DHS Data Logging System
6 Warner Road, Warner, NH 03278
(603) 456-2011 | madgetech.com DOC-1322009-00 REV 2 2020.04.27
Ordering Information
DHS Data Logging System PN 900002-00 HiTemp140-M12 data logger, a 36 inch flexible fiberglass insulated RTD probe, and the ThermoVault Max extreme
temperature thermal barrier
For Quantity Discounts call (603) 456-2011 or email sales@madgetech.com
TIME VS. TEMPERATURE CHART
Ambient Temperature Exposure Time in Air
-200 °C (-328 °F) 128 minutes
-180 °C (-292 °F) 137 minutes
-160 °C (-256 °F) 148 minutes
-140 °C (-220 °F) 163 minutes
-120 °C (-184 °F) 183 minutes
-100 °C (-148 °F) 213 minutes
-80 °C (-112 °F) 263 minutes
-60 °C (-76 °F) 368 minutes
-40 °C to +140 °C
(-40 °F to +284 °F) Indefinitely
150 °C (302 °F) 601 minutes
160 °C (320 °F) 468 minutes
170 °C (338 °F) 396 minutes
180 °C (356 °F) 348 minutes
190 °C (374 °F) 313 minutes
200 °C (392 °F) 286 minutes
210 °C (410 °F) 265 minutes
220 °C (428 °F) 247 minutes
230 °C (446 °F) 233 minutes
240 °C (464 °F) 220 minutes
250 °C (482 °F) 209 minutes
260 °C (500 °F) 200 minutes
270 °C (518 °F) 192 minutes
280 °C (536 °F) 184 minutes
290 °C (554 °F) 178 minutes
300 °C (572 °F) 172 minutes
310 °C (590 °F) 166 minutes
320 °C (608 °F) 161 minutes
330 °C (626 °F) 157 minutes
340 °C (644 °F) 153 minutes
350 °C to 400 °C
(662 °F to 752 °F) 60 minutes
Disclaimer & Terms of Use
Listed specifications can be used to determine maximum allowable
exposure times for the HiTemp140 with ThermoVault Max. The barrier
extends the operating temperature of the logger up to, but not
exceeding, the measurement range. Please consult the measurement
range of the probe for temperatures above 250 °C (482 °F).
Both the data logger and ThermoVault Max must be at ambient
temperature, approximately 25 °C (77 °F) before being placed in the
extreme temperature environment. Immediately following exposure
to high temperature, the data logger should be removed from the
ThermoVault Max, using appropriate precautions, as it could be VERY
hot. Failing to remove the data logger may allow heat trapped in the
ThermoVault Max to continue to heat the data logger to potentially
unsafe levels.
The ThermoVault Max may take hours to fully cool down. Even if the
exterior of the ThermoVault Max is cool to the touch, the interior of the
barrier and its contents may still be VERY hot.
The ThermoVault Max is primarily intended for use in dry air
environments, but with the addition of the TMAX Wet Seal Kit, the
ThermoVault Max may also be used in liquids and steam environments.
If your application involves a ramp up to a temperature above 150 °C
(302 °F) and/or any complex temperature profile that isn’t a constant
process, please contact MadgeTech to determine whether the
HiTemp140 with ThermoVault Max is suitable for the application.
To determine if the HiTemp140 with ThermoVault Max is suitable for
the application, please provide MadgeTech with a detailed description
of your process, including temperatures, durations, ramp times and
process media such as air, steam, oil or water. If MadgeTech is unable to
definitively calculate the suitability of our product for your application, a
test unit outfitted with a high temperature indicator can be provided.