- 20 -
NOTE:
[1] You can used the MS3202 Socket according to the capacitor pin size.
. The pin( marked ”Cx/Lx/Temp”) of MS3202 Socket must be pointed
to the “VΩ”input terminal.
If the capacitor pin is thick ( i.e, Diameter of pin >1 mm),
the insert to the MS3202 socket is disabled. In this case ,
directly insert the capacitor under test to the “COM”and
“VΩ”terminals, or using the leads(probe) to measuring
capacitance of capacitor.
[2] To increase the accuracy of capacitance measurement when measuring
under 5nF capacitance, you can connecting 0.15uF Capacitor in parallel
with the capacitor being
measured , or use relative measurement mode to automatically subtract the
capacitance build-in equalized capacitance and residual capacitance of
PCB lead wire from the result. .
5-5. Frequency & Duty cycle measurement
The normal measurement ranges are 22.00Hz, 220.0Hz, 22.000KHz,
220.00KHz,
2.2000MHz., 22.000MHz and 220.00MHz.
1. Set the rotary switch to ‘Hz/%’position,and the‘Hz’sign shown
on the LCD display.
2. Insert the red test lead into the “ VΩ”input terminal and the black
- 21 -
3. test lead into the “COM”terminal.
4. Connect the test leads across with the circuit under testing. The
measured value shown.
on the LCD display.
5. To switch to the duty cycle mode , press “VAHz/%”. Push key.
The duty cycle mode range is within 5.0% to 94.9%. The minimum
resolution is 0.1%.
If the sorce frequency duty cycle is smaller than 5.0%, the “UL”sign
will be shown on
The LCD display, and then, if the duty cycle is larger than 94.9%, the
“OL”sign will be shown on the LCD display.
NOTE: Input signal level must be higher than 0.7V (it is sensitivity).
5-6. DC uA/AC uA measurement
1. Set the rotary switch to the uA position.
2. Select the DC uA or AC uAwith “FUNC”push button. The meter
default to DC current measurement mode.
3. Turn off power to the circuit. Break the circuit point to be measured.
4. Connect the two test lead to complete the broken circuit.
5. If the measured current is too high, the display will indicate “OL”. In
this care, the higher current range (mA orA) should be selected.
5-7. DC/AC mA measurement
1. Turn off power to the circuit. Set the rotary switch to the mA position.
2. The meter default to DC current measurement mode. SelectAC mA
measurement mode by “FUNC”push button.
3. Break the current path to be tested. Connect the red test lead to the