI-7700 User manual

LI-COR

LI-COR LI-7700 User manual

LI-COR LI-8100A User manual

LI-COR

LI-COR LI-8100A User manual

LI-COR LI-200R User manual

LI-COR

LI-COR LI-200R User manual

LI-COR LI-600N User manual

LI-COR

LI-COR LI-600N User manual

LI-COR LI-8100A User manual

LI-COR

LI-COR LI-8100A User manual

LI-COR LI-600 User manual

LI-COR

LI-COR LI-600 User manual

LI-COR LI7200RS User manual

LI-COR

LI-COR LI7200RS User manual

LI-COR Eddy Covariance User manual

LI-COR

LI-COR Eddy Covariance User manual

LI-COR LI-870 Guide

LI-COR

LI-COR LI-870 Guide

LI-COR LI-3000C User manual

LI-COR

LI-COR LI-3000C User manual

LI-COR LI-8150 Multiplexer Installation and operating instructions

LI-COR

LI-COR LI-8150 Multiplexer Installation and operating instructions

Popular Measuring Instrument manuals by other brands

Blue-White ProSeries-M MS-6 Series Installation and operating manual

Blue-White

Blue-White ProSeries-M MS-6 Series Installation and operating manual

PeakTech 5165 Operation manual

PeakTech

PeakTech 5165 Operation manual

TPI 597 quick start guide

TPI

TPI 597 quick start guide

Tektronix WVR7200 Installation and safety manual

Tektronix

Tektronix WVR7200 Installation and safety manual

Joulescope JS220 user guide

Joulescope

Joulescope JS220 user guide

Anritsu Company VectorStar ME7838A4 quick start guide

Anritsu Company

Anritsu Company VectorStar ME7838A4 quick start guide

GOWIN SUG100 user guide

GOWIN

GOWIN SUG100 user guide

THORLABS Blueline Series Operation manual

THORLABS

THORLABS Blueline Series Operation manual

Tektronix TLA500 Series Technical reference

Tektronix

Tektronix TLA500 Series Technical reference

AEMC instruments SR701 user manual

AEMC instruments

AEMC instruments SR701 user manual

CAI 700 CLD Series Operator's manual

CAI

CAI 700 CLD Series Operator's manual

Rohde & Schwarz ZNL4 user manual

Rohde & Schwarz

Rohde & Schwarz ZNL4 user manual

Tegam 253 instruction manual

Tegam

Tegam 253 instruction manual

Badger Meter 2100R+ manual

Badger Meter

Badger Meter 2100R+ manual

Algodue ELETTRONICA UPM3100 Use and Programming Manual

Algodue ELETTRONICA

Algodue ELETTRONICA UPM3100 Use and Programming Manual

Topcon FC-2600 Operator's manual

Topcon

Topcon FC-2600 Operator's manual

Keller LEO 2 manual

Keller

Keller LEO 2 manual

ROOTECH Accura MD-GAS user guide

ROOTECH

ROOTECH Accura MD-GAS user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.