MadgeTech HiTemp150FP-TSK User manual

HiTemp150FP-TSK
Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289
To view the full MadgeTech product line,
visit our website at www.madgetech.com.
Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE,
FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH, PENETRATE OR INCINERATE. BATTERY MAY
LEAK OR EXPLODE IF HEATED ABOVE 150 °C (302 °F).
Specifications subject to change.
See MadgeTech’s terms and conditions at www.madgetech.com
MadgeTech, Inc.
6 Warner Road lWarner, NH 03278
Phone 603.456.2011 lFax 603.456.2012
www.madgetech.com l[email protected]
DOC-1197035-00 REV 8 2014.09.30
HiTemp150FP-TSK
Extended Range Temperature Data Logger
with Flexible Probe and Thermal Shield
Description HiTemp150FP-TSK
Temperature Sensor 100 ΩPlatinum RTD
Temperature Range -40 °C to +150 °C
Probe Measurement
Range -40 °C to +250 °C
Temperature Resolution 0.05 °C
Calibrated Accuracy +0.5 °C (over 100 °C span)
Memory 32,767 readings
Reading Rate 1 reading every second up to 1 reading every 12 hours
Required Interface
Package IFC110 or IFC200
Baud Rate 2,400
Typical Battery Life 1 year
Operating Environment -200 °C to +250 °C (Time Limited),
0 %RH to 95 %RH (non-condensing)
Material HiTemp150: 316 Stainless Steel
Thermal Shield and Cable: TEFLON®
Dimensions
Body: 1.1 in x 1.75 in dia. (28 mm x 45 mm dia.)
Shield: 2.1 in x 2.6 in dia. (53.34 mm x 66 mm dia.)
Probe: 4 in x 1/8 in dia. (102 mm x 4 mm dia.)
Weight 9 oz (255 g)
Approvals CE
Product Information Card
Product Notes
Getting Started
The HiTemp150FP-TSK consists of a HiTemp150FP data logger in a thermal enclosure. The
thermal enclosure allows the device to withstand higher temperatures for certain durations
of time. The chart located below outlines the time versus temperature durations.
To start the logger, unscrew the thermal enclosure and separate the thermal enclosure lid,
from the body of the data logger. Unscrew the knurled nut to access the communication
port. Screw the knurled nut back on.
The logger must be removed from the barrier immediately after removal from the heat
environment. Be careful, the logger may be very hot.
Operating Environment
The HiTemp150FP-TSK has a time limited operating environment of between -160 °C to
+250 °C, and 0 %RH to 95 %RH, non-condesing.
O-Rings
O-ring maintenance is a key factor when properly caring for the HiTemp150FP-TSK. The
o-rings ensure a tight seal and prevent liquid from entering the inside of the device.
Please refer to the application note “O-Rings 101: Protecting Your Data”, found on the
MadgeTech website, for information on how to prevent O-ring failure.
Installation Guide
Installing the Interface cable
- IFC200
Insert the device into a USB port. The drivers will install automatically.
- IFC110
Plug the serial cable into the port and verify it is secure.
- USB-1 or USB-101
Install the USB drivers from the USB software stick provided in the kit, then plug the USB
cable into the computer and the serial cable into the serial port.
Product Information Card
HiTemp150FP-TSK
Installing the software
Insert the Software USB into an open port. If the autorun does not appear, locate the drive
on the computer and double click on Autorun.exe. Follow the instructions provided in the
Wizard.
Device Operation
Connecting and Starting the data logger
- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer.
- The device will appear in the Connected Devices list, highlight the desired data logger.
- For most applications, select “Custom Start” from the menu bar and choose the desired
start method, reading rate and other parameters appropriate for the data logging
application and click “Start”. (“Quick Start” applies the most recent custom start options,
“Batch Start” is used for managing multiple loggers at once, “Real Time Start” stores the
dataset as it records while connected to the logger.)
- The status of the device will change to “Running”, “Waiting to Start” or “Waiting to
Manual Start”, depending upon your start method.
- Disconnect the data logger from the interface cable and place it in the environment to
measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device
cannot be restarted until it has been re-armed by the computer.
Downloading data from a data logger
- Connect the data logger to the interface cable.
- Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “Download”. You will
be prompted to name your report.
- Downloading will offload and save all the recorded data to the PC.
Device Maintenance
Battery Replacement
The HiTemp150FP-TSK must be sent to MadgeTech for battery replacement.
Recalibration
The HiTemp150FP standard calibration is two points at 50 °C and 150 °C.
Pricing:
Recalibration traceable to NIST $70.00
Recalibration $40.00
Prices and specifications subject to change. See MadgeTech’s terms and conditions at www.madgetech.com.
To send the devices back, visit www.madgetech.com, select Services then RMA Process.
Ambient Temperature Time in Air to Max Internal Temp
(150 °C/302 °F)
160 °C (320 °F) 75 minutes
170 °C (338 °F) 63 minutes
180 °C (356 °F) 55 minutes
190 °C (374 °F) 50 minutes
200 °C (392 °F) 45 minutes
210 °C (410 °F) 42 minutes
220 °C (428 °F) 39 minutes
230 °C (446 °F) 36 minutes
240 °C (464 °F) 34 minutes
250 °C (482 °F) 32 minutes

Other MadgeTech Data Logger manuals

MadgeTech QuadTCTempV2 Specification sheet

MadgeTech

MadgeTech QuadTCTempV2 Specification sheet

MadgeTech UltraShock Series User manual

MadgeTech

MadgeTech UltraShock Series User manual

MadgeTech UltraShock Specification sheet

MadgeTech

MadgeTech UltraShock Specification sheet

MadgeTech AVS140-1 Specification sheet

MadgeTech

MadgeTech AVS140-1 Specification sheet

MadgeTech QuadRTD User manual

MadgeTech

MadgeTech QuadRTD User manual

MadgeTech LyoTemp Specification sheet

MadgeTech

MadgeTech LyoTemp Specification sheet

MadgeTech Bridge101A User manual

MadgeTech

MadgeTech Bridge101A User manual

MadgeTech HiTemp140X2 Series Specification sheet

MadgeTech

MadgeTech HiTemp140X2 Series Specification sheet

MadgeTech RFpHTemp101A User manual

MadgeTech

MadgeTech RFpHTemp101A User manual

MadgeTech MicroTemp Specification sheet

MadgeTech

MadgeTech MicroTemp Specification sheet

MadgeTech RHTEMP1000IS Specification sheet

MadgeTech

MadgeTech RHTEMP1000IS Specification sheet

MadgeTech RHTemp1000IS-KR Specification sheet

MadgeTech

MadgeTech RHTemp1000IS-KR Specification sheet

MadgeTech Bridge101A-30 User manual

MadgeTech

MadgeTech Bridge101A-30 User manual

MadgeTech RHTemp101 User manual

MadgeTech

MadgeTech RHTemp101 User manual

MadgeTech Logicbus X Series Specification sheet

MadgeTech

MadgeTech Logicbus X Series Specification sheet

MadgeTech QuadTemp2000 User manual

MadgeTech

MadgeTech QuadTemp2000 User manual

MadgeTech X Series Specification sheet

MadgeTech

MadgeTech X Series Specification sheet

MadgeTech Bridge120 User manual

MadgeTech

MadgeTech Bridge120 User manual

MadgeTech OctVolt Specification sheet

MadgeTech

MadgeTech OctVolt Specification sheet

MadgeTech Cryo-Temp Specification sheet

MadgeTech

MadgeTech Cryo-Temp Specification sheet

MadgeTech Shock101-50 Specification sheet

MadgeTech

MadgeTech Shock101-50 Specification sheet

MadgeTech Volt101A-30V Specification sheet

MadgeTech

MadgeTech Volt101A-30V Specification sheet

MadgeTech Shock101-EB Specification sheet

MadgeTech

MadgeTech Shock101-EB Specification sheet

MadgeTech RTDTemp2000 User manual

MadgeTech

MadgeTech RTDTemp2000 User manual

Popular Data Logger manuals by other brands

Jeulin Tooxy quick start guide

Jeulin

Jeulin Tooxy quick start guide

TESTO Saveris 2 Guide

TESTO

TESTO Saveris 2 Guide

Ebro EBI-85 A manual

Ebro

Ebro EBI-85 A manual

ELPRO Hamster-E ET1 operating manual

ELPRO

ELPRO Hamster-E ET1 operating manual

National Instruments USB-9472 USER GUIDE AND SPECIFICATIONS

National Instruments

National Instruments USB-9472 USER GUIDE AND SPECIFICATIONS

Dickson SK500 Getting started

Dickson

Dickson SK500 Getting started

Measurement Computing USB-2404-UI user guide

Measurement Computing

Measurement Computing USB-2404-UI user guide

MONI LOG EnDal smart instruction manual

MONI LOG

MONI LOG EnDal smart instruction manual

HT SOLAR-02 user manual

HT

HT SOLAR-02 user manual

Geokon GeoNet 8000 Series Installing

Geokon

Geokon GeoNet 8000 Series Installing

TESTO 623 instruction manual

TESTO

TESTO 623 instruction manual

Cutting Edge GL2100 user manual

Cutting Edge

Cutting Edge GL2100 user manual

Hafele Dialock MDU 100 operating instructions

Hafele

Hafele Dialock MDU 100 operating instructions

CHCNAV HCE320 user guide

CHCNAV

CHCNAV HCE320 user guide

Soil Instruments GTecLink user manual

Soil Instruments

Soil Instruments GTecLink user manual

Gulf Coast Data Concepts X200-5 user manual

Gulf Coast Data Concepts

Gulf Coast Data Concepts X200-5 user manual

Xylem ebro EBI 40 manual

Xylem

Xylem ebro EBI 40 manual

mIoT Captis Multi installation guide

mIoT

mIoT Captis Multi installation guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.