3736
>OR FILL IN THE FORM AND RETURN TO LPG SYSTEMS®
7
WARRANTYWARRANTY
Name: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Address: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Email: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HUBER 360® EVOLUTION serial number: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Date: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Through the form above, LPG SYSTEMS, which is the data controller, operates a processing
of your personal data, based on the sales agreement of your HUBER 360®EVOLUTION
device, in order to manage the guarantees that apply to it. LPG SYSTEMS only stores your
data for the time required by the operations for which it was collected and in compliance
with the regulations in force.
In accordance with the applicable regulations, you have the right to access, rectify and
delete your personal data, a right of limitation or opposition to the treatment for legitimate
reasons, a right of define general and specific guidelines on how you intend to exercise
your rights after your death, the right to portability of your data, and the right to lodge a
complaint with a supervisory authority .
You can exercise your rights by email at dataprivacy@lpgsystems.com or by mail at: LPG
SYSTEMS, Customer Service Department, Technoparc de la Plaine, 30 rue Dr Abel, CS
90035 - 26000 VALENCE (FRANCE), with a photocopy of an identity document bearing your
signature.
APPENDIX: ELECTROMAGNETIC COMPATIBILITY
8
>APPENDIX: ELECTROMAGNETIC COMPATIBILITY
TABLE 1: DIRECTIVES AND MANUFACTURER DECLARATION – ELECTROMAGNETIC EMISSIONS
HUBER 360®EVOLUTION is intended for use in the electromagnetic environment specied below. HUBER 360®EVOLUTION customer or patient
should ensure that it is used in such an environment.
Emissions test Conformity Electromagnetic Environment - Directives
RF emissions CISPR 11 Group 1 HUBER 360®EVOLUTION uses RF energy only for its internal
functions. Therefore, its RF emissions are very low and unlikely
to cause interference in nearby electronic devices.
RF emissions CISPR 11 Class B HUBER 360®EVOLUTION may be used in all establishments,
including domestic sites and sites that are directly connected
to the low voltage public power grid, which supplies domestic
buildings.
Harmonic emissions IEC 61000-3-2 Class A
Voltage uctuations and icker IEC 61000-3-3 Conforms
TABLE 2: DIRECTIVES AND MANUFACTURER DECLARATION – ELECTROMAGNETIC IMMUNITY – EMISSIONS TESTS
HUBER 360®EVOLUTION is intended for use in the electromagnetic environment specied below. HUBER 360®EVOLUTION customer or patient
should ensure that it is used in such an environment.
Emissions test Conformity Conformity Electromagnetic Environment - Directives
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV on contact
± 8 kV in the air
± 6 kV on contact
± 8 kV in the air
Flooring should be made of wood, concrete, or ceramic tile.
If the oor is covered with synthetic material, relative humidity
should be at least 30%.
Fast transient/burst
IEC 61000-4-4
± 2 kV for electrical
power lines
± 1 kV for input/output
lines
± 2 kV for electrical
power lines
± 1 kV for input/
output lines
The quality of the electrical power network should be equivalent
to that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV between
phases
± 2 kV between phase
and earth
± 1 kV between
phases
± 2 kV between phase
and earth
The quality of the electrical power network should be equivalent to
that of a typical commercial or hospital environment.
Voltage dips, short
interruptions, and
voltage variations
IEC 61000-4-11
<5% UT
(>95% voltage dip)
during 0.5 cycle
40% UT
(60% UT voltage dip)
during 5 cycles
70% UT
(30 % UT voltage dip)
during 25 cycles
<5% UT
(>95% voltage dip)
during 5s
<5% UT
(>95% voltage dip)
during 0.5 cycle
40 % UT
(60% UT voltage dip)
during 5 cycles
70% UT
(30 % UT voltage dip)
during 25 cycles
<5% UT
(>95% voltage dip)
during 5s
The quality of the electrical power network should be equivalent to
that of a typical commercial or hospital environment.
If HUBER 360®EVOLUTION patient requires operation to continue
during power outages, HUBER 360®EVOLUTION should be powered
through an uninterruptible power supply or a battery.
Power frequency
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should have levels that are
characteristic of a typical hospital or commercial environment.
NOTE: UT is the AC voltage prior to the application of the test level.