marsden M-510 User manual

USER MANUAL
M-510
Please take time to read these
instructions before starting
to use the scale
Version 1.1 10/20
Please keep this instruction manual to hand for future reference

Contents
Introduction 3
Product Specification 3
INTENDED USE 3
Safety Instructions 4
Explanation of Graphic Symbols 5
Power Supply and Low Battery 6
Operation: Basic Functions 7
Operation: Advanced Functions 9
Using the Scale with a Printer 13
Using the Scale with Bluetooth 14
EMC Guidance and Manufacturer’s Declaration 15
Recommended Separation Distance 17
Error Messages 19
Manufacturer’s Declaration of Conformity 20
Page 2

Thank you for purchasing a Marsden professional medical scale. This is a
precision Class III Weighing Instrument and considerate use will result in
many years of accurate weighing.
The scale has a maximum load capacity of 250kg which must not be
exceeded.
Product Specification
Model
M-510
Accuracy Class
Class III
Capacity/Division
250kg x 100g
Weight of Scale
Approximately 6kg
Units of Measure
Kg
Function Keys
ON/OFF, ZERO, HOLD, TARE, BMI, UNIT,
PRINT,
0-9
Stabilization Time
1-2 Seconds
Operating Temperature
0 to 40˚C
Power Supply
Rechargeable battery pack
6 x AA batteries*
Mains Adaptor
Adaptor Type
2UE24WV-120100SPA & UE24WB-
120100SPA
Indicator Display
2.5 cm LCD display with 5 active digits
Dimensions
310mm x 310mm x 90mm
Warranty
4 years
*contact Marsden for details
Page 3

Safety Instructions
Before putting the device into use, please read with care the information given in this user manual,
which contains important instructions for proper installation, use and maintenance of the device.
Marsden/the manufacturer shall not be liable for damages arising from failure to heed the following
instructions:
•When using electrical components under increased safety requirements, always comply with
appropriate regulations.
•Inappropriate installation/use will render the warranty null and void.
•Ensure the voltage marked on the power supply unit matches your mains supply.
•This device is designed for use indoors.
•Observe the permissible ambient temperatures for use.
•The device meets the requirements for electromagnetic capability. Do not exceed the
maximum values specified in the applicable standards.
•Batteries should be kept away from small children. If swallowed, promptly seek medical
assistance.
If you have any problems, contact Marsden/your local dealer/your service partner.
Cleaning
•We recommend using alcohol-based wipes or similar when cleaning the scale.
•Please do not use corrosive liquids, large amounts of water or high pressure washers.
•Always disconnect the scale from the mains power supply before cleaning.
Maintenance
•The scale does not require any routine maintenance. However, we recommend checking the
scale’s accuracy at regular intervals. If any inaccuracies occur, please contact your local dealer
or service partner.
•Marsden can provide service contracts for your weighing scales, to keep them accurate and
reliable. For more information about Marsden service contracts call 01709 364296.
Disposing of the Scale
•This product should not be treated as regular household waste, but should be handed in to an
electrical/electronic equipment recycling centre.
•You can obtain further details from your local council, your municipal waste disposal company
or from where you purchased the product.
Intended use
•This scale is intended for use to determine the weight of patients, supported by professional
personnel and in rooms intended for carrying out healthcare. The weighing value can be read
after a stable weighing value has been obtained. Before use, the scale must be checked by an
authorised person to ensure it's in a suitable condition.
Page 4

Explanation of Graphic Symbols
SN-21300100
Designation of the serial number of
every device.
(Number as an example)
“Please note the accompanying
documents” or “Observe operating
instructions”
Charder Electronic Co. Ltd
No.103 Guozhong Rd, Dali Dist,
Taichung City 412, Taiwan (R.O.C)
Identification of manufacturer of
medical product including address.
Type B Applied Part.
Dispose of old appliances separately
from your household waste.
This product must be disposed of at a
communal collection point.
Carefully read this operation manual
before setup and commissioning, even if
you are already familiar with Marsden
scales.
Page 5

Power Supply & Low Battery
The indicator uses a rechargeable battery pack, or can be powered from the mains via the AC adaptor.
Make sure the battery pack is installed in the battery box of the indicator. Alternatively, plug the AC
adaptor (12V 2A) into the port on the side of the indicator.
Installing & Replacing the Battery Pack
1. Take out the battery housing.
2. The rechargeable battery pack will slide into, or out of, the housing.
3. Check that the housing pin is connecting to the right point inside the indicator.
4. Place the housing back in the back of the indicator, and close the battery housing cover.
Optional battery
holder (to fit AA sized
batteries –contact
your dealer)
Page 6

Operation: Basic Functions
Switching on the Scale
Press the ON/OFF button firmly.
The scale will first test all of the display
segments.
The scale will now show its current
software version number.
The scale will now go into weighing mode
and should show 0.0 on the display.
Page 7

Switching off the Scale
Press the ON/OFF button when the scale is
turned on. The scale will now power down.
Setting the Scale to Zero
If for any reason the scale shows a reading
other than 0.0 it can be reset to zero.
Press ZERO once.
The scale will return to 0.0.
Page 8

Operation: Advanced Functions
Hold Function
Press the HOLD button once.
Allow the patient to get on the scale.
After a few seconds the scale will lock on
the person’s weight. When the patient
leaves the scale, the weight will remain on
the display.
Press HOLD again to disable the Hold
function and return the scale to 0.0.
Page 9

Body Mass Index (BMI) Function
In normal mode, press BMI to enter BMI
mode.
The display will show the last height
entered and the extreme left digit will
flash. Enter the height by using the
numeric keys. Press ZERO to confirm the
height. (NB: There will always be an active
flashing digit in the height display, unless
HOLD is pressed).
Weigh the patient as normal. The display
will show the weight, height and BMI
value. At this time, the weight and height
can be freely changed, and the BMI value
will be automatically calculated according
to the changed weight and height.
Press the BMI button to return to normal
weighing mode.
Page 10

Tare and Pre-Set Tare Functions
Press TARE for three seconds to enter Pre-
set Tare setting mode.
The display will show the last pre-set tare
entered and the extreme left digit will
flash.
Enter the pre-set tare value by using the
numeric keys, then press TARE again to
confirm the value.
Press ZERO to return to normal weighing
mode.
To use the Tare function, add the item you
wish to tare off to the scale, and press
TARE. The display will show zero, and then
a minus number when the item is removed
from the scale.
Page 11

Setting the Date
Press HOLD for three seconds to access the time setting mode. The time period digit
that is flashing can be changed by entering the appropriate number using the numeric
keys. The time period to be edited is selected by pressing HOLD.
E.g. To input 25 December 2008, 8:00 a.m.:
Enter the year. Press HOLD to confirm
and access the date editing field.
Enter the date. E.g. “12.25” for
December 25th. Press HOLD to confirm
and access the time clock editing field.
Enter the time (24 hour clock only).
Press HOLD and the display shows:
YYYY→MM.DD→HH:SS
Press HOLD to return to normal
weighing mode.
Page 12

Using the Scale with a Printer
An optional Marsden external printer (model TP-2100) is available for all models. With the
printer fitted, the patient’s weight, height, and BMI result can be printed.
Once the person has been weighed and their BMI calculated, simply press PRINT to produce
the following ticket:
Connecting the TP-2100 Thermal Printer
Plug the cable to the printer, and then connect its 9D connector to the indicator.
GROSS WEIGHT 60.00kg
TARE WEIGHT 30.00kg
NET WEIGHT 30.00kg
PATIENT HEIGHT 100.0cm
PATIENT B.M.I 37. 5
29/12/2008 17:00
Page 13

Using the Scale with Bluetooth
If your scale has Bluetooth connectivity, the universal Bluetooth symbol will be on the main
indicator display.
Bluetooth Connection
Long press ZERO for three seconds to enter the Setting mode
and then display the A-OFF menu.
Press TARE twice, and then press HOLD once to enter the
Bluetooth setting mode.
Using the HOLD button, select “ON” (enable) or “OFF”
(disable).
Press TARE to confirm the setting.
Note: Disabling the Bluetooth function when not in use
will reduce battery power consumption.
Display the “bluEt” menu. Press TARE once.
Press HOLD to return to normal mode.
Search for the scale in your computer or device’s Bluetooth
settings (procedure may vary depending on device or system)
The scale will appear on the Bluetooth device list as
“MARSDEN BT”.
Connect your device to “MARSDEN BT”, and the scale is ready
to transmit data wirelessly via Bluetooth.
Page 14

EMC Guidance and Manufacturer’s Declaration
Guidance and manufacturer’s declaration-electromagnetic emissions
The MEDICAL SCALE M-510 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the MEDICAL SCALE M-510 should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic environment-
guidance
RF emissions CISPR 11
Group 1
The MEDICAL SCALE M-510 uses RF
energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The MEDICAL SCALE M-510 is suitable
for use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions IEC 61000-
3-2
Class A
Voltage fluctuations /flicker
emissions IEC 61000-3-3
Compliance
Guidance and manufacturer’s declaration-electromagnetic immunity
The MEDICAL SCALE M-510 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the MEDICAL SCALE M-510 should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment-guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8
kV, ± 15 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%
Electrical fast
transient/burst IEC
61000-4-4
± 2kV for power
supply lines + 1kV
for input/output
lines
± 2kV for power
supply lines Not
applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-
5
± 1kV line(s) to
line(s) ± 2kV line(s)
to earth
± 1kV differential
mode Not applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
Page 15

Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
0% UT for 0,5 cycle
0% UT for 1 cycle
70% UT(30% dip in
UT) for 25 cycles
0% UT for 5 s
<5% UT(>95% dip in
UT) for 0,5 cycle 40%
UT(60% dip in UT)
for 5 cycles 70%
UT(30% dip in UT)
for 25 cycles <5%
UT(>95% dip in UT)
for 5 s
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the MEDICAL SCALE M-510
requires continued operation
during power mains
interruptions, it is
recommended that the
MEDICAL SCALE M-510 be
powered from an
uninterruptible power supply
or a battery.
Power
frequency(50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
3 A/m
The MEDICAL SCALE M-510
power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration-electromagnetic immunity
The MEDICAL SCALE M-510 is intended for use in the electromagnetic environment specified
below.
The customer or the user of the MEDICAL SCALE M-510 should assure that is used in such and
environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment-
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 KHz to 80 MHz
6 V in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
3 Vrms
Portable and mobile RF
communications equipment
should be used no closer to any
part of the MEDICAL SCALE M-
510 including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d = 1,2 √
P
d = 1,2 √
P
80MHz to 800 MHz
d = 2,3 √
P
800MHz to 2,7 GHz
Where
P
is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
d
is the recommended separation
Page 16

Radiated RF IEC
61000-4-3
3 V/m 80MHz to 2,7
GHz
3 V/m
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site surveya,
should be less than the
compliance level in each
frequency rangeb.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the MEDICAL SCALE M-510 is used
exceeds the applicable RF compliance level above, the MEDICAL SCALE M-510 should be
observed to verify normal operation. If abnormal performance is observed, additional measures
my be necessary, such as re-orienting or relocating the MEDICAL SCALE M-510.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be les than 3 V/m.
Page 17

Recommended separation distance between portable and mobile RF communications
equipment and the MEDICAL SCALE
The MEDICAL SCALE M-510 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the MEDICAL SCALE M-510
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the MEDICAL SCALE M-510
as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d =1,2√
P
80 MHz to 800 MHz
d =1,2√
P
800 MHz to 2,7 GHz
d =2,3√
P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where p is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Page 18

Error Messages
Low Battery
The scale’s alkaline AA type batteries are flat;
please replace the batteries.
Overload
This indicates that the scale’s load sensor(s)
have been overloaded. Reduce the loading
and retry.
Counting Error
1. The signal from the load cells is too
high. Please remove any weight from
the scale and try to power on again. If
the scale continues to show the error
message, it indicates a fault with the
electronics or wiring.
2. The signal from the load cells is too
low. Please remove any weight from
the scale and try again. If the scale
continues to show the error message,
it indicates a fault with the electronics
or wiring.
High/Low Zero Count
1. The scale is above its zero range.
Please remove any weight from the
scale and power on again. If the scale
continues to show the error message,
it indicates a fault with the
electronics.
2. The scale is below its zero range.
Check there is nothing jammed
underneath the scale and power on
again. If the scale continues to show
the error message, it indicates a fault
with the electronics.
EEPROM Error
This indicates there is a fault with the scale’s
software and is normally caused by a fault
with the load cell or wiring. Contact your
local service representative.
Page 19

Manufacturer’s Declaration of Conformity
This product has been manufactured in accordance with the harmonized European standards,
following the provisions of the below stated directives:
93/42/EEC as amended by 2007/47/EC
Medical Device Directive
2014/31/EU
Non-automatic Weighing Instruments
Directive
Please see separate document showing on sticker of device for above CE marking.
Authorized EU Representative:
DISTRIBUTOR:
Manufactured by:
Charder Electronic Co., Ltd.
No.103, Guozhong Rd., Dali Dist.,
Taichung City 412,Taiwan (R.O.C.)
Page 20
Unit 1, Genesis Business Park,Sheffield Road, Rotherham, UK S60 1DX
Tel: +44 (0) 1709 364296 Fax: +44 (0) 1709 364293
E-mail: sales@marsdengroup.co.uk
Table of contents
Other marsden Scale manuals

marsden
marsden M-930 User manual

marsden
marsden V-100 User manual

marsden
marsden C-100 User manual

marsden
marsden V-24 User manual

marsden
marsden MPPS-250 User manual

marsden
marsden M-999 User manual

marsden
marsden MSS-I-400 User manual

marsden
marsden MS-150 User manual

marsden
marsden M-925 User manual

marsden
marsden M-400 User manual

marsden
marsden V-22 User manual

marsden
marsden M-545 User manual

marsden
marsden M-160 User manual

marsden
marsden B-100 User manual

marsden
marsden BS-250BT User manual

marsden
marsden M-550BT User manual

marsden
marsden B-200 User manual

marsden
marsden M-610 User manual

marsden
marsden BAS-250 User manual

marsden
marsden V-25 User manual