Light Progress UV-STICK E2 User manual

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 1/16
UV-STICK E2/75H-ST-NX-NT
Germicidal UV-C Device for Disinfection by Direct Irradiation
Portable on wheels
MODELS:
UV-STICK E2/75H-ST-NX-NT
UV-STICK E2/75H-P-ST-NX -NT

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 2/16
SECURITY WARNING
Our Company has ensured that every care is taken in the design and production of its machinery. In order to guar-
antee, as far as possible, that it is safe when used correctly. However, the laws on accident prevention require all users
of the machinery to ensure that whoever is responsible for installation, maintenance, use and sale of these products is
shown the instruction provided by our company and is made aware of the precautions to be taken.
According to the circumstances, the appropriate recommendations made in these instructions must be made available,
together with any other relative information, to any company or person involved in the installation of LIGHT PRO-
GRESS products.
Only use this appliance for its intended purpose as described in this booklet.
If the appliance is wrongly operated for a purpose other than that for which it is intended, no liability can be accept-
ed for any possible damage. The manufacturer cannot be responsible for any damage to people, animals and objects
caused by use or operation of the appliance contrary to these instructions.
As with all electrical appliances the safety information and precautions contained in this booklet must be carefully
observed, including following:
Before maintenance or cleaning ensure that the appliance is switched off and unplugged from the main supply.
Should the appliance become faulty please contact our authorized Service Centre direct and in case of repairs ask
for genuine parts.
Improper repairs may damage the appliance and place the user at serious risk.
WARNING
Avoid exposition to UV-C rays emitted by germicidal lamps,
even for few seconds, it may cause severe conjunctivitis and
erythema
Plastic or painted surfaces exposed to direct UV-C rays may
progressively yellow, as after a long exposure to sunlight.
The intended use is as a germicidal lamp for sterilization, any
other use is improper and dangerous
The lamp, as supplied, can not be modified or designed for use
with accessories or tools
We disclaim all responsibilities and void the warranty in case of
tampering or lack of maintenance
• Read carefully the following instruction before use
• In case of opening of the UV section, must be provided a shut-off of the UVC de-
vices in order to avoid exposure to UVC light emitted by the germicidal lamp; It
may cause severe conjunctivitis and erythema.
• Remove the protection film from the device before light-on the UV-C lamp.
• This device can be used by children with no less than 8 years and by persons with
reduced physical, sensory or metal capacities or with no experience or knowledge,
as long as under supervision or often they have receiv3ed the related instructions
about the safe use of the device and the related risks. Kids do not have to play
with this device.

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 3/16
CONVENTIONAL SYMBOLS NOTICED ON DEVICES
SUBSTITUTED ITEMS MUST BE PERFORMED RESPECTING THE LOCAL LAWS.
RECYCLE THE STUFFS THAT COMPOSES THIS PRODUCT.
OBLIGATION TO CONSULT THESE INSTRUCTIONS BEFORE USE THIS PRODUCT
DANGER OF ELECTRICAL SHOCK; REMOVE TENSION TO THE SYSTEM BEFORE PERFORMING
ANY OPERATION.
GENERIC DANGER.
DANGER OF EXPOSITION TO ULTRAVIOLET UV-C RAYS.
THIS DEVICE RESPECTS THE ROHS NORMS.
USE PROTECTIVE GLOVES
WEAR PROTECTIVE GLASSES
STORAGE OF THIS MANUAL
ATTENTION!
This instruction manual must be preserved and must always be kept available for consultation; it must follow
the device in case of transfer of ownership, and accompany the device until its demolition.
To request the updated instruction manual and to ask for all other information needed, it’s possible to contact Light
Progress’ technical office.
DEMOLITION AND DISPOSAL
Do not dispose of these products as unsorted municipal waste, but collect them separately as a special
waste and entrust them to specialized waste disposal companies. Disposal should be in accordance with
Legislative Decree No. 151/2005 (WEEE - Italy), or the law of the country where disposal will occur
(WEEE - EC). In case of failure to comply with laws or illegal disposal heavy penalties are provided. The
apparatus consists of plastic parts, electronic components, metal, UV lamps containing mercury (sub-
stances dangerous for the environment and human health). The dealer is available to the withdrawal of the apparatus
used.

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 4/16
DEVICE DESCRIPTION
UV STICK E2/75H-ST-NX-NT is a two lamps germi-
cidal device , placed on a stand, with four wheels for
easy mobility. It works on direct radiation at 360 de-
grees and sterilizes irradiated air and surfaces. The
UV-C lamps, with emission peak at 235.7 nm. (na-
nometers), have a strong germicidal power against all
micro-organisms (molds, bacteria and viruses).
All equipment is composed of (picture 1):
A) Metal stand: painted steel base plate, with 4 swiv-
eling wheels.
B) Control Unit: stainless steel (AISI 304), equipped
with bipolar magneto-thermal switch, two rotary knobs
for ignition delay and treatment duration setting,
start/stop button, hour meter for UV lamps replace-
ment.
C) UV unit: with 2 UV-lamps, protected by stainless
steel grid (AISI 304). The body is made of stainless
steel with mirror polished aluminium reflector, which
increases performance reflecting a higher percentage
of UV-C rays.
APPLICATIONS AND RESULTS
Each UV-STICK series can be installed in any type of
environment to achieve a deep disinfection of air and
surfaces, even those difficult to be reached by tradi-
tional systems.
When turned on, it gets a strong reduction of microbes
in the air and on surfaces. For example, in 8 minutes at
a distance of 3 meters, it achieves a 99% reduction of
bacteria such as: Bacillus coli, Clostridium, Legionella,
Vibrio, Salmonella, Pseudomonas, Staphylococcus,
Streptococcus, etc..
Thanks to natural air circulation, even microbes located
in hidden areas are continually drawn towards the irra-
diated area, so that their overall level progressively de-
creases.
The use is permitted only in absence of staff or if radia-
tion is adequately shielded.
The device can operate during personnel break from
work. Timer can be used to manage the switching off
immediately before the start of work, so that workers
can have perfectly disinfected room.
Switching on and off is triggered by dedicated control unit, equipped with two timers to manage ignition delay and
treatment duration.
Pic.1

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 5/16
BENEFITS AND ADVANTAGES
a) DEEP AND CONTINUOUS DISINFECTION
This device can be switched on continuously without people being present (i.e. during the night for 2-3 hours). The dis-
infection of air, machinery, and everything that is contained inside the room allows beginning the work day in ideal hy-
gienic conditions. In this way, the level of microbial load in one site is maintained constantly low.
b) PHYSICAL ACTION AND ECOLOGICAL PROTECTION
Treatment with UV-C rays is purely physical, always allowing treatment with the same efficiency; there is no danger of
overdosing. In contrast, many chemical treatment methods involve the use of products that are risky and difficult to bi-
odegrade, with consequent hazards to human health; besides, the use of chemical products, in contrast to UV-C rays,
could develop resistant microbial forms.
c) PRACTICALITY AND ECONOMY
Treatment is immediate and ready to use. Maintenance is minimal with low costs of both energy consumption and
maintenance.
d) IMMEDIATE EFFECT
Effective treatment (99.9% bacteria reduction) takes just a few minutes.
e) NO CHEMICAL/PHYSICAL CHANGES
Treatment is cold and dry; it causes no organoleptic or qualitative changes in the food.
f) NO DANGER OF GLASS DISPERSION
With special UVLON® protection, there is no danger of glass fragments dispersion resulting from any breakage of UV-
C lamps.
DEVICE CODES

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 6/16
RECEPTION AND ASSEMBLY
Handle with care as fragile material. Verify the integrity of the packaging, and please be careful not to damage the con-
tents when using sharp blades or cutters to open the box.
Remove the unit from the box, then inspect it immediately to ensure that no damage has occurred during transport.
Before installation and use of the device please READ CAREFULLY THE SECURITY WARNINGS and all others in-
structions that follows.
For packaging and transport reasons, the entire unit
comes in two parts: metal stand with wheels, and UV
unit . It is necessary to assemble the UV unit to the
base with the supplied screw kit (picture 2) .
We recommend to follow the following assembly
procedure:
- Place the base plate vertically, like shown on
picture 3, with semicircular side upwards. Align
the top hole with the corresponding hole on the
unit, insert the hex head screw then tighten by
hand or with the aid of a key. Repeat with the
bottom hole (picture 4).
- On the other side of the plate, insert the two hexagonal socket head cap screws into the holes at the bottom of
the control unit (picture 5), then tighten it with the self-locking nuts (picture. 6).
- Firmly tighten the four screws. It is now possible to place the device on its wheels in vertical position, and to
connect the power cable to the dedicated socket on the back of the control unit.
Pic. 2
Pic.3
Pic.5
Pic.4
Pic.6

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 7/16
INSTALLATION
Take the device to the room to be disinfected, and place it almost at central position. In case of more than one device,
you should place them so that UV irradiation is uniformly distributed, avoid creating shaded areas.
The area/space covered by the UV-STICK devices is shown in Table 1. Final disinfection results are in any case de-
pendent on the amount of germs coming in from the external environment.
WARNING !!!
Remeber to switch on the device ONLY IF there is a concrete shield and no accidental ex-
position to the rays by the personnel.
To avoid exposition to UV-C rays please provide a total shielding of the modules, or automatic turning OFF devices at
the entry doors in the environment treated. There must be no direct irradiation of the personnel.
To shield the devices, materials with opaque surfaces such as stainless steel or aluminium can be used, but also
transparent materials such as glass or Lexan®; to have further information please contact us.
To improve safety we suggest you to add light signs or signboards to report the potential danger.
SCHEDULING
UV-STICK -E2/75H-ST-NX-NT
With timer for ignition delay and treatment duration setting.
Insert the plug into the wall socket and turn the device on using the switch on the back of the control unit. The green
control light of the switch will confirm the activation of main power source.
It is possible to set treatment duration and ignition delay, for taking time to comfortably step back from the device and
avoid accidental expositions to UV-C rays.
The ranges of both adjustments can vary from a minimum of a second to a maximum of 50 hours.
Two settings must be made:
IGNITION DELAY DURATION (before steriliza-
tion starts)
TREATMENT DURATION
Proceed with the following steps: (see picture 7)
1) First of all, desired range (scale) of the two
chronometers must be set. To set the range of the
chronometers, act on the Phillips screws located
at the bottom left on the chronometer. For ignition
delay , act on screw 1 (Delay/Ritardo).
For treatment duration, act on screw 2 (Timer).
2) The desired durations can now be set acting
on the setting head of each chronometer. To set
ignition delay duration, act on left setting head
(Delay- Ritardo).
To set treatment duration, act on right setting
head (Timer). We recommend to choose a value
between 20 min. and 2-3 hours.
3) Once setting is completed, push START/STOP
button to begin the program. The button will light
up and the program will start.
4) After the end of the cycle press START/STOP
again (the light will turn off ) to finish the opera-
tion.
1
2
Fig 7

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 8/16
ELECTRICAL CONNECTIONS
We suggest you to let a qualified electrician install this device, following these instructions and the electrical scheme,
as well as CEI norms.
Before connecting the device to the electrical line, make sure that the labelled data on the device are the same as
the ones of the electrical line; we remind you that the electrical supply power of this device is 220-240V, 50/60Hz.
Make sure that the electrical system is capable of providing the supply current required by the device(please read
the label).
The connection to the electrical line needs a multipolar switch.
The connection to the electrical line must be made by the supplied cable.
The use of junction devices in which connection is made via screws, nuts, or other means, is necessary to connect
the device to the electrical line.
Protect the power supply cable from high temperature, sharp corners, water, oils.
MAINTENANCE
UV LAMP
Lamp maintenance is very important because UV-C light emit-
ted by the source is not visible to the eye, so, although you
see the lamp ON, its germicidal effect may not be sufficient to
performs disinfection treatment.
Before reaching the lamp, ALWAYS switch off power supply
If UV-C lamp gets dirty (or gets dusty) clean it with a clean cloth
and alcohol, and avoid touching the glass part with hands. To check
UVC lamp life number of hours, correct product code of the spare
UV lamp please refer to table 1. The lamp, once reached the timing
for replacement, must be replaced, this operation is carried out as
follows:
• Loosen the nuts on the lamp housing and let them slide
along the lamp; grab the lamp at the 2 ends and pull it in a
direction parallel to the lamp.
• Repeat the steps in reverse to insert the new lamp.
TROUBLESHOOTING
UV LAMP
See MAINTENANCE.
BALLAST
All UV-DIRECT models are supplied with electronic ballasts which guarantees an emission of 15% more if compared
with traditional ballasts. Furthermore, lamps are not influenced by environment temperature that can goes until -20°C.
Finally, the switching on is immediate, with no starter, and for this reason lamps last longer.
To preserve the original features of original IP protection and isolation class, and to maintain all the characteristics that
guarantee a long lasting life of this product, the ballast replacement must be done by qualified technicians verified by
Light Progress. In case of necessity, please, send the device back to our facility for any kind of repairs.
If this is not possible, you can order the entire power supply conduit for replacement (see Table 1 for code).
USE
PROTECTIVE
GLOVES
WEAR
PROTECTIVE
GLASSES

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 9/16
TECHNICAL FEATURES
TABELLA n°1
UV-STICK
E2/75H-ST-NX-NT
AVERAGE LAMP LIFETIME (h)*
≤ 18.000
POWER (W)
2x75
DIMENSIONS LxSxH (mm.)
1420 x 404 x 336
WEIGHT (Kg.)
7
TREATED AIR VOLUME (m3)
50 to 100
TREATED SURFACE (m2)
15 to 30
SPARE PARTS
UV LAMP Code
n°2 CHS-90WH
POWER SUPPLY CONDUIT Code (complete)
n°1 CE-2/75H
* continuous operation
• Selective UV-C lamp (at 253.7 nm.) with high efficiency lighting.
• Stainless steel body AISI 304 .
• All materials used are tested for resistance to intense UV-C rays.
• Waterproof and dust-proof (IP 55).
• Control Console (IP 20) grounded.
• Powered with electronic ballasts specific for UV-C rays lamps
• Reflector in extremely pure mirror bright aluminum.
• Direct protection of the lamp with a stainless steel grid
• CE mark (LVD - EMC - MD - RoHS).
• Noise: under 70 dB(A).
• Vibration: under 0.5 m/s2.
LABEL AND SYMBOLS
SYMBOL
MEANING
Warning! Check the at-
tached documentation
GROUND TERMINAL
(The unit must always be connected to the
ground)
Device with CE certifica-
tion.

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 10/16
WARRANTY
All products made by LIGHT PROGRESS are manufactured with the highest quality materials only, they are subjected
to rigorous testing, and are guaranteed to be free of factory faults according to current law and regulations.
Minimum requirements for warranty validity:
Warranty will be considered valid only in case of possession of the purchase document.
Warranty limits:
Warranty excludes:
Damages caused by not following the instructions that accompany the product at purchase.
Periodical controls, maintenance, repairs or replacements of parts due to normal wear.
Consumable parts.
Use of incompatible spare parts or consumables; Compatibility is ensured only by use of products pur-
chased from or recommended by Light Progress.
Damages caused by modification of device / system without Light Progress’ previous consent.
Improper use of the product, falling or crashes.
Modifications or repairs made by unauthorized personnel.
For defective materials return, you must request the return merchandise authorization (RMA) sending an email to in-
[email protected] or calling the number +39.0575.749255.
Once you receive the module, it has to be filled up in every part and sent via email to [email protected] or via fax to
+39.0575 789929 to obtain the RMA number that will authorize the shipments of the goods to Light Progress ware-
house. Only authorized goods will be accepted, otherwise it will be refused and sent back to the sender on ex-works
conditions. The packages must be sent on DDP (Delivery Duty Paid) with packaging suitable for transport. RMA num-
ber is valid up to 10 working days from release date.
MANUFACTURER RESPONSABILITY AND COMMITMENTS
Light Progress do not take any responsibility for any damage to persons or property, due to failure to follow these
instructions. We recommend you to carefully read all the given instructions.
The manufacturer agrees with the buyer to provide, upon request, schemes of circuits, component parts lists, instruc-
tions for calibration of lamps and other information, which are useful to repair those parts that the manufacturer consid-
ers repairable.
The company reserves itself the right to make changes without prior notification or public notice.

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 11/16
DECLARATION OF COMPLIANCE
We, LIGHT PROGRESS s.r.l., hereby declare under our own responsibility that the following units of own production:
UV-STICK E2/75H-ST-NX-NT series
are in accordance with EEC directive 2014/30/EU (Electromagnetic Compatibility)
are in accordance with EEC Machinery Directive dispositions 2006/42/EU
are in accordance with EEC Low Voltage Directive 2014/35/EU
are in accordance with EEC (RoHS) 2011/65/EU (D.Lgs 4 Marzo 2014, n.27)
TECHNICAL STANDARDS APPLIED
UNI EN ISO 12100:2010
Safety of Machinery - Basic Concepts, General Principles for Design - Risk
assessment and risk reduction
UNI EN ISO 13857:2008
Safety of Machinery - Safety Distances to prevent danger zones being reached
by the upper and lower limbs (2008)
ISO 14120:2015
Safety of Machinery - Guards - General Requirements for the Design and
construction of fixed and movable guards
UNI EN ISO 13849-1:2016
Safety of Machinery - Parts of the Control System related to the Safety – Part 1:
General Design Principles
UNI EN ISO 14119:2013
Safety of Machinery - Interlocking devices associated with guards - Principles for
design and selection
CEI EN 60204-1/2016
Safety of Machinery - Electrical Equipment of Machines. Part 1: General Rules
(2010)
EN 61439-1:2011
Low-voltage Switchgear and Control Gear Assemblies. Part 1: General rules
FURTHER TECHNICAL STANDARDS APPLIED:
IEC EN 60335-1 "Safety of household electrical appliances and similar”
Electronic Ballasts for the control of the lamps in accordance with the standard EN 60928.
Germicidal UV-C Lamps in accordance with EN 61199.
Electrical Protective Measures in accordance with IEC 70-1, EN 60229.
Anghiari, 2.7.2019
Responsible for Standards: Dr. Aldo Santi
LIGHT PROGRESS S.r.l. Loc. San Lorenzo, 40 - 52031 ANGHIARI (AR) - ITALY - http://www.lightprogress.com

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 12/16
Circuit scheme: UV-STICK E2/75H-ST-NX-NT

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 13/16
Notes:
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 14/16
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________

LIGHT PROGRESS STICK-ST [eng]
feb-2019 Pag. 15/16
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________
____________________________________________________