GF13702: ADVANCED PATIENT ALARM
SETUP & OPERATION INSTRUCTIONS
PLEASE SAVE THESE INSTRUCTIONS FOR FUTURE USE
GF13702-INS-LAB-RevB14 1
www.grahamfield.com
GF Health Products, Inc. is not responsible for typographical errors. Packaging, warranties, products and specifications
are subject to change without notice. Graham-Field and Lumex are registered trademarks of GF Health Products, Inc.
Made in China
© 2014 GF Health Products, Inc. • 2935 Northeast Parkway • Atlanta, GA 30360
tel 800-347-5678 • fax 800-721-0601
ONE-YEAR LIMITED WARRANTY
SCOPE OF WARRANTY
GF Health Products, Inc. (“GF”) warrants to the original purchaser only that it will replace or repair components, at GF’s sole discretion,
that are defective in material or workmanship under normal use and service. All warranties are conditioned upon the proper use of
the products strictly in accordance with good commercial practice and applicable GF instructions and manuals, including proper use
and maintenance. To the extent that a component is warranted by a third party, GF conveys all of its rights under that warranty to the
original purchaser, to the extent permitted. This limited warranty shall only apply to defects that are reported to GF’s customer service
team within the applicable warranty period and which, upon examination by GF or its authorized representative, prove to be a warranty
item. This limited warranty is not transferable. The warranted components and time period are set forth below:
GF13702: Advanced Patient Alarm
The applicable warranty period shall commence from date of shipment to the original customer, unless there is an expiration date on
the component in which case the warranty shall expire on the earlier of warranty period or the expiration date.
OBTAINING WARRANTY SERVICE
A GF Customer Service Representative must authorize warranty service. Please contact the GF Customer Service department by
calling 678-291- 3207, sending a fax request to 770-368-2386 or by e-mailing a request to cs@grahameld.com. Specic directions will be
provided by the Customer Service Representative. Failure to abide by the specic directions will result in denial of the warranty claim.
EXCLUSIONS
The warranty does not cover and GF shall not be liable for the following:
1)Defects, damage, or other conditions caused, in whole or in part, by misuse, abuse, negligence, alteration, accident, freight damage,
tampering or failure to seek and obtain repair or replacement in a timely manner;
2)Products which are not installed, used, or properly cleaned and maintained as required in the ofcial manual for the applicable product;
3)Products considered to be of a non-durable nature including, but not limited to: casters, lters, fuses, gaskets, lubricants, and charts;
4)Accessories or parts not provided by GF;
5)Charges by anyone for adjustments, repairs, replacement parts, installation or other work performed upon or in connection with such
products which are not expressly authorized in writing, in advance, by GF;
6)Any labor or shipping charges incurred in the replacement part installation or repair;
7)Costs and expenses of regular maintenance and cleaning; and
8)Representations and warranties made by any person or entity other than GF.
ENTIRE WARRANTY, EXCLUSIVE REMEDY AND CONSEQUENTIAL DAMAGES DISCLAIMER
THIS WARRANTY IS GF’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. GF MAKES NO IMPLIED
WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
IF ANY MODEL OR SAMPLE WAS SHOWN TO THE CUSTOMER, SUCH MODEL OR SAMPLE WAS USED MERELY TO ILLUSTRATE THE
GENERAL TYPE AND QUALITY OF THE PRODUCT AND NOT TO REPRESENT THAT THE PRODUCT WOULD NECESSARILY CONFORM TO
THE MODEL OR SAMPLE IN ALL RESPECTS.
THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF THE DEFECTIVE PARTS. GF SHALL NOT BE LIABLE FOR AND
HEREBY DISCLAIMS ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT
NOT LIMITED TO: DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE, DOWNTIME, COVER, OR EMPLOYEE OR INDEPENDENT
CONTRACTOR WAGES, PAYMENTS AND BENEFITS.
The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document,
and supersede all prior negotiations, agreements and understandings with respect thereto. The recipient of this document hereby
acknowledges and represents that it has not relied on any representation, assertion, guarantee, warranty, collateral contract or other
assurance, except those set out in this document.
For additional information on this product or this warranty, please contact a GF Customer Service Representative.
NOTES:
1)Additional terms and conditions may apply.
2)Freight claims must be notated on the Bill of Lading and must be made with immediacy. The ICC regulations govern specic
requirements for freight claims. Failure to abide by those regulations may result in a denial of the freight claim. GF will assist you in
ling the freight claim.
3)Claims for any short shipment must be made within thirty (30) days of the invoice date.
GF13702-INS-LAB-RevB14 4
Note: The most current version of these instructions can be found online at www.grahamfield.com
IMPORTANT SAFETY INFORMATION: PLEASE READ
WARNING:
Important! Read and understand these instructions before installing or using the
GF13702 Advanced Patient Alarm Setup. If you do not understand any part of these instructions,
contact a healthcare professional for direction in the use of this product. If the patient alarm is not
properly set up and adjusted, personal injury and/or damage to the patient alarm could result.
WARNING: If components are damaged or missing, contact your dealer immediately. DO NOT use
substitute parts. Use only Lumex®replacement parts. Use ONLY the included magnet and pull-cord
with this alarm; DO NOT use other types or lengths of cord. Non-Lumex®replacement parts could
cause personal injury and/or damage to patient alarm.
Notice for California Customers-
California Proposition 65 WARNING: This product contains a chemical known to the State of
California to cause cancer and reproductive or developmental harm.
WARNING: GF Health Products, Inc. assumes no responsibility for any damage or injury caused by
improper assembly or use of this product.
UNPACKING
Check for any obvious damage to the carton or its contents. If damage is evident, DO NOT USE. Contact
carrier / distributor for further instruction.
INTENDED USE
The intended use of the GF13702 Advanced Patient Alarm is to alert caregivers when a patient attempts to exit
a wheelchair, chair or bed without assistance. Removal of pressure from the sensor pad beneath the patient
activates the alarm.
Contraindications
WARNING: This device is not a substitute for routine visual monitoring by caregiver, and is
intended to be used in conjunction with a comprehensive fall management program.
WARNING: No claim is made that this device will prevent patient falling or reduce the likelihood of
injury as a result. The caregiver is responsible for ensuring that this product is properly installed,
tested, and functioning before each use.
CAUTION: This device is not suitable for all patients, including but not limited to, those wishing to
defeat the alarm system.
CAUTION: This product is designed for individuals weighing 74 lb or more.
CAUTION: Seat cushions, sling seats, and other environmental factors may negatively affect alarm
operation.
Manufactured for
GF Health Products, Inc.
Atlanta, GA 30360
GF13702-INS-LAB-RevB14 3GF13702-INS-LAB-RevB14 2
ALARM SETUP
Fig. BFig. A Fig. C Fig. D
Delay
Switch
Volume
Switch
Low
Battery
Indicator
Advanced Patient Alarm Monitor
AC Adapter
Jack
Sensor Mat
Jack
Battery
Compartment
Strain
Relief
Nurse Call
Button
Reset
Button
Monitor
Positioning
Clip
1. Remove the screw that holds the battery compartment cover in place.
2. Inside the battery compartment (see Fig. B):
a. Set the volume switch to LO, HI, or RE (remote) volume. For description of the remote alarm feature, see
next page.
b. Set the delay switch to either 0 or 2 seconds. Selecting 2-second alarm delay may reduce the likelihood
of false alarms when used with restless patients.
c. Install a new 9V alkaline battery (not included).
d. Replace battery compartment cover and reinstall screw.
Alarm Testing
WARNING: Test alarm before each use.
1. Insert the sensor pad’s telephone-style plug into the monitor’s sensor mat jack (see Fig. C) until it clicks
into place.
2. To protect the cable plug end from accidental pull force, route the cable through the strain relief (see Fig. C).
3. Use your hand to apply pressure to the sensor pad. The monitor will automatically sense the pressure and
beep when activated.
4. Remove the pressure and the alarm will sound. To silence the alarm, press the reset button on the side of
the monitor (see Fig. D).
Patient Setup
1. Follow this procedure before each use.
a. If using a chair sensor pad, center the chair sensor pad under the patient’s buttocks.
b. If using a bed sensor pad, center the bed sensor pad under the patient’s buttocks.
2. The monitor will automatically sense the pressure and beep when activated.
3. Have the patient attempt to stand to remove pressure from the sensor pad. The alarm will sound. To silence
the alarm, press the reset button on the side of the monitor (see Fig. D).
4. Place monitor out of reach of patient.
WARNING: If alarm fails to sound, immediately discontinue use and review alarm setup.
Note: False alarms are often the result of poor pad placement. For instance, patients may position
themselves differently on a bed with a raised head section than on a completely flat mattress.
To reduce the likelihood of false alarms, the caregiver must ensure that the pad is correctly placed for each
patient’s unique positioning needs.
Remote Alarm (Optional)
The remote alarm option allows the caregiver to silence the local alarm and sound the alarm remotely through
the nurse call system.
1. Select the remote alarm option: Open the battery compartment and move the volume switch to “RE”
(remote) position.
2. Connect the monitor to the nurse call system using an adapter cable and splitter (not included).
WARNING: Test the alarm in the nurse call system before each use. If alarm fails to sound,
immediately discontinue use and review alarm setup.
Note: Optional accessories (adapter cable and AC adapter) are available through GF Health Products,
Inc.
FLOOR MAT MONITOR SETUP & USE
1. Remove the tamper-resistant screw, which holds the battery compartment cover in place.
2. Inside the battery compartment (See Figure B):
a. Select the volume level “Lo, Hi or Remote”.
b. Set the alarm “Delay” time to 0 seconds.
c. Set the Pad – Floor switch to Floor
d. Install a new 9V alkaline battery (Not included)
e. Replace battery compartment cover and reinstall tamper-resistant screw.
3. Bench test monitor and sensor pad prior to use.
a. Insert the Floor Mat’s telephone style plug into “Sensor Mat” jack on the monitor until you hear or feel
the “click” of the plug locking in place. (See Figure C)
b. Route the cable through the strain relief to protect the plug end from accidental pull force applied to the
cable. (See Figure C)
c. Step on the sensor mat with your foot to apply pressure to the sensor. The monitor will alarm
immediately. To silence the alarm, press the reset button located on the side of the alarm. (See Figure D)
MAINTENANCE
WARNING: Test the alarm before each use. If alarm fails to sound, remove it from service
immediately.
Check before each use that all alarm components are functional and battery is charged. To replace battery, see
ALARM SETUP section on previous page.
Note: The low battery indicator on the front of the alarm will flash when battery is low and needs to be
replaced. Replace battery immediately if indicator flashes (see Fig. A).
Cleaning
1. To clean: Gently wipe alarm components with a soft, clean cloth.
CAUTION: Do not use cleansers that could damage the alarm's nish. Do not submerge any alarm
components in liquid.

Popular Security System manuals by other brands

Whelen Engineering Company RB6PAPS installation guide

Whelen Engineering Company

Whelen Engineering Company RB6PAPS installation guide

Eaton i-on50EX installation guide

Eaton

Eaton i-on50EX installation guide

Bulldog Security 781 Installation and owner's guide

Bulldog Security

Bulldog Security 781 Installation and owner's guide

Speaks z wave WWA-01 instruction manual

Speaks z wave

Speaks z wave WWA-01 instruction manual

Schlage keypad entry with built-in alarm user guide

Schlage

Schlage keypad entry with built-in alarm user guide

ADT Safewatch Plus Enterpreneur Security Systems user manual

ADT

ADT Safewatch Plus Enterpreneur Security Systems user manual

Super Circuits BLK-SY10 Series Systems Setup guide

Super Circuits

Super Circuits BLK-SY10 Series Systems Setup guide

Esse-ti Helpy L200 user manual

Esse-ti

Esse-ti Helpy L200 user manual

Sabik Marine Vega VLB-44X Series product manual

Sabik Marine

Sabik Marine Vega VLB-44X Series product manual

Lorex SG17LD800 Frequently asked questions

Lorex

Lorex SG17LD800 Frequently asked questions

Aruba LS-BT1 installation guide

Aruba

Aruba LS-BT1 installation guide

DA System Aplique 6/14COM installation guide

DA System

DA System Aplique 6/14COM installation guide

Tracon CO218A user manual

Tracon

Tracon CO218A user manual

jablotron JA-16A RB manual

jablotron

jablotron JA-16A RB manual

Eldes ESIM264 manual

Eldes

Eldes ESIM264 manual

Gemini 953 Installation and use manual

Gemini

Gemini 953 Installation and use manual

First Alert One Link SA520B Specifications

First Alert

First Alert One Link SA520B Specifications

jablotron JA-151A quick start guide

jablotron

jablotron JA-151A quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.