manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. MadgeTech
  6. •
  7. Data Logger
  8. •
  9. MadgeTech RHTemp1000IS-KR Specification sheet

MadgeTech RHTemp1000IS-KR Specification sheet

This manual suits for next models

1

Other MadgeTech Data Logger manuals

MadgeTech QuadRTDTempV2 Specification sheet

MadgeTech

MadgeTech QuadRTDTempV2 Specification sheet

MadgeTech RHTEMP1000EX Specification sheet

MadgeTech

MadgeTech RHTEMP1000EX Specification sheet

MadgeTech Event101A User manual

MadgeTech

MadgeTech Event101A User manual

MadgeTech LynxPro User manual

MadgeTech

MadgeTech LynxPro User manual

MadgeTech Volt101A User manual

MadgeTech

MadgeTech Volt101A User manual

MadgeTech Temp1000P User manual

MadgeTech

MadgeTech Temp1000P User manual

MadgeTech PR1000EX-30A Specification sheet

MadgeTech

MadgeTech PR1000EX-30A Specification sheet

MadgeTech Shock101 Specification sheet

MadgeTech

MadgeTech Shock101 Specification sheet

MadgeTech IC-RFCO2RH Specification sheet

MadgeTech

MadgeTech IC-RFCO2RH Specification sheet

MadgeTech QuadState User manual

MadgeTech

MadgeTech QuadState User manual

MadgeTech HiTemp140-FP-6 Specification sheet

MadgeTech

MadgeTech HiTemp140-FP-6 Specification sheet

MadgeTech QuadTemp2000 User manual

MadgeTech

MadgeTech QuadTemp2000 User manual

MadgeTech VFC2000 User manual

MadgeTech

MadgeTech VFC2000 User manual

MadgeTech TSR101-EB User manual

MadgeTech

MadgeTech TSR101-EB User manual

MadgeTech HiTempl40-PT-5 Specification sheet

MadgeTech

MadgeTech HiTempl40-PT-5 Specification sheet

MadgeTech Titan S8 Specification sheet

MadgeTech

MadgeTech Titan S8 Specification sheet

MadgeTech Rain101A Specification sheet

MadgeTech

MadgeTech Rain101A Specification sheet

MadgeTech RFC1000 User manual

MadgeTech

MadgeTech RFC1000 User manual

MadgeTech Volt101A-160 Specification sheet

MadgeTech

MadgeTech Volt101A-160 Specification sheet

MadgeTech QuadTemp2000 User manual

MadgeTech

MadgeTech QuadTemp2000 User manual

MadgeTech PR140 Specification sheet

MadgeTech

MadgeTech PR140 Specification sheet

MadgeTech Shock101-100 Specification sheet

MadgeTech

MadgeTech Shock101-100 Specification sheet

MadgeTech State101A Specification sheet

MadgeTech

MadgeTech State101A Specification sheet

MadgeTech Therm-A-lert Specification sheet

MadgeTech

MadgeTech Therm-A-lert Specification sheet

Popular Data Logger manuals by other brands

UGCS SkyHub user manual

UGCS

UGCS SkyHub user manual

Avisaro 4.0 Series user manual

Avisaro

Avisaro 4.0 Series user manual

DDB C3 manual

DDB

DDB C3 manual

TESTO 176 instruction manual

TESTO

TESTO 176 instruction manual

HBM SOMAT XR CX22B-R operating manual

HBM

HBM SOMAT XR CX22B-R operating manual

Agtrix 380i manual

Agtrix

Agtrix 380i manual

Beijing ART Tech PCI2320 user manual

Beijing ART Tech

Beijing ART Tech PCI2320 user manual

Athena GET SL1 user manual

Athena

Athena GET SL1 user manual

Motec Systems C185 user manual

Motec Systems

Motec Systems C185 user manual

Beijing ART Tech USB2832 user manual

Beijing ART Tech

Beijing ART Tech USB2832 user manual

HT Italia XL421 user manual

HT Italia

HT Italia XL421 user manual

Racelogic VBOX LapTimer quick start guide

Racelogic

Racelogic VBOX LapTimer quick start guide

Edu-Logger 900-240 Guide

Edu-Logger

Edu-Logger 900-240 Guide

Elma DT173 manual

Elma

Elma DT173 manual

Comet U0121M user guide

Comet

Comet U0121M user guide

pico Technology PT-104 user guide

pico Technology

pico Technology PT-104 user guide

Sensus DIAVASO user manual

Sensus

Sensus DIAVASO user manual

Soil Instruments GTecLink user manual

Soil Instruments

Soil Instruments GTecLink user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

RHTEMP1000IS
Intrinsically Safe Temperature
and Humidity Data Logger
PRODUCT
USER GUIDE
TABLE OF CONTENTS
2Product Overview
2Installation Guide
2Device Operation
3Device Maintenance
4Need Help?
www.calcert.com [email protected]1.888.610.7664 0
5
10
15
20
25
30
Product User Guide | 2
RHTemp1000IS
PRODUCT USER GUIDE
Product Overview
The RHTemp1000IS carries hazardous location, intrinsically
safe certification in accordance with the latest issue of:
FM3600, FM3610, CAN/CSA-C22.2 No. 60079-0:15,
CAN/CSA-C22.2 No. 60079-11:14
Certified Intrinsically Safe for:
• Class 1 Division 1 Group ABCD
• Class 1 Division 2 Group ABCD
• Temperature Class: T4A
Operational Warnings
• When used in hazardous locations, the RHTemp1000IS
is to be installed prior to the location becoming
hazardous and removed only after the area is no longer
hazardous.
• The maximum allowed ambient temperature for the
RHTemp1000IS (under any circumstances) is 80 °C. The
minimum rated operating temperature is -40 °C.
• The RHTemp1000IS is approved for use only with the
Tadiran TL-2150/S battery. Replacement with any other
battery will void the safety rating.
• Batteries are user replaceable, but are to be removed or
replaced only in locations known to be non-hazardous.
• Tampering or replacement of non-factory components
may adversely affect the safe use of the product, and is
prohibited. Except for replacement of the battery, the
user may not service the RHTemp1000IS. MadgeTech,
Inc. or an authorized representative must perform all
other service to the product.
Ordering Information
• 902213-00 — RHTemp1000IS
• 902218-00 — RHTemp1000IS-KR (key ring end cap)
• 900319-00 — IFC400
• 900325-00 — IFC406
• 901745-00 — Battery Tadiran TL-2150/S
Installation Guide
Installing the Software
The Software can be downloaded from the MadgeTech
website at madgetech.com. Follow the instructions
provided in the Installation Wizard.
Installing the USB Interface Drivers
IFC400 or IFC406 — Follow the instructions provided in
the Installation Wizard to install the USB Interface Drivers.
Drivers can also be downloaded from the MadgeTech
website at madgetech.com.
Device Operation
Connecting and Starting the Data Logger
1. Once the software is installed and running, plug the
interface cable into the docking station (IFC400 or
IFC406).
2. Connect the USB end of the interface cable into an
open USB port on the computer.
3. Place the data logger into the docking station (IFC400
or IFC406).
4. The data logger will automatically appear under
Connected Devices within the software.
5. For most applications, select Custom Start from
the menu bar and choose the desired start method,
reading rate and other parameters appropriate for the
data logging application and click Start. (Quick Start
applies the most recent custom start options, Batch
Start is used for managing multiple loggers at once,
Real Time Start stores the dataset as it records while
connected to the logger.)
6. The status of the device will change to Running or
Waiting to Start, depending upon your start method.
7. Disconnect the data logger from the interface cable
and place it in the environment to measure.
Note: The device will stop recording data when the end of memory is
reached or the device is stopped, unless user selectable memory wrap
is enabled. At this point the device cannot be restarted until it has been
re-armed by the computer.
www.calcert.com [email protected]1.888.610.7664 0
5
10
15
20
25
30
Product User Guide | 3
Device Operation (cont’d)
Downloading Data from a Data Logger
1. Place the logger into the docking station (IFC400 or
IFC406).
2. Highlight the data logger in the Connected Devices
list. Click Stop on the menu bar.
3. Once the data logger is stopped, with the logger
highlighted, click Download.
4. Downloading will offload and save all the recorded
data to the PC.
Device Maintenance
Battery Replacement
Materials: Replacement Battery (Tadiran TL-2150/S)
1. Move the device to a non-hazardous location before
replacing battery.
2. Observe Operational Warnings when removing and
replacing the battery.
3. Unscrew the bottom of the data logger and remove the
battery.
4. Place the new battery into the logger. Caution: Observe
correct battery polarity when installing.
5. Screw the cover onto the data logger.
RHTemp1000IS
PRODUCT USER GUIDE
O-Rings
O-ring maintenance is a key factor when properly caring
for the RHTemp1000IS. The O-rings ensure a tight
seal and prevent liquid from entering the inside of the
device. Please refer to the application note “O-Rings 101:
information on how to prevent O-ring failure.
Recalibration
Additional Services:
Custom calibration and verification point options
available, please call for pricing.
Communication
To ensure desired operation of the RHTEMP1000IS, please keep the surface clear of any foreign objects
or substances. The RHTEMP1000IS’s data is downloaded through external contact with the IFC400 or
IFC406 docking station. Covering the surface with foreign objects (i.e. Calibration Labels) can prevent the
communication and/or downloading process.
Protecting Your Data”, for
www.calcert.com [email protected]1.888.610.7664 0
5
10
15
20
25
30